Article

Development of a Water Matrix Certified Reference Material for Volatile Organic Compounds Analysis in Water

State Environmental Protection Key Laboratory of Environmental Pollutant Metrology and Reference Materials, Institute for Environmental Reference Materials, Environmental Development Centre of the Ministry of Ecology and Environment, Beijing 100029, China
*
Author to whom correspondence should be addressed.
Submission received: 2 June 2021 / Revised: 9 July 2021 / Accepted: 12 July 2021 / Published: 20 July 2021
(This article belongs to the Special Issue Environmental Analysis of Organic Pollutants)

Abstract

:
Water matrix certified reference material (MCRM) of volatile organic compounds (VOCs) is used to provide quality assurance and quality control (QA/QC) during the analysis of VOCs in water. In this research, a water MCRM of 28 VOCs was developed using a “reconstitution” approach by adding VOCs spiking, methanol solution into pure water immediately prior to analysis. The VOCs spiking solution was prepared gravimetrically by dividing 28 VOCs into seven groups, then based on ISO Guide 35, using gas chromatography-mass spectrometry (GC-MS) to investigate the homogeneity and long-term stability. The studies of homogeneity and long-term stability indicated that the batch of VOCs spiking solution was homogeneous and stable at room temperature for at least 15 months. Moreover, the water MCRM of 28 VOCs was certified by a network of nine competent laboratories, and the certified values and expanded uncertainties of 28 VOCs ranged from 6.2 to 17 μg/L and 0.5 to 5.3 μg/L, respectively.

1. Introduction

Volatile organic compounds (VOCs), due to their toxicity and persistence in the environment, are one group of particularly important pollutants [1,2]. Many of these substances are toxic, and some are considered to be carcinogenic, mutagenic or teratogenic [3,4]. As reported, VOCs have been widely detected at trace levels in surface water and groundwater [2,5]. Therefore, during VOCs detection in water, references with VOCs are needed to calibrate the instrument, verify analysis procedure or control the analysis quality according to QA/QC (quality assurance and quality control) guidelines. Validation of the entire analytical procedure requires the use of an matrix certified reference material (MCRM), which is a homogeneous and well-defined matrix that contains known amounts of the target compounds [6,7,8]. However, there is currently no MCRM available for the measurement of VOCs in water. Therefore, in this research, we prepared and characterized a water MCRM of 28 VOCs.
There are several reasons for the current absence of MCRMs for organic materials in water. Specifically, the relatively low concentrations of most organic contaminants in water necessitates that large volumes of water be manipulated and transferred for analysis, which may lead to problems such as cross-contamination or improper and inaccurate dilution. Additionally, there is a tendency for target compounds to be adsorbed onto suspended or colloidal particles because of their low water solubility or their high lipophilic characteristics as expressed by the high octanol-water partition coefficient (Kow). Furthermore, the instability of some organic materials causes problems, as they may have a tendency to undergo hydrolysis, metabolization or photochemical reactions. Finally, the residual biological activity of samples can cause severe stability problems [9].
At present, the only commercially available water matrix reference material (MRM) is IRMM-428, released by the Institute for Reference Materials and Measurements (IRMM). IRMM-428 is prepared by spiking PFASs in methanol into a known volume of drinking water. Other attempts have been made to prepare proficiency testing (PT) samples for organics in water employing the widely used “reconstitution” approach. In this approach, a solution of analytes of interest in an organic solvent (miscible with water) is spiked into a water sample in the laboratory immediately prior to analysis [8,10,11] or at the producer’s premises just before shipping [12,13]. The rest of the approach is the ‘‘immobilization’’ of analytes (in this case, pesticides) on a solid-phase extraction (SPE) cartridge [8]. Water MRMs that consisted of pesticides stored on SPE cartridges were used in a collaborative study including 15 laboratories, and the observed reproducibility was 26.7% [10]. However, this approach is not sound from a metrological point of view because it was not possible to confirm that the true value equaled the initial concentration in the percolated water sample.
To the best of our knowledge, no water MRMs for VOCs are currently available for quality control. Therefore, in this research, we prepared and certified a water MCRM containing 28 VOCs. The water MCRM was prepared using a “reconstitution” approach by spiking the solution of VOCs in methanol into pure water in the laboratory immediately prior to analysis. The preparation and characterization of 28 VOCs spiking solutions were comprehensively studied, and the water MCRM containing all 28 VOCs was certified using a network of nine competent laboratories.

2. Results and Discussion

2.1. Methodology Study

The 28 VOCs in the spiking solution were determined by GC-MS, and the total ion chromatography is shown in Figure 1. With the exception of p-xylene and m-xylene, 26 other VOCs, IS1 and IS2 were separated well on the chromatographic column.
The precision of the method was investigated by analyzing each reference material of 10 µg/mL 28 VOCs in methanol 10 times, and the results were presented in Table 1. The within laboratory RSDs of the method were between 0.14% and 2.55%, indicating good instrumental repeatability that could meet the requirements of homogeneity and stability studies. The detection limits of 28 VOCs in methanol were calculated based on three standard deviations, and the results were between 0.014 and 0.225 μg/mL. These levels were obviously lower than the concentrations of the 28 VOCs in the spiking solution, which ranged from 6.0 to 18 μg/mL. With the exception of hexachloro-1,3-butadiene, which had an r value of 0.9989, the linear correlation r values of the other 27 VOCs were higher than 0.9990, indicating good linear relationships.

2.2. Purity Test

The purity was used to correct the gravimetric preparation of the standard solutions, and therefore ensure the metrological traceability of the water MCRM. Detailed information regarding the manufacturer, labeled purities and the uncertainties of the 28 commercial standards are given in Table 1. The labeled purities of the commercial standards were verified by GC-FID using the peak area normalization method, and the results are presented in Table 1. The measured purities of all commercial standards were within the range of their labeled uncertainties. Therefore, the labeled purities and uncertainties were used during gravimetric preparation and calculation of the uncertainty in the characterization study of water MCRM, respectively, in consideration of the limitations of the peak area normalization method.
The purity test results of the batch of methanol showed that the methanol purity was acceptable. Additionally, no VOCs were detected, indicating that methanol was suitable for use as the solvent for the 28 VOCs spiking solution.

2.3. Homogeneity Assessment

Homogeneity is an important property of a reference material. Nevertheless, it is a relative concept closely related to the distribution of components in the material, sample size and the number of samples that have been selected to measure homogeneity [14].
In the present study, homogeneity was assessed by selecting 15 ampoules of the VOCs spiking solution using a stratified random samplings scheme covering the entire batch, after which three sub-samples of each ampoule were analyzed. The homogeneity study was evaluated using ANOVA [15], and the results are presented in Figure 2. All calculated F values were below or equal to the critical value F0.05(14,30) = 2.04, indicating no significant difference within bottles. The homogeneity analysis confirmed that the batch had a good agreement among its units (ampoules) for each of the 28 analytes in methanol, and the VOCs spiking solution was regarded as homogeneous.
It should be noted that the calculated F value of vinyl chloride was equal to the critical value F0.05(14, 30) = 2.04. The relatively high F value of vinyl chloride might relate to its high volatility and the relatively poor method repeatability using GC-MS, leading to the calculated MSamong being relatively higher than the MSwithin.

2.4. Stability Assessment

Stability testing is crucial to the certification of reference materials. The long-term stability study was based on linear regression [16], and the results measured after 0, 1, 3, 6, 9, 12 and 15 months of storage in different temperatures are shown in Figure 3 and Table 2, Table 3 and Table 4. At a confidence level of 95%, the observed slope for all results was |b1| < t0.95,n-2 × s(b1), indicating that the 28 analytes in the VOCs spiking solution stored at room temperature, 4 and −18 °C were stable after 15 months of storage. Therefore, the spiking solution, including the 28 VOCs, could be stored at room temperature for convenience, and the shelf-life of the VOCs spiking solution was at least 15 months.
Interestingly, the dispersions of the stability of the 28 VOCs under the same storage conditions were quite different. The stability of vinyl chloride and m-xylene after storage at room temperature for different lengths are shown as examples in Figure 3. The stability of vinyl chloride fluctuated with storage time around the preparation value in a relatively large range, although vinyl chloride in methanol was confirmed to be stable. However, the stability of m-xylene fluctuated closely around the preparation value with storage time. The obvious differences in stability results between vinyl chloride and m-xylene might be related to their characterizations. In the former methodology study, the within-laboratory RSD of vinyl chloride was 2.55%, while that of m-xylene was 0.43%. The lower boiling point of vinyl chloride than m-xylene might lead to significant differences in method precision [17], which could then lead to the different dispersions of stable results. Similar results were also observed for the other 26 VOCs, with the stable values of VOCs having higher boiling points generally being closer to their certified values than those having lower boiling points.

2.5. Characterization and Uncertainty Study

The certification of the water MCRM was conducted by nine competent laboratories. Laboratory 4 used the headspace GC-MS method for certification, while the remaining eight laboratories used the purge-and-trap GC-MS method. The reported values of the 28 VOC analytes were expressed after diluting 1000 times in water. Statistical analysis was conducted for the received data using Grubb’s test, the Cochran test and Dixon’s test, and the mean values of the retained data were calculated as certified values. Figure 4 shows the reported mean values and standard deviations of the 28 VOCs in water from the nine laboratories. Most of the values were close to the certified values and within their expended uncertainties. However, some values from one or two laboratories showed obvious deviations from the certified values, such as the results reported from laboratory 3. Because both headspace GC-MS and purge-and-trap GC-MS are commonly used for the determination of VOCs in water, there were no obvious discrepancies in the results from the nine laboratories using the two different detection methods.
The results of the certified values and the expanded uncertainties of the water MCRM expressed at 95% confidence (with the coverage factor k = 2) are presented in Table 5. The certified values of the 28 VOCs in the water MCRM were in the range of 6.162 to 17.37 μg/L. The uncertainty of the water MCRM was calculated by combining the uncertainty of inhomogeneity (ubb), instability (ulst) and characterization (uchar) [18], and the ubb and ulst were those of VOCs spiking solution. Some of the ubb were calculated through u b b = s b b = M S among M S within n when MSamong > MSwithin for some VOCs, while others were calculated through u b b = M S within n 2 v M S within 4 when MSamong < MSwithin. As shown in Table 5, the ubb of the 28 VOC analytes ranged from 0.18% to 3.73%, which was lower than the reported CRM of BTEX in methanol [19]. The uchar and ulst of the 28 VOC analytes ranged from 0.87% to 9.44% and 0.91% to 11.4%, respectively. The expanded uncertainty of the 28 VOC analytes in the water MCRM ranged from 0.5 to 5.3 μg/L with a coverage factor k = 2 under an approximately 95% confidence level. The expanded uncertainty of o-xylene as 0.5 μg/L was lowest, and the expanded uncertainty of vinyl chloride as 5.3 μg/L was highest. The relatively high expanded uncertainty of vinyl chloride might result from two aspects. On the one hand, the highly volatile nature and early elution time on chromatography could result in a relatively high operation and instrument effect on quantification. On the other hand, the tested value of commercial vinyl chloride standard solution was not consistent with the labeled value, and large discrepancies existed among commercial standard solutions from different producers, making precise quantification difficult.

3. Materials and Methods

The Institute for Environmental Reference Materials, Ministry of Environmental Protection (IERM) has a quality management system based on ISO Guide 35 and ISO/IEC 17025, which is accredited by the China National Accreditation Service for Conformity Assessment (CNAS). The preparation and certification of the CRMs for environmental monitoring have been carried out according to the technical requirements of ISO Guide 35 [20].

3.1. Chemicals and Instruments

The commercial standards of 28 VOCs were purchased from several manufacturers. Detailed information regarding the manufacturers, purities and uncertainties of the purities are given in Table 6. Pesticide residue grade methanol was purchased from J.T. Baker, USA. The stock standard solutions were 27 mixed VOCs standard solution (IRMM, 100 µg/mL) and vinyl chloride standard solution (2000 µg/mL, Supelo, USA). The internal stock standard solutions were fluorobenzene and 1,4-dichlorobenzene D4 (IRMM, 1000 and 1000 µg/mL, respectively).
The VOCs spiking solution was prepared gravimetrically using a calibrated Mettler Toledo analytical balance (AE-240, 205 g capacity, resolution of 0.01 mg, Switzerland). The purities of the 28 VOC commercial standards were verified by a calibrated gas chromatography with flame ionization detection (Agilent 7890A GC-FID, USA). Homogeneity and stability studies of VOCs spiking solution were performed on a calibrated Agilent 7890A gas chromatograph coupled with an Agilent 5975C mass spectrometer (Agilent 7890A GC-5975C MS, USA). Analysis of the water MCRM was performed on a calibrated purge-and-trap Agilent 7890A GC-5975C MS.

3.2. Purity Test

The purities of the 28 VOC commercial standards were determined in-house using a GC-FID with a DB-1 column (30 m × 320 µm ID × 0.25 µm film). The oven temperature program started at 70 °C, then increased to 150 °C at 5 °C/min. The injection volume was 1 µL, and the injection was performed in split mode (30:1). The carrier gas was high purity nitrogen (1.0 mL/min), and the temperature of the injector and detector were 220 and 230 °C, respectively. The final purities of the 28 commercial standards were calculated by the peak area normalization method.
The purity of methanol (pesticides residue grade) selected as the solvent was checked by gas chromatography-mass spectrometry (GC-MS) prior to the preparation of CRM.

3.3. Determination of VOCs

3.3.1. Determination of VOCs in the Spiking Solution

Measurement of the 28 VOCs in the spiking solution was performed on a GC-MS [21] equipped with a DB-624 (60 m × 250 µm ID × 1.4 µm film, Agilent, Santa Clara, CA, USA) capillary column. The oven temperature was programmed as follows: 35 °C for 2 min, followed by 5 °C/min to 120 °C, then 10 °C/min to 220 °C, where it was held for 3 min. The injection volume was 1 µL, and the injection was performed in split mode (30:1). The carrier gas was helium (1.0 mL/min), and the temperature of the injector, transfer line and ion source was 220, 260 and 230 °C, respectively. Data acquisition was performed under selected ion monitoring (SIM) mode.

3.3.2. Determination of VOCs in Water MCRM

Analysis of the 28 VOCs in water MCRM was performed on a purge-and-trap GC-MS [21]. The conditions of the purge-and-trap were as follows: purge time: 11 min; purge rate: 40 mL/min; dry purge time: 1 min; desorption time: 2 min; desorption temperature: 190 °C; baking time: 6 min; baking temperature: 200 °C. The conditions of GC-MS were the same as for the determination of 28 VOCs in the spiking solution.

3.4. Preparation of the VOCs Spiking Solution

The 28 VOCs spiking solution was prepared gravimetrically. Briefly, the 28 VOCs were divided into seven groups during weighing and dissolution, then mixed into a certain volume. Among the 28 VOC commercial standards, only vinyl chloride standard is gaseous at room temperature. Therefore, vinyl chloride was placed in its own group, while the remaining 27 VOCs were divided into six groups according to their characteristics (Table 7). According to the labeled purities of the 28 VOC commercial standards, the stock solution of vinyl chloride in methanol was prepared by drawing a moderate volume of vinyl chloride using an airtight syringe and then adding it to methanol. For the other six groups, the stock solution of each group was prepared gravimetrically in the sequence of their polarity from weak to strong. After all stock solutions were prepared, a moderate volume of stock solution from each group was transferred into the same 1 L flask and then diluted to 1 L with methanol. The mass fractions of target compositions of the spiked solution were between 5.0 and 20 mg/L. Approximately 1 L of the VOC mixture was subdivided into 2 mL amber glass ampoules with 1.2 mL per ampoule using an ampoule filling machine. During the process of subdivision, 15 ampoules were sampled for a homogeneity study using a stratified random sampling strategy. After confirmation that the batch was homogeneous, the sealed ampoules were packed and divided into three parts, then stored at room temperature, 4 and −18 °C.

3.5. Homogeneity Testing

The VOCs spiking solution was subjected to a homogeneity study in which both the homogeneity between and within ampoules was evaluated. Fifteen ampoules of VOC spiking solution were selected using a stratified random sampling scheme covering the whole batch, and three sub-samples were analyzed in each ampoule. The internal standard was spiked into the solution for the QA/QC process, and each sample was analyzed in triplicate by GC-MS. Measurement sequences were randomized to be able to minimize possible trends in both the filling sequence and the analytical sequence.
The homogeneity was evaluated by one-way analysis of variance (ANOVA) as described by ISO Guide 35 [20]. The between-bottle standard deviation (sbb) and within-bottle standard deviation (swb) were also calculated. Possible inhomogeneity was expressed as the uncertainty due to the between-bottle inhomogeneity of the material (ubb) and quantified as:
u b b = s b b = M S among M S within n
In cases in which MSamong < MSwithin (indicating that the study set-up and/or method repeatability were not sufficient), the maximum heterogeneity that could be hidden by the method variability, the influence of analytical variation on the standard deviation between units ubb′ was calculated and used for estimates of in-homogeneity. The ubb′ was calculated as:
u b b = M S within n 2 v M S within 4
where MSwithin is the mean square within groups determined from ANOVA, n is the number of replicates per bottle and v M S within represents the degrees of freedom of MSwithin.
The instrumental repeatability of the measurement of 28 VOCs was determined by conducting 10 replicate GC-MS analyses of 10 µg/mL reference material containing 28 VOCs.

3.6. Stability Testing

A long-term stability study was conducted to ensure the shelf-life of the VOCs spiking solution. A long-term stability study was conducted based on a classical stability study. Stability monitoring was performed for each analyte of the spiking solution after 0, 1, 3, 6, 9, 12 and 15 months of storage at room temperature, 4 and −18 °C. For each round of analysis, three ampoules were sampled randomly from those samples stored under each storage condition.
A linear regression model was utilized for processing data, namely assuming component values (Y) of time (X) varying linear equation as Y = b0 + b1X, where b0 and b1 are the regression coefficients. The estimated standard deviation of b1 is then given by:
S ( b 1 ) = s i = 1 n ( X i X ¯ ) 2
where
s 2 = i = 1 n ( Y i b 0 b 1 X i ) 2 n 2
and tcal is given by
t c a l = b 1 S b 1
According to ISO Guide 35, the long-term instability is estimated as ults = S(b1) × t, with ults being the uncertainty of long-term instability, S(b1) is the standard error of the slope and t is the selected duration.

3.7. Characterization and Uncertainty Study

The certified values of 28 VOCs in the water MCRM were determined by the average of the results obtained from nine competent laboratories that each received six ampoules selected at random. Before analysis, the samples were diluted 1000 times with pure water, and the average concentrations of analytes after dilution were used.
Data sets were checked to ensure they followed approximately normal distributions and that variances for each compound were homogeneous. Grubb’s test was applied to evaluate within laboratory parallel data. Between-laboratory outliers of variance were detected using the Cochran test, while outliers of average were detected using Dixon’s test. Generally, the outliers detected from Grubb’s test were retained. When the RSD of within-laboratory parallel data was smaller than 15%, the stragglers and outliers from Cochran’s test were retained. The outliers (95% confidence interval) detected from Dixon’s test were removed.
The standard uncertainty of the certified values included collaborating characterization uncertainty (uchar), between-bottle inhomogeneity uncertainty (ubb) and long-term instability uncertainty (ults). The expanded uncertainty (U) of the certified property value was calculated as: U = k × u b b 2 + u l t s 2 + u c h a r 2 , where k is the coverage factor (usually set k = 2, approximately the 95% confidence level), ubb is the uncertainty due to inhomogeneity of the material, uits is the uncertainty due to instability of the material and uchar is uncertainty in the characterization of the property value.

4. Conclusions

In conclusion, a water MCRM of 28 VOCs was developed using a “reconstitution” approach by adding the prepared VOCs spiking methanol solution into pure water directly prior to analysis. The 28 VOCs in methanol as a spiking solution was prepared gravimetrically by dividing the VOCs into seven groups. The batch of spiking solution was homogeneous and stable at room temperature, 4 and −18 °C for at least 15 months. The certification of the water MCRM was established in a study involving nine competent laboratories applying purge-and-trap GC-MS or headspace GC-MS. The certified values of the 28 VOC analytes in the MCRM ranged from 6.162 to 17.37 μg/L with expanded uncertainties in the range of 0.5 to 5.3 μg/L. The prepared water MCRM could be used for quality control during VOC analysis in water and for developing or verifying measurement methods for VOCs monitoring in water.

Author Contributions

Conceptualization, L.F. and W.T.; methodology, L.F. and G.Y.; software, B.L.; validation, Y.J., H.L., and Y.Z.; investigation, H.L. and L.F.; data curation, L.H.; writing—original draft preparation, L.F. and L.H.; writing—review and editing, L.F., L.H. and W.T.; visualization, L.H.; supervision, W.T.; project administration, W.T. All authors have read and agreed to the published version of the manuscript.

Funding

This study was funded by the National Key Scientific Instrument and Equipment Development Project, China (2012YQ-060027).

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Chary, N.S.; Fernandez-Alba, A.R. Determination of volatile organic compounds in drinking and environmental waters. Trac-Trend. Anal. Chem. 2012, 32, 60–75. [Google Scholar] [CrossRef]
  2. Marczak, M.; Wolska, L.; Chrzanowski, W.; Namieśnik, J. Microanalysis of volatile organic compounds (VOCs) in water samples–methods and instruments. Microchimica Acta 2006, 155, 331–348. [Google Scholar] [CrossRef]
  3. Feron, V.J.; Til, H.P.; de Vrijer, F.; van Bladeren, P.J. Toxicology of volatile organic compounds in indoor air and strategy for further research. Indoor Environ. 1992, 1, 69–81. [Google Scholar]
  4. Chang, C.T.; Chen, B.Y.J. Toxicity assessment of volatile organic compounds and polycyclic aromatic hydrocarbons in motorcycle exhaust. Hazard. Mater. 2008, 153, 1262–1269. [Google Scholar] [CrossRef]
  5. Altalyan, H.N.; Jones, B.; Bradd, J.; Long, D.N.; Alyazichi, Y.M.J. Removal of volatile organic compounds (VOCs) from groundwater by reverse osmosis and nanofiltration. Water Process Eng. 2016, 9, 9–21. [Google Scholar] [CrossRef] [Green Version]
  6. Emons, H.; Linsinger, T.P.J.; Gawlik, B.M. Trac-Trend. Reference materials: Terminology and use. Can’t one see the forest for the trees? Anal. Chem. 2004, 23, 442–449. [Google Scholar]
  7. Quevauviller, P. Challenges for achieving traceability of environmental measurements. Trac-Trend. Anal. Chem. 2004, 23, 171–177. [Google Scholar] [CrossRef]
  8. Deplagne, J.; Vial, J.; Pichon, V.; Lalere, B.; Hervouet, G.; Hennion, M.C. Feasibility study of a reference material for water chemistry: Long term stability of triazine and phenylurea residues stored in vials or on polymeric sorbents. J. Chromatogr. A. 2006, 1123, 31–37. [Google Scholar] [CrossRef]
  9. Bercaru, O.; Manfred, G.; Awlik, B.; Ulberth, F.; Vandecasteele, C.J. Reference materials for the monitoring of the aquatic environment-a review with special emphasis on organic priority pollutants. Environ. Monitor. 2003, 5, 697–705. [Google Scholar] [CrossRef]
  10. Mrabet, K.E.; Poitevin, M.; Vial, J.; Pichon, V.; Amarouche, S.; Hervouet, G.; Lalere, B.J. An interlaboratory study to evaluate potential matrix reference materials for herbicides in water. Chromatogr. A. 2006, 1134, 151–161. [Google Scholar] [CrossRef]
  11. Kreeke, J.V.D.; Calle, B.D.L.; Held, A.; Bercaru, O.; Ricci, M.; Shegunova, P.; Taylor, P. IMEP-23: The eight EU-WFD priority PAHs in water in the presence of humic acid. Trend. Anal. Chem. 2010, 29, 928–937. [Google Scholar] [CrossRef]
  12. Elordui-Zapatarietxe, S.; Fettig, I.; Philipp, R.; Gantois, F.; Lalère, B.; Swart, C.; Petrov, P.; Goenaga-Infante, H.; Vanermen, G.; Boom, G.; et al. Novel concepts for preparation of reference materials as whole water samples for priority substances at nanogram-per-liter level using model suspended particulate matter and humic acids. Anal. Bioanal. Chem. 2015, 407, 3055–3067. [Google Scholar] [CrossRef] [Green Version]
  13. Elordui-Zapatarietxe, S.; Fettig, I.; Richter, J.; Philipp, R.; Gantois, F.; Lalère, B.; Swart, C.; Emteborg, H. Interaction of 15 priority substances for water monitoring at ng L−1 levels with glass, aluminium and fluorinated polyethylene bottles for the containment of water reference materials. Accredit. Qual. Assur. 2015, 20, 447–455. [Google Scholar] [CrossRef] [Green Version]
  14. Guimarães, E.D.F.; Rego, E.C.P.D.; Cunha, H.C.M.; Rodrigues, J.M.; Figueroa-Villar, J.D. Certified reference material for traceability in environmental analysis: PAHs in toluene. J. Braz. Chem. Soc. 2013, 25, 351–360. [Google Scholar] [CrossRef]
  15. van der Veen, A.M.H.; Linsinger, T.P. Pauwels, Uncertainty calculations in the certification of reference materials. 2. Homogeneity study. J. Accredit. Qual. Assur. 2001, 6, 26–30. [Google Scholar] [CrossRef]
  16. van der Veen, A.M.H.; Linsinger, T.P.; Lamberty, A. Pauwels, Uncertainty calculations in the certification of reference materials 3. Stability study. J. Accredit. Qual. Assur. 2001, 6, 257–263. [Google Scholar] [CrossRef]
  17. Sun, S.P.; Duan, J.P.; Zhou, X.M.; Xing, D.R. Determination of VOCs in air with capillary gas chromatography. J. Environ. Health 2007, 24, 446–448. [Google Scholar]
  18. van der Veen, A.M.H.; Linsinger, T.P.J.; Schimmel, H.; Lamberty, A. Pauwels, Uncertainty calculations in the certification of reference materials 4. Characterisation and certification. J. Accredit. Qual. Assur. 2001, 6, 290–294. [Google Scholar] [CrossRef]
  19. Neves, L.A.; Almeida, R.R.; Rego, E.P.; Rodrigues, J.M.; de Carvalho, L.J.; Goulart, A.L.D.M. Certified reference material of volatile organic compounds for environmental analysis: BTEX in methanol. Anal. Bioanal. Chem. 2015, 407, 3225–3229. [Google Scholar] [CrossRef]
  20. Ellison, S.L.R. Homogeneity studies and ISO Guide 35:2006. Accredit. Qual. Assur. 2015, 20, 519–528. [Google Scholar] [CrossRef]
  21. Fang, L.P.; Yang, G.; Jiang, Y.; Wang, W.; Liu, H.P.; Lu, B.W.; Tian, W. Development of the certified reference material of volatile organic compounds. Environ. Chem. 2016, 35, 688–696. [Google Scholar]
Figure 1. Total ion chromatography of 28 VOCs determined by GC-MS (1. vinyl chloride; 2. 1,1-dichloroethylene; 3. dichloromethane; 4. trans-1,2-dichloroethene; 5. cis-1,2-dichloroethene; 6. trichloromethane; 7. carbon tetrachloride; 8. benzene; 9. 1,2-dichloroethane; 10. trichloroethylene; 11. bromodicloromethane; 12. toluene; 13. tetrachloroethylene; 14. chlorodibromomethane; 15. chlorobenzene; 16. ethylbenzene; 17,18. p-xylene/m-xylene; 19. o-xylene; 20. styrene; 21. bromoform; 22. cumene; 23. 1,4-dichlorobenzene; 24. 1,2-dichlorobenzene; 25. 1,3,5-trichlorobenzene; 26. 1,2,4-trichlorobenzene; 27. hexachloro-1,3-butadiene; 28. 1,2,3-trichlorobenzene; IS1. fluorobenzene; IS2. 1,4-dichlorobenzene D4).
Figure 1. Total ion chromatography of 28 VOCs determined by GC-MS (1. vinyl chloride; 2. 1,1-dichloroethylene; 3. dichloromethane; 4. trans-1,2-dichloroethene; 5. cis-1,2-dichloroethene; 6. trichloromethane; 7. carbon tetrachloride; 8. benzene; 9. 1,2-dichloroethane; 10. trichloroethylene; 11. bromodicloromethane; 12. toluene; 13. tetrachloroethylene; 14. chlorodibromomethane; 15. chlorobenzene; 16. ethylbenzene; 17,18. p-xylene/m-xylene; 19. o-xylene; 20. styrene; 21. bromoform; 22. cumene; 23. 1,4-dichlorobenzene; 24. 1,2-dichlorobenzene; 25. 1,3,5-trichlorobenzene; 26. 1,2,4-trichlorobenzene; 27. hexachloro-1,3-butadiene; 28. 1,2,3-trichlorobenzene; IS1. fluorobenzene; IS2. 1,4-dichlorobenzene D4).
Molecules 26 04370 g001
Figure 2. Results of homogeneity study of 28 VOCs in spiking solution.
Figure 2. Results of homogeneity study of 28 VOCs in spiking solution.
Molecules 26 04370 g002
Figure 3. Results of vinyl chloride and m-xylene in spiking solution stored at room temperature at intervals of 15 months.
Figure 3. Results of vinyl chloride and m-xylene in spiking solution stored at room temperature at intervals of 15 months.
Molecules 26 04370 g003
Figure 4. Measurement of 28 VOCs in water from nine laboratories. Thick red lines represent certified values, thin blue lines represent expended uncertainties.
Figure 4. Measurement of 28 VOCs in water from nine laboratories. Thick red lines represent certified values, thin blue lines represent expended uncertainties.
Molecules 26 04370 g004aMolecules 26 04370 g004b
Table 1. Precision and detection limits of 28 VOCs in methanol using GC-MS.
Table 1. Precision and detection limits of 28 VOCs in methanol using GC-MS.
No.CompositionRSD(%)Detection Limit/ (μg/mL)Linear Correlation r
1vinyl chloride2.550.11.000
21,1-dichloroethylene1.530.10.9999
3dichloromethane0.780.071.000
4trans-1,2-dichloroethene1.030.070.9999
5cis-1,2-dichloroethene0.340.031.000
6trichloromethane0.260.031.000
7carbon tetrachloride0.740.040.9998
8benzene0.440.021.000
91,2-dichloroethane0.500.041.000
10trichloroethylene0.720.21.000
11bromodicloromethane0.630.020.9998
12toluene0.530.030.9998
13tetrachloroethylene0.330.11.000
14chlorodibromomethane0.940.050.9997
15chlorobenzene0.450.020.9990
16ethylbenzene0.340.030.9999
17p-xylene0.430.040.9997
18m-xylene0.430.040.9997
19o-xylene0.430.030.9999
20styrene0.500.031.000
21bromoform0.390.030.9998
22cumene0.270.041.000
231,4-dichlorobenzene0.180.010.9991
241,2-dichlorobenzene0.140.020.9992
251,3,5-trichlorobenzene0.270.020.9992
261,2,4-trichlorobenzene0.280.010.9996
27hexachloro-1,3-butadiene 0.500.060.9989
281,2,3-trichlorobenzene0.310.020.9995
Table 2. Stability study of VOCs spiking solution stored at room temperature.
Table 2. Stability study of VOCs spiking solution stored at room temperature.
No.Compositionb0
(μg/mL)
b1
(μg/mL)
s(b1)
(μg/mL)
t0.95,n−2t0.95,n−2 × s(b1)
1vinyl chloride18.04−0.1650.2202.570.565
21,1-dichloroethylene12.74−0.0480.0802.570.207
3dichloromethane12.560.0110.0562.570.145
4trans-1,2-dichloroethene10.560.0210.0692.570.178
5cis-1,2-dichloroethene10.270.0000.0412.570.106
6trichloromethane13.720.1120.1602.570.410
7carbon tetrachloride12.000.0690.0742.570.190
8benzene8.654−0.0190.0472.570.120
91,2-dichloroethane10.830.0590.0672.570.172
10trichloroethylene14.000.0490.1092.570.281
11chlorodibromomethane14.490.1170.1002.570.257
12toluene8.580−0.0030.0232.570.059
13tetrachloroethylene12.210.0460.0612.570.158
14bromodicloromethane12.090.0510.1092.570.280
15chlorobenzene9.873−0.0410.0372.570.095
16ethylbenzene7.701−0.0280.0202.570.050
17p-xylene6.092−0.0070.0092.570.023
18m-xylene6.092−0.0070.0092.570.023
19o-xylene7.191−0.0200.0132.570.034
20styrene8.533−0.0210.0172.570.043
21bromoform15.430.0150.0552.570.141
22cumene7.752−0.0100.0052.570.012
231,4-dichlorobenzene10.33−0.0530.0292.570.074
241,2-dichlorobenzene9.287−0.0320.0262.570.066
251,3,5-trichlorobenzene9.433−0.0190.0292.570.073
261,2,4-trichlorobenzene9.275−0.0400.0402.570.102
27hexachloro-1,3-butadiene10.80−0.0400.0572.570.147
281,2,3-trichlorobenzene8.463−0.0060.0252.570.063
Table 3. Stability study for VOCs spiking solution stored at 4 °C.
Table 3. Stability study for VOCs spiking solution stored at 4 °C.
No.Compositionb0
(μg/mL)
b1
(μg/mL)
s(b1)
(μg/mL)
t0.95,n−2t0.95,n−2 × s(b1)
1vinyl chloride17.06 0.108 0.158 2.570.405
21,1-dichloroethylene12.11 0.059 0.050 2.570.128
3dichloromethane12.56 0.041 0.028 2.570.071
4trans-1,2-dichloroethene10.83 0.025 0.037 2.570.095
5cis-1,2-dichloroethene10.13 0.000 0.048 2.570.124
6trichloromethane14.77 0.048 0.044 2.570.113
7carbon tetrachloride12.11 0.079 0.041 2.570.105
8benzene8.330 0.010 0.017 2.570.045
91,2-dichloroethane11.10 0.057 0.024 2.570.063
10trichloroethylene14.64 0.021 0.068 2.570.174
11chlorodibromomethane15.03 0.075 0.063 2.570.163
12toluene8.543 0.002 0.025 2.570.064
13tetrachloroethylene12.36 0.037 0.033 2.570.084
14bromodicloromethane11.83 0.073 0.092 2.570.238
15chlorobenzene10.03 −0.049 0.035 2.570.090
16ethylbenzene7.805 −0.032 0.020 2.570.050
17p-xylene6.156 −0.007 0.010 2.570.027
18m-xylene6.156 −0.007 0.010 2.570.027
19o-xylene7.259 −0.023 0.011 2.570.029
20styrene8.410 −0.009 0.011 2.570.028
21bromoform15.75 −0.023 0.041 2.570.107
22cumene7.753 −0.021 0.008 2.570.022
231,4-dichlorobenzene10.21 −0.041 0.027 2.570.068
241,2-dichlorobenzene9.324 −0.034 0.019 2.570.049
251,3,5-trichlorobenzene9.432 −0.019 0.025 2.570.065
261,2,4-trichlorobenzene9.275 −0.040 0.040 2.570.102
27hexachloro-1,3-butadiene10.28 −0.008 0.041 2.570.105
281,2,3-trichlorobenzene8.578 −0.016 0.019 2.570.048
Table 4. Stability study for VOCs spiking solution stored at −18 °C.
Table 4. Stability study for VOCs spiking solution stored at −18 °C.
No.Compositionb0
(μg/mL)
b1
(μg/mL)
s(b1)
(μg/mL)
t0.95,n−2t0.95,n−2 × s(b1)
1vinyl chloride18.50 −0.018 0.139 2.570.358
21,1-dichloroethylene12.48 0.021 0.052 2.570.134
3dichloromethane12.79 0.021 0.035 2.570.090
4trans-1,2-dichloroethene10.99 0.022 0.035 2.570.091
5cis-1,2-dichloroethene10.18 0.023 0.031 2.570.079
6trichloromethane14.82 0.051 0.043 2.570.110
7carbon tetrachloride12.21 0.073 0.049 2.570.126
8benzene8.434 0.011 0.021 2.570.054
91,2-dichloroethane11.12 0.058 0.029 2.570.074
10trichloroethylene14.79 0.023 0.061 2.570.156
11chlorodibromomethane15.03 0.090 0.064 2.570.165
12toluene8.516 0.012 0.028 2.570.071
13tetrachloroethylene12.33 0.048 0.031 2.570.080
14bromodicloromethane11.61 0.091 0.060 2.570.154
15chlorobenzene10.093 −0.059 0.035 2.570.089
16ethylbenzene7.836 −0.034 0.020 2.570.051
17p-xylene6.168 0.001 0.011 2.570.028
18m-xylene6.168 0.001 0.011 2.570.028
19o-xylene7.243 −0.020 0.008 2.570.020
20styrene8.378 0.000 0.009 2.570.022
21bromoform15.64 −0.027 0.057 2.570.147
22cumene7.710 −0.012 0.007 2.570.017
231,4-dichlorobenzene7.739 −0.021 0.006 2.570.015
241,2-dichlorobenzene9.331 −0.032 0.018 2.570.047
251,3,5-trichlorobenzene9.508 −0.032 0.022 2.570.057
261,2,4-trichlorobenzene8.942 −0.018 0.023 2.570.060
27hexachloro-1,3-butadiene10.25 −0.013 0.042 2.570.107
281,2,3-trichlorobenzene8.628 −0.028 0.016 2.570.042
Table 5. Certified values and expended uncertainties of 28 VOCs in the water MCRM.
Table 5. Certified values and expended uncertainties of 28 VOCs in the water MCRM.
No.Compositionuchar
(%)
ubb
(%)
ulst
(%)
Certified Value
(μg/L)
uCRM (k = 2)
(μg/L)
1vinyl chloride9.44 3.7311.41.7 × 1015.3
21,1-dichloroethylene5.75 1.226.201.4 × 1012.5
3dichloromethane2.87 1.894.071.4 × 1011.4
4trans-1,2-dichloroethene2.04 0.844.751.1 × 1011.2
5cis-1,2-dichloroethene2.24 0.814.481.1 × 1011.1
6trichloromethane2.42 1.524.251.6 × 1011.6
7carbon tetrachloride2.61 1.715.801.3 × 1011.7
8benzene2.99 0.863.678.1 × 1000.8
91,2-dichloroethane4.49 1.023.771.2 × 1011.4
10trichloroethylene2.68 1.206.091.4 × 1011.9
11chlorodibromomethane0.87 1.406.171.1 × 1011.5
12toluene3.74 0.464.817.9 × 1001.0
13tetrachloroethylene2.18 0.823.691.2 × 1011.0
14bromodicloromethane2.23 1.237.381.5 × 1012.3
15chlorobenzene3.65 0.505.389.7 × 1001.3
16ethylbenzene4.65 0.303.947.2 × 1000.9
17p-xylene4.25 0.312.626.2 × 1000.6
18m-xylene4.25 0.312.626.2 × 1000.6
19o-xylene3.46 0.311.647.1 × 1000.5
20styrene5.44 0.341.569.2 × 1001.0
21bromoform1.87 1.415.541.5 × 1011.8
22cumene9.87 0.421.329.0 × 1001.8
231,4-dichlorobenzene3.15 0.180.919.5 × 1000.7
241,2-dichlorobenzene1.74 0.223.009.1 × 1000.7
251,3,5-trichlorobenzene1.56 0.253.609.3 × 1000.8
261,2,4-trichlorobenzene3.76 0.393.988.8 × 1001.0
27hexachloro-1,3-butadiene1.83 0.776.141.0 × 1011.4
281,2,3-trichlorobenzene4.72 0.312.918.2 × 1000.9
Table 6. Purities of 28 commercial VOC standards.
Table 6. Purities of 28 commercial VOC standards.
No.CompositionManufacturerLabeled Purity (%)Labeled Uncertainty (%)Measured Purity (%)
1vinyl chloridegmgas, China99.9990.599.99
21,1-dichloroethyleneChemService,USA99.50.599.83
3dichloromethaneChemService,USA99.50.599.74
4trans-1,2-dichloroetheneChemService,USA99.30.599.37
5cis-1,2-dichloroetheneChemService,USA99.50.599.78
6trichloromethaneChemService,USA99.50.599.34
7carbon tetrachlorideChemService,USA99.50.599.85
8benzeneChemService,USA99.50.599.95
91,2-dichloroethaneChemService,USA99.50.599.93
10trichloroethyleneChemService,USA99.50.599.83
11chlorodibromomethaneFluka,USA98.80.598.88
12tolueneChemService,USA99.50.599.99
13tetrachloroethyleneChemService,USA99.50.599.94
14bromodicloromethaneFluka,USA99.50.599.86
15chlorobenzeneChemService,USA99.50.599.91
16ethylbenzeneChemService,USA99.50.599.48
17p-xyleneChemService,USA,USA99.50.599.77
18m-xyleneChemService,USA99.40.599.83
19o-xyleneChemService,USA99.00.599.27
20styreneChemService,USA99.40.599.86
21bromoformChemService,USA99.50.599.53
22cumeneChemService,USA99.50.599.93
23hexachloro-1,3-butadieneAccuStandard,USA98.31.097.82
241,4-dichlorobenzeneChemService,USA99.50.599.98
251,2-dichlorobenzeneChemService,USA99.50.599.73
261,2,4-trichlorobenzeneChemService,USA99.50.599.58
271,2,3-trichlorobenzeneChemService,USA99.50.599.93
281,3,5-trichlorobenzeneChemService,USA99.50.599.11
Table 7. Grouping of 28 VOCs during preparation of VOCs spiking solution.
Table 7. Grouping of 28 VOCs during preparation of VOCs spiking solution.
GroupComposition
1vinyl chloride
2dichloromethane, trichloroethylene, bromoform, chlorodibromomethane
3carbon tetrachloride, bromodicloromethane, tetrachloroethylene
4trichloromethane, 1,2-dichloroethane, 1,4-dichlorobenzene, hexachloro-1,3-butadiene
5toluene, ethylbenzene, o-xylene, m-xylene, p-xylene, styrene, cumene
6chlorobenzene, 1,2-dichlorobenzene, 1,2,4-trichlorobenzene, 1,2,3-trichlorobenzene, 1,3,5-trichlorobenzene
71,1-dichloroethylene, benzene, trans-1,2-dichloroethene, cis-1,2-dichloroethene
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Fang, L.; Huang, L.; Yang, G.; Jiang, Y.; Liu, H.; Lu, B.; Zhao, Y.; Tian, W. Development of a Water Matrix Certified Reference Material for Volatile Organic Compounds Analysis in Water. Molecules 2021, 26, 4370. https://0-doi-org.brum.beds.ac.uk/10.3390/molecules26144370

AMA Style

Fang L, Huang L, Yang G, Jiang Y, Liu H, Lu B, Zhao Y, Tian W. Development of a Water Matrix Certified Reference Material for Volatile Organic Compounds Analysis in Water. Molecules. 2021; 26(14):4370. https://0-doi-org.brum.beds.ac.uk/10.3390/molecules26144370

Chicago/Turabian Style

Fang, Liping, Linyan Huang, Gang Yang, Yang Jiang, Haiping Liu, Bingwen Lu, Yaxian Zhao, and Wen Tian. 2021. "Development of a Water Matrix Certified Reference Material for Volatile Organic Compounds Analysis in Water" Molecules 26, no. 14: 4370. https://0-doi-org.brum.beds.ac.uk/10.3390/molecules26144370

Article Metrics

Back to TopTop