Next Article in Journal
Corroles-Porphyrins: A Teamwork for Gas Sensor Arrays
Next Article in Special Issue
The Development of Love Wave-Based Humidity Sensors Incorporating Multiple Layers
Previous Article in Journal
Detection of Steel Fatigue Cracks with Strain Sensing Sheets Based on Large Area Electronics
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Sensing Properties of Pd-Loaded Co3O4 Film for a ppb-Level NO Gas Sensor

1
National Institute of Advanced Industrial Science and Technology (AIST), Inorganic Functional Materials Research Institute, 2266-98, Anagahora, Shimo-Shidami, Moriyama-ku, Nagoya-shi 463-8560, Japan
2
Department of Mechanical Engineering, Aichi Institute of Technology, 1247, Yachigusa, Yakusa-cho, Toyota-shi 470-0392, Japan
*
Author to whom correspondence should be addressed.
Submission received: 16 February 2015 / Revised: 26 March 2015 / Accepted: 31 March 2015 / Published: 7 April 2015
(This article belongs to the Special Issue Gas Sensors—Designs and Applications)

Abstract

:
We prepared 0.1 wt%–30 wt% Pd-loaded Co3O4 by a colloidal mixing method and investigated the sensing properties of a Pd-loaded Co3O4 sensor element, such as the sensor response, 90% response time, 90% recovery time, and signal-to-noise (S/N) ratio, toward low nitric oxide (NO) gas levels in the range from 50 to 200 parts per billion. The structural properties of the Pd-loaded Co3O4 powder were investigated using X-ray diffraction analysis and transmission electron microscopy. Pd in the powder existed as PdO. The sensor elements with 0.1 wt%–10 wt% Pd content have higher sensor properties than those without any Pd content. The response of the sensor element with a 30 wt% Pd content decreased markedly because of the aggregation and poor dispersibility of the PdO particles. High sensor response and S/N ratio toward the NO gas were achieved when a sensor element with 10 wt% Pd content was used.

1. Introduction

Human breath contains many main gases such as N2, CO2, O2 and minor gases such as inflammable gases (CO, CH4, and H2), and volatile organic compounds (VOCs) [1,2,3,4,5]. Some gases in the human breath are considered as biomarkers for diseases. For example, the breath of lung cancer patients includes aldehydes such as nonanal [3], and nitric oxide (NO) in human breath increases in airway inflammatory disorders such as asthma in adults and children [4,5]. Some biomarker gases in human breath are present at very low concentrations in the range from several parts per billion (ppb) to parts per million (ppm). Thus, monitoring human breath is one of the best non-invasive screening tests for early diagnosis, but this process requires analysis equipment with sufficient accuracy at the ppb level.
NO in human breath has been measured according to the American Thoracic Society recommendations [6] using a chemiluminescence analyzer (model 280i NO analyzer; Sievers, Boulder, CO, USA) and is expressed in ppb. Because the analyzer is bulky and needs an ozone supply, effective, inexpensive, and miniaturized systems for the detection and quantification of NO are required [7]. Gas sensors have been considered promising candidates for NO measurement in human breath owing to their low cost, compact size, and direct electronic interface [8,9]. Many types of NO gas sensors consisting of semiconductors have been investigated [8,9,10,11,12,13,14,15]. Among the metal–oxide–semiconductors, the n-type semiconductors, specifically those based on highly sensitive WO3, are promising candidates that can be used to detect NO gas. We have reported the NO and NO2 sensing (0.5–5 ppm in air) properties of a p-type Co3O4 gas sensor [13]. The Co3O4 gas sensor showed low sensor resistance of below 1 kΩ and the NO sensor responses were higher than the NO2 sensor responses. This result suggested that the peripheral circuit used to measure the sensor resistance and the device cost can be reduced. Although the sensor responses of a Co3O4-based gas sensor at 0.5–5 ppm NO have been investigated, the sensor responses to several hundred ppb NO have not been studied. Moreover, the Co3O4-based gas sensor showed a low sensor response, even at 0.5 ppm NO and could not accurately detect NO concentrations in several hundred ppb. The addition of metal or metal–oxides in the metal–oxide–semiconductors is critical in improving the sensor response and response time of metal–oxide–semiconductor-based gas sensors [12,14]. Penza et al. reported that the Pd-doped WO3-based sensor showed higher sensor response to 440 ppm NO gas than the undoped, Pt-doped, and Au-doped WO3-based sensors [12]. In addition, the NO gas sensor must show high NO selectivity against other gases such as H2, CO, and CH4 in human breath. The H2 concentration in human breath has been measured at several tens of ppm level and is higher than that of other gases such as inflammable gases and VOCs [16].
In the present study, Pd-loaded Co3O4 was prepared by a colloidal mixing method [17] to improve the gas-sensing properties, and the sensor response, response time, recovery time, and signal-to-noise (S/N) ratio toward low NO concentration were investigated. To study the possibility of breath analysis using a 10 wt% Pd-loaded Co3O4 sensor, the sensor responses to 25 and 100 ppm H2 in air were investigated as an initial evaluation.

2. Experimental

Co3O4 powder with an average particle size of 20–30 nm was used as a NO gas-sensing material. The Co3O4 powder was mixed and stirred in ethanol solution. Then, Pd colloid suspension (particle size: 4 nm) was added to the Co3O4-mixed ethanol solution, and the mixed solution was stirred and dried at 50 °C. The Pd contents were 0 wt%, 0.1 wt%, 1 wt%, 5 wt%, 10 wt%, and 30 wt% relative to the Co3O4 content. The resulting powders were denoted as Co3O4, 0.1%Pd/Co3O4, 1%Pd/Co3O4, 5%Pd/Co3O4, 10%Pd/Co3O4, and 30%Pd/Co3O4, respectively. The powders were mixed with an organic dispersant, which consisted of a mixture of 10 wt% ethyl cellulose and 90 wt% terpineol, to obtain a paste suitable for air dispensing. The weight ratio of the powder to the organic dispersant was 1:16. The paste was dispensed on a surface-oxidized Si substrate, which consisted of 2.5 × 4 mm2 platinum interdigital electrodes with line and space definitions of 15 μm each, using an air-dispenser (FAD320s, Musashi Engineering, Tokyo, Japan) [13,17]. The substrate was baked at 400 °C for 2 h under air to obtain the sensor elements. The Co3O4 film thickness could be controlled to 2–4 μm by air-dispenser.
The obtained sample powders were characterized by X-ray diffraction (XRD) analysis and transmission electron microscopy (TEM). The XRD analysis was carried out using a RINT 2100V/PC instrument (Rigaku Corporation, Tokyo, Japan) equipped with a copper source (CuKα). TEM was performed using a JEM-2010 instrument (JEOL Ltd., Tokyo, Japan) equipped with an energy dispersive X-ray spectroscope (EDX).
To investigate the NO gas-sensing properties, the sensor element was placed in a test chamber heated to 100–300 °C in an electrical tube furnace. Air was introduced into the chamber for 15 min, and a gas mixture of NO in air was then injected for 15 min. Subsequently, the gas mixture flow was halted and replaced by air injected at a flow rate of 200 mL/min. The NO concentration was controlled to the values of 50, 100, 150, and 200 ppb in air. The resistance among the electrodes of the element in various gaseous atmospheres was measured by two-probe method at 10-s intervals using a K2700 digital multimeter (Keithley). We defined the sensor response value (S) using the following equation: S = Rg/Ra, where Rg denotes the resistance after the NO gas exposure for 15 min and Ra denotes the average resistance (μ) in air for 3 min before the NO gas exposure. The coefficient of variation (CV) of Ra was defined as CV(%) = μ/σ × 100, where σ denotes the Ra standard deviation in air for 3 min. The S/N ratio was defined as S/N = (Rg − μ)/σ. The H2 gas-sensing properties were investigated using the same NO gas-sensing measurement apparatus. The H2 gas concentration was controlled to 0, 25, and 100 ppm in air.

3. Results and Discussion

Figure 1 shows the XRD patterns of the Co3O4, 0.1%Pd/Co3O4, 1%Pd/Co3O4, 5%Pd/Co3O4, 10%Pd/Co3O4, and 30%Pd/Co3O4 powder after baking at 400 °C for 2 h. Because the peaks of Co3O4 (JCPDS card No. 74-2120) were observed in all patterns, no structural changes due to the oxidation or decomposition of Co3O4 were observed after baking at 400 °C for 2 h. The peaks at 34°, 42°, and 55° were assigned to PdO (JCPDS card No. 75-0584). The palladium, which was loaded by a colloidal mixing method process, was oxidized to PdO. A small hill at 34° was observed in the pattern of 1%Pd/Co3O4, and this became a peak in the patterns of 5%Pd/Co3O4, 10%Pd/Co3O4, and 30%Pd/Co3O4. The PdO peaks at 42° and 55° were observed in the pattern of 30%Pd/Co3O4. No PdO peaks were observed in the pattern of 0.1%Pd/Co3O4 owing to its lower Pd amount. The crystallite sizes (D) were calculated using the Scherrer equation: D = /βcosθ, where k (= 0.9) is the shape factor, λ is the X-ray wavelength, β is the full width at half maximum of the diffraction peak, and θ is the Bragg diffraction angle. The D values of the Co3O4 peaks at 37° and PdO peak at 34° in 30%Pd/Co3O4 were estimated to be 19.0 and 11.7 nm, respectively.
Figure 1. XRD patterns of (a) Co3O4; (b) 0.1%Pd/Co3O4; (c) 1%Pd/Co3O4; (d) 5%Pd/Co3O4; (e) 10%Pd/Co3O4; and (f) 30%Pd/Co3O4 powder after baking at 400 °C for 2 h.
Figure 1. XRD patterns of (a) Co3O4; (b) 0.1%Pd/Co3O4; (c) 1%Pd/Co3O4; (d) 5%Pd/Co3O4; (e) 10%Pd/Co3O4; and (f) 30%Pd/Co3O4 powder after baking at 400 °C for 2 h.
Sensors 15 08109 g001
Figure 2 shows the response of the sensor element using the Co3O4, 0.1%Pd/Co3O4, 1%Pd/Co3O4, 5%Pd/Co3O4, 10%Pd/Co3O4, and 30%Pd/Co3O4 powder during exposure to NO (50, 100, 150, and 200 ppb) in air at 200 °C. The resistance of all sensor elements began to increase when they were exposed to NO gas. In the case of a p-type semiconductor such as Co3O4 and NiO whose majority carrier is a hole, when the concentration of electrons on the semiconductor surface increases, the resistance of the semiconductor increases because the generated electrons recombine with the holes [18]. When a reductive NO gas is streamed on a p-type semiconductor, the NO gas reacts with the adsorbed oxygen ions on the semiconductor surface, releasing electrons back to the conduction band. Therefore, the resistance of the sensor elements increases toward the NO gas exposure. All sensor elements exhibited a distinct response to NO gas even at 50 ppb. The resistance of all sensor elements increased with the NO gas concentration. The resistance of the NO-exposed Co3O4 film did not reach Ra even after 15 min of air exposure (see Figure 2a). This result is considered to imply that the adsorbed NO on Co3O4 film surface did not desorb from the surface during air exposure for 15 min. On the other hand, the resistances of the NO-exposed Pd-loaded Co3O4 films reached Ra even after 15 min of air exposure (see Figure 2b to Figure 2f). The background resistance signal in Figure 2f was slightly noisy, which might be due to the 30 wt% Pd content. The sensor responses S(= Rg/Ra) were plotted with respect to the NO concentration of 50–200 ppb (see Figure 3). All sensor responses linearly increased with the NO gas concentration. The sensor responses increased with the Pd content up to 10 wt% but became low in the 30 wt% Pd content. Since the XRD pattern of the 30%Pd/Co3O4 showed large PdO crystallite size, the large PdO interfered with the adsorption of the NO onto the Co3O4, resulting in the decrease of the sensor response. The high sensor responses S = 1.16 at 50 ppb NO was obtained for the optimum composition 10%Pd/Co3O4. Figure 4 shows the sensor responses to 50 and 200 ppb NO of the sensor element using the 10%Pd/Co3O4 powder as a function of the operating temperature in the range from 100 to 300 °C. The NO sensor response of the 10%Pd/Co3O4 sensor was the highest at 200 °C.
Figure 2. Responses of the sensor element using (a) Co3O4; (b) 0.1%Pd/Co3O4; (c) 1%Pd/Co3O4; (d) 5%Pd/Co3O4; (e) 10%Pd/Co3O4; and (f) 30%Pd/Co3O4 powder during exposure to NO (50, 100, 150, and 200 ppb) in air at 200 °C.
Figure 2. Responses of the sensor element using (a) Co3O4; (b) 0.1%Pd/Co3O4; (c) 1%Pd/Co3O4; (d) 5%Pd/Co3O4; (e) 10%Pd/Co3O4; and (f) 30%Pd/Co3O4 powder during exposure to NO (50, 100, 150, and 200 ppb) in air at 200 °C.
Sensors 15 08109 g002
Figure 3. Sensor responses of the sensor element using Co3O4, 0.1%Pd/Co3O4, 1%Pd/Co3O4, 5%Pd/Co3O4, 10%Pd/Co3O4, and 30%Pd/Co3O4 powders as a function of the NO gas concentration.
Figure 3. Sensor responses of the sensor element using Co3O4, 0.1%Pd/Co3O4, 1%Pd/Co3O4, 5%Pd/Co3O4, 10%Pd/Co3O4, and 30%Pd/Co3O4 powders as a function of the NO gas concentration.
Sensors 15 08109 g003
Figure 4. Sensor responses of the sensor element using a 10%Pd/Co3O4 powder to 50 and 200 ppb NO as a function of the operating temperature.
Figure 4. Sensor responses of the sensor element using a 10%Pd/Co3O4 powder to 50 and 200 ppb NO as a function of the operating temperature.
Sensors 15 08109 g004
Table 1 shows the sensing properties of the Co3O4 and Pd/Co3O4 sensor elements upon exposure to 50 ppb NO in air at 200 °C. The 90% response time, defined as the time needed to reach 90% of the final steady value of Rg toward 50 ppb NO exposure, and the 90% recovery time, defined as the time needed to reach 90% of Ra, for the Co3O4 sensor element, were 12.7 and 13.5 min, respectively. On the other hand, the 90% response time and 90% recovery time for the Pd/Co3O4 sensor elements were approximately 5 min shorter than those for the Co3O4 sensor element. This result suggests that the loaded PdO provided some kinds of surface modification effects with Co3O4. Thus, the average resistance μ increased with the Pd content in the films due to this modification effects. Although μ increased due to the Pd loading, all sensor resistances, which were less than 600 Ω, were lower than the resistances of the sensor using WO3 and SnO2 in our earlier works [13,17].
Table 1. Sensing properties of the Co3O4 and Pd/Co3O4 sensor elements upon exposure to 50 ppb NO in air. The operating temperature was 200 °C, and the gas flow rate was 200 mL/min.
Table 1. Sensing properties of the Co3O4 and Pd/Co3O4 sensor elements upon exposure to 50 ppb NO in air. The operating temperature was 200 °C, and the gas flow rate was 200 mL/min.
Sample90% Response Time (min)90% Recovery Time (min)RaRg (Ω)S (Rg/Ra)S/N Ratio
μ (Ω)σ (Ω)CV (%)
Co3O412.713.552.40.0050.01053.61.02224
0.1%Pd/Co3O48.58.658.50.0180.03164.11.10307
1%Pd/Co3O49.39.858.20.0070.01162.81.08695
5%Pd/Co3O49.810.3121.10.0570.047137.01.13278
10%Pd/Co3O47.68.0233.00.0410.017269.31.16891
30%Pd/Co3O44.77.7478.51.1470.240502.21.0521
The Ra standard deviation σ and the CV value increased with the Pd content. The S/N ratio of the Co3O4 toward 50 ppb NO exposure was 64 and was lower than that of the Pd/Co3O4 with 0.1 wt%–10 wt% Pd content. The S/N ratio of the 30%Pd/Co3O4 was lower than that of the Co3O4 because of the noisy background resistance signal. The S/N ratio of the 5%Pd/Co3O4 was lower than that of the 1%Pd/Co3O4 or 10%Pd/Co3O4 owing to large CV value. The origin of the discontinuity of the S/N ratio is unclear at this stage. Further work is necessary to clarify the origin. Following the conventional definition in Kaiser’s work [19], the limit of detection value was defined as the NO concentration at which a sensor response corresponds to S/N = 3. Thus, the sensor element with a large S/N ratio can detect a low concentration of NO with high reliability. The limit of detection value (S/N = 3) of the 10%Pd/Co3O4 sensor was estimated to be 0.66 ppb considering S/N = 227 at 50 ppm NO, which was lower than that of the 12 ppb value reported in the literature [14]. The suitable Pd content of the sensing material Pd/Co3O4 was considered to be 10 wt% from the sensor response and S/N ratio.
To examine the reason why the sensor using the 10%Pd/Co3O4 powder had good sensor properties, TEM observations of the Co3O4, 10%Pd/Co3O4, and 30%Pd/Co3O4 powders were carried out. Figure 5 shows the TEM images and EDX spectrum of the Co3O4 powder after baking at 400 °C for 2 h. Pale phases were observed, and the grain size of the Co3O4 powder was approximately 20 nm. EDX spot analysis was carried out at the area inside the circle in Figure 5a. The EDX spectrum showed that the Co3O4 powder has no impurities such as noble metals.
Figure 6 shows the TEM images and EDX spectrum of the 10%Pd/Co3O4 powder after baking at 400 °C for 2 h. Although no difference was observed in the TEM images between the Co3O4 and 10%Pd/Co3O4 powder, Pd peak was confirmed in the EDX spectrum of the 10%Pd/Co3O4 powder. Matsushima et al. reported 0.3–5 wt% Pd-loaded SnO2 powder baked at 300 °C in air, and the Pd particles on the SnO2 surface became less visible in the TEM image [20], which was considered to be due to the existence of Pd on the SnO2 surface as PdO with a lower contrast to SnO2. Pd in the 10%Pd/Co3O4 powder existed as PdO and was not distinguishable from the TEM image.
Figure 5. (a,b) TEM images of the Co3O4 powder; (c) EDX analysis result of the area inside the circle in the (a) image. The powder is baked at 400 °C for 2 h.
Figure 5. (a,b) TEM images of the Co3O4 powder; (c) EDX analysis result of the area inside the circle in the (a) image. The powder is baked at 400 °C for 2 h.
Sensors 15 08109 g005
Figure 6. (a,b) TEM images of the 10%Pd/Co3O4 powder; (c) EDX analysis result of the area inside the circle in the (a) image. The powder is baked at 400 °C for 2 h.
Figure 6. (a,b) TEM images of the 10%Pd/Co3O4 powder; (c) EDX analysis result of the area inside the circle in the (a) image. The powder is baked at 400 °C for 2 h.
Sensors 15 08109 g006
Figure 7 shows the TEM images and EDX spectra of the 30%Pd/Co3O4 powder after baking at 400 °C for 2 h. Figure 7a shows that although no difference was observed among the TEM images with low magnification of the Co3O4, 10%Pd/Co3O4, and 30%Pd/Co3O4 powder, the intensity of the Pd peak in the EDX spectrum increased with the Pd content. Figure 7b shows small dark phases with 5-nm diameter, such as the area inside circle A, and pale phases, such as the area inside circle B. Figure 7d,e shows the EDX spectra of the area inside circles A and B in Figure 7b, respectively. The intensity of the Pd peak in Figure 7c was much lower than that in Figure 7d and higher than that in Figure 7e. This result showed that the aggregation of the PdO particles was formed at the area inside circle A, and the 30%Pd/Co3O4 powder had poor dispersibility of the PdO particles. Because of the aggregation and poor dispersibility of the PdO particles, the sensor element using 30%Pd/Co3O4 was considered to show a lower sensor response and S/N ratio than the others. No difference was shown between the Co3O4 crystallite size determined from the Scherrer equation and Co3O4 particle size obtained by TEM image. However, the size of the PdO aggregation in Figure 7b was smaller than the PdO crystallite size determined from XRD. Because the PdO in the 30%Pd/Co3O4 powder had a lower contrast to Co3O4, the PdO aggregation size obtained from the TEM image was smaller than that determined by XRD.
Figure 7. (a,b) TEM images of the 30%Pd/Co3O4 powder; (c) EDX analysis result of the area inside the circle in the (a) image; (d) EDX analysis result of the area inside circle A; (e) EDX analysis result of the area inside circle B. The powder is baked at 400 °C for 2 h.
Figure 7. (a,b) TEM images of the 30%Pd/Co3O4 powder; (c) EDX analysis result of the area inside the circle in the (a) image; (d) EDX analysis result of the area inside circle A; (e) EDX analysis result of the area inside circle B. The powder is baked at 400 °C for 2 h.
Sensors 15 08109 g007
Figure 8 shows the sensing responses of the sensor element using 10%Pd/Co3O4 powder toward 25 and 100 ppm H2 as a function of the operating temperature. When the 10%Pd/Co3O4 sensor was exposed to H2 gas, which is a typical reducing gas, the resistance of the sensor increased. The high sensor responses (S = 1.9 at 25 ppm H2 and S = 3.0 at 100 ppm H2) were obtained at 150 °C. This result showed that the 10%Pd/Co3O4 sensor is not immune to the interference of human breath level H2 gas. In addition, the sensor can possibly respond to other inflammable gases and VOCs because semiconductor sensors are sensitive. Therefore, the Pd-loaded Co3O4 sensor needs to have a gas separation membrane and a filter to remove interference such as H2, CH4, CO, and H2O.
Figure 8. Sensing responses of the sensor element using 10%Pd/Co3O4 powder to 25 and 100 ppm H2 as a function of the operating temperature.
Figure 8. Sensing responses of the sensor element using 10%Pd/Co3O4 powder to 25 and 100 ppm H2 as a function of the operating temperature.
Sensors 15 08109 g008
The Pd/Co3O4 sensor elements showed a good response to 50 ppb of NO in air. Further investigation for improvement of the sensor response is required to detect the sub-ppb levels of NO with sufficient accuracy. To enhance the sensor response, sensor properties using Co3O4 powder with noble metals such as Pt, Au, and Ag, will be investigated in the future.

4. Conclusions

In this paper, we have investigated the gas sensing properties of a sensor element using Pd-loaded Co3O4 prepared by a colloidal mixing method. The Pd particles in the Co3O4-based sensing materials existed as PdO particles. The sensor elements with 0.1 wt%–10 wt% Pd content showed shorter 90% response time and 90% recovery time and higher S/N ratio than those with no Pd content. However, the sensor response and S/N ratio of the sensor element with a 30 wt% Pd content markedly decreased because of the aggregation and poor dispersibility of the PdO particles. The sensor element with 10 wt% Pd content showed a high sensor response (S = 1.16) at 50 ppb NO.

Acknowledgments

This work was supported in part by the “Development Project for Extremely-Early Stage Diagnostics Technologies for Human Diseases” of Aichi Prefecture, Japan.

Author Contributions

T.A. performed the experiments and prepared the manuscript. T.I., N.I., and W.S. helped in designing the experiments and provided valuable guidance. K.S. contributed to the conception of this work and provided helpful feedback on the comments on the manuscript. All authors read and approved the final manuscript.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Behall, K.M.; Scholfield, D.J.; van der Sluijs, A.M.; Hallfrisch, J. Breath hydrogen and methane expiration in men and women after oat extract consumption. J. Nutr. 1998, 128, 79–84. [Google Scholar] [PubMed]
  2. Roberge, M.T.; Finley, J.W.; Lukaski, H.C.; Borberding, A.J. Evaluation of the pulsed discharge helium ionization detector for the analysis of hydrogen and methane in breath. J. Chromatogr. A 2004, 1027, 19–23. [Google Scholar] [CrossRef] [PubMed]
  3. Fucks, P.; Loeseken, C.; Schubert, J.K.; Miekisch, W. Breath gas aldehydes as biomarkers of lung cancer. Int. J. Cancer 2010, 126, 2663–2670. [Google Scholar] [PubMed]
  4. Saito, J.; Inoue, K.; Sugawara, A.; Yoshikawa, M.; Watanabe, K.; Ishida, T.; Ohtsuka, Y.; Munakata, M. Exhaled nitric oxide as a marker of airway inflammation for an epidemiologic study in schoolchildren. J. Allergy Clin. Immunol. 2004, 114, 512–516. [Google Scholar] [CrossRef] [PubMed]
  5. Grzelewski, T.; Pawel, M.; Joanna, J.; Łukasz, C.; Jan, K.; Włodzimierz, S.; Rafał, S.; Anna, J.; Aleksandra, G.; Iwona, S. The association between fractional exhaled nitric oxide (FeNO) and cat dander in asthmatic children. Nitric Oxide 2011, 25, 288–293. [Google Scholar] [CrossRef] [PubMed]
  6. Recommendations for standardized procedures for the online and offline measurement of exhaled lower respiratory nitric oxide and nasal nitric oxide in adults and children-1999. Am. J. Respir. Crit. Care Med. 1999, 160, 2104–2117.
  7. Fontijn, A.; Sabadell, A.J.; Ronco, R.J. Homogeneous chemiluminescent measurement of nitric oxide with ozone: Implications for continuous selective monitoring of gaseous air pollutants. Anal. Chem. 1970, 42, 575–579. [Google Scholar] [CrossRef]
  8. Fruhberger, B.; Stirling, N.; Grillo, F.G.; Ma, S.; Ruthven, D.; Lad, J.; Frederick, B.G. Detection and quantification of nitric oxide in human breath using a semiconducting oxide based chemiresistive microsensor. Sens. Actuators B Chem. 2001, 76, 226–234. [Google Scholar] [CrossRef]
  9. Sun, C.; Maduraiveeran, G.; Dutta, P. Nitric oxide sensors using combination of p- and n-type semiconducting oxides and its application for detecting NO in human breath. Sens. Actuators B Chem. 2013, 186, 117–125. [Google Scholar] [CrossRef]
  10. Ishihara, T.; Shiokawa, K.; Eguchi, K.; Arai, H. The mixed oxide Al2O3-V2O5 as a semiconductor gas sensor for NO and NO2. Sens. Actuators 1989, 19, 259–265. [Google Scholar] [CrossRef]
  11. Akiyama, M.; Tamaki, J.; Miura, N.; Yamazoe, N. Tungsten oxide-based semiconductor sensor highly sensitive to NO and NO2. Chem. Lett. 1991, 20, 1611–1614. [Google Scholar] [CrossRef]
  12. Penza, M.; Martucci, C.; Cassano, G. NOx gas sensing characteristics of WO3 thin films activated by noble metals (Pd, Pt, Au) layers. Sens. Actuators B Chem. 1998, 50, 52–59. [Google Scholar] [CrossRef]
  13. Akamatsu, T.; Itoh, T.; Izu, N.; Shin, W. NO and NO2 sensing properties of WO3 and Co3O4 based gas sensors. Sensors 2013, 13, 12467–12481. [Google Scholar] [CrossRef] [PubMed]
  14. Wang, X.; Miura, N.; Yamazoe, N. Study of WO3-based sensing materials for NH3 and NO detection. Sens. Actuators B Chem. 2000, 66, 74–76. [Google Scholar] [CrossRef]
  15. Lin, C.Y.; Fang, Y.Y.; Lin, C.W.; Tunney, J.J.; Ho, K.C. Fabrication of NOx gas sensors using In2O3-ZnO composite films. Sens. Actuators B Chem. 2010, 146, 28–34. [Google Scholar] [CrossRef]
  16. Shin, W. Medical applications of breath hydrogen measurements. Anal. Bioanal. Chem. 2014, 406, 3931–3939. [Google Scholar] [CrossRef] [PubMed]
  17. Itoh, T.; Nakashima, T.; Akamatsu, T.; Izu, N.; Shin, W. Nonanal gas sensing properties of platinum, palladium, and gold-loaded tin oxide VOCs sensors. Sens. Actuators B Chem. 2013, 187, 135–141. [Google Scholar] [CrossRef]
  18. Wetchakun, K.; Samerjai, T.; Tamaekong, N.; Liewhiran, C.; Siriwong, C.; Kruefu, V.; Wisitsoraat, A.; Tuantranont, A.; Phanichphant, S. Semiconducting metal oxides as sensors for environmentally hazardous gases. Sens. Actuators B Chem. 2011, 160, 580–591. [Google Scholar] [CrossRef]
  19. Kaiser, H. Zum Problem der Nachweisgrenze. Fresenius Z. Anal. Chem. 1965, 209, 1–18. [Google Scholar] [CrossRef]
  20. Matsushima, S.; Tamaki, J.; Muira, N.; Yamazoe, N. TEM observation of the dispersion state of Pd on SnO2. Chem. Lett. 1989, 18, 1651–1654. [Google Scholar] [CrossRef]

Share and Cite

MDPI and ACS Style

Akamatsu, T.; Itoh, T.; Izu, N.; Shin, W.; Sato, K. Sensing Properties of Pd-Loaded Co3O4 Film for a ppb-Level NO Gas Sensor. Sensors 2015, 15, 8109-8120. https://0-doi-org.brum.beds.ac.uk/10.3390/s150408109

AMA Style

Akamatsu T, Itoh T, Izu N, Shin W, Sato K. Sensing Properties of Pd-Loaded Co3O4 Film for a ppb-Level NO Gas Sensor. Sensors. 2015; 15(4):8109-8120. https://0-doi-org.brum.beds.ac.uk/10.3390/s150408109

Chicago/Turabian Style

Akamatsu, Takafumi, Toshio Itoh, Noriya Izu, Woosuck Shin, and Kazuo Sato. 2015. "Sensing Properties of Pd-Loaded Co3O4 Film for a ppb-Level NO Gas Sensor" Sensors 15, no. 4: 8109-8120. https://0-doi-org.brum.beds.ac.uk/10.3390/s150408109

Article Metrics

Back to TopTop