Review

Laser Analgesia Associated with Restorative Dental Care: A Systematic Review of the Rationale, Techniques, and Energy Dose Considerations

1
Leicester School of Pharmacy, De Montfort University, Gateway House, Leicester LE1 9BH, UK
2
School of Dental Medicine, University of Nevada Las Vegas, Las Vegas, NV 89154, USA
*
Author to whom correspondence should be addressed.
Submission received: 4 September 2020 / Revised: 30 October 2020 / Accepted: 10 November 2020 / Published: 12 November 2020
(This article belongs to the Special Issue Lasers in Dentistry: Hard and Soft Tissues)

Abstract

:
It is a common experience amongst laser dentists and patients that mid-IR wavelength application in cavity preparation may be achieved without causing any associated pain. The erbium family of lasers (Er,Cr:YSGG 2780 nm and Er:YAG 2940 nm) are frequently used without employing injectable local anesthesia as an adjunct: the phenomenon arising from the application of these devices is known as laser analgesia. This review seeks to apply a systematic approach to the examination of appropriate published studies but also to highlight the need for much more structured clinical investigations that consolidate photonic dose and methodology. A search of published data using PRISMA criteria was carried out to examine clinical trials into laser analgesia in conjunction with restorative dentistry, applying inclusion and exclusion criteria. From this, 10 published articles were selected for analysis. Suitability assessment was carried out, using a modified Cochrane risk of bias methodology. In 8/10 of the included studies, laser-induced analgesia is claimed to be better and effective, while in 2/10 of the studies, no difference was exhibited compared to the control group. Statistical analysis of three split mouth studies concluded that only one of these investigations reviewed demonstrated a significant analgesic effect for laser treatment while the other two did not support this observation. From this data, it is inconclusive to assess the predictability of laser analgesia in cavity preparation. A possible rationale and laser operating parametry has been discussed. Successful implementation of this treatment modality remains technique sensitive and subject to further investigation.

1. Introduction

During the development of laser applications in dentistry, several wavelengths have been used in an attempt to perform caries removal and cavity preparation. However, the first devices (i.e., CO2 10,600 nm and Nd:YAG 1064 nm) had insufficient ability to penetrate into dental hard tissues and produced major thermal side-effects, such as melting, carbonisation, cracks in the surrounding dental tissues, and increases in pulpal temperature [1,2,3,4]. However, the introduction of erbium family lasers (Er,Cr:YSGG 2780 nm and Er:YAG 2940 nm) has now allowed a more effective removal of oral hard tissues (enamel, dentine, cementum, and bone) and this major benefit is coupled with a low risk of thermal damage.
The use of erbium lasers has been reported to be a more comfortable method of cavity preparation than a bur, with a significantly reduced requirement for local anaesthesia [5,6,7,8]. Cozean et al. demonstrated that the Er:YAG wavelength can be efficiently used for painless caries removal and cavity preparation, and that less than 2% of the patients involved required local anaesthesia [9]. Hibst et al. [6] found that only 6% of 103 patients (206 cavities) required injected local anaesthesia and the vast majority rated the laser preparation as “more comfortable” when compared to bur tissue removal.
During the last two decades and in order to address and suppress the potential for pain associated with dental treatment, every dental practitioner has been using injectable local anaesthesia; in itself, such adjunctive treatment might represent a further source of anxiety and dental fear, with the ability to trigger further phobic reactions. Possible alternatives that have been attempted or used to bypass this anxiety have been hypnotic analgesia and conscious sedation with nitrous oxide, although it is not possible to suppress pain perception up to a point where the patient may experience a painless restorative procedure. Topical anaesthetics and electrically induced anaesthesia have also been tested in several investigations, albeit with contrasting results. The use of intra-venous sedation or general anaesthesia in restorative dentistry is considered to be reserved for patients who are completely uncooperative; thus, it is not usually included amongst these alternatives [10,11,12,13,14].
Laser analgesia is a non-invasive, non-destructive, and non-thermal bio-modulating technique with the ability to reduce or suppress painful sensations; it is obtainable in a low-energy-level irradiation form. The causes of this effect are not completely clear, even though several theories have been proposed. Possible explanations are represented by the photo-acoustic effect and its actions within the gate control pathway, direct and indirect influences of laser energy on nerves and nociceptors, modifications of the Na+-K+ pump systems, and the bio-resonance and biochemical modifications that laser energy is able to induce. Techniques are available that result in a laser-induced analgesic effect, and methods for computing the energy dose. Notwithstanding, it is not a profound anaesthesia since it is unable to suppress all sensations. The aim of this laser application is the suppression of nerve and nociceptor activities, processes creating a lack of pain perception and diminishing dental pulp and soft tissue reactivity [15,16,17].
It is appropriate to record that laser–tissue interaction has a potential for thermal rise, through the very nature of photothermal energy transfer; it is necessary for the clinician to apply precise laser and air/water co-axial operating parameters in order to avoid any unwanted dental and pulpal damage, both during the induction of laser analgesia and consequent cavity preparation [18,19].
Several wavelengths have been investigated in order to obtain laser analgesia: visible (445–670 nm), diodes (810–980 nm), Nd:YAG (1064 nm), erbium family lasers (Er,Cr:YSGG 2780 nm and Er:YAG 2940 nm), and CO2 (10,600 nm).
Reflecting elevations in fear and perceived fear in relation to dental care, patients are therefore prone to seek alternative therapeutic experiences and many of them prefer to avoid traditional restorative treatments, since there is a strong association between their fear level and the use of the traditional turbine and bur, along with its associated noise, smell, and vibrations [20,21,22,23].
According to previous reports, in relation to receiving restorative dental treatment, the percentage of anxious subjects appears to be on the increase. Data relating to US and European investigation can be helpful in providing some context to the prevalence of odontophobia; in 1983, it was estimated to be in the range of 10–12% [21]; later, in young (child and adolescent) patients, it was determined to range from 6% to as much as 52% in 1992 [21]; and finally, it attained a figure of about 50% in 2009 [22].
The same trend appears to occur amongst odontophobic patients; this group represents the most unwilling to cooperate and difficult to treat subjects, estimated to be 3–5% in 1983 [21], up to the most recent data available (2004) of 4–16% of adults, and 7–20% of children [23]. Indeed, we may consider that there is an increasing trend of anxiety and phobia related to dental treatments.
There is considerable dichotomy between the predominately anecdotal and patient-centred approach to laser analgesia, compared to the constraints of rigorous randomised clinical trials and therein lies a good deal of the challenge to establish a basis for investigation or to successfully examine published data. The introduction of the first laser within dentistry in 1989 (American Dental Laser DLase 300 Nd:YAG 1064 nm) was heralded by considerable promotion, albeit based upon anecdote and unsubstantiated claims. Early published reports [24,25,26,27,28] were evidently supported by the manufacturer and unfortunately, many of these early studies formed part of the product promotion. A significant randomized trial in 1995 [29] was unsuccessful in determining predictable and statistically significant advantages of laser analgesia as adjunct to tooth cavity preparation, despite recording patient acceptance of the technique over needle local anaesthesia.
In order to seek further clarity, a systematic review was carried out to examine the effectiveness of laser-induced analgesia in restorative dentistry. As an example, one approach in treating patients requiring restorative dentistry is detailed as an example of the approach, and is detailed in the discussion.

2. Materials and Methods

2.1. Search Strategy

An electronic search was conducted relating to laser analgesia applications in all fields of dentistry from 17 July to 19 July 2020. Databases used were PubMed, Cochrane, and Google Scholar with the following MeSH terms, keywords, and their combinations: Laser AND (analgesia OR pain) AND (caries OR cavity preparation).
Duplicate articles were excluded and the initial number of 103 was reduced to 79.
After applying the additional filters (published within the last 15 years, only clinical trials in humans, and only English language reports), the preliminary number of 79 articles was reduced to 16.
Titles and abstracts of the above articles were independently screened by two reviewers via application of the following criteria. In case of any disagreements arising, these were satisfactorily resolved by discussions.
Inclusion criteria:
  • Laser used;
  • At least 10 patients per group;
  • Only clinical trials; and
  • Laser use applicable to dental procedure to be carried out.
Exclusion criteria:
  • Duplicates or studies with the same ethical approval number;
  • Pain control studies where non-laser analgesia was employed;
  • Less than 10 patients per group; and
  • No clinical trials or study protocols or pilot studies.
After screening and implementation of the eligibility criteria, a total of 9 articles were retained.
In accordance with the PRISMA statement [30], details of the selection criteria are presented in Figure 1.

2.2. Data Extraction

Having reached a consensus regarding the selection of included articles, the four reviewers involved subsequently extracted data regarding:
  • Citation (first author and publication year);
  • Type of study/number of patients;
  • Test/control group;
  • Aim/approach;
  • Laser parameters applied; and
  • Outcome.

2.3. Quality Assessment

Subsequent to data extraction, articles were further evaluated by assessing their risk of bias assessment. The Cochrane risk of bias tool [31] was modified according to the requirements of this systematic review.
The risk of bias was determined according to the number of “yes” or “no” responses to the parameters provided below, which were allocated to each study:
  • Randomization;
  • Sample size calculation and required number included;
  • Baseline situation similar;
  • Blinding;
  • Parameters of laser use described appropriately and calculations correct;
  • Power meter used;
  • Numerical results available (statistics);
  • Outcome data complete; and
  • Correct interpretation of data.
The classification was performed according to the total number of “yes” answers to the above questions. For the current study, the degree of bias was computed according to the score limits provided below:
  • High risk: 0–3;
  • Moderate risk: 4–6; and
  • Low risk: 7–9.

3. Results

3.1. Primary Outcome

The primary goal of this review was to explore the effectiveness of laser-induced analgesia in clinical dental practice.

3.2. Data Presentation

The results of the included studies are presented in Table 1.

3.3. Quality Assessment Presentation

The risk of bias evaluation is presented in Table 2.
In total, 3/10 articles showed a low risk of bias:
  • One article [10] scoring 8/9; and
  • Two articles [32,33] scoring 7/9.
In total, 4/10 articles showed a medium risk of bias:
  • Three articles [35,36,39] scoring 5/9; and
  • One article [34] scoring 4/9.
In total, 3/10 articles showed a high risk of bias:
  • Three articles [37,38,40] scoring 3/10.
It can be seen that only four articles [10,33,35] had sufficient and reproducible description of the laser protocol applied, while only in one study [10] was a power meter used, and in two others [10,32], sample size calculation was done before the clinical trial and subsequently the required number of patients was included.

3.4. Analysis of Data

According to the type of study, the included articles can be divided into two categories and subsequently critically appraised regarding the primary outcome:
  • In split-mouth RCTs, 3/5 [33,36,39] studies showed that the laser group achieved statistically significant better results compared to the control group while 2/5 [32,35] studies showed no difference between these groups.
  • In the parallel RCT study [35], results showed a significant difference in pain perception regarding the laser group.
  • In clinical trials, the range of positive treatment outcomes was between 66.7% and 89.5% in the remaining five studies [10,34,37,38,40].
Statistical analysis was performed on a combination of three split-mouth studies in which it was possible to classify the responses as pain experience or no pain, and these data are provided in Table 3. Firstly, the Cochran–Mantel–Haenszel (CMH) test was applied to the complete dataset in order to determine if the relative proportions of pain or no pain experienced following the laser treatment and corresponding control groups are consistent or not within all 3 sets of 2 × 2 contingency tables. Hence, the null hypothesis was that there is a consistent difference in these proportions across all three studies explored here. However, the CMH test revealed a high level of inconsistency between the results acquired in these studies (p = 6.04 × 10–5).
Secondary 2 × 2 contingency table analysis of each study dataset independently showed that for the Sarnadi [32] and Chan [10] studies, there was no significant association between the proportions of participants experiencing pain above a selected threshold and the treatments applied (laser vs. control), χ2 test (with Yates’ correction for continuity) p values of 0.39 and 0.51 (ns), respectively. However, for the Liu study, this p value was 2.36 × 10−6, and data therein clearly demonstrated that the laser system applied exerted a major analgesic effect, which was very highly statistically significant. Therefore, it may be concluded that only one of these investigations reviewed demonstrated a significant analgesic effect for laser treatment, but two other corresponding studies performed did not support this observation.

4. Discussion

The systematic review offers some insight to the approach when employing laser-assisted analgesic techniques. Considering the impact of the object of the technique among a large potential population of patients, there is some regret that the quantity and specificity of objective investigations remains small and not without some risk of bias. It is difficult to understand the lack of quality investigations into technique, operating parameters, and predictability relative to patient acceptance, when it is beyond question that the impact on delivering painless dentistry remains an important aspect of gaining patient confidence.
The search criteria in this literature review allowed some limited consistency in results to enable possible techniques and laser operating parameters to be explored. It is safe to reflect on the wide range of techniques and laser dose in terms of energy per pulse, frequency, and length of application time.
From the studies examined, the ones that offered detail in terms of technique and laser irradiance values have either adopted a dedicated pre-operative analgesia stage [33,35,36,37,38] or utilised effective but tolerable laser ablation power and draw comparisons with a similar treatment using a rotary bur [10,32,38,39,40]. Two of the studies analysed utilised NIR laser wavelengths to achieve laser analgesia alone [33,38]. Of the Erbium family, five studies used Er:YAG 2940 nm [10,32,38,39,40], two used the Er,Cr:YSGG 2780 nm wavelength [34,40], and one study used both wavelengths [37].
Of those studies with statistical analysis, five [10,32,33,36,39] offered a statistically significant benefit of laser use over alternative modalities, whereas another was equivocal in outcome with no statistical significance [35]. In the Poli et al. study [34], utilising both methods, it was discovered that amongst 30 patients exposed to laser-assisted restorative dentistry, 80% perceived no pain, and none required local anaesthesia; the use of pain scales to assess laser benefit provided a general level of acceptance over conventional treatment alternatives. Although free running pulsed (FRP) lasers exert significant effects on pain, researchers have yet to reach a definitive conclusion over the mechanisms of such effects.
Only recently has a protocol focused on achieving a predictable onset of analgesia using an erbium family laser system been proposed [34,41].
Two specific techniques have been described in order to attain a reliable level of analgesia in restorative dentistry [34,41]. The “rabbit” technique, also known as the “hare” technique, assumes the use of high energy settings at the beginning of treatment. Key to any success with either technique is the specificity of applied fluence and to achieve this, and one possible approach is the use of a “spacer” to preserve the exact “tip to tissue” distance, as illustrated in Figure 2, Figure 3 and Figure 4. The parameters set are then immediately able to create a tooth cavity, but initially, the handpiece is kept at a distance of 6–10 mm from the target, in order to obtain a de-focused beam and a low fluence that is not perceived as pain (Figure 2 and Figure 3).
The tip is moved all around the tooth neck, and following several tens of seconds (up to two minutes in total), the tip is placed closer to the tooth surface, and when it is at 0.5–1 mm from that site, the ablation commences. If the patient then feels any discomfort, the handpiece can be moved aside and the analgesia phase starts again. When the dentine is penetrated, the parameters are lowered, or the tip is placed further away to reduce energy density.
The second technique is known as the “turtle” (or “tortoise”) strategy. Differently from the previous approach, at first, the energy and power settings are low to create a low level of irradiation. The tip is then placed at an ablative distance (0.5–1 mm) from the surface, but the formation of the cavity does not occur until the energy settings increase and overcome the ablation threshold of the specific tissue targeted. When the dentin is reached, the parameters are then modified accordingly (Figure 4).
With both techniques, primarily, the irradiation process is always considered a low-level laser therapy (LLLT). With the rabbit technique, however, the risk to have a greater sensitivity is higher, since the energy is immediately set at ablative values and hence, if the distance is reduced, the impact of the beam may be perceived by the patient. On the contrary, with the “turtle” technique, it is easier to obtain the laser analgesia, and the risk of discomfort is generally lower [41]. It is also possible to combine the two different techniques to take advantage of both and also increase the likelihood of attaining a painless restorative treatment [34].
The settings necessary to induce laser analgesia have already been investigated through published studies [34,41]; however, further research is required in order to determine the exact amount of medium-IR energy, and the correct parameters necessary to produce a more predictable protocol and a reliable laser analgesia-based treatment regimen. In the absence of more data, it might be appropriate to consider the general dose range in the treatment of dento-alveolar pain associated with tissue injury and orthodontic pain. In this way, according to Bjordal et al. [42], the optimal dose on acute pain in the first 72 h following the injury should be 7.5 J/cm2. Other authors proposed greater levels of energy, such as Lizarelli [43], who indicated a dose within the range 5–20 J/cm2 for severe pain. A fluence of 35 J/cm2 was suggested by another group to reduce orthodontic pain, since a level of only 5 J/cm2 was not found to be effective [43]. In the current available literature, similar protocols reported conflicting results [17,44].
In examining possible mechanisms that may contribute to the positive patient experience during laser therapy, these are most likely additive in nature rather than being mutually exclusive.
According to the bio-psycho-social model of pain [45], many different aspects can influence pain perception. Environment, pain behaviour, psychological distress, attitudes and beliefs, pain modulation, health conditions, nerve and nociceptor activity, and/or neuropathy are all possible elements that can interact with each other. Therefore, it seems likely that lasers have different effects on one or many of these factors [46]. The limited pulp temperature increase [47,48,49] appears to represent a further element that limits the pain experienced, both during and after the applied treatment.
FRP laser systems reduce the annoyance factor, which is a combination of the pressure applied to the tooth, the vibrations and noise recorded through the bones of the skull, and the heat and smell generated at the interface between the tooth and cutting instrument.
Numerous host response elements seem to be combined and involved in modulation and suppression of nociception. All of them are able to reduce, modify, or block nerve impulse transmission and conduction from several nociceptors. These morphological and physiological neuronal changes are, however, reversible after photobiomodulation [50]. Points of consideration may be summarised as follows:
  • The gate control pathway [51,52];
  • Indirect influence on the nervous system [53,54,55,56,57];
  • Direct inhibition of the nervous system [15,16,35,58,59,60];
  • Effects of laser energy on oral environment biochemistry [17,61,62,63,64,65,66,67,68,69,70,71,72,73];
  • Photo-acoustic effect of pulsed lasers [46];
  • Techniques to induce laser analgesia [34,41]; and
  • Energy dose considerations [17,34,41,43,44,61].

5. Conclusions

From this review, it can be pointed out that laser-induced analgesia can be considered a possibly effective alternative treatment modality in affecting pain perception. An important role is also attributed to the altered patient responses due to the limited invasiveness of laser treatments and consequently to the minor level of anxiety.
Although none of the studies declared any adverse effect, the lack of consistent laser operating parameters and inconsistent statistical significane regarding outcomes over conventional analgesic and anaesthetic techniques does not allow firm conclusions to be formed as to which laser wavelength is superior or what parameters might be employed to achieve predictability. The inconsistencies of non-standardised study design, varying parametry or lack of essential laser operating values, and lack of statistical analysis together have failed to underline a strong protocol of how the mid-IR wavelengths may be applied in achieving laser analgesia.
Above all, laser analgesia remains a profoundly technique-sensitive modality to achieve success. As such, this application can be considered safe as long as the correct parameters are applied, although it is clear that more clinical studies with standardised protocol should be performed in order to deeply explore and understand this procedure.

Author Contributions

Conceptualization, R.P., S.P.; methodology, R.P.; validation, S.P.; formal analysis, S.P., E.A., V.M., M.G.; investigation, R.P.; data curation, R.P., S.P., E.L.; statistical analysis, M.G.; writing-original draft preparation, R.P.; writing-review and editing, S.P., E.A., V.M., M.G. and E.L.; visualization, S.P., E.L.; supervision, S.P., E.L. and M.G.; project administration, S.P. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Tavares de Oliveira, M.; Galvão Arrais, C.A.; Aranha, A.C.; de Paula Eduardo, C.; Miyake, K.; Allen Rueggeberg, F.; Giannini, M. Micromorphology of resin-dentin interfaces using one-bottle etch & rinse and self-etching adhesive systems on laser-treated dentin surfaces: A confocal laser scanning microscope analysis. Lasers Surg. Med. 2010, 42, 662–670. [Google Scholar] [CrossRef]
  2. Frentzen, M.; Koort, H.J. Lasers in dentistry: New possibilities with advancing laser technology? Int. Dent. J. 1990, 40, 323–332. [Google Scholar] [PubMed]
  3. Wigdor, H.; Abt, E.; Asrafi, S.; Walsh, J.T. The effect of lasers on dental hard tissues. J. Am. Dent. Assoc. 1993, 124, 65–70. [Google Scholar] [CrossRef] [PubMed]
  4. Israel, M.; Cobb, C.M.; Rossmann, J.A.; Spencer, P. The effects of CO2, Nd:YAG and Er:YAG lasers with and without surface coolant on tooth root surfaces. An in vitro study. J. Clin. Periodontal. 1997, 24, 595–602. [Google Scholar] [CrossRef] [PubMed]
  5. Ceballos, L.; Osorio, R.; Toledo, M.; Marshall, G.W. Microleakage of composite restorations after acid or Er-YAG laser cavity treatments. Dent. Mater. 2001, 17, 340–346. [Google Scholar] [CrossRef]
  6. Keller, U.; Hibst, R.; Geurtsen, W.; Schilke, R.; Heidemann, D.; Klaiber, B.; Raab, W.H.M. Erbium:YAG laser application in caries therapy. Evaluation of patient perception and acceptance. J. Dent. 1998, 26, 649–656. [Google Scholar] [CrossRef]
  7. Pelagalli, J.; Gimbel, C.B.; Hansen, R.T.; Swett, A.; Winn, D.W. Investigational study of the use of Er:YAG laser versus dental drill for caries removal and cavity preparation—Phase I. J. Clin. Laser Med. Surg. 1997, 15, 109–115. [Google Scholar] [CrossRef]
  8. Matsumoto, K.; Nakamura, Y.; Mazeki, K.; Kimura, Y. Clinical dental application of Er:YAG laser for class V cavity preparation. J. Clin. Laser Med. Surg. 1996, 14, 123–127. [Google Scholar] [CrossRef]
  9. Cozean, C.; Arcoria, C.J.; Pelagalli, J.; Powell, G.L. Dentistry for the 21st century? Erbium:YAG laser for teeth. J. Am. Dent. Assoc. 1997, 128, 1080–1087. [Google Scholar] [CrossRef]
  10. Chan, A.; Armati, P.; Moorthy, A.P. Pulsed Nd:YAG laser induces pulpal analgesia: A randomized clinical trial. J. Dent. Res. 2012, 91, 79S–84S. [Google Scholar] [CrossRef]
  11. Holroyd, I. Conscious sedation in pediatric dentistry. A short review of the current UK guidelines and the technique of inhalation sedation with nitrous oxide. Pediatr. Anaesth. 2008, 18, 13–17. [Google Scholar] [CrossRef] [PubMed]
  12. Houpt, M.I.; Limb, R.; Livingstone, R.L. Clinical effects of nitrous oxide conscious sedation in children. Pediatr. Dent. 2004, 26, 29–36. [Google Scholar] [PubMed]
  13. Ryding, H.A.; Murphy, H.J. Use of Nitrous Oxide and Oxigen for Conscious Sedation to Manage Pain and Anxiety. J. Can. Dent. Assoc. 2007, 73, 7–11. [Google Scholar] [PubMed]
  14. Zacny, J.P.; Hurst, R.J.; Graham, L.; Janiszewski, D.J. Preoperative dental anxiety and mood changes during nitrous oxide inhalation. J. Am. Dent. Assoc. 2002, 133, 82–88. [Google Scholar] [CrossRef] [PubMed]
  15. Chow, R.T.; David, M.A.; Armati, P.J. 830 nm laser irradiation induces varicosity formation, reduces mitochondrial membrane potential and blocks fast axonal flow in small and medium diameter rat dorsal root ganglion neurons: Implications for the analgesic effects of 830 nm laser. J. Periph. Nerv. Syst. 2007, 12, 28–39. [Google Scholar] [CrossRef]
  16. Chow, R.T.; Armati, P.J. Photobiomodulation: Implications for anesthesia and pain relief. Photomed. Laser Surg. 2016, 34, 599–609. [Google Scholar] [CrossRef]
  17. Cronshaw, M.; Parker, S.; Arany, P. Feeling the Heat: Evolutionary and microbial basis for the analgesic mechanisms of photobiomodulation therapy. Photobiomodul. Photomed. Laser Surg. 2019, 37, 517–526. [Google Scholar] [CrossRef]
  18. Colucci, V.; Lucisano Botelho do Amaral, F.; Pécora, J.D.; Palma-Dibb, R.; Corona, S. Water flow on erbium:yttrium–aluminum–garnet laser irradiation: Effects on dental tissues. Lasers Med. Sci. 2009, 24, 811–818. [Google Scholar] [CrossRef]
  19. Bello-Silva, M.S.; Wehner, M.; de Paula Eduardo, C.; Lampert, F.; Poprawe, R.; Hermans, M.; Esteves-Oliveiraet, M. Precise ablation of dental hard tissues with ultra-short pulsed lasers. Preliminary exploratory investigation on adequate laser parameters. Lasers Med. Sci. 2013, 28, 171–184. [Google Scholar] [CrossRef]
  20. Bedi, R.; Sutcliff, P.; Donnan, P.T.; McConnachie, J. The prevalence of dental anxiety in a group of 13- and 14-year-old Scottish children. Int. J. Paediatr. Dent. 1992, 2, 17–24. [Google Scholar] [CrossRef]
  21. Ayer, W.A.; Domoto, P.K.; Gale, E.N.; Joy, E.D., Jr.; Melamed, B.G. Overcoming dental fear: Strategies for its prevention and management. J. Am. Dent. Assoc. 1983, 107, 18–27. [Google Scholar] [CrossRef] [PubMed]
  22. Caprioglio, A.; Mariani, L.; Tettamanti, L. A pilot study about emotional experiences by using CFSS-DS in young patients. Eur. J. Paediatr. Dent. 2009, 10, 121–124. [Google Scholar] [PubMed]
  23. Lundgren, J.; Berggren, U.; Carlsson, S.G. Psychophysiological reactions in dental phobic patients with direct vs. indirect fear acquisition. J. Behav. Ther. Exp. Psychiatry 2004, 35, 3–12. [Google Scholar] [CrossRef] [PubMed]
  24. Myers, T.D.; McDaniel, J.D. The pulsed Nd:YAG dental laser: Review of clinical applications. J. Calif. Dent. Assoc. 1991, 19, 25–30. [Google Scholar]
  25. White, J.M.; Goodis, H.E.; Daniels, T.E. Effects of Nd:YAG laser on pulps of extracted teeth. Lasers Life Sci. 1991, 4, 191–200. [Google Scholar]
  26. The Institute for Laser Dentistry. Clinical Procedures Performed with the American Dental Laser; The Institute for Laser Dentistry: Hawaii, HI, USA, 1991. [Google Scholar]
  27. Parkins, F.; Miller, R. Nd:YAG laser analgesia of dentin. J. Dent. Res. 1992, 71, 162. [Google Scholar]
  28. Goodis, H.E.; White, J.M.; Harlan, L. Absence of pulpal response from Nd:YAG laser exposure on enamel. J. Dent. Res. 1992, 71, 162. [Google Scholar]
  29. Whitters, C.J.; Hall, A.; Creanor, S.L.; Moseley, H.; Gilmour, W.H.; Strang, R.; Saunders, W.P.; Orchardson, R. A clinical study of pulsed Nd: YAG laser-induced pulpal analgesia. J. Dent. 1995, 23, 145–150. [Google Scholar] [CrossRef]
  30. Moher, D.; Liberati, A.; Tetzlaff, J.; Altman, D.G.; The Prizma Group. Preferred Reporting Items for Systematic Reviews and Meta-Analyses: The PRISMA Statement (Reprinted from Annals of Internal Medicine). PLoS Med. 2009, 6, e1000097. [Google Scholar] [CrossRef] [Green Version]
  31. Higgins, J.; Savović, J.; Page, M.; Elbers, R.; Sterne, J. Assessing risk of bias in a randomized trial. In Cochrane Handbook for Systematic Reviews of Interventions, 2nd ed.; Higgins, J., Thomas, J., Chandler, J., Cumpston, M., Li, T., Page, M., Welch, V., Eds.; John Wiley & Sons: Chichester, UK, 2019; pp. 205–228. [Google Scholar] [CrossRef]
  32. Sarmadi, R.; Andersson, E.V.; Lingström, P.; Gabre, P. A Randomized Controlled Trial Comparing Er:YAG Laser and Rotary Bur in the Excavation of Caries—Patients’ Experiences and the Quality of Composite Restoration. Open Dent. J. 2018, 12, 443–454. [Google Scholar] [CrossRef] [Green Version]
  33. Liang, R.; George, R.; Walsh, L.J. Pulpal response following photo-biomodulation with a 904-nm diode laser: A double-blind clinical study. Lasers Med. Sci. 2016, 31, 1811–1817. [Google Scholar] [CrossRef] [PubMed]
  34. Poli, R.; Parker, S. Achieving dental analgesia with the Erbium Chromium Yttrium Scandium Gallium Garnet Laser (2780 nm): A protocol for painless conservative treatment. Photomed. Laser Surg. 2015, 33, 364–371. [Google Scholar] [CrossRef] [PubMed]
  35. Belcheva, A.; Shindova, M. Subjective Acceptance of Pediatric Patients during Cavity Preparation with Er:YAG Laser and Conventional Rotary Instruments. J. IMAB 2014, 20, 634–637. [Google Scholar] [CrossRef] [Green Version]
  36. Tanboga, I.; Eren, F.; Altinok, B.; Peker, S.; Ertugral, F. The effect of low level laser therapy on pain during dental tooth-cavity preparation in children. Eur. Arch. Paediatr. Dent. 2011, 12, 93–95. [Google Scholar] [CrossRef]
  37. Genovese, M.D.; Olivi, G. Laser in paediatric dentistry: Patient acceptance of hard and soft tissue therapy. Eur. J. Paediatr. Dent. 2008, 9, 13–17. [Google Scholar] [PubMed]
  38. Matsumoto, K.; Wang, X.; Zhang, C.; Kinoshita, J.I. Effect of a Novel Er:YAG Laser in Caries Removal and Cavity Preparation: A Clinical Observation. Photomed. Laser Surg. 2007, 25, 8–13. [Google Scholar] [CrossRef]
  39. Liu, J.F.; Lai, Y.L.; Shu, W.Y.; Lee, S.Y. Acceptance and efficiency of Er:YAG laser for cavity preparation in children. Photomed. Laser Surg. 2006, 24, 489–493. [Google Scholar] [CrossRef]
  40. Boj, J.; Galofre, N.; Espana, A.; Espasa, E. Pain Perception in pediatric patients undergoing laser treatments. J. Oral Laser Appl. 2005, 5, 85–89. [Google Scholar]
  41. Chen, W. The Clinical applications for the Er,Cr:YSGG laser system. Chen Laser Inst. 2011, 12, 42–86. [Google Scholar]
  42. Bjordal, J.M.; Johnson, M.I.; Iversen, V.; Aimbire, F.; Lopes-Martins, R. Photoradiation in acute pain: A systematic review of possible mechanisms of action and clinical effects in randomized placebo-controlled trials. Photomed. Laser Surg. 2006, 24, 158–168. [Google Scholar] [CrossRef] [Green Version]
  43. Angelieri, F.; Sousa, M.V.D.S.; Kanashiro, L.K.; Siquera, D.F.; Maltagliati, L.A. Effects of low intensity laser on pain sensitivity during orthodontic movement. Dent. Press J. Orthod. 2011, 16, 95–102. [Google Scholar] [CrossRef] [Green Version]
  44. Figueiredo Deiana, N.; Zaror, C.; Sandoval, P.; Alves, N. Effectiveness of Low-Level Laser Therapy in reducing orthodontic pain: A systematic review and meta-analysis. Pain Res. Manag. 2017, 2017, 1–18. [Google Scholar] [CrossRef]
  45. Moseley, G.L.; Butler, D.S. Fifteen years of explaining pain: The past, present, and future. J. Pain 2015, 16, 807–813. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  46. Walsh, L.J. Laser analgesia with pulsed infrared lasers: Theory and practice. J. Oral Laser Appl. 2008, 8, 7–16. [Google Scholar]
  47. Hoke, J.A.; Burkes, E.J., Jr.; Gomnes, E.D.; Wolbarsht, M.L. Erbium:YAG (2.94 μm) Laser Effects on Dental Tissues. J. Laser Appl. 1990, 2, 61–65. [Google Scholar] [CrossRef]
  48. Cavalcanti, B.N.; Lage-Marques, J.L.; Rode, S.M. Pulpal temperature increases with Er:YAG laser and high-speed handpieces. J. Prosthet. Dent. 2003, 90, 447–451. [Google Scholar] [CrossRef]
  49. Attrill, D.C.; Davies, R.M.; King, T.A.; Dickinson, M.R.; Blinkhorn, A.S. Thermal effects of the Er:YAG laser on a simulated dental pulp: A quantitative evaluation of the effects of a water spray. J. Dent. 2004, 32, 35–40. [Google Scholar] [CrossRef]
  50. Yan, W.; Chow, R.; Armati, P.J. Inhibitory effects of visible 650-nm and infrared 808-nm laser irradiation on somatosensory and compound muscle action potentials in rat sciatic nerve: Implications for laser-induced analgesia. J. Peripher. Nerv. Syst. 2011, 16, 130–135. [Google Scholar] [CrossRef]
  51. Melzack, R.; Wall, P. Pain mechanisms: A new theory. Science 1965, 150, 171–179. [Google Scholar] [CrossRef]
  52. Testani, E.; Le Pera, D.; Del Percio, C.; Miliucci, R.; Brancucci, A.; Costanza, P.; De Armas, L.; Babiloni, C.; Rossini, P.M.; Valeriani, M. Cortical inhibition of laser pain and laser-evoked potentials by non-nociceptive somatosensory input. Eur. J. Neurosci. 2015, 42, 2407–2414. [Google Scholar] [CrossRef]
  53. Orchardson, R.; Peacock, J.M.; Whitters, C.J. Effect of pulsed Nd:YAG laser radiation on action potential conduction in isolated mammalian spinal nerves. Lasers Surg. Med. 1997, 21, 142–148. [Google Scholar] [CrossRef]
  54. Orchardson, R.; Peacock, J.M.; Whitters, C.J. Effects of pulsed Nd:YAG laser radiation on action potential conduction in nerve fibres inside teeth in vitro. J. Dent. 1998, 26, 421–426. [Google Scholar] [CrossRef]
  55. Orchardson, R.; Whitters, C.J. Effect of HeNe and pulsed Nd:YAG laser irradiation on intradental nerve responses to mechanical stimulation of dentine. Lasers Surg. Med. 2000, 26, 241–249. [Google Scholar] [CrossRef]
  56. Zeredo, J.L.; Sasaki, K.M.; Fujiyama, R.; Okada, Y.; Toda, K. Effects of low power Er:YAG laser on the tooth pulp-evoked jaw-opening reflex. Lasers Surg. Med. 2003, 33, 169–172. [Google Scholar] [CrossRef] [PubMed]
  57. Zeredo, J.L.; Sasaki, K.M.; Takeuchi, Y.; Toda, K. Antinociceptive effect of Er:YAG laser irradiation in the orofacial formalin test. Brain Res. 2005, 1032(1-2), 149–153. [Google Scholar] [CrossRef]
  58. Snyder-Mackler, L.; Bork, C.E. Effect of helium-neon laser irradiation on peripheral sensory nerve latency. Phys. Ther. 1988, 68, 223–225. [Google Scholar] [CrossRef]
  59. Sono, T.; Kasai, S.; Sakamoto, T.; Mito, M. Cord dorm potentials suppressed by low power laser irradiation on a peripheral nerve in the cat. J. Clin. Laser Med. Surg. 1993, 11, 115–118. [Google Scholar] [CrossRef]
  60. Elbay, Ü.Ś.; Tak, Ö.; Elbay, M.; Ugurluel, C.; Kaya, C. Efficacy of Low-level laser therapy in the management of postoperative pain in children after primary teeth extraction: A randomized clinical trial. Photomed. Laser Surg. 2016, 34, 171–177. [Google Scholar] [CrossRef]
  61. Hamblin, M.R. The role of nitric oxide in low level light therapy. Proc. SPIE 2008, 6846, 1–14. [Google Scholar] [CrossRef]
  62. Rochkind, S. Phototherapy in peripheral nerve regeneration: From basic science to clinical study. Neurosurg. Focus 2009, 26, E8. [Google Scholar] [CrossRef] [Green Version]
  63. Kneebone, W.J. Practical applications of low level laser therapy. Pract. Pain Manag. 2006, 6, 34–40. [Google Scholar]
  64. Martin, R. Laser-accelerated inflammation/pain reduction and healing. Pract. Pain. Manag. 2003, 3, 20–25. [Google Scholar]
  65. Farivar, S.; Malekshahabi, T.; Shiari, R. Biological effects of low level laser therapy. J. Lasers Med. Sci. 2014, 5, 58–62. [Google Scholar] [PubMed]
  66. Laasko, E.; Cabot, P.J. Nociceptive scores and endorphin-containing cells reduced by low-level laser therapy (LLLT) in inflamed paws of Wistar rat. Photomed. Laser Surg. 2005, 23, 32–35. [Google Scholar] [CrossRef]
  67. Bruehl, S.; Burns, J.W.; Chung, O.Y.; Chont, M. What do plasma beta-endorphin levels reveal about endogenous opioid analgesic function? Eur. J. Pain 2012, 16, 370–380. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  68. Moriyama, Y.; Nguyen, J.; Akens, M.; Moriyama, E.H.; Lilge, L. In vivo effects of low-level laser therapy on inducible nitric oxide synthase. Lasers Surg. Med. 2009, 41, 227–231. [Google Scholar] [CrossRef] [PubMed]
  69. Cidral-Filho, F.J.; Mazzardo-Martins, L.; Martins, D.F.; Santos, A.R.S. Light-emitting diode therapy induces analgesia in a mouse model of postoperative pain through activation of peripheral opioid receptors and the L-arginine/nitric oxide pathway. Lasers Med. Sci. 2014, 29, 695–702. [Google Scholar] [CrossRef]
  70. Aimbire, F.; Albertini, R.; Pacheco, M.T.T.; Castro-Faria-Neto, H.C.; Bjordal, J.M. Low-level laser therapy indices dose-dependent reduction of TNFa levels in acute inflammation. Photomed. Laser Surg. 2006, 24, 33–37. [Google Scholar] [CrossRef]
  71. Fulop, M.A.; Dimmer, S.; Deluca, J.R.; Johanson, D.D.; Lenz, R.V.; Patel, K.B.; Douris, P.C.; Enwemeka, C.S. A meta-analysis of the efficacy of laser phototherapy on pain relief. Clin. J. Pain 2010, 26, 729–736. [Google Scholar] [CrossRef]
  72. Boschi, E.S.; Leite, C.E.; Saciura, V.C.; Caberlon, E.; Lunardelli, A.; Bitencourt, S.; Melo, D.A.S.; Oliveira, J.R. Anti-inflammatory effects of low-level laser therapy (660 nm) in the early phase in carrageenan-induced pleurisy in rat. Lasers Surg. Med. 2008, 40, 500–508. [Google Scholar] [CrossRef]
  73. Doukas, A.G.; Flotte, T.J. Physical characteristics and biological effects of laser-induced stress waves. Ultrasound Med. Biol. 1996, 22, 151–164. [Google Scholar] [CrossRef]
Figure 1. PRISMA flow-chart of selected criteria for the included article reports.
Figure 1. PRISMA flow-chart of selected criteria for the included article reports.
Dentistry 08 00128 g001
Figure 2. Example of tip kept at 10 mm for the rabbit technique of laser-induced analgesia. Irradiation at tooth neck level.
Figure 2. Example of tip kept at 10 mm for the rabbit technique of laser-induced analgesia. Irradiation at tooth neck level.
Dentistry 08 00128 g002
Figure 3. Example of the tip kept at 1 mm of distance from the tooth neck as in the turtle technique.
Figure 3. Example of the tip kept at 1 mm of distance from the tooth neck as in the turtle technique.
Dentistry 08 00128 g003
Figure 4. Rabbit technique: irradiation from the occlusal side of the tooth.
Figure 4. Rabbit technique: irradiation from the occlusal side of the tooth.
Dentistry 08 00128 g004
Table 1. Data evaluation of the included studies.
Table 1. Data evaluation of the included studies.
Citation
[Ref]
Type of Study/Number of PatientsTest/Control GroupAim/ApproachLaser ParametersOutcome
Sarmadi et al. (2018)
[32]
Split-mouth RCT/25 patients with at least 2 primary caries of equal size and same location (occlusal or interproximal)2940 nm (28 cavities)/rotary bur (28 cavities)Discomfort and Pain (VAS)Enamel preparation:
250–300 mJ, VSP pulse, 30 Hz, water/air
Dentin preparation:
200–300 mJ, VSP/SP pulse, 10–20 Hz, water/air
Excavation dentin caries:
200–300 mJ, VSP/SP pulse, 20–30 Hz, water/air
Excavation deep caries:
150–250 mJ, SP pulse, 5–15 Hz, water/air
Laser group:
10/28 patients required LA
Bur group:
15/28 patients required LA
No difference immediately after tx
(p = 0.881)
Liang et al. (2016)
[33]
Split-mouth RCT/30 patients, healthy maxillary first premolars examined (one tooth/group)PBM with 904 nm (30 teeth)/placebo effect (30 teeth)Pulpal response (EPT)
2 min prior and immediately after irradiation
PBM:
Average Power 30 mW, 25 Hz, 50% duty cycle, buccal surface for 60 s, 3.6 J, 1 cm2 spot size
Significant difference p < 0.0001
Poli et al. (2015)
[34]
Clinical trial/30 patients with a single cavity2780 nmPulpal response (EPT) for pre-, intra- and post-treatment evaluation
Pain (VAS)
ANALGESIA:
The delivery tip used was the MGG6 sapphire tip, diam-
eter 0.600 mm, length, 6 mm (beam spot area = 0.283 mm2).
Power of 0.1 W and then 0.2 W (energy per pulse 10 and 20 mJ), for 30 s each (without air/water spray), 10 Hz, tip at 10 mm from the tooth using a spacer. Subsequently, the power was increased to 0.5 and then to 1 W (33 and 67 mJ) for 60 s each, with a spray of 15% water (*10 mL/min) and 20% air, 15 Hz, keeping the same distance.
Hard tissues were preconditioned with 2 W for 30 s, spray 50% water (*20 mL/min) and 80% air, 15 Hz, tip at 1 mm from the tooth. The laser was defocused if the patient felt discomfort
VAS:
0–1 values at 24/30 (80%) of patients
EPT values variations:
No difference (p > 0.05)
Chan et al. (2012)
[10]
Split-mouth RCT/44 patients having a cavity by bur at bilateral premolars (one tooth/group)Nd:YAG + sham EMLA (44 teeth)/EMLA + sham laser (44 teeth)Pain (EPT and VAS)ANALGESIA:
150 μsec, AP 1.1 W, 15 Hz; 60–87 mJ, 0.3–0.45 W/cm2, 73–107
J/cm2, total energy, 211- 312 J, 320 μm fiber, scanning motion (speed at 3 mm/s) approximately 1 mm (spot diameter = 6 mm) from the buccal and lingual/palatal cervical areas for 240 s
Both groups achieved analgesic effects
No significant difference in EPT and VAS scores between groups
Belcheva et al. (2014) [35]Parallel-group RCT/90 patients. 1 or more dentine carious lesions without pulp involvement or pain. Occlusal or proximal surface of a primary or
a permanent molar
Er:YAG (45 patients)/Rotary bur (45 patients) Pain (universal pain assessment tool)200–300 mJ/20 Hz, water 8 for the permanent teeth.
100–200 mJ/
20 Hz, water 8 for the primary teeth.
Laser group significant lower pain (p = 0.005)
Laser group significant difference in low degree of pain (p < 0.005) and in severe degree (p < 0.01) compared to control group, respectively
Tanboga et al. (2012)
[36]
Split-mouth RCT/10 patients (6–9 y.o) with primary molars (one tooth/group)LLLT (Er:YAG) + Er:YAG prep (10 teeth)/Er:YAG prep (10 teeth)Pain (VAS)ANALGESIA:
at a distance of 2 mm from the tooth surface on the gingival margin and slowly moved for 2 min, 60 mJ, 20 Hz, 250 ms pulse, additional water spray
Test group significant better
p = 0.004
Genovese et al. (2008)
[37]
Clinical trial/50 patients (6–12 y.o) required both hard and soft tissue therapy, without anaesthesiaEr:YAG and Er,Cr:YSGG
Cavity prep:
20 permanent molars and 30 deciduous molars
soft tissues:
23 frenectomies, 12 gingivectomies, 15 operculectomies
Patient’s experience (Wong-Baker modified facial image scale)ANALGESIA:
0.5–2.5 W, 20 Hz, 75 mJ, Air 20%, Water 15%, at a distance of 3 mm from the tooth surface, in defocused mode on the gingival margin (1–3 mm), and slowly moved for 2 min
Pain values 1–2 at 44/50 (88%) of patients tested for cavity preparation
Matsumoto et al. (2007)
[38]
Clinical trial/45 patients (95 teeth) with primary carious lesions in vital teethCavity preparation in enamel and dentin with Er:YAG (95 teeth)Pain (4-point scale)CAVITY PREP:
100–700 mJ, 80–700 ms pulse, 8–20 Hz, water 12 mL/min
Pain values 1–2 at 85/95 (89.5%) of patients
Liu et al. (2006)
[39]
Split-mouth RCT/40 patients (4–12 y.o) with two maxillary anterior carious teeth, same type of lesion and approximately equal-sized cavitiesEr:YAG (40 teeth)/rotary bur (40 teeth)Pain (simple modified face scale)CAVITY PREP:
700 mJ, 10 Hz, non-contact 1 mm distance, 800 μm tip, water 24 mL/min
Laser group significantly better than control
p < 0.001
Boj et al. (2005)
[40]
Clinical trial/33 patients (8–16 y.o.) required restorations in permanent teethEr,Cr:YSGG used for restorations in permanent teethPain (Wong-Baker facial image scale)Manufacturer’s recommendations Pain values 1–2 at 22/33 (66.7%) of patients
Table 2. Risk of bias evaluation of the included studies.
Table 2. Risk of bias evaluation of the included studies.
Citation
[Ref]
RandomizationSample Size Calculation and Required Number IncludedBaseline Situation SimilarBlindingParameters of Laser Use Described Appropriately and Calculations CorrectPower Meter
Used
Numerical Results Available (Statistics)No Missing Outcome DataCorrect Interpretation
of Data
Total Score/
9
Sarmadi et al. (2018)
[32]
yesyesyesyesnonoyesyesyes7
Liang et al. (2016)
[33]
yesnoyesyesyesnoyesyesyes7
Poli et al. (2015)
[34]
nonononoyesnoyesyesyes4
Chan et al. (2012)
[10]
yesnoyesyesyesyesyesyesyes8
Belcheva et al. (2014) [35]yesnoyesnononoyesyesyes5
Tanboga et al. (2012)
[36]
yesnoyesnononoyesyesyes5
Genovese et al. (2008)
[37]
nonononononoyesyesyes3
Matsumoto et al. (2007)
[38]
nonononononoyesyesyes3
Liu et al. (2006)
[39]
yesnoyesnononoyesyesyes5
Boj et al. (2005)
[40]
nonononononoyesyesyes3
Table 3. Statistical analysis was performed on a combination of three split-mouth studies in which it was possible to classify the responses as pain experience or no pain. Data extracts allowed comparative patient numbers to be computed and assessed.
Table 3. Statistical analysis was performed on a combination of three split-mouth studies in which it was possible to classify the responses as pain experience or no pain. Data extracts allowed comparative patient numbers to be computed and assessed.
StudyPts Total/Cavities by GroupTreatmentResponse
PainNo Pain
Sarnadi [32]125/28Control1312
25/28Laser916
Chan [10]244/44Control1826
44/44Laser1430
Liu [39]340/40Control2911
40/40Laser733
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Poli, R.; Parker, S.; Anagnostaki, E.; Mylona, V.; Lynch, E.; Grootveld, M. Laser Analgesia Associated with Restorative Dental Care: A Systematic Review of the Rationale, Techniques, and Energy Dose Considerations. Dent. J. 2020, 8, 128. https://0-doi-org.brum.beds.ac.uk/10.3390/dj8040128

AMA Style

Poli R, Parker S, Anagnostaki E, Mylona V, Lynch E, Grootveld M. Laser Analgesia Associated with Restorative Dental Care: A Systematic Review of the Rationale, Techniques, and Energy Dose Considerations. Dentistry Journal. 2020; 8(4):128. https://0-doi-org.brum.beds.ac.uk/10.3390/dj8040128

Chicago/Turabian Style

Poli, Riccardo, Steven Parker, Eugenia Anagnostaki, Valina Mylona, Edward Lynch, and Martin Grootveld. 2020. "Laser Analgesia Associated with Restorative Dental Care: A Systematic Review of the Rationale, Techniques, and Energy Dose Considerations" Dentistry Journal 8, no. 4: 128. https://0-doi-org.brum.beds.ac.uk/10.3390/dj8040128

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop