Article

Electrochemical Detection of Electrolytes Using a Solid-State Ion-Selective Electrode of Single-Piece Type Membrane

1
Department of Mechanical Engineering, National Chung-Hsing University, Taichung 40227, Taiwan
2
Graduate Institute of Biomedical Engineering, National Chung-Hsing University, Taichung 40227, Taiwan
3
Regenerative Medicine and Cell Therapy Research Center, Kaohsiung Medical University, Kaohsiung 80708, Taiwan
*
Author to whom correspondence should be addressed.
Submission received: 1 March 2021 / Revised: 22 March 2021 / Accepted: 4 April 2021 / Published: 7 April 2021

Abstract

:
This study aimed to develop simple electrochemical electrodes for the fast detection of chloride, sodium and potassium ions in human serum. A flat thin-film gold electrode was used as the detection electrode for chloride ions; a single-piece type membrane based solid-state ion-selective electrode (ISE), which was formed by covering a flat thin-film gold electrode with a mixture of 7,7,8,8-tetracyanoquinodimethane (TCNQ) and ion-selective membrane (ISM), was developed for sodium and potassium ions detection. Through cyclic voltammetry (CV) and square-wave voltammetry (SWV), the detection data can be obtained within two minutes. The linear detection ranges in the standard samples of chloride, sodium, and potassium ions were 25–200 mM, 50–200 mM, and 2–10 mM, with the average relative standard deviation (RSD) of 0.79%, 1.65%, and 0.47% and the average recovery rates of 101%, 100% and 96%, respectively. Interference experiments with Na+, K+, Cl, Ca2+, and Mg2+ ions demonstrated that the proposed detection electrodes have good selectivity. Moreover, the proposed detection electrodes have characteristics such as the ability to be prepared under relatively simple process conditions, excellent detection sensitivity, and low RSD, and the detection linear range is suitable for the Cl, Na+ and K+ concentrations in human serum.

1. Introduction

The chloride, sodium and potassium ions in the human body can maintain the physiological functions of water metabolism, the neuromuscular system, acid-base balance and osmotic pressure in the body. Therefore, the metabolic functions of the human body can be evaluated based on the measurements of the chloride, sodium and potassium ions concentrations in the body [1,2,3,4].
The concentrations of chloride, sodium, and potassium ions in normal human serum are 98–106 mM [2], 135–145 mM [3] and 3.5–5 mM [3], respectively. Too-high concentrations in the human body will cause hyperchloremia, hypernatremia, and hyperkalemia, while the occurrence of concentrations lower than the normal concentration will cause hypochloremia, hyponatremia, and hypoxemia potassium. Another common chloride-related disease is cystic fibrosis, which is an autosomal recessive genetic disease that affects mucus-producing cells in the body and causes mucus accumulation in the lungs, trachea, and digestive tract, leading to the degradation of breathing functions [5]. The general clinical diagnosis of this disease is based on measuring the chloride ion content in sweat. People with the disease usually have a chloride ion content of over 60 mM in their sweat [2], while the chloride ion content in normal human sweat is below 40 mM [5].
The laboratory autoanalyzer is a general electrolyte-detecting instrument that can provide doctors with reliable results. However, the specimen must be delivered to a specific laboratory and diluted before indirect ion-selective electrodes (ISEs) are used for detection. For patients with a severe or life-threatening degree of the illness, the blood gas analyzer (BGA) is usually used clinically to achieve rapid detection. The BGA uses direct ISE for direct analysis without the need to first dilute the sample, and the results can be obtained within 1 to 5 min [6,7,8,9,10]. The BGA can be divided into two types: benchtop and portable. The benchtop BGA detection cartridge has a service life of about 28–30 days; hence, it is suitable for mass detection applications. However, the instrument needs regular calibration and maintenance. The portable BGA usually uses a disposable detection cassette, which can reduce the frequency of instrument calibration and maintenance; however, since the cassette cannot be reused, the portable BGA is only suitable for small-volume detection applications [11]. In recent years, as the utilization rate of point-of-care (POC) testing has significantly increased, the economic value of POC-related products has also increased [12]. Therefore, there are many portable and small BGAs and electrolyte analyzer-related products in the market, such as Abbott i-STAT [13], Siemens epoc® blood analysis system [12], and Arkray Spotchem EL SE-1520 [14].
At present, the electrochemical potentiometric method [13,14,15,16,17,18] is the main method for electrolyte ions quantification. Other detection methods include the conductivity method [19,20], cyclic voltammetry (CV) [21,22,23,24,25,26], and coulometry [27,28]. The potentiometric method uses liquid-state ISEs that need to be regularly replaced. The operating temperature is limited, and the electrodes cannot be effectively miniaturized [29,30,31]. Moreover, the practical application of the conductance method is not yet mature [21]. Therefore, in the current study, a novel detection method combining high-sensitivity voltammetry and solid-state ISE is developed. The solid-state ISE does not need to be filled with a solution. The ion-selective membrane directly contacts the electrode, and the solid contact material of the intermediate layer is added to enhance the effect of ion and electron conversion. Therefore, the sensor can be more easily miniaturized, and its durability and operating temperature range are improved.
Commonly used solid contact materials include high-molecular-weight conductive polymers such as polypyrroles [32], poly(3-octylthiophene) [33], and poly(3,4-ethylenedioxythiophene) (PEDOT) [34]. Nanomaterials used as solid contact materials include carbon nanotubes [17], graphene [35], platinum nanoparticles [36], and noble metal nanostructures [37]. Electroactive substances are also used, such as tetrathiafulvalene [38] and 7,7,8,8-tetracyanoquinodimethane (TCNQ) [39]. Conductive polymers such as PEDOT need to be synthesized via electrochemical polymerization [34], while nanomaterials need to be produced via electrochemical deposition [37]. The manufacturing processes of both are relatively cumbersome. Therefore, this study uses the solid-state contact conduction mechanism of high redox capacitors [29] and adopts the electroactive material TCNQ as the solid-state contact material.
TCNQ is a powerful organic electron acceptor with excellent electrical properties and high electron affinity (2.88 eV) [40]; thus, it can easily donate electron acceptors to react with transition metals or organic compounds and form numerous TCNQ anion–based salt complexes, such as CuTCNQ [40], AgTCNQ [40], KTCNQ [39], and NaTCNQ [39]. In addition, the use of the TCNQ involves a simple process; the TCNQ only needs to be dropped on the electrode substrate through drop-casting [39].
In this study, electrodes for the effective and fast detection of electrolytes in human serum were developed. A flat thin-film gold electrode was used as the detection electrode for chloride ions and a solid-state ISE, which was formed by covering a flat thin-film gold electrode with a mixture of TCNQ and ion-selective membrane (ISM), was developed for sodium and potassium ions detection, respectively.

2. Materials and Methods

2.1. Materials and Reagents

TCNQ, tetrahydrofuran (THF), potassium chloride, sodium chloride, calcium chloride, magnesium chloride, human serum, potassium ionophore I, poly(vinyl chloride) and potassium phosphate monobasic were purchased from Sigma-Aldrich (St. Louis, MO, USA). Potassium ferricyanide and potassium ferrocyanide were purchased from SHOWA (Tokyo, Japan). Potassium tetrakis (4-chlorophenyl)borate, 2-nitrophenyl octyl ether, and 4-tert-butylcalix[4]arenetetraacetic acid tetraethyl ester were purchased from Thermo (Darmstadt, Germany). Sodium phosphate dibasic was purchased from J. T. Baker (USA). Acetone was purchased from ECHO Chemical (Miaoli, Taiwan). Distilled deionized water (>18 MΩ) was obtained from ELGA LabWater (London, UK).

2.2. Electrode Fabrication

Scheme 1 illustrates the fabrication process of the chloride, potassium, and sodium ions detection electrodes. First, a thin gold layer was sputtered onto a 3 × 5 mm2 polyethylene terephthalate (PET) membrane (Scheme 1A). Then, the gold thin-film–sputtered PET membrane was packaged to form the detection chip (Scheme 1B). For the potassium and sodium ions detection electrodes, 3 μL of a TCNQ and ISM mixture solution was dripped onto the electrode (Scheme 1C). Table 1 presents the compositions and contents of potassium and sodium ISMs.

2.3. Apparatus

Cyclic voltammetry and square-wave voltammetry (SWV) were conducted with an SP-150 potentiostat (BioLogic, Seyssinet-Pariset, France) connected to a computer. A three-electrode system, consisting of a working electrode, a platinum counter electrode, and an Ag/AgCl reference electrode, was applied. Field-emission scanning electron microscopy (FE-SEM, Zeiss UltraPlus, Potsdam, Germany) was used for the electrode surface characterization. The analysis system and the electrodes are shown in Figure 1.

3. Results and Discussion

3.1. Optimization of Detection Electrodes

Two parameters of the detection electrode need to be optimized—the organic solvent to dissolve TCNQ and the volume ratio of TCNQ to ISM. Regarding the organic solvent, a suitable solvent for TCNQ was selected from two organic solvents: acetone and THF. The TCNQ solution was mixed with the corresponding ISM (Na+ and K+) with a volume ratio of 1:1 to determine which was the more suitable solvent. After the suitable TCNQ solvent was selected, TCNQ/ISM mixed solutions with different volume ratios (1:1, 1:2, 2:1, 3:1) were prepared to optimize the volume ratio.
Figure 2A,B show the cyclic voltammograms for Na+ (140 mM NaCl) and K+ (5 mM KCl) detections, respectively. The acetone-dissolved TCNQ resulted in peak oxidation currents at 0.46 and 0.50 V for Na+ and K+ detections, respectively. The peak oxidation currents were much higher than those obtained from the THF-dissolved TCNQ. Therefore, acetone was selected as the TCNQ solvent. The cyclic voltammograms for Na+ and K+ detections with different TCNQ/ISM volume ratios are shown in Figure 2C,D, respectively. The results show that for the Na+ detection, the TCNQ/ISM volume ratio of 2:1 resulted in the maximum peak oxidation current at 0.45 V, while for the K+ detection, the ratio of 3:1 resulted in the maximum peak oxidation current at 0.43 V. In Figure 2C,D, as the TCNQ content increases, the oxidation potentials of Na+ and K+ shift to the left. Therefore, using a lower potential can drive the oxidation reaction and reduce the scanning voltage range during CV.

3.2. Characterization of Fabricated Electrodes

Figure 3 displays the scanning electron microscopy (SEM) images of the fabricated electrodes. The top view shows many typical TCNQ black blocks and a few white flake structures. The white flake structures are TCNQ erected under the influence of gravity [41]. The cross-sectional view (Figure 3B) shows that the thickness of the deposited Na+ ISM and K+ ISM layers: about 24.4 and 14.6 µm, respectively.

3.3. The Electrochemical Behavior of Cl, Na+, and K+ Electrodes

The electron diffusion rate can be used to study the characteristics of electrochemical electrodes. The oxidation and reduction potentials, as well as the reaction current of the analyte, can be obtained based on a CV analysis of the reaction between the analyte and the electrode. The results can be used to determine whether the reaction is diffusion-controlled.
Figure 4 shows the CV results of the proposed different electrodes at scan rates ranging from 100 to 500 mV/s in 50 mM KCl, 140 mM NaCl, and 1 mM KCl solutions. The cyclic voltammograms displayed in Figure 4(A1–C1) show that the absolute value of the oxidation and reduction peak currents increased with the increase in scan rate. Generally, the Randles–Sevcik equation (Equation (1) is used to evaluate the diffusion-controlled characteristic of a sensing electrode.
i p = 2.69 × 10 5 × n 3 / 2 × A × C × D 1 / 2 × v 1 / 2
where ip is the peak current, n is the number of electrons appearing in a half-reaction for the redox couple, A is the electrode area (cm2), C is the analyte concentration (mole/cm3), D is the analyte diffusivity (cm2/s), and v is the scan rate of potential (V/s). For stationary n, A, C, and D, ip is a linear function of the square root of the scanning rate. The corresponding linear relationship between the peak current and the square root of the scanning rate of the proposed electrodes for the cyclic voltammograms shown in Figure 4A1–C1 are given in Figure 4A2–C2, respectively. The R2 values were calculated to be 0.9986, 0.9981, and 0.9989, respectively. The results demonstrate that the proposed electrodes possess a typical diffusion-controlled electrochemical characteristic; hence, the electrodes are suitable for practical quantitative analyses.

3.4. Detection of Cl, Na+, and K+

Cyclic voltammetry was used for the detection of Cl (KCl) because it can be easily oxidized by gold with a suitable oxidation potential. Square-wave voltammetry was adopted for the detections of Na+ (NaCl) and K+ (KCl) because the oxidation process involved the ISM.
(1)
Cl detection
Figure 5 shows the detection results of Cl with various concentrations (25, 50, 100, 150, and 200 mM). Clear oxidation and reduction current peaks can be observed in Figure 5A. The absolute value of the oxidation and reduction peak currents increased with the increase in Cl concentration. Figure 5B displays the corresponding calibration curve for Cl detection, with the normal concentration range of Cl in human blood identified by a bar icon. The linear detection range, sensitivity, correlation coefficient, relative standard deviation (RSD), and recovery rate of the electrode were calculated to be 5–200 mM, 0.27 mA mM−1 cm−2, 0.9988, 0.79%, and 101%, respectively.
(2)
Na+ detection
Figure 6 shows the SWV detection results of Na+ with various concentrations (50, 100, 150, and 200 mM). Figure 6A displays clear oxidation current peaks, which increase with the increase in the Na+ concentrations. Figure 6B shows the corresponding calibration curve for Na+ detection, with the normal Na+ concentration range in human blood marked by a bar icon. The linear detection range, sensitivity, correlation coefficient, RSD, and recovery rate of the electrode were calculated to be 50–200 mM, 0.125 μA·mM−1cm−2, 0.9786, 1.65% and 100%, respectively.
(3)
K+ detection
The SWV detection results of K+ with various concentrations (2, 3.5, 5, 7.5, and 10 mM) are shown in Figure 7. In Figure 7A, the oxidation current peaks increase with increasing K+ concentrations. Figure 7B shows the corresponding quadratic calibration curve for K+ detection, which is more appropriate for fitting the measured data. The detection range, sensitivity, correlation coefficient, RSD, and recovery rate of the electrode were measured to be 2–10 mM, 0.56–7.98 μA·mM−1 cm−2, 0.9778, 0.47%, and 96%, respectively.
(4)
Interference detection
We used normal concentrations of Na+, K+, Cl, Ca2+, and Mg2+ in human blood to examine the influence of interference substances on the oxidation current response of different electrodes. The experimental results are depicted in Figure 8.
The results of the Cl detection interference experiments (Figure 8) showed that the oxidation currents of four chlorides (KCl, NaCl, CaCl2, and MgCl2) having the same concentration (100 mM) were between 1.84 and 1.94 mA, with an average current of 1.89 mA. The RSD was calculated to be 2.35%. The experimental results indicate that the Cl detection electrode was not interfered by cations.
The results of the Na+ detection interference experiments (Figure 8) showed that the current response of the Na+ detection electrode in the NaCl solution was 5.55 µA. After KCl, CaCl2, and MgCl2 were added to the KCl solution, the current interferences were measured to be 1.08%, 7.31%, and 1.68%, respectively. Among the introduced compounds, CaCl2 had the greatest impact on the current response. However, there was no statistically significant difference between the current response of the NaCl solution containing any of the three interference substances and that of the pure NaCl solution. The experimental results show that the proposed Na+ detection electrode has good selectivity.
As shown in Figure 8, the current response of the K+ detection electrode in pure KCl solution was measured to be 8.92 µA. The current interferences after the addition of NaCl, CaCl2, and MgCl2 were measured to be 5.23%, 1.01%, and 4.21%, respectively. Similar to the results of the interference experiment of the Na+ detection electrode, there was no statistically significant difference between the current response of the KCl solution containing any of the three interference substances and that of the pure KCl solution. The experimental results show that the proposed K+ detection electrode possesses good selectivity.

3.5. Performance Comparisons

In recent years, the electrochemical techniques for measuring electrolyte concentration have been rapidly increasing. The materials for Cl detection electrode are mostly nano- and micro-structures of carbon, silver, and platinum. Sodium-ion detection electrode materials are mainly manganese oxide–modified carbon and iron sulfate–modified glass carbon. Potassium ion electrode materials include manganese oxide nanorod–modified carbon electrode as well as carrier- and DNA-modified gold electrode. Table 2 compares the detection performances of the proposed detection electrodes and the existing technologies. The proposed detection electrode has characteristics such as the ability to be prepared under relatively simple process conditions, excellent detection sensitivity, and low RSD, and the detection linear range is suitable for the chloride, sodium, and potassium ions concentrations in human serum.

4. Conclusions

In this study, Cl, Na+, and K+ detection electrodes were successfully developed through a simple process to effectively improve the electrolyte detection speed and obtain an appropriate linear detection range, good sensitivity, RSD and recovery rate. Through CV and SWV, the detection data can be obtained in two minutes. The linear detection ranges in the standard samples of Cl, Na+, and K+ were measured to be 25–200 mM, 50–200 mM, and 2–10 mM, with the average RSDs of 0.79%, 1.65%, and 0.47%, respectively, and the average recovery rates of 101%, 100% and 96%, respectively. The proposed detection electrode has characteristics such the ability to be prepared under relatively simple process conditions, excellent detection sensitivity, and low RSD; moreover, the detection linear range is suitable for the Cl, Na+, and K+ concentrations in human serum.

Author Contributions

Conceptualization, G.-J.W. and L.-D.C.; methodology, G.-J.W. and L.-D.C.; validation, G.-J.W. and L.-D.C.; formal analysis, G.-J.W. and W.-J.W.; data curation, L.-D.C.; writing—original draft preparation, G.-J.W.; writing—review and editing, G.-J.W. and W.-J.W.; supervision, G.-J.W.; funding acquisition, G.-J.W. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the Ministry of Science and Technology of Taiwan, grant number MOST-109y-002.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The data that support the findings of this study are available from the corresponding author upon reasonable request.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Berend, K.; van Hulsteijn, L.H.; Gans, R.O. Chloride: The queen of electrolytes? Eur. J. Intern. Med. 2012, 23, 203–211. [Google Scholar] [CrossRef]
  2. Cinti, S.; Fiore, L.; Massoud, R.; Cortese, C.; Moscone, D.; Palleschi, G.; Arduini, F. Low-cost and reagent-free paper-based device to detect chloride ions in serum and sweat. Talanta 2018, 179, 186–192. [Google Scholar] [CrossRef]
  3. Pohl, H.R.; Wheeler, J.S.; Murray, H.E. Sodium and potassium in health and disease. In Interrelations between Essential Metal Ions and Human Diseases; Sigel, A., Sigel, H., Sigel, R., Eds.; Springer: Dordrecht, The Netherlands, 2013; Volume 13, pp. 29–47. [Google Scholar]
  4. Yan, Y.; Shapiro, J.I. The physiological and clinical importance of sodium potassium ATPase in cardiovascular diseases. Curr. Opin. Pharmacol. 2016, 27, 43–49. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  5. Taghizadeh-Behbahani, M.; Hemmateenejad, B.; Shamsipur, M.; Tavassoli, A. A paper-based length of stain analytical device for naked eye (readout-free) detection of cystic fibrosis. Anal. Chim. Acta 2019, 1080, 138–145. [Google Scholar] [CrossRef] [PubMed]
  6. Altunok, I.; Aksel, G.; Eroğlu, S.E. Correlation between sodium, potassium, hemoglobin, hematocrit, and glucose values as measured by a laboratory autoanalyzer and a blood gas analyzer. Am. J. Emerg. Med. 2019, 37, 1048–1053. [Google Scholar] [CrossRef]
  7. Budak, Y.U.; Huysal, K.; Polat, M. Use of a blood gas analyzer and a laboratory autoanalyzer in routine practice to measure electrolytes in intensive care unit patients. BMC Anesthesiol. 2012, 12, 17. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  8. Gibbons, M.; Klim, S.; Mantzaris, A.; Dillon, O.; Kelly, A. How closely do blood gas electrolytes and haemoglobin agree with serum values in adult emergency department patients: An observational study. Emerg. Med. Australas. 2018, 31, 241–246. [Google Scholar] [CrossRef]
  9. Uysal, E.; Acar, Y.A.; Kutur, A.; Cevik, E.; Salman, N.; Tezel, O. How reliable are electrolyte and metabolite results measured by a blood gas analyzer in the ED? Am. J. Emerg. Med. 2016, 34, 419–424. [Google Scholar] [CrossRef]
  10. Yi, H.; Shi, W.; Zhang, Y.; Zhu, X.; Yu, Y.; Wang, X.; Dai, Z.; Lin, Y. Comparison of electrolyte and glucose levels measured by a blood gas analyzer and an automated biochemistry analyzer among hospitalized patients. J. Clin. Lab. Anal. 2020, 34, e23291. [Google Scholar] [CrossRef]
  11. Gonzalez, A.L.; Waddell, L.S. Blood Gas Analyzers. Top. Companion Anim. Med. 2016, 31, 27–34. [Google Scholar] [CrossRef]
  12. Chen, J.; Gorman, M.; Oreilly, B.; Chen, Y. Analytical evaluation of the epoc® point-of-care blood analysis system in car-diopulmonary bypass patients. Clin. Biochem. 2016, 49, 708–712. [Google Scholar] [CrossRef]
  13. I-STAT POCT System. International Institute for Running Medicine. Available online: https://racemedicine.org/wp-content/uploads/2019/08/iSTAT_Presentation.pdf (accessed on 31 January 2021).
  14. Electrolyte Measuring System SPOTCHEMTM EL SE-1520|Operating Manual, ARKRAY, Inc. Available online: https://www.woodleyequipment.com/docs/el_operating_manual.pdf (accessed on 31 January 2021).
  15. Gao, W.; Emaminejad, S.; Nyein, H.Y.Y.; Challa, S.; Chen, K.; Peck, A.; Fahad, H.M.; Ota, H.; Shiraki, H.; Kiriya, D.; et al. Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis. Nat. Cell Biol. 2016, 529, 509–514. [Google Scholar] [CrossRef] [Green Version]
  16. Jin, J.-H.; Kim, J.; Lee, S.; Choi, S.; Park, C.; Min, N. A fully integrated paper-microfluidic electrochemical device for simul-taneous analysis of physiologic blood ions. Sensors 2018, 18, 104. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  17. Parrilla, M.; Ortiz-Gómez, I.; Cánovas, R.; Salinas-Castillo, A.; Cuartero, M.; Crespo, G.A. Wearable Potentiometric Ion Patch for On-Body Electrolyte Monitoring in Sweat: Toward a Validation Strategy to Ensure Physiological Relevance. Anal. Chem. 2019, 91, 8644–8651. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  18. Sempionatto, J.R.; Nakagawa, T.; Pavinatto, A.; Mensah, S.T.; Imani, S.; Mercier, P.; Wang, J. Eyeglasses based wireless electrolyte and metabolite sensor platform. Lab Chip 2017, 17, 1834–1842. [Google Scholar] [CrossRef]
  19. Braiek, M.; Djebbi, M.A.; Chateaux, J.-F.; Jaffrezic-Renault, N. A conductometric sensor for potassium detection in whole blood. Sens. Actuators B Chem. 2016, 235, 27–32. [Google Scholar] [CrossRef]
  20. Day, C.; Søpstad, S.; Ma, H.; Jiang, C.; Nathan, A.; Elliott, S.; Frankl, F.K.; Hutter, T. Impedance-based sensor for potassium ions. Anal. Chim. Acta 2018, 1034, 39–45. [Google Scholar] [CrossRef] [Green Version]
  21. Ahn, M.-S.; Ahmad, R.; Yoo, J.-Y.; Hahn, Y.-B. Synthesis of manganese oxide nanorods and its application for potassium ion sensing in water. J. Colloid Interface Sci. 2018, 516, 364–370. [Google Scholar] [CrossRef]
  22. Chen, Z.; Guo, J.; Zhang, S.; Chen, L. A one-step electrochemical sensor for rapid detection of potassium ion based on structure-switching aptamer. Sens. Actuators B Chem. 2013, 188, 1155–1157. [Google Scholar] [CrossRef]
  23. Cunha-Silva, H.; Arcos-Martinez, M.J. Development of a selective chloride sensing platform using a screen-printed plat-inum electrode. Talanta 2019, 195, 771–777. [Google Scholar] [CrossRef] [PubMed]
  24. Cuartero, M.; Crespo, G.A.; Bakker, E. Ionophore-Based Voltammetric Ion Activity Sensing with Thin Layer Membranes. Anal. Chem. 2016, 88, 1654–1660. [Google Scholar] [CrossRef]
  25. Zhang, M.; Wang, C.; Zhang, Z.; Ye, J.; Fang, P. A novel carbon paste electrode for sensitive, selective and rapid electro-chemical determination of chloride ion based on three-dimensional graphene. Sens. Actuators B Chem. 2019, 299, 126951. [Google Scholar] [CrossRef]
  26. Machini, W.B.; Martin, C.S.; Martinez, M.T.; Teixeira, S.R.; Gomes, H.M.; Teixeira, M.F. Development of an electro-chemical sensor based on nanostructured hausmannite-type manganese oxide for detection of sodium ions. Sens. Actuators B Chem. 2013, 181, 674–680. [Google Scholar] [CrossRef]
  27. Han, T.; Mattinen, U.; Bobacka, J. Improving the Sensitivity of Solid-Contact Ion-Selective Electrodes by Using Coulometric Signal Transduction. ACS Sens. 2019, 4, 900–906. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  28. Kondratyeva, Y.O.; Tolstopjatova, E.G.; Kirsanov, D.O.; Mikhelson, K.N. Chronoamperometric and coulometric analysis with ionophore-based ion-selective electrodes: A modified theory and the potassium ion assay in serum samples. Sens. Actuators B Chem. 2020, 310, 127894. [Google Scholar] [CrossRef]
  29. Hu, J.; Stein, A.; Bühlmann, P. Rational design of all-solid-state ion-selective electrodes and reference electrodes. TrAC Trends Anal. Chem. 2016, 76, 102–114. [Google Scholar] [CrossRef]
  30. El-Rahman, M.K.A.; Zaazaa, H.E.; Abbas, S.S.; El-Zeany, B.; El-Sherif, Z.A.; El-Haddad, D.A. A comparative study of liquid and solid inner contact roxatidine acetate ion-selective electrode membranes. Chin. Chem. Lett. 2015, 26, 714–720. [Google Scholar] [CrossRef]
  31. Van De Velde, L.; D’Angremont, E.; Olthuis, W. Solid contact potassium selective electrodes for biomedical applications—A review. Talanta 2016, 160, 56–65. [Google Scholar] [CrossRef] [PubMed]
  32. Yu, K.; He, N.; Kumar, N.; Wang, N.; Bobacka, J.; Ivaska, A. Electrosynthesized polypyrrole/zeolite composites as solid contact in potassium ion-selective electrode. Electrochim. Acta 2017, 228, 66–75. [Google Scholar] [CrossRef]
  33. Xu, K.; Cuartero, M.; Crespo, G.A. Lowering the limit of detection of ion-selective membranes backside contacted with a film of poly(3-octylthiophene). Sens. Actuators B Chem. 2019, 297, 126781. [Google Scholar] [CrossRef]
  34. Emaminejad, S.; Gao, W.; Wu, E.; Davies, Z.A.; Nyein, H.Y.Y.; Challa, S.; Ryan, S.P.; Fahad, H.M.; Chen, K.; Shahpar, Z.; et al. Autonomous sweat extraction and analysis applied to cystic fibrosis and glucose moni-toring using a fully integrated wearable platform. Proc. Natl. Acad. Sci. USA 2017, 114, 4625–4630. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  35. Liu, Y.; Liu, Y.; Meng, Z.; Qin, Y.; Jiang, D.; Xi, K.; Wang, P. Thiol-functionalized reduced graphene oxide as self-assembled ion-to-electron transducer for durable solid-contact ion-selective electrodes. Talanta 2020, 208, 120374. [Google Scholar] [CrossRef] [PubMed]
  36. Paczosa-Bator, B.; Cabaj, L.; Pięk, M.; Piech, R.; Kubiak, W.W. Carbon-Supported Platinum Nanoparticle Solid-State Ion Selective Electrodes for the Determination of Potassium. Anal. Lett. 2015, 48, 2773–2785. [Google Scholar] [CrossRef]
  37. Criscuolo, F.; Taurino, I.; Stradolini, F.; Carrara, S.; De Micheli, G. Highly-stable Li+ ion-selective electrodes based on noble metal nanostructured layers as solid-contacts. Anal. Chim. Acta 2018, 1027, 22–32. [Google Scholar] [CrossRef]
  38. Xu, J.; Li, F.; Tian, C.; Song, Z.; An, Q.; Wang, J.; Han, D.; Niu, L. Tubular Au-TTF solid contact layer synthesized in a mi-crofluidic device improving electrochemical behaviors of paper-based potassium potentiometric sensors. Electrochim. Acta 2019, 322, 134683. [Google Scholar] [CrossRef]
  39. Paczosa-Bator, B.; Pięk, M.; Piech, R. Application of Nanostructured TCNQ to Potentiometric Ion-Selective K+and Na+Electrodes. Anal. Chem. 2015, 87, 1718–1725. [Google Scholar] [CrossRef]
  40. Mohammadtaheri, M.; Ramanathan, R.; Bansal, V. Emerging applications of metal-TCNQ based organic semiconductor charge transfer complexes for catalysis. Catal. Today 2016, 278, 319–329. [Google Scholar] [CrossRef]
  41. Yuan, B.; Xu, C.; Zhang, R.; Lv, D.; Li, S.; Zhang, D.; Liu, L.; Fernandez, C. Glassy carbon electrode modified with 7,7,8,8-tetracyanoquinodimethane and graphene oxide triggered a synergistic effect: Low-potential amperometric detection of reduced glutathione. Biosens. Bioelectron. 2017, 96, 1–7. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  42. Suherman, A.L.; Lin, M.; Rasche, B.; Compton, R.G. Introducing insertive stripping voltammetry: Electrochemical deter-mination of sodium ions using an iron (III) phosphate-modified electrode. ACS Sens. 2020, 5, 519–526. [Google Scholar] [CrossRef]
  43. Kumbhat, S.; Singh, U. A potassium-selective electrochemical sensor based on crown-ether functionalized self assembled monolayer. J. Electroanal. Chem. 2018, 809, 31–35. [Google Scholar] [CrossRef]
  44. Chai, H.; Ma, X.; Meng, F.; Mei, Q.; Tang, Y.; Miao, P. Electrochemical aptasensor based on a potassium ion-triggered DNA conformation transition and self-assembly on an electrode. New J. Chem. 2019, 43, 7928–7931. [Google Scholar] [CrossRef]
Scheme 1. Fabrication process of the chloride, potassium, and sodium ions detection electrodes: (A) sputtering of Au thin-film on a polyethylene terephthalate (PET) membrane; (B) packaging; (C) dropping of 7,7,8,8-tetracyanoquinodimethane (TCNQ) + ion-selective membrane (ISM) mixture solution (potassium and sodium ions detection electrodes).
Scheme 1. Fabrication process of the chloride, potassium, and sodium ions detection electrodes: (A) sputtering of Au thin-film on a polyethylene terephthalate (PET) membrane; (B) packaging; (C) dropping of 7,7,8,8-tetracyanoquinodimethane (TCNQ) + ion-selective membrane (ISM) mixture solution (potassium and sodium ions detection electrodes).
Biosensors 11 00109 sch001
Figure 1. The analysis system: (A) experimental setup; (B) electrodes for the experiment.
Figure 1. The analysis system: (A) experimental setup; (B) electrodes for the experiment.
Biosensors 11 00109 g001
Figure 2. Cyclic voltammograms for TCNQ solvent selection and TCNQ/ISM volume ratio optimization: (A,B) cyclic voltammograms for Na+ (140 mM NaCl) and K+ (5 mM KCl) detections, respectively; (C,D) cyclic voltammograms for Na+ and K+ detections with different TCNQ/ISM volume ratios, respectively.
Figure 2. Cyclic voltammograms for TCNQ solvent selection and TCNQ/ISM volume ratio optimization: (A,B) cyclic voltammograms for Na+ (140 mM NaCl) and K+ (5 mM KCl) detections, respectively; (C,D) cyclic voltammograms for Na+ and K+ detections with different TCNQ/ISM volume ratios, respectively.
Biosensors 11 00109 g002
Figure 3. Scanning electron microscopy (SEM) images of electrode surface: (A) TCNQ/ion-selective membrane (ISM) deposited electrode (top view); (B) cross-sectional view.
Figure 3. Scanning electron microscopy (SEM) images of electrode surface: (A) TCNQ/ion-selective membrane (ISM) deposited electrode (top view); (B) cross-sectional view.
Biosensors 11 00109 g003
Figure 4. Electrochemical characteristics of the proposed electrode for electrolyte detection: (A1,B1,C1) cyclic voltammograms of the proposed different electrodes at scan rates ranging from 100 to 500 mV/s in 50 mM KCl, 140 mM NaCl, and 1 mM KCl solutions, respectively; (A2,B2,C2) the corresponding Randles–Sevcik curves of the cyclic voltammograms in (A1,B1,C1), respectively.
Figure 4. Electrochemical characteristics of the proposed electrode for electrolyte detection: (A1,B1,C1) cyclic voltammograms of the proposed different electrodes at scan rates ranging from 100 to 500 mV/s in 50 mM KCl, 140 mM NaCl, and 1 mM KCl solutions, respectively; (A2,B2,C2) the corresponding Randles–Sevcik curves of the cyclic voltammograms in (A1,B1,C1), respectively.
Biosensors 11 00109 g004aBiosensors 11 00109 g004b
Figure 5. Cl detection using cyclic voltammetry (CV): (A) cyclic voltammograms (100 mV/s) for various Cl concentrations (25–200 mM); (B) linear calibration curve representing the relation between the stable oxidization peak current and concentration; scanning range: 0–1.6 V, N = 3.
Figure 5. Cl detection using cyclic voltammetry (CV): (A) cyclic voltammograms (100 mV/s) for various Cl concentrations (25–200 mM); (B) linear calibration curve representing the relation between the stable oxidization peak current and concentration; scanning range: 0–1.6 V, N = 3.
Biosensors 11 00109 g005
Figure 6. Na+ detection using square-wave voltammetry (SWV): (A) cyclic voltammograms (100 mV/s) for various Na+ concentrations (50–200 mM); (B) linear calibration curve representing the relation between the stable oxidization peak current and concentration; scanning range: 0–0.5 V, pulse height 100 mV, pulse width 200 ms, step height 20 mV, N = 3.
Figure 6. Na+ detection using square-wave voltammetry (SWV): (A) cyclic voltammograms (100 mV/s) for various Na+ concentrations (50–200 mM); (B) linear calibration curve representing the relation between the stable oxidization peak current and concentration; scanning range: 0–0.5 V, pulse height 100 mV, pulse width 200 ms, step height 20 mV, N = 3.
Biosensors 11 00109 g006
Figure 7. SWV detection results of K+ with various concentrations: (A) cyclic voltammograms (100 mV/s) for various K+ concentrations (2–10 mM); (B) quadratic calibration curve representing the relation between the stable oxidization peak current and concentration; scanning range: 0.2–0.6 V, pulse height 100 mV, pulse width 200 ms, step height 20 mV, N = 3.
Figure 7. SWV detection results of K+ with various concentrations: (A) cyclic voltammograms (100 mV/s) for various K+ concentrations (2–10 mM); (B) quadratic calibration curve representing the relation between the stable oxidization peak current and concentration; scanning range: 0.2–0.6 V, pulse height 100 mV, pulse width 200 ms, step height 20 mV, N = 3.
Biosensors 11 00109 g007
Figure 8. Interferences detection. Cl detection: 100 mM KCl, 100 mM NaCl, 50 mM CaCl2, and 50 mM MgCl2; Na+ detection: 140 mM NaCl separately mixed with 4 mM KCl, 2.5 mM CaCl2, and 1 mM MgCl2; K+ detection: 4 mM KCl separately mixed with 140 mM NaCl, 2.5 mM CaCl2, and 1 mM MgCl2.
Figure 8. Interferences detection. Cl detection: 100 mM KCl, 100 mM NaCl, 50 mM CaCl2, and 50 mM MgCl2; Na+ detection: 140 mM NaCl separately mixed with 4 mM KCl, 2.5 mM CaCl2, and 1 mM MgCl2; K+ detection: 4 mM KCl separately mixed with 140 mM NaCl, 2.5 mM CaCl2, and 1 mM MgCl2.
Biosensors 11 00109 g008
Table 1. Compositions and contents of potassium and sodium-ion-selective membrane.
Table 1. Compositions and contents of potassium and sodium-ion-selective membrane.
ReagentsTarget Ion
K+Na+
Potassium ionophore I14 mg0
4-tert-Butylcalix[4]arenetetraacetic acid tetraethyl ester09.9 mg
Potassium tetrakis (4-chlorophenyl)borate3 mg2.5 mg
Polyvinyl chloride (high-molecular-weight)328 mg329 mg
2-Nitrophenyl octyl ether0.63 mL0.63 mL
Table 2. Performance comparison of the proposed ion detection electrode and existing technologies.
Table 2. Performance comparison of the proposed ion detection electrode and existing technologies.
IonsElectrodesSensitivity
(μA·mM−1)
Linear Range
(mM)
RSD
(%)
Reference
ClAg/SPE1.9810–200˂10[2]
AgGNP/SPE0.00001930.0005–0.092.22[22]
SPPtE−24.1470.76–1505.8 [23]
3D-GN/CPE0.0440.5–1000-[25]
This study19.125–2000.79
Na+Mn3O4/CPE19.50.02–0.2-[26]
FePO4/GCE0.01525–250-[42]
This study0.12650–2001.65
K+MnO2 nanorods/GCE0.00040.002–0.09-[21]
4-aminobenzo-18-Crown-6 ether/11-Mercaptoundecanoic acid/Au-0.5–7-[43]
DNA/Au11310−7–100˂5[44]
This study0.56–7.982–100.47
Note: Normal Cl, Na+, K+ concentrations in human blood: 98–106 mM, 135–145 mM, and 3.5–5 mM, respectively.
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Chen, L.-D.; Wang, W.-J.; Wang, G.-J. Electrochemical Detection of Electrolytes Using a Solid-State Ion-Selective Electrode of Single-Piece Type Membrane. Biosensors 2021, 11, 109. https://0-doi-org.brum.beds.ac.uk/10.3390/bios11040109

AMA Style

Chen L-D, Wang W-J, Wang G-J. Electrochemical Detection of Electrolytes Using a Solid-State Ion-Selective Electrode of Single-Piece Type Membrane. Biosensors. 2021; 11(4):109. https://0-doi-org.brum.beds.ac.uk/10.3390/bios11040109

Chicago/Turabian Style

Chen, Li-Da, Wei-Jhen Wang, and Gou-Jen Wang. 2021. "Electrochemical Detection of Electrolytes Using a Solid-State Ion-Selective Electrode of Single-Piece Type Membrane" Biosensors 11, no. 4: 109. https://0-doi-org.brum.beds.ac.uk/10.3390/bios11040109

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop