Review

Hybrid Impedimetric Biosensors for Express Protein Markers Detection

by
Nikita Sitkov
1,2,*,
Andrey Ryabko
3,*,
Vyacheslav Moshnikov
1,
Andrey Aleshin
3,
Dmitry Kaplun
4,5 and
Tatiana Zimina
1,2
1
Department of Micro and Nanoelectronics, Saint Petersburg Electrotechnical University “LETI”, 197022 Saint Petersburg, Russia
2
Engineering Centre for Microtechnology and Diagnostics, Saint Petersburg Electrotechnical University “LETI”, 197022 Saint Petersburg, Russia
3
Laboratory of Nonequilibrium Processes in Semiconductors, Ioffe Institute, 26 Politekhnicheskaya, 194021 Saint Petersburg, Russia
4
Artificial Intelligence Research Institute, China University of Mining and Technology, 1 Daxue Road, Xuzhou 221116, China
5
Department of Automation and Control Processes, Saint Petersburg Electrotechnical University “LETI”, 197022 Saint Petersburg, Russia
*
Authors to whom correspondence should be addressed.
Submission received: 31 December 2023 / Revised: 22 January 2024 / Accepted: 23 January 2024 / Published: 25 January 2024
(This article belongs to the Special Issue Intelligent Approaches in Biosensing)

Abstract

:
Impedimetric biosensors represent a powerful and promising tool for studying and monitoring biological processes associated with proteins and can contribute to the development of new approaches in the diagnosis and treatment of diseases. The basic principles, analytical methods, and applications of hybrid impedimetric biosensors for express protein detection in biological fluids are described. The advantages of this type of biosensors, such as simplicity and speed of operation, sensitivity and selectivity of analysis, cost-effectiveness, and an ability to be integrated into hybrid microfluidic systems, are demonstrated. Current challenges and development prospects in this area are analyzed. They include (a) the selection of materials for electrodes and formation of nanostructures on their surface; (b) the development of efficient methods for biorecognition elements’ deposition on the electrodes’ surface, providing the specificity and sensitivity of biosensing; (c) the reducing of nonspecific binding and interference, which could affect specificity; (d) adapting biosensors to real samples and conditions of operation; (e) expanding the range of detected proteins; and, finally, (f) the development of biosensor integration into large microanalytical system technologies. This review could be useful for researchers working in the field of impedimetric biosensors for protein detection, as well as for those interested in the application of this type of biosensor in biomedical diagnostics.

1. Introduction

Proteins are vitally important biomolecules, present in all living organisms, which form many essential biological compounds such as enzymes, hormones, antibodies, etc., that are involved in various vital processes in the body, such as catalysis, transport, signaling, regulation, defense, structure maintenance, and so on. Proteins also could serve as biological markers of diseases, that is, molecules indicating the presence or absence of normal or pathological conditions in the body. That is why the detection and quantification of proteins in biological fluids is very important for the diagnosis and monitoring of various diseases such as cancer, infections, autoimmune diseases, allergies, etc. [1,2,3,4,5,6,7,8]. It is noted that for accurate and timely identification of the disease, it is necessary to determine a diagnostically important group of protein markers [9,10,11], which allows specialists to choose efficient treatment tactics. There are a number of methods for determining proteins including enzyme-linked immunosorbent assay (ELISA), fluorescence immunoassay (FIA), surface plasmon resonance (SPR), chromatography, spectroscopy, electrophoresis, etc. [12,13]. However, at present, these methods are mostly implemented in the form of stationary laboratory instruments, which have a number of disadvantages, such as relatively high cost, complex and multistage procedures, the need for specialized equipment and reagents for sample preparation, and the use of labels that can affect the properties and activity of proteins. Therefore, there is a need to develop new devices for detecting proteins that would be simpler, faster, more sensitive, and selective. In addition, at present, it is quite difficult to miniaturize these techniques in order to create low-cost portable devices for the rapid diagnosis of diseases at the point of care. These problems can be solved using the biosensing technologies approach.
Biosensors are analytical devices that use specific biochemical reactions to detect chemical compounds, usually using electrical, thermal, or optical detection [14,15,16]. In addition to medical diagnostics, biosensors are also used in ecology, the food industry, and agriculture, as they allow for fast, sensitive, selective, and cheap analysis of biological samples. The accuracy of the information obtained from biosensors depends on the composition of the analyte, the biologically active component, the design of the biosensor, and the physical characteristics of the signal transducer. Biosensors play an important role in biomedical research, as they can provide diagnosis and monitoring of various diseases, as well as study the mechanisms of biological processes at the molecular and cellular levels [17].
In the classification of biosensors according to the principles of signal transduction, electrochemical biosensors occupy a special place, since they have a number of advantages, such as simplicity of design, the possibility of miniaturization and automation, as well as a low level of interference from the environment [18]. They have the potential to provide continuous monitoring and can be implanted into the human body [19]. Electrochemical biosensors can be classified according to the type of electrical parameter being measured into potentiometric, amperometric, and impedimetric [20]. Impedimetric biosensors are based on measuring changes in the electrical impedance of an electrochemical cell upon binding of biomolecules to the surface of electrodes that are modified with specific biorecognition elements. Unlike amperometric and potentiometric biosensors, impedimetric biosensors do not require labels to detect the analyte [21,22], which significantly reduces the cost and simplifies their production.
Impedimetric biosensors have a number of characteristics that make them suitable for protein detection. Firstly, they allow the measurement of frequency dependencies of electrical impedance, which correlates with various parameters of the binding process of proteins and biorecognition elements of the sensor, such as kinetics, thermodynamics, mass transfer, and structural changes to be carried out [23]. Secondly, they allow electrodes to be modified with various nanomaterials and nanostructures, which increase the surface area of the electrodes, improve electron transfer, and enhance the biosensor signal [24]. Thirdly, they allow the use of different types of biorecognition agents, such as antibodies, aptamers, peptides, receptors, or enzymes, which provide specific and stable binding to target proteins [25,26]. Fourthly, they allow the creation of miniature hybrid integrated biosensor matrices coupled with microfluidic systems that reduce sample and reagent volumes, increase the speed and accuracy of analysis, and implement multichannel and parallel measurement [27]. Fifthly, they allow the use of new signal processing, information technology, and artificial intelligence techniques to improve resolution, reduce noise, optimize parameters, classify data, and automate analysis [28]. A general scheme of the hybrid impedimetric biosensor for protein detection development is presented in Figure 1.
At the same time, the development of new impedimetric biosensors for detecting proteins poses a number of challenges for researchers, such as (i) the selection of optimal materials and nanostructures for electrodes, (ii) the development of effective methods for immobilizing biorecognition elements for the formation of stable specific binding groups on the surface of electrodes, (iii) increasing specificity and sensitivity of biosensors, (iv) reducing the influence of interference and nonspecific binding, (v) adapting biosensors to real samples and operating conditions, and (vi) expanding the spectrum of detected proteins. To successfully solve these problems, it is necessary to apply an interdisciplinary approach, combining knowledge and skills in the fields of chemistry, physics, biology, materials science, nanotechnology, electronics, information technology, and medicine. This review will discuss the operation principles of impedimetric sensors, methods of modification and nanomodification of electrode surfaces, types of biorecognition agents and methods of their immobilization, and the integration of impedimetric biosensors into microfluidic systems. We hope that this review will be useful for researchers working in the field of impedimetric biosensors for protein detection, as well as for those interested in the application of such biosensors in biomedical diagnostics.

2. Basic Principles of Impedimetric Measurements

Electrochemical impedance spectroscopy (EIS) is a technique that measures the complex impedance, combining the effects of resistance and reactance in a circuit, when an alternating voltage of varying frequency is applied to electrodes. The resistance depends on the physicochemical processes occurring at the electrode–electrolyte interface, such as charge transfer, diffusion, adsorption, desorption, and biological interaction. Impedimetric biosensors belong to label-free biosensors, which are based on a measurement of the impedance value change in the sensor system as a result of the specific binding of the biorecognition element and the target analyte, depending on its concentration. These biosensors are based on two types of EIS: faradaic and non-faradaic. Biosensors for faradaic and non-faradaic EIS have differences not only in the measurement methodology but also in the materials used to modify the electrodes and in design.
Faradaic EIS is sensitive to the presence of a redox-active probe in the solution, which participates in an electrochemical reaction on the surface of the electrode. This reaction involves electron transfer between the probe and the electrode, which can be hindered by the binding of the analyte to the bioreceptor. To carry out the faradaic EIS measurements, as a rule, it is necessary to use a redox couple in solution, which is usually a ferro/ferri electrolyte solution (K3Fe(CN)6 + K4Fe(CN)6·3H2O) [29,30,31]. Before performing EIS, biosensors are usually immersed in a solution (or biological fluid) for immobilization of the target analyte, usually washed, and then immersed in the solution for measurements [32,33,34]. The electrochemical cell, in addition to the biosensor electrode, also contains a counter electrode (CE) and a reference electrode (RE). The biosensor plays the role of a working electrode (WE). A typical faradaic EIS spectrum is presented in Figure 2a as a Nyquist plot. It contains a semicircle at high frequencies due to the charge transfer resistance (Rct) associated with the oxidation of the redox pair and the capacitance of the electrical double layer (Cdl). Also, the Nyquist diagram contains a straight line at low frequencies, which is associated with mass transfer, including diffusion [35].
Thus, the electrochemical system for faradaic EIS is typically represented as a Randle equivalent circuit, which contains the electrolyte resistance Rs between the biosensor (WE) and the reference electrode (RE), the electrical double-layer capacitance Cdl (or constant phase element (CPE)), charge transfer resistance Rct, and the Warburg element (ZW) (Figure 2b). Charge transfer resistance (Rct) is the most important parameter for faradaic EIS, as it reflects the activation barrier to electron exchange between the redox probe and the electrode. As shown in Figure 2a, the diameter of the semicircle is determined by the value of the charge transfer resistance Rct, the change in which is used to construct the concentration dependence of the biosensor signal [36,37,38].
To ensure a good level of electronic conductivity between the working electrode and the ferri–ferro electrolyte, it is useful to provide a sufficiently low charge transfer resistance Rct for the initial electrode [39,40,41]. The binding of biorecognition molecules (for example, anti-CYFRA 21-1) to the nanocomposite leads to an increase in Rct. Blocking nonspecific binding sites with bovine serum albumin (BSA) further reduces the electron transfer process between the working electrode and the electrolyte [42]. Finally, specific immobilization of the analyte leads to a significant blocking of charge transfer for the oxidation of the redox pair and, accordingly, an increase in Rct. Therefore, the analyte concentration significantly affects the diameter of the semicircle on the Nyquist diagram. Thus, it is possible to set up a calibration concentration dependence on the change in Rct. It can be linearly approximated, which is quite convenient for automating the determination of biomarker concentration. However, to determine Rct, preliminary mathematical processing of Nyquist diagrams and spectroscopy in a wide frequency range of alternating voltage is necessary.
It should also be noted that the electrochemical system for faradaic EIS allows a cyclic voltammetry (CV) measurement to be carried out. During this process, when the analyte concentration increases, the peak current values decrease. However, concentration dependences are usually determined relative to ΔRct. The initial modified working electrode of the biosensor must have sufficiently good electrical conductivity to ensure the process of charge transfer between the electrode and the electrolyte; therefore, conductive materials are used in the formation of the electrode or electrode-modifying material [43,44,45].
Unlike faradaic EIS, non-faradaic EIS does not use redox reactions, that is, measurements are carried out in a buffer solution or biological fluid without a specific electrolyte and without additional reference (RE) and counter (CE) electrodes. The processes of analyte immobilization and spectroscopy are often carried out in the same solution [46,47,48,49]. In non-faradaic EIS, there is no electrochemical reaction on the electrode surface, but only the formation of an electrical double layer, which can be modeled as a capacitor. The double-layer capacitance (Cdl) depends on the dielectric properties of the medium, the electrode area, and the distance between the capacitor plates. When the analyte binds the biorecognition element on the electrode surface, these parameters change, which leads to a change in the capacitance of the double layer. Thus, non-faradaic EIS measures the change in capacitance between the double layer and the electrode interface, which reflects the change in the dielectric behavior of the sensor surface. The biosensor itself, instead of one electrode, must contain two electrodes (for example, interdigitated once, increasing the area between the electrodes) between which there is a material to which immobilization is ensured. Dielectric and high-resistivity semiconductor materials are used as such materials.
For non-faradaic EIS, it is necessary to use an insulating layer on the electrode to prevent electrochemical reactions and make the interface capacitance the dominant parameter. For an ideal insulator or in the absence of redox compounds, the leakage resistance is theoretically infinite [50]. In order for this resistance to tend to infinity in practice, the electrodes can be electrically isolated from the immobilization material. Thus, in [51], a ZnO/CuO nanofilm is separated from the electrodes by borosilicate glass.
Non-faradaic EIS spectra are usually presented as Bode diagrams, which show the dependence of the impedance modulus (|Z|) and phase shift (φ) on the logarithm of frequency (Figure 3). For non-faradaic EIS, when there is no redox-active probe in the solution, the Bode diagram is preferable, since it allows one to study the frequency dependence of the impedance. In addition, other parameters such as capacitance (C) and conductance (Y) can be used to represent non-faradaic EIS results. In general, the Bode diagram also makes it possible to identify the frequency at which it is preferable to determine the biosensor response. When choosing such a frequency, you can proceed, for example, from the maximum signal-to-noise ratio [52,53]. The ability to use only one frequency in non-faradaic impedimetric biosensors is very important for their practical application since this significantly simplifies the required equipment when used in the analysis [54].
In general, it should be noted that non-faradaic impedimetric biosensors can be much simpler for practical use, since they do not require additional electrodes or the addition of a redox pair and can operate at a single frequency, while faradaic biosensors, as a rule, analyze over a fairly large frequency range and require a slightly more complex preparation (electrochemical cell). However, faradaic impedimetric biosensors can generally demonstrate significantly greater analytical sensitivity. Also, for a general overview of faradaic and non-faradaic label-free impedance biosensors, one can refer to the review articles [50,55,56].

3. Materials for Electrodes and Their Modifications

Impedimetric biosensors can be made from various materials, both to form the electrodes and to modify them. In faradaic EIS biosensors, in addition to the working electrode, a counter electrode and a reference electrode are also used. To create a working electrode, which, in fact, is the sensor itself, various materials are used, both for the conductive electrode and its modification for subsequent functionalization. In non-faradic EIS biosensors, two electrodes are used for measurements, between which another material is located, serving for subsequent functionalization. In this section, we will highlight the main groups of materials for the formation of electrodes, which are directly functionalized to create an impedimetric biosensor. We will also review some examples of impedimetric biosensors from the main groups of electrode materials.

3.1. Metals

Impedimetric biosensors with a metal working electrode typically contain gold electrodes that are functionalized with molecules to form self-assembled monolayers (SAMs) for subsequent binding [57,58,59,60,61,62,63,64,65,66,67,68].
In [68], an impedimetric immunosensor was developed based on gold screen-printed electrodes (SPE, DropSens, Oviedo-Asturias, Spain) for the detection of surfactant protein B (SPB) in human amniotic fluid (AF) samples. A gold electrode was used as CE, and a pseudo-reference electrode made of silver was used as RE. The biofunctionalization of the working electrode (biosensor electrode) is presented in Figure 4.
The surface of the gold electrode was precleaned in isopropanol and deionized water and dried under N2 flow. The gold surface was incubated in Sulfo-LC-SPDP (sulfosuccinimidyl 6-(3′-(2pyridyldithio)propionamido)hexanoate) to form a SAM. Anti-SPB antibodies were immobilized for 2 h at room temperature. Finally, the surface was incubated in BSA solution (1%) for 20 min to block the exposed groups to prevent nonspecific binding. SPB solutions of varying concentrations were applied to the immobilized electrode for 90 min at room temperature to capture the antigen. After thorough washing, they were placed in redox probe solution (5 mM [Fe(CN)6]3−/4− in PBS). The authors demonstrated that with increasing antigen concentration, Rct increases, i.e., the process of charge transfer is reduced. The immunosensor showed a linear dynamic range from 2 ng/mL to 2000 ng/mL and a limit of detection (LOD) of 0.1 ng/mL.
It should be noted that the mass production of such miniature measuring electrochemical platforms makes possible the widespread use of impedimetric biosensors and also simplifies and speeds up their development. For example, a similar sensor platform with gold SPE (Metrohm DropSens, Oviedo-Asturias, Spain) was used in [65] to develop an impedimetric biosensor for Alpha-1-B glycoprotein (A1BG), a marker of endometriosis. To form SAMs on a gold electrode, 3-Mercaptopropionic acid (3-MPA) was used in this work. The SAM was then stirred into the mixture of 1-Ethyl-3-(3-dimethylaminopropyl)carbodiimide (EDC) and N-Hydroxysuccinimide (NHS) to allow conversion of the terminal carboxyl group (-COOH) of the SAM to the NHS transition ester.
In [57], interdigitated microelectrode (IDWμE) arrays were used for functionalization with a self-assembled monolayer of thioctic acid (TA) to immobilize the antigen. The developed immunosensor was aimed at detecting rheumatoid factor–immunoglobulin M (IgM-RF). For surface modification, the IDWμEs were immersed in a solution of TA in ethanol, where a SAM-TA was formed by adsorption on the electrodes. After modification of the TA-SAM electrodes, they were washed to remove unbound thiol moieties. The covalent immobilization of human IgG-Fc fragments was carried out to bind the target analyte. The sensor showed an LOD of 0.6 IU/mL in the presence of a redox probe and 0.22 IU/mL in human serum.
Gold, as a material for immobilization, is also used in the form of nanoparticles, which can provide an additional increase in the specific surface area of the electrode [69,70,71,72]. Basic examples of impedimetric biosensors using metal electrodes are given in Table 1.

3.2. Conductive Oxides

It is obvious that, in addition to metals, a wide class of conductive materials can be used to make electrodes, including transparent conductive films based on metal oxides [73,74,75,76,77,78,79]. Transparent conductive oxides (TCO) are widely used for solar panels, touch screens, and displays. Among the TCOs, the most widely used are indium tin oxide (ITO) and second-doped tin oxide (FTO). Less commonly used are zinc oxides doped with aluminum or gallium (AZO, GZO). Transparent conductive oxides are used in the production of impedimetric biosensors using the same method as metal ones, that is, SAMs are formed on the surface from molecules containing functional groups. However, their binding to TCO is carried out via hydroxyl groups on the surface.
A typical example of the functionalization and immobilization of a TCO-based electrode, which was developed for the detection of p21-activated kinase 2 (PAK 2), is presented in Figure 5 [76]. The first step is to treat the ITO electrode to form hydroxyl groups on the surface. Then, the surface was silanized in a solution of 3-glycidoxypropyltrimethoxysilane (GPTMS), forming a monolayer connected to the electrode through Si-O-Si bonds. The next steps include immobilization of the recognition layer using anti-PAK 2 and blocking unbound groups with BSA. The final step of the process presented in Figure 5 is the binding of the analyte to the immunosensor. The immunosensor had a linear calibration curve of response to PAK 2 capture in the range of 0.005–0.075 pg/mL with an LOD of 1.5 fg/mL. The immunosensor demonstrated high stability, good reproducibility, and stability.
The binding of SAM to OH groups is also carried out via the formation of P-O bonds. Thus, using an ITO electrode, an impedimetric immunosensor was developed to detect the cancer marker interleukin-8 (IL-8) in human serum and saliva samples [73]. Prewashed ITO-coated substrates were immersed in a solution of NH4OH:H2O2:H2O (1:1:5) for 90 min to form hydroxyl groups on the ITO surface. The hydroxylated ITO electrode was then functionalized with a carboxyl-terminated phosphonic acid (PHA) SAM. At this stage, the end groups of SAM-PHA were covalently bonded to the hydroxyl groups, and P–O bonds were formed. The biorecognition layer was formed using immobilization of biorecognition elements, anti-IL8 antibodies, via the carboxyl groups of SAM-PHA, which were chemically activated using EDC/NHS. EDC and NHS acted as a coupling agent and an activator, respectively. After immobilization with anti-IL8, BSA was used to block the carboxyl groups of PHA and other proteins. The impedimetric biosensor showed a wide detection range of IL8 from 0.02 pg/mL to 3 pg/mL and an LOD of 6 fg/mL.
Similarly, a biosensor for the detection of interleukin 1α (IL-1α) was developed in [74], where the surface of FTO was modified with carboxyalkylphosphonic acid (PHP) and immobilized with anti-IL-1α antibodies. IL-1α antigen concentration was determined over a linear detection range of 0.02 to 2 pg/mL. The LOD value of the proposed immunosensor was 6 fg/mL. The FTO electrode surface was also hydroxylated with NH4OH/H2O2/H2O solution. The FTO electrode was then immersed in PHP solution to form a layer, and after washing, the carboxyl groups in the NHS/EDC solution were activated. Immobilization was also carried out by immersing the electrodes in a phosphate-buffered solution containing antibodies to IL-1α. Anti-IL-1α binding is mediated by the amino groups of antibodies and the carboxyl groups of PHP.
In [79], a coating was applied to vertically oriented silicon nanowires to increase the specific area of an impedimetric biosensor and to expand the concentration range of ITO. This gave the linear detection range of human cardiac troponin I (cTnI) in the range from 1.760 to 1760 ng/mL.
Basic examples of impedimetric biosensors using conductive oxides are given in Table 2.

3.3. Carbon Materials

Various carbon materials have good electrical conductivity and are widely used as materials to produce electrodes like TCO electrodes, Au electrodes, and other metal electrodes. Electrodes based on carbon materials are also applicable for chemical modification of the surface to provide selective coupling of analytes. And since carbon-based nanomaterials not only have good electrical conductivity but also provide a high specific surface area and the possibility of modification, they are also often used to form nanocomposites. This section will review the applications of various carbon materials with chemically modified surfaces for analyte biorecognition.
Carbon electrodes, as in the case of Au electrodes, are applied using the screen-printing method [80,81,82]. Moreover, to increase the specific area, screen-printed carbon electrodes (SPCE) based on graphite can be modified with carbon nanotubes [81,82,83,84]. It should be noted that other types of electrodes could also be modified with carbon nanotubes, such as, for example, the gold electrodes described in [85] or the ITO electrodes described in [86].
Senturk et al. [81] presented an impedimetric aptasensor for the detection of lysozyme (LYZ), in which the peptide aptamer was immobilized on the surface of screen-printed electrodes made of carbon nanofibers (DropSens, Oviedo-Asturias, Spain). During the passive adsorption of reagents, the formation of peptide bonds between the carboxyl groups of CNF and the NH2 groups of the aptamer occurred. When the aptamer complexed with LYZ, the reaction blocks electron transfer at the interface. Thus, an increase in concentration led to an increase in Rct, from which the quantitative response of the biosensor was determined. Under optimal conditions, the LOD of the aptasensor is 0.36 μg/mL. The aptasensor was also used to determine LYS in fetal serum samples of cattle with an LOD value of 1.89 μg/mL.
In [87], a working electrode made of multi-walled carbon nanotubes was modified with graphene quantum dots and gold nanoparticles (AuNPs@MWCNTs-GQDs). This working electrode was used to create an impedimetric immunosensor for the detection of a prostate cancer biomarker in human serum—a prostate-specific antigen (PSA). Activation of the carboxylate group of the nanocomposite was carried out in a solution of EDC and NHS. The AuNPs@MWCNTs-GQD electrode was incubated for antibody binding for 45 min. Blocking of nonspecific binding sites was performed using BSA. Impedance gain was linearly related to PSA concentration over a wide range from 1 to 10,000 pg/mL with an LOD of 0.48 pg/mL. The authors note that modification of the electrodes not only leads to an increase in the measured signal level due to an increase in the rate of electron transfer but also increases the ability of the electrode surface to adsorb bioactive substances. Carbon nanotube MWCNTs were also modified with AuNPs in [86].
Glassy carbon [38,48] and graphite paper [88,89,90] are also used as electrodes based on carbon materials. And in [90], graphite paper was modified with AuNPs. In [59], in addition to glassy carbon, boron-doped diamond was also used as an electrode, as well as in [91] for the detection of SARS-CoV-2 S1.
In [88], a highly sensitive immunosensor based on a disposable graphite paper (GP) electrode modified with C60 fullerene was developed to determine the suppression of tumorigenicity 2 (ST2) in human serum. GP electrodes were coated with a solution of C60 fullerene (in toluene) and thoroughly dried with pure argon gas. To break the carboxyl groups on the electrode surface, it was incubated with an H2SO4 solution overnight in the dark. The electrodes were then incubated in EDC/NHS in phosphate buffer for 1 h. To immobilize the electrodes, they were immersed in anti-ST2 solutions at room temperature for 60 min. BSA was used as a blocking agent. The overall design of the ST2 immunosensor is presented in Figure 6. The ST2 impedimetric immunosensor demonstrated excellent repeatability, reproducibility, and a wide detection range (0.1 fg/mL to 100 fg/mL). The LOD and limit of quantitation (LOQ) values were 0.124 fg/mL and 0.414 fg/mL, respectively.
In [92], graphene foam was used as a flexible electrode in a microfluidic laboratory-on-a-chip to detect interleukin 10 (IL-10). In this work, the conductivity of graphene was maintained by noncovalent π–π functionalization with pyrene carboxylic acid (PCA). Graphene foam Gii-sense® (Integrated Graphene Ltd., Eurohouse, UK) electrode working areas were directly grown on polyimide substrates without the use of a catalyst and a transfer process. A solution of PCA in dimethylformamide (DMF) was used to functionalize the working electrode. The electrode was immobilized with a polyclonal anti-IL-10 antibody. Nonspecific sites were blocked with ethanolamine. The PCA antibody modification showed that the graphene foam withstands the incubation, measurement, and washing performed in a microfluidic device. Also, when studying the working electrode using the contact angle method, it was found that the wettability of the graphene surface gradually increased after each modification step. The Gii-sense®/PCA/anti-IL-10 biosensor was tested with artificial saliva samples containing various concentrations (from 10 to 100 fg/mL) of IL-10. Automated detection measurements performed in a 3D-printed microfluidic device showed a linear relationship between 10 fg/mL and 100 fg/mL with an LOD value of 7.89 fg/mL in artificial saliva. Specificity was assessed against interleukin 6, TNF-α, and BSA.
Graphene oxide is widely used to modify the electrodes of impedimetric biosensors [93,94,95,96,97,98]. Filip et al. [93] studied the anchoring of the lectin concanavalin A (ConA) on the surface of electrochemically reduced graphene oxide (ErGO) with thionine (Thi) using glutaraldehyde (GA) as a crosslinker and the use of this method for impedance detection of invertase glycoprotein (INV), which has an affinity to ConA. It has been established that the application of ConA/GA complexes to the ErGO/Thi surface allows the biosensor to achieve linear characteristics in the INV concentration range from 10−14 to 10−8 mol. These results indicated the possibility of attaching biorecognition molecules to weakly reduced or unreduced graphene oxide. In [94], a nanocomposite was prepared from reduced graphene oxide (rGO) and gold nanoparticles (AuNP) to create a disposable neurobiosensory probe for the determination of the Tau-441 protein, which is a marker of Alzheimer’s disease. The surface of the nanocomposite (rGO-AuNP) was modified with 11-mercaptoundecanoic acid (11-MUA), acting as a covalent anchor to improve the sensitivity of the assay. Activation of the end groups was carried out in an EDC/NHS solution. The electrodes were incubated in an anti-Tau solution for 60 min at room temperature. To block nonreactive ends, electrodes were incubated in a BSA solution. The immunoreaction of Tau-441 with anti-Tau was monitored in real time using single-frequency impedance (SFI). The developed immunosensor demonstrated a linear response in the concentration range of 1–500 pg/mL and an LOD value of 0.091 pg/mL.
In addition to the use of boron-doped diamond, a nonstandard carbon nanomaterial was used to create a biosensor in [99]. It reports a new impedimetric immunosensor for the detection of protein D in purified and bacterial (Haemophilus influenzae, Hi) samples. Recognition was based on the immobilization of anti-protein D antibodies (apD) onto maze-like boron-doped carbon nanowall electrodes (B:CNW). The SEM image of the sensor is shown in Figure 7. The B:CNW electrodes were prepared by CVD using diborane (B2H6) as a precursor for acceptor doping. Other precursors were H2 and CH4 gases. The resulting B:CNW electrodes were modified in an electrochemical cell in a diazonium salt solution to produce B:CNW/C6H5COOH. Electrografting on B:CNW was carried out with polarization from 0 V to 1 V 11 times. To immobilize ligands on the B:CNW/C6H5COOH electrodes, an apD solution together with a carboxyl group activator (EDC/NHS) was applied to the electrode surface. The immunosensor demonstrated a linear antigen detection range. The LOD of protein D detection was 2.39 × 102 fg/mL with a linear range from 3.37 × 10−11 to 3.37 × 10−3 μg/mL (in the purified sample).
An interesting heteronanostructure (HsGDY@NDs) was obtained by mixing nanodiamonds (ND) and hydrogen-substituted graphdiyne (HsGDY) to create an impedimetric aptasensor for the detection of biomarkers: myoglobin (Myo) and cardiac troponin I (cTnI) [100]. HsGDY@ND (Figure 8) consists of nanospheres of 200–500 nm size, in which NDs are embedded in the HsGDY network. The HsGDY@NDs nanostructure combines the good chemical stability and 3D porous network morphology of HsGDY. The resulting sensor provided an LOD of 6.29 fg/mL and 9.04 fg/mL for cTnI and Myo, respectively, high selectivity, stability, and reproducibility.
Examples of impedimetric biosensors using carbon materials are given in Table 3.

3.4. Polymers

Functionalization of conductive electrodes can also be carried out by modifying them with polymers [101,102,103,104,105,106,107,108,109,110]. This approach is often used to modify conductive electrodes directly with macromolecules containing functional groups to form SAMs. In this case, thin polymer layers deposited on the electrodes must have a low enough resistance to allow the charge transfer process to occur for the faradaic EIS mode.
The work [109] describes an impedimetric immunosensor for the determination of CCR4, a biomarker for prostate cancer. The sensor was formed by deposition of acid-substituted poly(pyrrole) P(Pyr-Pac) polymer on an ITO electrode with subsequent immobilization of anti-CC chemokine receptor 4 (CCR4) antibodies. P(Pyr-Pac) contained many carboxyl groups for antibody attachment. To create the ITO sensor, the electrodes were hydroxylated (in a H2O:NH4OH:H2O2 solution). Then, the electrode was immersed in a solution of the P(Pyr-Pac) polymer to form self-assembled polymer monolayers with the formation of an ester bond. Immobilization of anti-CCR4 antibodies was achieved using a binding reagent (NHS) and a crosslinking reagent (EDC). The electrode was modified with BSA molecules to block the carboxyl end groups. The developed immunosensor demonstrated a linear EIS response range to CCR4 concentration ranging from 0.02 pg/mL to 8 pg/mL and an LOD of 6.4 fg/mL. A slightly different approach is described in [106], where a polypyrrole polymer containing epoxy active side groups (PPCE) was conjugated to an ITO electrode to detect interleukin 6 (IL-6). For ITO coating, the synthesized pyrrole monomer containing an epoxy active side group (PCE) was subjected to electropolymerization, after which the PPCE-modified electrode was immersed in IL-6 receptor solution and then into BSA solution to prevent nonspecific binding. The biosensor demonstrated a linear range of IL-6 of 0.02–16 pg/mL, an LOD of 6.0 fg/mL, and good selectivity for other interference biomarkers.
Carboxylated polypyrrole was also used to create biosensors in [107,108]. Also described is the use of polypyrrole with epoxy groups in the form of a polymer composite with acetylene black [111]. In [112], overoxidized polypyrrole modified with gold nanoparticles was used. In [113], polypyrrole in the form of nanotubes was used to create a sensor. In addition to polypyrrole, a number of conducting polymers, such as polythiophene [110,114], polyaniline [115], PEDOT [116], and PEDOT:PSS [117], are used to create impedimetric biosensors. At the same time, the small thickness of the polymer layers used to create a biosensor allows the less conductive polymers to be used for modification as well.
In [103], gold electrodes were coated with polytyramine. Polymerization of 25 mM polytyramine in 0.3 M NaOH was carried out on the working electrode using cyclic voltammetry. The potential was varied twice from 0 to 1.6 V and then back to 0 V at a scan rate of 200 mV/s. The resulting polytyramine-modified gold electrodes were used to create an impedimetric biosensor for fibroblast growth factor receptor 3 (FGFR3). To create the biosensor, affimer reagents were isolated, and affimer proteins were synthesized that exhibit strong binding to FGFR. Biotinylated non-affimers were attached to the electrode surface via neutravidin–biotin. The developed impedimetric biosensors showed the ability to detect sub-pM to nM concentrations of recombinant FGFR3 protein in phosphate-buffered solution, as well as in synthetic urine.
A similar architecture of an impedimetric biosensor is presented in [104]. The authors proposed an impedimetric biosensor for the specific detection of Alzheimer’s beta-amyloid oligomers (AβO). The biorecognition element was a fragment of the cellular prion protein (PrPC). The coating of the electrodes in this work was carried out using electropolymerization. To deposit a copolymer of polytyramine/3-(4-hydroxyphenyl) propionic acid (POPA), sonicated electrodes were immersed in methanol containing 18.75 mM tyramine, 6.25 mM 3-(4-hydroxyphenyl) propionic acid, and also 0.3 M NaOH. After electropolymerization, the POPA-modified gold electrodes were washed with water and dried under an argon flow. Impedance spectroscopy using electrodes with POPA showed a decrease in Cdl and Rct to approximately half of the corresponding values for polytyramine, which reduces losses. Biotinylated PrPC was linked to the POPA copolymer also through a biotin/neutravidin bridge. This made it possible to attach biotinylated PrPC with high affinity (Kd~10−15 M) as a bioreceptor. The developed biosensor showed specificity for the detection of synthetic AβO, demonstrating a linear response.
Hassan-Nixon et al. [102] presented a label-free affinity impedimetric biosensor using the zwitterionic polymer pCBMA for the quantitative detection of interleukin-8 (IL-8) present in nasal epithelial fluid (NELF). The entire sensor formation procedure is presented in Figure 9. The biosensor showed a wide detection range of the target protein, ranging from 55 fM to 55 nM with an LOD of 10 fM. The biosensor showed high selectivity for IL-8 and 90% reproducibility when using untreated NELF samples. The time to obtain the result is 15 min.
To provide a wide detection range of a polymer-based biosensor, nanocomposites are synthesized using nanomaterials [111,114,118,119]. Thus, Aydın et al. [118] presented a simple and reliable immunosensor for detecting the smallest change in cytokeratin fragment subunit 19 (CYFRA 21-1), a biomarker for lung cancer. The proposed immunosensor was fabricated using a C45-PTNH2 (carbon black C45/polythiophene containing amino-terminal groups) nanocomposite, which has excellent biocompatibility, is low cost, and has an electrically conductive surface. The biorecognition molecules of anti-CYFRA 21-1 were attached to the electrode utilizing the amino-thermal groups of the PTNH2 polymer via a relatively simple procedure. The concentration range for Rct was 0.03 pg/mL to 90 pg/mL. The LOD and LOQ of the proposed system were 4.7 fg/mL and 14.1 fg/mL, respectively.
Biopolymers can also be used in the development of impedimetric biosensors to modify the working electrode. In [120], electrodes based on graphite paper were modified with polyglutamic acid using electropolymerization and activated in an EDC/NHS solution to create a leptin biosensor. Xia et al. [121] used biotinylated phenylalanine (amino acid) monomers to modify the electrode and self-assemble into various nanostructures by controlling the pH of the solution. In particular, the strong interaction between biotin and streptavidin (SA) linker on the electrode surface initiated the formation of networks. The method was successfully used to monitor caspase-3 activity in HeLa cells treated with various anticancer reagents. In [117], to create a flexible, fully organic, and biodegradable impedimetric biosensor, the electrode was formed using a biocomposite based on photoreactive silk sericin combined with a conductive polymer. Functional electrodes were printed on flexible fibrion substrates. The biosensor was developed for the analysis of vascular endothelial growth factor (VEGF) by immobilizing anti-VEGF on a conductive matrix. In [122], to detect the S1 protein of the SARS-CoV-2 virus spike, a carbon electrode was modified with gelatin.
In addition to antibodies, other types of biorecognition elements, such as aptamers, peptides, molecular imprinted polymers (MIPs), etc., are used. They can be synthesized in the presence of a template (analyte) and are a promising alternative to biomolecules in the design of biosensors [119]. Since the principle of creating biosensors based on MIPs differs from the immobilization of working electrodes based on other materials, including polymers [113,119,123,124,125,126,127], in this section, we highlight MIPs as a material for electrode modification.
As an example of the use of MIPs to create an impedimetric biosensor, let us consider a biosensor for determining Tau protein (p-Tau-441), a biomarker of Alzheimer’s disease [123]. A MIP was deposited on carbon SPE by electropolymerization of 3-aminophenol (AMP) in the presence of a protein matrix (p-Tau). After washing the resulting AMP polymer with p-Tau, incubation was carried out in a solution of proteinase K to cleave p-Tau (for 2.5 h). After protein removal and washing, the electrochemical stabilization step was carried out. The detection limit of p-Tau-441 for MIP-based sensors was 0.02 pM in PBS buffer pH 5.6. The linear response range of the biosensor was between 2.18 pM and 2.18 nM.
In [113], a dual-template molecular imprinted polymer (DMIP) biosensor was fabricated by electropolymerization to detect carcinoembryonic antigen (CEA) and alpha-fetoprotein (AFP) as biomarkers of lung cancer. The protein matrices (CEA and AFP antigens) were removed by washing in sodium hydroxide solution. The polymer used was polypyrrole (PPy) with the addition of methyl orange (MO), which increased the conductivity of PPy and caused the formation of rectangular-shaped nanotubes (PPy-MO), as shown in Figure 10. The LOD values of 1.6 pg/mL and 3.3 pg/mL were obtained for CEA and AFP, respectively. It has been shown that the self-assembly of PPy-MO nanotubes made it possible to increase the specific surface area of the biosensor and thus further expand the response range.
A specific member of the synthetic organic compound polymers should be mentioned in this section. Thus, in [128], to create a biosensor for D-dimer, the carbon electrode was modified with negatively charged synthetic phospholipid-dihexadecyl phosphate and gold nanoparticles.
In [129,130], dendrimers were used to modify electrodes and for subsequent immobilization. For example, Erdem et al. [129] presented an aptasensor based on a graphite electrode coated with a polyamidoamine dendrimer (PAMAM) with 16 carboxyl groups on the surface (PAMAM G2). A DNA aptamer selective for activated protein C (APC) was immobilized onto the dendrimer. The sensitivity of the system was 0.74 μg/mL (0.46 pmol in 35 μL of sample) in buffer solution and 2.03 μg/mL (1.27 pmol in 35 μL of sample) in serum. In [131], fifth-generation polyamidoamine dendrimers PAMAM (G5) were used as part of a nanocomposite, which was used to detect Fetuin-A (HFA) in real blood samples. To manufacture the biosensor, a coating was formed on the surface of the gold electrode by sequentially applying a self-assembly monolayer (SAM) of 4-aminothiophenol (4-ATP), fullerene, and PAMAM-NH2 (G5, generation 5). As can be seen from Figure 11, when fullerenes are deposited, a large increase in the specific surface area of the electrode is observed. Before applying polyhydroxyl fullerenes (C120O30(OH)30), a fullerene solution was incubated in HEPES buffer with the addition of EDC and NHS to activate OH groups. To obtain the final structure, a solution of fifth-generation polyamidoamine dendrimers PAMAM(G5) in phosphate buffer pH = 6.8 was applied to the AuE/4-ATP/fullerene electrode and incubated for 1 h. The activation of amino groups of AuE/4-ATP/fullerene/PAMAM was carried out in a glutaraldehyde solution. Anti-Fetuin-A antibodies were incubated for 2 h. Open groups of glutaraldehyde were blocked with the amino acid glycine. The LOD and LOQ of the biosensor were calculated as 0.48 ng/mL and 1.46 ng/mL, respectively.
Examples of impedimetric biosensors using polymeric materials are given in Table 4.

3.5. Metal Complexes of Porphyrins, Phthalocyanines, and Metal–Organic Frameworks

In addition to polymers, dendrimers, and SAMs, we can highlight the use of metal–organic frameworks (MOFs) [132,133,134,135], as well as metal complexes of porphyrins [136] and phthalocyanines [137,138].
In [136], an impedimetric biosensor based on a zirconium porphyrin complex was developed for the detection of neuron-specific enolase (NSE). A complex based on Zr(III) and tetra(4-aminophenyl)porphyrin (Zr-TAPP) provided a strong affinity for the antibody (anti-NSE). To create the biosensor, the Zr-TAPP dispersion was deposited on a Au electrode (AE) and dried. The resulting Zr-TAPP/AE was incubated in an anti-NSE solution at 4 °C for 40 min. Anti-NSE/Zr-TAPP/AE was incubated with an NSE solution for further electrochemical measurements. A porphyrin-based Zr-MOF immunosensor (PCN-224) was similarly constructed. The developed biosensor demonstrated an NSE LOD value of 7.1 fg/mL and a linear range from 10.0 fg/mL to 2.0 ng/mL.
Centane et al. [137] developed a biosensor for human epidermal growth factor receptor 2 (HER2) and studied the use of cobalt tetraphenylacetic acid phthalocyanine and cobalt tetraphenylpropionic acid phthalocyanine, which, after activation in DCC/NHS, were used as a platform to immobilize the HB5 aptamer. Also, to increase the specific area of the sensor, phthalocyanines were associated with nanoparticles based on cerium oxide. The aptasensors showed good LOD values for HER2 (all less than 0.2 ng/mL) with high stability.
Vijayaraghavan et al. [134] developed an impedimetric immunosensor based on bimetallic amine functionalized FeCo metal–organic frameworks uniformly grown on porous solid nickel substrates (FeCo-MOF/NF) as a transducer for the detection of oral squamous cell carcinoma (OSCC). The functional group providing selectivity and rapid sensitivity to the target analyte is IL-1RA. The antibody-immobilized FeCo-MOF/NF electrode was the working electrode. To fabricate the sensor, commercial nickel foam (NF) was cut into 3 × 3 cm pieces. The growth of FeCo-MOF nanoarrays on a porous nanofilter (denoted as FeCo-MOF/NF) was performed using a hydrothermal reaction. The original nickel foam and the resulting FeCo-MOF/NF are shown in Figure 12. The immunosensor showed a wide linear dynamic range for IL-1RA detection (10 fg/mL to 10 ng/mL) with an LOD value of 7.30 fg/mL in buffer and 7.22 fg/mL in serum conditions and an LOQ value of 22.14 fg/mL in PBS and 21.88 fg/mL in serum.
Gupta et al. [133] used Cu3(BTC)2-based metal–organic frameworks (MOFs) embedded in a conductive polymer (PANI) as a coating on the electrode surface. This composite was used to develop an impedance sensor for the cardiac biomarker troponin I (cTnI). An electrically conductive composite thin layer was coupled with anti-cTnI antibodies. Cu3(BTC)2 (where BTC = benzene-1,3,5-tricarboxylate) was synthesized at room temperature from H3BTC (benzene-1,3,5-tricarboxylic acid) and Cu(NO3)2 using the solvothermal method at room temperature as a solution in distilled water, ethanol, and dimethylformamide with the addition of triethylamine. The resulting immunosensor made it possible to selectively register cTnI in a clinically significant concentration range from 1 to 400 ng/mL. The cTnI analysis time was 5 min.
Examples of impedimetric biosensors using metal complexes of porphyrins, phthalocyanines, and metal–organic frameworks are given in Table 5.

3.6. Metal Oxides and Dielectrics

Metal oxides (MO) and nanomaterials based on MO are often used as substances for the functionalization of surfaces and subsequent functionalization. Metal oxides are typically highly electrically resistive semiconductors and are primarily used for biosensors operating in the non-faradaic EIS mode, in which the main contribution to the impedance spectrum is the capacitance change. Nanomaterials made of metal oxides, like other nanomaterials, provide a high specific surface area for immobilization. Among metal oxides, nanoparticles and nanostructures of TiO2, ZnO, WO3, and others are most often used [51,139,140,141,142,143,144,145]. The charge transfer resistance Rct in non-faradaic EIS biosensors can tend to infinity if dielectric materials are used for immobilization [146,147,148]. For the same purpose, electrodes can be separated from semiconductor nanomaterials by a dielectric layer, such as glass.
The work [149] presents a biosensor for biomarkers of sepsis-procalcitonin (PCT) and C-reactive protein (CRP) in human serum and blood. The biosensor is based on recording the impedance of the reaction between the target analyte and a specific probe on the surface of zinc oxide (ZnO). Interdigitated gold electrodes on a polyimide substrate were used as electrodes in the biosensor. To increase sensitivity, a thin ZnO film was deposited on gold electrodes using RF magnetron sputtering. Thiol-based crosslinker molecule dithiobis (succinimidyl propionate) (DSP) in DMSO was incubated at room temperature in the dark for 90 min for effective thiol functionalization. Each sensor was individually functionalized with a PCT- or CRP-specific capture probe. Finally, 10 μL of Superblock was added to the sensor and incubated for 15 min to hydrolyze any unbound binding sites. A frequency of 10 Hz was selected for the calibration dose response of PCT in human serum and whole blood, and a frequency of 100 Hz was selected for CRP in human serum and whole blood. The LOD values for PCT and CRP were 0.10 ng/mL and 0.10 μg/mL, respectively.
In [141], a titanium dioxide nanotube (TNT) matrix was used to create an impedimetric sensor for the detection of human interleukin 6 (IL-6), interleukin 8 (CXCL8, IL-8), and tumor necrosis factor (TNFα). Antibodies were used as recognition elements. To form TNT, titanium foil was sonicated in acetone, ethanol, and distilled water and dried in nitrogen. The formation of oxide layers of nanotubes was carried out by electrochemical anodization in solutions of ethylene glycol with NH4F. The resulting titanium dioxide nanotube array is shown in Figure 13. Antibodies were immobilized onto TNT by physical adsorption. The immunosensor demonstrated good selectivity and high sensitivity up to 5 pg/mL, which is the standard concentration of analyte proteins in human blood. The detection range for the three cytokines was 5–2500 pg/mL in 0.01 M phosphate buffer (pH 7.4).
Chen et al. [51] developed a biosensor to detect caspase-9, an enzyme that is involved in apoptosis or programmed cell death. In this work, metal oxides were used in the form of nanoparticles, and they were separated from the electrodes by a thin cover glass. To prepare the biosensor, a suspension of ZnO nanoparticles and CuO nanoparticles was applied to the cover glasses by dripping and then dried in an oven at 50 °C for 4 h. The resulting coating is shown in Figure 14a. For immobilization, an antibody solution (rabbit anti-cleaved caspase-9 IgG) was dropped onto the sensor area and incubated for 3 h at room temperature. The biosensor was positioned over a pair of coplanar electrodes formed from 35 μm copper on a printed circuit board. A cross-section of the biosensor is shown in Figure 14b. The sensor response was assessed using readings obtained at a frequency of 6 MHz. As a result, the sensor demonstrated an LOD value of about 0.07 U/mL caspase-9 concentration. The average analysis time was 45 min, including preparation. A similar biosensor structure based on ZnO/CuO nanoparticles was used in [150] to create a biosensor for C-reactive protein.
ZnO and CuO nanoparticles were also used to create a biosensor for C-reactive protein (CRP) (CRP) in [151]. However, in this work, a 3D membrane was fabricated based on nitrocellulose membranes and ZnO and CuO nanoparticles. The use of a 3D membrane aims at more efficient capturing of antibodies by monoclonal mouse anti-human C-reactive protein (4C28). Cellulose membranes were immersed in nanoparticle suspensions and treated with ultrasound to form a nanoparticle-based 3D membrane. The sonicated nano-ZnO and nano-ZnO/CuO nitrocellulose membranes showed detection limits of 27 and 16 pg/mL, respectively. The authors noted the ease of fabrication of this sensor, the lack of contact between the immobilized nanomaterial and the electrodes, and the absence of the need to use a redox active species for measurement.
In [147], immobilization was carried out directly on a glass substrate. An impedimetric biosensor was fabricated with Au electrodes for the quantification of human serum albumin (HSA). The glass sensing surface between two adjacent Au electrodes was modified with 3-aminopropyltriethoxysilane (APTES) to silanize and subsequently bind to anti-human serum albumin (AHSA). Before APTES modification, the samples were cleaned and treated with a solution of H2SO4 and H2O2 (1:3) to functionalize the surface with hydroxyl groups. The HSA LOD value limit of the developed biosensor was about 2 × 10–4 mg/mL.
In [146], a biosensor for detecting HSA was fabricated based on a dielectric, silicon nitride Si3N4, and anti-HAS recognition agent. In general, the electrode had a Si-p/SiO2/Si3N4 structure. The Si3N4 sample was similarly treated with a Piranha solution of H2SO4 and H2O2 (1:3) for modification with reactive OH groups. Silanization was carried out using triethoxysilane aldehyde (TEA) in the vapor phase. The detection limit of HSA was 10–14 M.
Examples of impedimetric biosensors implemented using metal oxides and dielectrics are given in Table 6.

3.7. Other Nanomaterials

This section describes nanomaterials that are somewhat less common but are also successfully used to create impedimetric biosensors. Such materials include colloidal quantum dots [152,153], which during the synthesis process are formed with a shell of ligand molecules, two-dimensional nanomaterials from other semiconductor compounds [154,155,156], and, finally, magnetic nanoparticles [157,158].
In [153], a biosensor for detecting Zika virus envelope protein (EP-ZIKV) was developed based on an electrode with carboxylated CdTe QDs. Quantum dots were obtained in the form of an aqueous colloidal dispersion using mercaptosuccinic acid as a stabilizing agent (ligand molecules). The size of CdTe QDs was estimated to be 3.3 ± 0.4 nm. CdTe QDs were deposited on a polymer-coated electrode. A film of polymer with amine groups 2-(1Hpyrrol-1-yl)ethanamine (Pyam) was deposited onto the electrode electrochemically. QDs were immobilized on the surface of Pyam in the presence of EDC/NHS to promote covalent bonding between the carboxyl groups of the QDs and the amine groups of the PPyam polymer film. Anti-EP ZIKV antibodies were immobilized on QDs through carboxyl groups activated in the previous step. The general preparation scheme of the electrochemical immunosensor is shown in Figure 15. It is reported that Anti-EP ZIKV antibodies were effectively immobilized on the surface of PPyam/QD even after 2 months of electrode storage. The biosensor had high sensitivity with an LOD value of 0.1 ng/mL and demonstrated good specificity.
A 2D semiconductor material was used to create a biosensor in which MoS2 nanoflower (nf) was used to create a substrate surface for direct immobilization of antibodies for binding tumor necrosis factor-α (TNF-α) [154]. The synthesis of MoS2 nfs was carried out by a one-step hydrothermal method for 24 h. Then, MoS2 was washed with a mixture of ethanol and water to separate impurities, dried at 60 °C for 5 h to obtain a black powder, and ground to obtain a homogeneous powder. The morphology of the resulting MoS2 was determined from electron microscopy (Figure 16). Then, a colloidal solution of MoS2 nfs was prepared in acetonitrile under ultrasonication. And a hydrolyzed electrode was deposited on the ITO at a potential of 50 V using a platinum plate as the second electrode. As a result, a uniform layer of MoS2 nfs was formed. Anti-TNF-α antibodies were then immobilized on the electrodes by incubation for 6 h. The morphology of the nanoflowers facilitated the immobilization of antibodies through physical adsorption without the use of linkers. The BSA/anti-TNF-α/MoS2 nf/ITO immunoelectrode demonstrated the value of LOD of 0.202 pg/mL and a linear detection range of TNF-α of 1–200 pg/mL with fairly high selectivity.
Mi et al. [155] used a hybrid of AuNPs and Ti3C2-MXenes to fabricate aptasensors for cTnI and Myo. Before fabricating the biosensor, the ITO electrode was amino-functionalized using APTMS, and a hybrid suspension was synthesized. To fabricate the biosensor, a hybrid AuNP/Ti3C2-MXene suspension was applied to the surface of an APTMS/ITO electrode and left for 30 min at room temperature, followed by washing and drying in an N2 stream. The AuNP/Ti3C2-MXene nanohybrid was anchored on APTMS/ITO via a Au–N bond. Next, the electrode was incubated in the aptamer solution. For the cTnI aptasensor, blocking of unbound regions was performed using MCH. For the aptasensor to Myo, blocking of unbound regions was performed with BSA. AuNP/Ti3C2-MXene captured a cTnI-specific, thiol-functionalized DNA aptamer via Au-S self-assembly, and the Myo-aptamer was bound via adsorption and metal–halate interaction between phosphate groups and titanium. The developed aptasensors demonstrated specificity and good sensitivity. The LOD for cTnI was 0.14 fg/mL and for Myo was 0.2 ng/mL.
In [157], particles with a magnetic core and a gold shell were used to create a biosensor for PSA. Ferrite cores Mn0.61Zn0.42Fe1.97O4 were synthesized by hydrothermal method and then encapsulated into silica. The Au nanoshell was deposited using the seed-and-growth method. MP@silica@Au and the Au electrode surface were modified using carboxybetaine aryldiazonium derivative (CB) to obtain the SAM layer. For clarity, all stages are presented in Figure 17. At the first stage, a SAM layer carrying carboxybetaine was formed using cyclic voltammetry, which subsequently served for covalent immobilization of the lectin. The bottom row shows the MP@silica@Au composite spontaneously modified with a CB derivative and then with an antibody (Ab, anti-fPSA, chemically treated). After fPSA enrichment from a human sample using MP@silica@Au/CB/Ab, a sandwich was prepared using a permanent magnet, and the signal was assessed electrochemically. The developed biosensor demonstrated the ability to detect PSA down to 1.2 fM.
Examples of impedimetric biosensors using other nanomaterials and dielectrics are given in Table 7.

4. Biorecognition Elements for Protein Detection

Since impedimetric biosensors respond to the binding of an analyte to a biorecognition element (or ligand), these recognition elements must have the following set of qualities: (i) high affinity and specificity for the target protein to ensure reliable and sensitive detection, (ii) high stability for maintaining activity and functionality over long periods of time and under different operating conditions, (iii) simplicity and efficiency of immobilization, that is, the ability to be easily and effectively attached to the electrode surface with minimal impact on their activity and functionality to ensure optimal exposure of binding sites and minimization of undesirable side reactions, and (iv) low cost for mass production with low costs of materials and technology.
There are several classes of recognition elements that can be used for impedimetric detection of proteins, such as antibodies, aptamers, peptides, enzymes, affimers, affibodies, nanobodies, and molecularly imprinted polymers [159,160,161,162,163,164,165]. The binding of the analyte to the recognition element leads to a change in the electrical properties on the electrode surface. These changes may be due to various factors, including the following:
  (i).
Change in thickness and dielectric constant of the recognition element layer. The binding of the analyte to the recognition element increases the thickness and dielectric constant of the layer, which leads to a change in capacitance [166] and an increase in resistance at the electrode surface [167]. This effect is observed when using thin layers of recognition elements such as antibodies or aptamers [168,169].
 (ii).
Change in charge and conductivity of the recognition element layer. The binding of the analyte to the recognition element can change the charge and conductivity of the layer, which leads to a change in the potential and current at the electrode surface [170]. This effect is observed when using charged recognition elements such as nucleic acids [171] or peptides [172].
(iii).
Change in the conformation of the recognition element. The binding of an analyte to a recognition element can cause a change in its conformation, which leads to a change in capacitance and resistance to direct current on the surface of the electrode [173,174]. This effect is observed when using recognition elements that can change their structure when binding to the analyte, such as aptamers or enzymes [29,175,176].
Biorecognition elements for impedimetric biosensors can be classified according to various criteria, including the following:
 (i).
Origin. Recognition elements can be natural or synthetic [177]. Natural recognition elements such as antibodies or enzymes usually have higher specificity and selectivity but are also more expensive, difficult to obtain and store, and are subject to degradation [178]. Synthetic recognition elements, such as aptamers, peptides, and nanobodies, generally have a lower cost, greater ease of synthesis and modification, and higher stability and regenerability [179].
(ii).
Size and molecular weight. Recognition elements can have different sizes, from several nanometers to several hundred nanometers, and different molecular weights. The size of the recognition element affects its properties, such as affinity, binding kinetics, mass transfer, electron transfer, and signal-to-noise ratio [180,181]. In general, the smaller size and molecular weight of the biorecognition element lead to faster and more sensitive detection due to the greater number of binding sites immobilized on the electrode surface [182]. In addition, the larger size and molecular weight of the biorecognition element increases its steric specificity, thereby reducing the likelihood of unwanted interactions with other molecules. This is especially important when studying the binding abilities of peptides and aptamers.
When choosing a biorecognition element and its characterization, an important parameter is the method of its immobilization. Recognition elements can be immobilized on the electrode surface in various ways, such as physical adsorption, covalent binding, biotin–avidin interaction, electropolymerization, self-assembled monolayers, etc. [33,183,184,185,186]. The first two methods are the most common. During physical adsorption, biorecognition agents are attached to the surface of the transducer due to physical forces, such as van der Waals forces, hydrophobic interactions, electrostatic forces, etc. [187]. With covalent binding, biorecognition elements are attached to the surface of the transducer by forming covalent bonds with the surface of the electrodes. This method provides high stability, regeneration ability, and orientation of biorecognition elements but may require complex and expensive procedures, such as surface activation, selection of reagents, and control of reaction conditions [188]. Biotin–avidin interaction is a method in which a ligand is attached to the surface of a transducer using biotin and avidin (or streptavidin) [189]. Biotin is a small molecule that can be synthetically attached to a ligand, and avidin (or streptavidin) is a protein that has a high affinity for biotin. This method provides high specificity, stability, and regenerability of the biorecognition element but requires additional steps, such as biotinylation of the biorecognition element, immobilization of avidin (or streptavidin) on the surface of the transducer, etc. [190]. Electropolymerization is the electrochemical deposition of a polymer film material on the surface of an electrode with the simultaneous inclusion of biorecognition elements in the polymer matrix [191,192]. This method has advantages such as simplicity, stability, selectivity, and the ability to control the properties of the polymer film. However, this method also has disadvantages, such as a limited selection of biorecognition elements and a possible decrease in their activity [193]. Monolayer self-assembly is a method in which a ligand is attached to the surface of a transducer using molecules that spontaneously form an ordered monomolecular layer on the surface [194,195]. These molecules have two functional groups: one group binds to the surface of the transducer, and the other group binds to the biorecognition element. This method provides high orientation, stability, and regenerability of the ligand but requires careful selection of SAM molecules, control of concentration, temperature, time, etc. [196].
The immobilization method affects the density, orientation, activity, and stability of the recognition element, as well as the electrical properties of the electrode [184]. In general, covalent binding provides higher density, stability, and reproducibility of biorecognition elements on electrodes but also lower activity and electron transfer than physical adsorption [185,186]. Therefore, the choice of ligand immobilization method depends on the specific goals and conditions of the analysis, as well as on the type of biorecognition element and transducer. In some cases, it may be advisable to use a combination of different immobilization techniques to achieve optimal results. For example, one can use physical sorption to create a first ligand layer and then chemical sorption to create a second ligand layer of a different type or orientation.
Next, we will consider the features of using various biorecognition elements for detecting proteins using impedimetric biosensors.

4.1. Impedimetric Detection of Proteins Using Antibodies

One of the most common biorecognition elements used in biosensors are antibodies. Antibodies are glycoproteins that are produced by the immune system in response to foreign agents (antigens) such as bacteria, viruses, toxins, and others [197,198,199]. Therefore, biosensors that use antibodies as a recognizer are often called immunosensors [200,201]. Antibodies have a Y-shaped structure of four polypeptide chains that are able to bind a specific antigen and form an antibody–antigen complex. Antibodies are capable of recognizing and binding to specific epitopes on the surface of an antigen with high specificity and affinity. Depending on how many epitopes of a single-antigen molecule they can recognize, antibodies can be either mono- or polyclonal [202]. Antibodies are widely used in immunosensors because they can detect a variety of protein analytes with high sensitivity and selectivity. However, antibodies have some disadvantages, such as high synthesis cost, difficulty in production, low stability, and the possibility of aggregation, degradation, and denaturation when immobilized on the electrode [203].
When creating impedimetric immunosensors, it is important to prevent possible nonspecific binding of the analyte to nontarget components, which introduces a spurious signal into the impedance measurements. For this purpose, free spaces on the surface of the chip that were not functionalized with antibodies are blocked with BSA [204], casein [205], twin-surfactants [206], commercial solutions such as SmartBlock™ [207], and other compounds.
The efficiency of antibody binding with the target protein may depend on the amount of immobilized ligands on the surface of the electrodes [208,209]. It was noted in [78] that excess antibody molecules to the Aβ42 protein bound on the surface created a steric barrier for the protein molecules to approach the antibody molecules in the correct orientation. Therefore, to achieve the required sensitivity, the authors had to use a fairly long incubation period of about 60 min.
Another important property of immobilized ligands that determines the efficiency of binding is their uniform orientation on the surface of the electrodes. Santos et al. [210] proposed a structure based on polypyrrole nanotubes (Figure 18) linked via Ni(OH)2 ligation to an engineered antibody heavy chain antigen-binding (VHH) moiety called Sb#15. This fragment is capable of selectively binding the receptor (RBD) of SARS-CoV-2. The developed structure allows the correct orientation of antibody immobilization for sensitive detection of the SARS-CoV-2 antigen. The position of the His tag on a protein can be controlled by genetic engineering, resulting in the uniform orientation of the antibody during immobilization. The developed sensor requires a small volume of saliva for analysis and provides results in 15 min without additional sample preparation with a sensitivity of up to 0.01 pg/mL. In [114], to detect the specific spike receptor protein (RBD), a structure based on a thiophene monomer with a substituted epoxy functional group was fabricated. It was electrodeposited onto a disposable indium tin oxide (ITO) platform in the presence of acetylene black. The resulting composite provided the correct attachment points for antibody binding and also supported the biosensor structure. The detection limit was 0.58 fg/mL. Adesina et al. [211] used a self-assembled monolayer (SAM) of 4-mercaptophenylboronic acid (4-MPBA) to orient immobilized anti-C-reactive protein (CRP) antibodies. Controlling ligand orientation was key to eliminating false positives and negatives during sample analysis. The resulting sensor made it possible to achieve a detection limit of 0.10 μg/mL.
One of the common problems of immunosensors is the storage stability of manufactured structures with immobilized antibodies [212,213,214]. Manufacturers and researchers are trying to increase it by creating new biosensor structures. Zhang et al. [215] presented a biosensor for the detection of apolipoprotein-A1 (Apo-A1), a biomarker for bladder cancer. The sensor is based on a glassy carbon electrode (GCE) coated with molybdenum and graphene nanoparticles on which antibodies to Apo-A1 are attached. The sensor also uses other lead and gold nanoparticles that are anchored with other Apo-A1 antibodies. When these antibodies bind to Apo-A1, they form a sandwich structure that changes the resistance of the electrode when an electrical current is passed through. The sensor showed a good dependence of resistance on Apo-A1 concentration in the range from 1 pg/mL to 1 μg/mL. To check the stability of the biosensor, the relative change in impedance %ΔRet was regularly measured. At a storage temperature of 4 °C, %ΔRet remained almost unchanged for a month and then gradually decreased and remained at the level of 90.1% after 2 months. The work [216] considers an immunosensor for detecting brain natriuretic peptide (BNP), which is a biomarker of cardiovascular diseases. The sensor consists of a graphite electrode coated with gold nanoparticles and graphene with chitosan, to which antibodies to BNP are covalently attached. The sensor showed high sensitivity and specificity for BNP in the range from 0.01 to 1000 pg/mL. The sensor was consistent with the commercial method and retained its activity for 60 days when stored under low-temperature conditions. The sensor did not respond to other proteins such as BSA or IgG. The sensor could be restored using an alkaline solution or pyranic acid.
Lee et al. [167] described the development of a sensor to detect ubiquitin C-terminal hydrolase L1 (UCH-L1), which is a biomarker of traumatic brain injury. The sensor is based on the use of antibodies immobilized on a self-assembled monolayer (SAM), which can be lyophilized and frozen with added sugar for long-term storage. The sensor has a disposable screen-printed electrode, which increases measurement reproducibility. The sensor operates on the principle of impedance spectroscopy with one optimal frequency, which simplifies the system and reduces measurement time to 5 min. The sensor has a low LOD value of 1 pM and a wide dynamic range from 1 pM to 10 pM, which makes it possible to distinguish between normal and elevated levels of UCH-L1 in the blood. The sensor does not require the separation of bound and free antibodies or washing. In artificial serum, the sensor is less sensitive due to the high salt concentration, which affects the solution resistance and the flow of the redox probe (ferricyanide). To overcome this problem, it is necessary to improve the reproducibility of sensor manufacturing.

4.2. Aptamer-Based Impedimetric Biosensors

Another common group of ligands for impedimetric biosensors is the aptamers. Aptamers are short, single-stranded nucleic acids (DNA or RNA) that can bind to a variety of molecular targets, such as proteins, peptides, amino acids, nucleotides, ions, and small molecules, with high specificity and affinity [26,217]. Aptamers are obtained using the systematic evolution of ligands with the exponential enrichment (SELEX) method and methods based on it [218,219]. Aptamers can also change their conformation upon binding to an analyte, which can lead to a change in electrical impedance at the electrode surface. Impedimetric aptasensors are fairly simple and fast diagnostic devices that do not require the use of labels and lengthy sample preparation [220].
Aptamers have a number of advantages over antibodies, such as low cost, chemical synthesizability, high thermal stability, and the possibility of modification and regeneration. The authors of [25] conducted a comparative study of two types of bioreceptors—an aptamer and an antibody—to detect the cancer marker HER2 on a glass–carbon electrode. The electrode is modified with graphene quantum dots with gold nanoparticles and a porphyrin binuclear framework. The aptamer and antibody are attached to them via an amide bond. Aptasensors showed better detection limit values than immunosensors. The developed aptasensor is also superior to the immunosensor in terms of repeatability and storage, while the immunosensor showed poor regeneration capabilities. Ramanathan et al. [221] presented a biosensor for diagnosing SARS-CoV-2 using aptasensing based on a gold interdigitated electrode (AuIDE) with nanodiamonds. The aptasensor can detect the nucleocapsid protein (NCP) of SARS-CoV-2 with high sensitivity and specificity. The detection limit of the aptasensor is 0.389 fM, and the linear range is from 1 fM to 100 pM.
Chemical modifications can be used to improve the sensitivity of aptasensors and improve the immobilization of ligands on the electrode surface. Januarie et al. [152] proposed an aptasensor for the detection of interferon-gamma, a biomarker of tuberculosis, based on SnTeSe metal dichalcogenide quantum dots (Figure 19). The surface of the quantum dots is modified with short chain L-cysteine peptides to improve stability and solubility. The LOD for the developed sensor was 0.151 pg/mL, and the response time was about 8 min. The aptasensor also had good selectivity for interferon-gamma in the presence of interference. When evaluated on a clinical sample, the aptasensor showed a good recovery range of 98–105%, indicating its suitability for infectious disease monitoring.
The work [222] describes a method for determining the protein marker of breast cancer MUC1 using an electrode coated with nanofibers, nanotubes, and gold nanoparticles with an attached aptamer that binds to MUC1. The method can detect MUC1 in the range from 5 to 115 nM with a detection limit of 2.7 nM. Aptamers are thermally and chemically stable and remain active after several cycles of denaturation/renaturation. The stability of the modified electrode was tested after storage in the refrigerator for 4 and 8 weeks. The selectivity of the aptamer has been confirmed in the presence of other biological molecules such as has, PSA, urea, glucose, and HIgG.
Thus, to create new effective impedimetric aptasensors, it is necessary to design new modified electrodes using nanotechnology and chemical modification. It is also necessary to develop new methods for selecting aptamers sequences.

4.3. Peptide “Aptamers”

Peptides are short chains of amino acids that can have specificity and affinity for certain proteins. Peptides can be obtained using chemical or biological synthesis, as well as using combinatorial chemistry or phage display [223,224]. Peptides can be modified with various functional groups such as thiols, amines, carboxyls, biotin, streptavidin, etc., as well as various electroactive labels [225] to ensure specific and stable binding to the electrode [226]. Peptides can be targeted on the electrode surface using various strategies, such as the use of spatial probes, multivalent ligands, selective receptors, etc., to ensure optimal exposure of binding sites [227].
Peptides have a number of advantages over more widely used antibodies and aptamers, such as low cost, ease of synthesis and modification, high stability and regenerability, low immunogenicity and toxicity, and the ability to be designed as needed [228]. However, peptides also have a number of disadvantages, such as low affinity and specificity compared to antibodies, low catalytic activity compared to enzymes, and the potential for denaturation, aggregation, and degradation under certain conditions. However, the advantages of peptides often outweigh their disadvantages, especially when compared to more complex immunoassays.
Peptide biosensors often do not require modification of electrodes using nanomaterials. An interesting example of an impedimetric peptide-based biosensor for detecting the S-protein of the SARS-CoV-2 virus is presented in [229]. The biosensor does not require nanomaterials or complex techniques but uses gold electrodes with a thiolated peptide whose interaction with protein S is measured by EIS. The biosensor showed high sensitivity and reproducibility, and the detection limit was 18.2 ng/mL for S-protein and 0.01 copies/mL for viral particles. The biosensor successfully identified protein S in samples from infected patients in 15 min without labels. The biosensor also has long-term stability (73.6% of the initial signal after 20 days of storage) and high repeatability and reproducibility (RSD 4.1% and 2.2%). Cho et al. [230] presented a biosensor for the detection of neutrophil gelatinase-associated lipocalin (NGAL), which is a marker of diabetic nephropathy. A peptide with the sequence DRWVARDPASIF, found using phage display, selectively bound to the target protein. The sequence was further modified with a cysteine at the C-terminus to form a self-assembled thiol monolayer, as well as a linker (-GGGGS-) for molecular flexibility on the gold surface. The affinity of the resulting peptide was measured using square wave voltammetry (SWV) and electrochemical impedance spectroscopy EIS. The LOD value by EIS was 1.74 ng/mL, and by SWV, it was 3.93 ng/mL. The performance of this biosensor is comparable to commercial ELISA tests, which was verified using real patient plasma samples (Figure 20).
The improvement of this technique for detecting NGAL is presented in [124], where a sensor based on a film of molecularly imprinted polymer (MIP) with a specific binding peptide (BP1) was developed on a gold quartz electrode. The authors used a surface imprinting method with photopolymerization and a PDMS template with hemispherical depressions. Using EIS (Figure 21), the values of LOD 0.07 μg/mL and LOQ 0.24 μg/mL were achieved. They determined that the selectivity coefficient (k*) of NGAL protein on BP1 peptide-imprinted film was 3.5–5.8 relative to BSA over the entire low concentration range, indicating high affinity between BP1 peptide and NGAL protein.
Additionally, peptide-based dendrimers can be used to detect target proteins. Thus, Matsubara et al. [231] developed a biosensor to detect human and avian influenza virus (IFV) using gold electrodes with peptides that mimic receptors for viral hemagglutinins (HA). The authors modified the density and structure of peptides on boron-doped diamond (BDD) electrodes to improve virus binding and measure charge transfer resistance using electrochemical impedance spectroscopy. The ligands used were dendrimers having pentapeptide units mimicking the sialyl glycoconjugate receptor for uptake of IFV HA. The electrode with a peptide coating based on BDD had a fairly high sensitivity (3−400 pfu/sample), as did the electrode with antibodies based on BDD (5−10 pfu/sample).

4.4. Other Types of Biorecognition Elements Used in Impedimetric Biosensors for Protein Detection

Enzymes. Enzymes are proteins that catalyze chemical reactions in living organisms. Enzymes can be used as recognition elements to detect their substrates, products, or inhibitors [232,233]. Enzymes have high specificity and activity and can also amplify the signal by generating a large number of reaction products. However, enzymes also have some disadvantages, such as being expensive, difficult to store, and susceptible to denaturation. However, impedance measurement is rarely used in enzyme-based biosensors compared to amperometry or potentiometry due to the relatively long time required to obtain the full impedance spectrum over a wide frequency range [234]. One of the most famous works in this direction is the biosensor from Chan et al. [235] for the determination of L-lactate based on a bioselective membrane of lactate dehydrogenase (LDH) and pyruvate oxidase (PyrOx). The biosensor is made by coating and crossing enzymes using glutaraldehyde vapor. The principle of operation of the biosensor is to measure the change in impedance caused by the formation of charged ions (CH3COO, H+, and HCO3) during the oxidation of pyruvate under the influence of PyrOx, pyruvate is formed during the oxidation of L-lactate under the action of LDH + nicotinamide adenine dinucleotide (NAD+). The detection limit was 17 and 20 μM for the resistance values at the interface between the two enzyme layers, RIM, and the corresponding capacitance, represented as CPEIM.
Peptide nanotubes. Peptide nanotubes are nanostructures that are formed from short peptides that self-assemble into tubular shapes [236]. Peptide nanotubes have unique electrical, mechanical, and biological properties that make them suitable for impedimetric biosensors. In [125], a biosensor was developed for the determination of interleukin 6 (IL-6), an important cytokine, based on a film of molecularly imprinted polymer with polydopamine (MIP(pDa)) on peptide nanotubes (PNT) deposited on screen-printed electrodes (SPEs). The authors used IL-6 as a template to self-polymerize dopamine without an initiator, enzyme, or crosslinker. They found that IL-6 could be detected in the range from 1 to 200 pg/mL with a good correlation between the redox couple and the logarithm of IL-6 concentration. The MIP(pDa)/PNT film had selective binding capacity for IL-6 with high recovery values in the urine sample.
Molecularly imprinted polymers. Molecularly imprinted polymers are synthetic polymers that have specific molecular imprints matching the structure and size of the analyte [237]. They are also distinguished as a separate type of biorecognition element. Molecularly imprinted polymers can be prepared through an imprinting process where the analyte is used as a template for the polymerization of monomers and then removed from the polymer matrix. Molecularly imprinted polymers have high selectivity, stability, and regenerability and can also be quite easily synthesized and modified. Choi, D. Y. et al. [127] demonstrated a biosensor imprinted with IL-1β protein, a cytokine associated with inflammation, on printed carbon electrodes (SPCEs) (Figure 22). They deposited a bilayer of poly(o-phenylenediamine) and poly(chromotrope 2R) with IL-1β as a template using cyclic voltammetry (CV) to form an MIP film. The authors studied the electrochemical properties of the sensor using CV and electrochemical impedance spectroscopy (EIS) methods and confirmed the imprinting effect on the MIP film. They showed that the MIP film sensor has high sensitivity to trace amounts of IL-1β (several pg/mL) with an LOD of 0.23 pg/mL and LOQ of 0.78 pg/mL. Balayan et al. [165] developed an electrochemical biosensor for the detection of sepsis in newborns based on MIPs selective for serum amyloid A (SAA) protein. They coated the screen-printed electrode with multi-walled carbon nanotubes (MWCNTs), manganese oxide nanospheres (MnO2NSs), and cobalt oxide nanoparticles (Co3O4NPs) to obtain a synergistic effect and high conductivity. They then polymerized MIP, specifically synthesized for SAA, onto the modified electrode. The created biosensor operates in the range from 0.01 pM to 1 µM with a detection limit of 0.01 pM and has a fairly long service life (42 days).
Affimers. Affimers are scaffolding proteins that are not antibodies but are capable of binding to other molecules with high affinity and specificity [238]. Affimers are obtained using the phage display method, in which those that have the highest affinity for the analyte of interest are selected from a library of random protein sequences [239]. Affimers have a small size (about 14 kDa) [240], high affinity (from subnano- to femtomolar binding constants), low immunogenicity, and the possibility of modification and functionalization. In addition, their advantage is their relatively low cost and ease of synthesis compared to antibodies. Shamsuddin et al. [105] presented an electrochemical biosensor for the detection of colorectal carcinoembryonic antigens (CEAs) based on polyoctopamine (POct), a nonconducting polymer with amine groups. The authors electropolymerized POct as a transmitter layer that allowed covalent binding of different bioreceptors, such as antibodies and synthetic affimer proteins, without surface activation. Affimers were fixed on the surface of gold screen-printed electrodes through the heterobifunctional crosslinking agent sulfo-SMCC (Sulfosuccinimidyl 4-(N-maleimidomethyl) cyclohexane-1-carboxylate). The close location of the bioreceptors to the transmitter layer significantly increased the sensitivity of detection. The sensitivity of the small bioreceptor (affimer, 12.6 kDa) to CEA was comparable to the large antibody (150 kDa) with a detection limit of 11.76 fM, much lower than the clinical level of 25 pM. However, the affimer-based biosensor had a narrower dynamic range than the immunosensor (1–100 fM vs. 1 fM–100 nM). All electrochemical measurements were performed in less than 5 min with a small sample volume (10 μL).
Affibodies. Affibodies are engineered proteins that are produced from the Z domain of protein A from Staphylococcus aureus [241]. Affibodies have a small size (about 6 kDa), high thermal stability, salt resistance, and the ability to regenerate [242]. Affibodies can be synthesized using phage display or recombinant DNA technology [243]. Affibodies can be engineered to bind to a variety of target molecules, including proteins. Ravalli et al. [244] created an electrochemical biosensor for the detection of HER2, a cancer biomarker, based on affibodies, which serve as bioreceptors, and gold nanostructured printed graphite electrodes. They anchored a cysteine-terminated affibody onto gold nanoparticles. They showed that the disposable biosensor has good analytical performance for HER2 detection in the range of 0 to 40 μg/L with a detection limit of 6.0 μg/L.
Nanobodies. Nanobodies are antibody fragments consisting of one monomeric variable domain of the antibody heavy chain [245,246]. They have a high ability to bind antigens and are the smallest functional fragments derived from natural immunoglobulins. Nanobodies are found in animals of the camelid family, in which the antibodies do not contain a light chain. Nanobodies have unique physicochemical and structural characteristics that make them ideal candidates for the development of diagnostic tests. Sánchez-Salcedo et al. [247] presented an electrochemical biosensor for the detection of IL-6 (Figure 23). The authors used a cheap, thin, eight-channel gold sensor array on a flexible substrate as electrodes. They used gold surfaces modified with an anti-IL-6 nanobody (anti-IL-6 VHH) or a specific IL-6 aptamer. In the first system, the nanobody was covalently linked to the gold surface using a self-assembled bilayer monolayer of 6-mercapto-1-hexanol (MCH) and 11-mercaptoundecanoic acid. In the second system, the aptamer was chemically adsorbed onto the electrode surface in a mixed monolayer with MCH. Using nanobodies, the authors detected IL-6 in the range from 10 pg/mL to 10 ng/mL. They noticed that the capacitance of the electrodes drifted monotonically during prolonged incubation in the buffer. This could mask a specific response, especially in a wearable format where long measurement times are required. Using aptamers, the authors achieved the detection of IL-6 in the range from 10 to 10,000 pg/mL. In contrast to the nanobody-based platform, the aptasensor was less susceptible to background drift, which stabilized after 90 min.
The work [248] describes two nanobiocomposite platforms based on synthetic nanobodies, which allow label-free electrochemical detection of epidermal growth factor receptor (EGFR). To do this, either NiO particles or a layer of poly(thiophene acetic acid) (PTAA) are applied to screen-printed carbon electrodes, to which an EGFR-specific nanobody with dual functionality—a 6xHis tag and lysine—is attached. These nanobodies serve as bioreceptors for impedance sensing in the presence of EGFR. The platforms are characterized by various methods and show good sensitivity and specificity for EGFR ranging from 0.25 to 50 μg/mL with detection limits of 0.46 μg/mL and 1.14 μg/mL, respectively.
Impedimetric biosensors can be used to detect and quantify various proteins, which are important biomarkers for the diagnosis, monitoring, and treatment of various diseases such as cancer, infections, inflammation, allergies, etc. However, impedimetric biosensors also face several problems and limitations, such as instability, degradation, nonspecific binding, interference, drift, etc. Therefore, the development and implementation of new types of recognition elements that can improve the performance, reliability, and functionality of impedimetric biosensors is an urgent and promising task.

5. Integration of Impedimetric Biosensors into Microfluidic Systems

In addition to providing high sensitivity and selectivity, modern biosensor systems must provide rapid detection, be cost-effective, and also be miniaturized, portable, and have small sample volumes. Problems associated with miniaturization, namely reducing the volume of the analyte and increasing analysis productivity, are of significant interest from the point of view of using methods and approaches of microfluidics to solve them. Microfluidics is a field of science and technology that studies and controls the flow of liquids on a microscale [249,250]. Microfluidics is of great importance in integration with biosensor systems, as it allows the manipulation of biological samples and reagents with high precision, efficiency, and speed. Microfluidic systems can integrate microchannels, micropumps, microshutters, micromixers, microreactors, and other elements [251,252,253], which can be made of various materials, such as silicone, polymers, glass, metals, etc. [254,255,256].
Integrating impedimetric biosensors with microfluidic systems provides several advantages, such as (i) reduced consumption of biological samples and reagents, which reduces cost and risk of contamination; (ii) increasing the sensitivity and selectivity of biosensors, as microfluidics provides better control over mass transport, temperature, pH, etc.; (iii) faster analysis and reaction times as microfluidics reduces diffusion distance and mixing time; and (iv) the ability to automate and integrate various bioanalytical functions such as sample preparation, separation, concentration, detection, etc., in a single microfluidic platform (e.g., lab-on-a-chip).
In addition, microfluidics is facilitating the development of point-of-care (POC) impedimetric biosensor systems that can be used to quickly, easily, and accurately diagnose and monitor various diseases at the point of patient care, such as at home or in the clinic [257,258,259].
There are various examples of the use of impedimetric biosensors in combination with microfluidic systems for protein detection [260,261,262,263,264,265]. Alsabbagh et al. [266] present a microfluidic circuit for electrochemical impedance spectroscopy (EIS), which can serve as a biosensor for label-free detection of cardiac troponin I. The biosensor system consists of a glass plate, sputtered electrodes, and a polydimethylsiloxane (PDMS) microchannel. Functionalization methods have been developed for electrodes that take into account the possible charge transfer through the sensitive layer, the specific binding of the analyte to the corresponding antibodies, and the reduction in nonspecific protein adsorption. The use of a 1000 ng/mL human serum albumin (HSA) sample did not result in a noticeable change in impedance, while the use of a 1 ng/mL cardiotroponin I sample caused a significant shift in the Nyquist plot.
Wu et al. [267] developed a microfluidic device with two opposing electrodes and a channel to create immunosensors for the peanut allergen Ara h 1 (Figure 24). With optimal fluid management, the device can obtain a better immune response by pumping 80 μL of Ara h 1 solution supplemented with 2.5 mM Fe(CN)63−/4− mediator at a rate of 50 μL/min for 2.5 min. A single pass of the sample through the microfluidic system provides a linear detection range of 1 pg/mL–10 ng/mL and an LOD of 3.9 fg/mL, respectively. The developed device can determine the Apa h 1 concentration with good recovery power (95−103%) for a diluted supernatant isolated from standard peanut butter. The microfluidic system creates a sealed sample chamber with a predetermined volume to ensure standardization of measurements. Bhardwaj et al. [268] presented an aptamer-based microfluidic biosensor for in-line monitoring of ranibizumab in bioreactors. The aptamer with the best affinity for the analyte was immobilized on gold microelectrodes on a microfluidic chip, which was fabricated using a glass base and PDMS top. The linear detection range and detection limit were 25–100 nM and 25 nM, respectively, which were significantly better than the HPLC-based detection method (about 240 nM). The device did not require preconcentration or sample pretreatment and resulted in detection times of about 30 min compared to several hours for HPLC.
Using microfluidic systems makes it convenient to combine several different functional zones in one device [269,270]. Muhsin et al. [271] present a microfluidic biosensor device based on a microelectromechanical system (MEMS) for detecting pathological prions of chronic wasting disease (CHD) in deer. The device consists of functional zones (Figure 25) for the concentration, capture, and detection of the prion. The detection zone contains an array of electrodes coated with a monoclonal antibody against pathological prions. Testing can be completed in less than 1 h with high sensitivity and selectivity. The biosensor detected the engineered prion antigen at a dilution of 1:24, while the ELISA detected the same antigen at a dilution of 1:8. The relative detection limit of the biosensor was 1:1000 dilution of a known strong positive sample, while the ELISA showed a 1:100 dilution.
In addition, microfluidics allows the creation of multiplex impedimetric biosensor systems that can simultaneously detect and measure multiple targets in a single sample [272,273,274]. This improves the information content and reliability of the analysis and also saves time and resources. Multiplex impedimetric biosensors can be implemented by dividing the electrodes into different regions, each with its own receptor, or by using different frequencies to excite and measure impedance on a single electrode with multiple receptors. However, there is a fairly small number of works in the direction of multiplex microfluidic impedance biosensing. Article [275] presents a 2 × 4 interdigitated electrode (IDE) array to improve the sensitivity of immunoassays using microparticles as a label. To amplify the signal, the system is integrated into a microfluidic channel. The developed impedimetric biosensor includes all the necessary components for bioassays in one system, which consists of a gold chip with an IDE array, an eight-channel impedance analyzer with software for semi-real-time data acquisition, and a PDMS microfluidic chip for sample delivery and washing. Therefore, the creation of microfluidic chips for impedance detection of several proteins at once is a relevant and important task. This will help to create cheap biosensor panels for point-of-care diagnostics of a number of diseases.

6. Conclusions

Recently, the biosensor market has expanded significantly due to the increased demand for specialized sensors that can provide fast and accurate results in various scientific fields. In this article, we reviewed the basic principles, methods, and applications of impedimetric biosensors for protein detection. The material demonstrated that these biosensors have several advantages, such as simplicity, speed, sensitivity, selectivity, cost-effectiveness, and the ability to be integrated into microfluidic systems. Impedimetric biosensors represent a powerful and promising tool for studying and monitoring biological processes associated with proteins and can contribute to the development of new approaches to the diagnosis and treatment of various diseases.
The impedance measurement technique is one of the important factors in the creation of impedimetric biosensors. Impedimetric biosensors can be faradaic or non-faradaic, depending on the presence or absence of a redox-active probe in the solution. Non-faradaic biosensors may be simpler for practical use since they do not require additional electrodes or the addition of a redox couple and can operate at a single frequency. However, faradaic biosensors can demonstrate greater analytical sensitivity.
An important goal is to improve the operational characteristics of impedimetric biosensors. Its solution requires a deep understanding of the properties of materials for creating and functionalizing electrodes and methods for their manufacture. Current analysis identifies the main groups of materials that are widely used for the formation of electrodes of impedimetric biosensors and their modification. The main materials used to create electrodes are gold (Au), carbon (graphite), and conductive metal oxides. Electrodes based on Au and carbon materials are typically formed by screen printing together with a counter electrode and a reference electrode. This technology is highly scalable for mass production. It is also compatible with the use of substrates (for example, PET) with coatings of conductive metal oxides (for example, ITO), including the possibility of additional irradiation of the sensor from the substrate. Screen-printing electrodes are modified with various nanomaterials, such as graphene, carbon nanotubes, polymers, gold nanoparticles, metal oxide nanoparticles, quantum dots, and others. The sensor platforms with screen-printed electrodes and modifying nanomaterials for faradaic EIS, as well as interdigitated electrodes for non-faradaic EIS, are currently commercially available. This certainly contributes to an increase in the amount of research in the field of impedimetric biosensors, their clinical trials, and their applications. However, research into the application of various nanomaterials, new polymers, and biopolymers, including MIPs, as well as various molecular structures, remains urgently relevant.
The need and choice of electrode-modifying materials are primarily determined by the choice of the linker used or the immobilization method (covalent binding or physical adsorption) and biorecognition element, biocompatibility, as well as the required response range and detection limit to values significant for clinical studies. Thus, obviously, for many biomarkers, obtaining the required wide range of analytical responses is ensured by increasing the specific electrode area with nanomaterials, nanostructures, and the use of composite materials. New modifying materials, their decoration, and the creation of compositions based on them also help to amplify the signal to the required sensitivity. Also, the choice of electrode materials is determined by the ease of its synthesis and implementation in the sensor system, the ability to scale the technology, and its durability. As we can see from the review, the use of new materials to create impedimetric biosensors looks promising.
The selection of the required highly specific biorecognition elements and the technology for their immobilization is also an important component of modern impedimetric biosensors. The choice of the bioligand immobilization method depends on the specific goals and conditions of the analysis, as well as on the type of biorecognition element and transducer. In some cases, it may be advisable to use a combination of different immobilization methods to achieve the optimal steric configuration of the biorecognition element. In this way, the probability of affinity binding of the ligand to the target protein can be significantly improved. Along with traditional antibodies, aptamers, and enzymes, the popularity of biosensors based on peptides, as well as other new bioligands, is increasing. The key challenge in creating new biorecognition elements is to achieve high sensitivity and selectivity while achieving low-cost mass production to reduce the overall cost of the analysis.
Integration of impedimetric biosensors with microfluidic systems reduces sample volume and detection time and also holds great promise for creating point-of-care systems that can be portable and used in emergency care. To do this, it is necessary to create new biosensors for multiplex determination of various protein markers, which can be implemented in the format of a miniature chip.
The creation of impedimetric biosensors for protein detection is a complex and multifaceted task that includes various aspects, including the following:
  (i).
Selection and synthesis of suitable materials and nanostructures for electrodes that should provide good electrochemical activity, resistance to corrosion and degradation, and the ability to functionalize the surface for immobilization of bioligands.
 (ii).
Development and optimization of methods for immobilizing biorecognition agents on the surface of electrodes, which should ensure specific, stable, and oriented binding to target proteins, as well as preservation of their activity and structure.
(iii).
Improving the specificity and sensitivity of biosensors, which should be able to detect low concentrations of proteins in complex biological matrices such as blood, serum, plasma, saliva, and urine, as well as distinguish between proteins with high degrees of homology or post-translational modifications.
 (iv).
Reduction of the influence of interference and nonspecific binding, which can distort the biosensor signal and degrade its performance. Interference can be caused by electromagnetic fields, temperature, pH, salt composition, oxidizing or reducing agents, and other biomolecules present in the sample.
  (v).
Adaptation of biosensors to real samples and operating conditions, which may differ from laboratory ones. For example, biosensors must be able to operate in a wide range of temperatures, humidities, pressures, and illuminations and also be resistant to mechanical damage and contamination.
 (vi).
Expanding the range of detectable proteins that can be associated with various diseases, such as cancer, infections, autoimmune diseases, allergies, and others. To achieve this, it is necessary to develop new biorecognition agents that can specifically bind to proteins of interest and also take into account their structural and functional characteristics.
Indeed, the task of creating new effective biosensor systems is complex and multifactorial. To solve it successfully, it is necessary to apply an interdisciplinary approach, combining knowledge and skills in the fields of chemistry, physics, biology, materials science, nanotechnology, electronics, computer science, and medicine.

Author Contributions

Conceptualization, N.S. and A.R.; formal analysis, V.M., A.A., D.K. and T.Z.; writing—original draft preparation, N.S. and A.R.; writing—review and editing, N.S., A.R. and T.Z.; visualization, N.S. and A.R.; supervision, V.M., A.A., D.K. and T.Z.; project administration, N.S.; funding acquisition, T.Z. All authors have read and agreed to the published version of the manuscript.

Funding

N.S. acknowledges the support of the Russian Science Foundation (Project No. 23-79-01136).

Data Availability Statement

Not applicable.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Sidransky, D. Emerging Molecular Markers of Cancer. Nat. Rev. Cancer 2002, 2, 210–219. [Google Scholar] [CrossRef]
  2. Kaigorodova, E.; Bogatyuk, M. Heat Shock Proteins as Prognostic Markers of Cancer. Curr. Cancer Drug Targets 2014, 14, 713–726. [Google Scholar] [CrossRef] [PubMed]
  3. Wang, X.; Huang, J.; Chen, W.; Li, G.; Li, Z.; Lei, J. The Updated Role of Exosomal Proteins in the Diagnosis, Prognosis, and Treatment of Cancer. Exp. Mol. Med. 2022, 54, 1390–1400. [Google Scholar] [CrossRef] [PubMed]
  4. Dufour, J.-F. C-Reactive Protein, a Prognostic Marker in Hepatocellular Carcinoma. Hepatology 2013, 57, 2103–2105. [Google Scholar] [CrossRef] [PubMed]
  5. Schindler, M.; Walter, N.; Maderbacher, G.; Sigmund, I.K.; Alt, V.; Rupp, M. Novel Diagnostic Markers for Periprosthetic Joint Infection: A Systematic Review. Front. Cell. Infect. Microbiol. 2023, 13, 1210345. [Google Scholar] [CrossRef] [PubMed]
  6. Fischer, P.; Grigoras, C.; Bugariu, A.; Nicoara-Farcau, O.; Stefanescu, H.; Benea, A.; Hadade, A.; Margarit, S.; Sparchez, Z.; Tantau, M.; et al. Are Presepsin and Resistin Better Markers for Bacterial Infection in Patients with Decompensated Liver Cirrhosis? Dig. Liver Dis. 2019, 51, 1685–1691. [Google Scholar] [CrossRef] [PubMed]
  7. Tumani, H.; Brettschneider, J. Biochemical Markers of Autoimmune Diseases of the Nervous System. Curr. Pharm. Des. 2012, 18, 4556–4563. [Google Scholar] [CrossRef] [PubMed]
  8. Ogulur, I.; Pat, Y.; Ardicli, O.; Barletta, E.; Cevhertas, L.; Fernandez-Santamaria, R.; Huang, M.; Bel Imam, M.; Koch, J.; Ma, S.; et al. Advances and Highlights in Biomarkers of Allergic Diseases. Allergy 2021, 76, 3659–3686. [Google Scholar] [CrossRef]
  9. Klebes, A.; Kittel, A.-S.; Verboket, R.D.; von Stetten, F.; Früh, S.M. Multianalyte Lateral Flow Immunoassay for Simultaneous Detection of Protein-Based Inflammation Biomarkers and Pathogen DNA. Sens. Actuators B Chem. 2022, 355, 131283. [Google Scholar] [CrossRef]
  10. Mousavi, S.M.; Hashemi, S.A.; Kalashgrani, M.Y.; Gholami, A.; Omidifar, N.; Babapoor, A.; Vijayakameswara Rao, N.; Chiang, W.-H. Recent Advances in Plasma-Engineered Polymers for Biomarker-Based Viral Detection and Highly Multiplexed Analysis. Biosensors 2022, 12, 286. [Google Scholar] [CrossRef]
  11. Van Gool, A.; Corrales, F.; Čolović, M.; Krstić, D.; Oliver-Martos, B.; Martínez-Cáceres, E.; Jakasa, I.; Gajski, G.; Brun, V.; Kyriacou, K.; et al. Analytical Techniques for Multiplex Analysis of Protein Biomarkers. Expert Rev. Proteom. 2020, 17, 257–273. [Google Scholar] [CrossRef]
  12. Song, J.G.; Baral, K.C.; Kim, G.-L.; Park, J.-W.; Seo, S.-H.; Kim, D.-H.; Jung, D.H.; Ifekpolugo, N.L.; Han, H.-K. Quantitative Analysis of Therapeutic Proteins in Biological Fluids: Recent Advancement in Analytical Techniques. Drug Deliv. 2023, 30, 2183816. [Google Scholar] [CrossRef]
  13. Swami, S.; Kayenat, F.; Wajid, S. SPR Biosensing: Cancer Diagnosis and Biomarkers Quantification. Microchem. J. 2024, 197, 109792. [Google Scholar] [CrossRef]
  14. Karunakaran, R. Merve Keskin Biosensors: Components, Mechanisms, and Applications. In Analytical Techniques in Biosciences; Academic Press: Cambridge, MA, USA, 2022; pp. 179–190. [Google Scholar] [CrossRef]
  15. Zimina, T.M.; Sitkov, N.O.; Gareev, K.G.; Fedorov, V.; Grouzdev, D.S.; Koziaeva, V.V.; Gao, H.; Combs, S.E.; Shevtsov, M. Biosensors and Drug Delivery in Oncotheranostics Using Inorganic Synthetic and Biogenic Magnetic Nanoparticles. Biosensors 2022, 12, 789. [Google Scholar] [CrossRef] [PubMed]
  16. Singh, A.K.; Mittal, S.; Das, M.; Saharia, A.; Tiwari, M. Optical Biosensors: A Decade in Review. Alex. Eng. J. 2023, 67, 673–691. [Google Scholar] [CrossRef]
  17. Machado, M.; Oliveira, A.M.L.; Silva, G.A.; Bitoque, D.B.; Tavares Ferreira, J.; Pinto, L.A.; Ferreira, Q. Graphene Biosensors—A Molecular Approach. Nanomaterials 2022, 12, 1624. [Google Scholar] [CrossRef] [PubMed]
  18. Fethi, A.; Hicham, M. Carbon Electrodes as Emerging Platforms for Miniaturization of Electrochemical Biosensors. In Miniaturized Biosensing Devices; Springer: Singapore, 2022; pp. 39–55. [Google Scholar] [CrossRef]
  19. Khan, H.; Shah, M.R.; Barek, J.; Malik, M.I. Cancer Biomarkers and Their Biosensors: A Comprehensive Review. TrAC Trends Anal. Chem. 2023, 158, 116813. [Google Scholar] [CrossRef]
  20. Sanko, V.; Kuralay, F. Label-Free Electrochemical Biosensor Platforms for Cancer Diagnosis: Recent Achievements and Challenges. Biosensors 2023, 13, 333. [Google Scholar] [CrossRef] [PubMed]
  21. Nemati, M.; Farajzadeh, M.A.; Afshar Mogaddam, M.R.; Pourali, A. Recent Advances in Impedimetric Biosensors Focusing on Myocardial Infarction Diagnosis. Critical Rev. Anal. Chem. 2022, 1–14. [Google Scholar] [CrossRef]
  22. Antiochia, R. Electrochemical Biosensors for SARS-CoV-2 Detection: Voltametric or Impedimetric Transduction? Bioelectrochemistry 2022, 147, 108190. [Google Scholar] [CrossRef]
  23. Ullah, S.F.; Moreira, G.; Datta, S.P.A.; McLamore, E.; Vanegas, D. An Experimental Framework for Developing Point-of-Need Biosensors: Connecting Bio-Layer Interferometry and Electrochemical Impedance Spectroscopy. Biosensors 2022, 12, 938. [Google Scholar] [CrossRef]
  24. Dong, T.; Matos Pires, N.M.; Yang, Z.; Jiang, Z. Advances in Electrochemical Biosensors Based on Nanomaterials for Protein Biomarker Detection in Saliva. Adv. Sci. 2022, 10, e2205429. [Google Scholar] [CrossRef] [PubMed]
  25. Centane, S.; Nyokong, T. Aptamer versus Antibody as Probes for the Impedimetric Biosensor for Human Epidermal Growth Factor Receptor. J. Inorg. Biochem. 2022, 230, 111764. [Google Scholar] [CrossRef] [PubMed]
  26. Sinitsyna, V.V.; Vetcher, A.A. Nucleic Acid Aptamers in Nanotechnology. Biomedicines 2022, 10, 1079. [Google Scholar] [CrossRef]
  27. Wang, L.; Zhu, W.; Zhang, J.; Zhu, J.-J. Miniaturized Microfluidic Electrochemical Biosensors Powered by Enzymatic Biofuel Cell. Biosensors 2023, 13, 175. [Google Scholar] [CrossRef]
  28. Shimizu, F.M.; de Barros, A.; Braunger, M.L.; Gaal, G.; Riul, A., Jr. Information Visualization and Machine Learning Driven Methods for Impedimetric Biosensing. TrAC Trends Anal. Chem. 2023, 165, 117115. [Google Scholar] [CrossRef]
  29. Strong, M.E.; Richards, J.R.; Torres, M.; Beck, C.M.; La Belle, J.T. Faradaic Electrochemical Impedance Spectroscopy for Enhanced Analyte Detection in Diagnostics. Biosens. Bioelectron. 2021, 177, 112949. [Google Scholar] [CrossRef]
  30. Uygun, Z.O.; Ertuğrul Uygun, H.D. A Short Footnote: Circuit Design for Faradaic Impedimetric Sensors and Biosensors. Sens. Actuators B Chem. 2014, 202, 448–453. [Google Scholar] [CrossRef]
  31. Vivier, V.; Orazem, M.E. Impedance Analysis of Electrochemical Systems. Chem. Rev. 2022, 122, 11131–11168. [Google Scholar] [CrossRef]
  32. Castro Gutierrez, H.; Panigrahi, S. Selective Impedimetric Interdigitated Electrode for Sensing Gaseous, Biological, and Inorganic Targets: A State-of-The-Art Review. IEEE Sens. J. 2023. [Google Scholar] [CrossRef]
  33. Robinson, C.; Juska, V.B.; O’Riordan, A. Surface Chemistry Applications and Development of Immunosensors Using Electrochemical Impedance Spectroscopy: A Comprehensive Review. Environ. Res. 2023, 237, 116877. [Google Scholar] [CrossRef]
  34. Mariani, F.; Gualandi, I.; Schuhmann, W.; Scavetta, E. Micro- and Nano-Devices for Electrochemical Sensing. Microchim. Acta 2022, 189, 459. [Google Scholar] [CrossRef]
  35. Štukovnik, Z.; Bren, U. Recent Developments in Electrochemical-Impedimetric Biosensors for Virus Detection. Int. J. Mol. Sci. 2022, 23, 15922. [Google Scholar] [CrossRef] [PubMed]
  36. Haeverbeke, M.V.; Stock, M.; De Baets, B. Equivalent Electrical Circuits and Their Use Across Electrochemical Impedance Spectroscopy Application Domains. IEEE Access 2022, 10, 51363–51379. [Google Scholar] [CrossRef]
  37. Hui, Y.; Huang, Z.; Alahi, M.E.E.; Nag, A.; Feng, S.; Mukhopadhyay, S.C. Recent Advancements in Electrochemical Biosensors for Monitoring the Water Quality. Biosensors 2022, 12, 551. [Google Scholar] [CrossRef]
  38. Shlosberg, Y.; Sepunaru, L. Advantages of Imprinted Polymer Electrodes for Electrochemical Pathogen Detection. Curr. Opin. Electrochem. 2022, 36, 101123. [Google Scholar] [CrossRef]
  39. Pänke, O.; Balkenhohl, T.; Kafka, J.; Schäfer, D.; Lisdat, F. Impedance Spectroscopy and Biosensing. In Biosensing for the 21st Century; Advances in Biochemical Engineering/Biotechnology; Springer: Berlin/Heidelberg, Germany, 2007; pp. 195–237. [Google Scholar] [CrossRef]
  40. Sheen, H.-J.; Panigrahi, B.; Kuo, T.-R.; Hsu, W.-C.; Chung, P.-S.; Xie, Q.-Z.; Lin, C.-Y.; Chang, Y.-S.; Lin, C.-T.; Fan, Y.-J. Electrochemical Biosensor with Electrokinetics-Assisted Molecular Trapping for Enhancing C-Reactive Protein Detection. Biosens. Bioelectron. 2022, 210, 114338. [Google Scholar] [CrossRef]
  41. Kemp, N.T. A Tutorial on Electrochemical Impedance Spectroscopy and Nanogap Electrodes for Biosensing Applications. IEEE Sens. J. 2021, 21, 22232–22245. [Google Scholar] [CrossRef]
  42. Mwanza, D.; Adeniyi, O.; Tesfalidet, S.; Nyokong, T.; Mashazi, P. Capacitive Label-Free Ultrasensitive Detection of PSA on a Covalently Attached Monoclonal Anti-PSA Antibody Gold Surface. J. Electroanal. Chem. 2022, 927, 116983. [Google Scholar] [CrossRef]
  43. Otero, F.; Magner, E. Biosensors—Recent Advances and Future Challenges in Electrode Materials. Sensors 2020, 20, 3561. [Google Scholar] [CrossRef]
  44. Huang, X.; Zhu, Y.; Kianfar, E. Nano Biosensors: Properties, Applications and Electrochemical Techniques. J. Mater. Res. Technol. 2021, 12, 1649–1672. [Google Scholar] [CrossRef]
  45. Nasrin, F.; Khoris, I.M.; Chowdhury, A.D.; Boonyakida, J.; Park, E.Y. Impedimetric Biosensor of Norovirus with Low Variance Using Simple Bioconjugation on Conductive Polymer-Au Nanocomposite. Sens. Actuators B Chem. 2022, 369, 132390. [Google Scholar] [CrossRef]
  46. Dhamu, V.N.; Poudyal, D.C.; Samson, M.; Paul, A.; Muthukumar, S.; Prasad, S. Review—Environmental Biosensors for Agro-Safety Based on Electrochemical Sensing Mechanism with an Emphasis on Pesticide Screening. ECS Sens. Plus 2023, 2, 024601. [Google Scholar] [CrossRef]
  47. McCrae, L.E.; Ting, W.-T.; Howlader, M.M.R. Advancing Electrochemical Biosensors for Interleukin-6 Detection. Biosens. Bioelectron. X 2023, 13, 100288. [Google Scholar] [CrossRef]
  48. Subramani, I.G.; Perumal, V.; Gopinath, S.C.B.; Mohamed, N.M.; Joshi, N.; Ovinis, M.; Sze, L.L. 3D Nanoporous Hybrid Nanoflower for Enhanced Non-Faradaic Redox-Free Electrochemical Impedimetric Biodetermination. J. Taiwan Inst. Chem. Eng. 2020, 116, 26–35. [Google Scholar] [CrossRef]
  49. Ibau, C.; Arshad, M.K.M.; Gopinath, S.C.B.; Nuzaihan, M.N.M.; Fathil, M.F.M.; Shamsuddin, S.A. Immunosensing Prostate-Specific Antigen: Faradaic vs Non-Faradaic Electrochemical Impedance Spectroscopy Analysis on Interdigitated Microelectrode Device. Int. J. Biol. Macromol. 2020, 162, 1924–1936. [Google Scholar] [CrossRef] [PubMed]
  50. Daniels, J.S.; Pourmand, N. Label-Free Impedance Biosensors: Opportunities and Challenges. Electroanalysis 2007, 19, 1239–1257. [Google Scholar] [CrossRef]
  51. Chen, B.; Kiely, J.; Williams, I.; Luxton, R. A Non-Faradaic Impedimetric Biosensor for Monitoring of Caspase 9 in Mammalian Cell Culture. SSRN Electron. J. 2023, 153, 108456. [Google Scholar] [CrossRef]
  52. Neshani, S.; Momeni, K.; Chen, D.J.; Neihart, N.M. Highly Sensitive Readout Interface for Real-Time Differential Precision Measurements with Impedance Biosensors. Biosensors 2023, 13, 77. [Google Scholar] [CrossRef]
  53. Buscaglia, L.A.; Oliveira, O.N.; Carmo, J. Roadmap for Electrical Impedance Spectroscopy for Sensing: A Tutorial. IEEE Sens. J. 2021, 21, 22246–22257. [Google Scholar] [CrossRef]
  54. Pinkova Gajdosova, V.; Lorencova, L.; Blsakova, A.; Kasak, P.; Bertok, T.; Tkac, J. Challenges for Impedimetric Affinity Sensors Targeting Protein Detection. Curr. Opin. Electrochem. 2021, 28, 100717. [Google Scholar] [CrossRef]
  55. Guan, J.-G.; Miao, Y.-Q.; Zhang, Q.-J. Impedimetric Biosensors. J. Biosci. Bioeng. 2004, 97, 219–226. [Google Scholar] [CrossRef] [PubMed]
  56. Bhand, S.; Bacher, G. Impedimetric Sensors in Environmental Analysis: An Overview. In Environmental, Chemical and Medical Sensors; Springer: Singapore, 2017; pp. 67–85. [Google Scholar] [CrossRef]
  57. Chinnadayyala, S.R.; Park, J.; Abbasi, M.A.; Cho, S. Label-Free Electrochemical Impedimetric Immunosensor for Sensitive Detection of IgM Rheumatoid Factor in Human Serum. Biosens. Bioelectron. 2019, 143, 111642. [Google Scholar] [CrossRef]
  58. Cecchetto, J.; Carvalho, F.C.; Santos, A.; Fernandes, F.C.B.; Bueno, P.R. An Impedimetric Biosensor to Test Neat Serum for Dengue Diagnosis. Sens. Actuators B Chem. 2015, 213, 150–154. [Google Scholar] [CrossRef]
  59. Białobrzeska, W.; Ficek, M.; Dec, B.; Osella, S.; Trzaskowski, B.; Jaramillo-Botero, A.; Pierpaoli, M.; Rycewicz, M.; Dashkevich, Y.; Łęga, T.; et al. Performance of Electrochemical Immunoassays for Clinical Diagnostics of SARS-CoV-2 Based on Selective Nucleocapsid N Protein Detection: Boron-Doped Diamond, Gold and Glassy Carbon Evaluation. Biosens. Bioelectron. 2022, 209, 114222. [Google Scholar] [CrossRef] [PubMed]
  60. Bryan, T.; Luo, X.; Bueno, P.R.; Davis, J.J. An Optimised Electrochemical Biosensor for the Label-Free Detection of C-Reactive Protein in Blood. Biosens. Bioelectron. 2013, 39, 94–98. [Google Scholar] [CrossRef]
  61. Singh, K.V.; Bhura, D.K.; Nandamuri, G.; Whited, A.M.; Evans, D.; King, J.; Solanki, R. Nanoparticle-Enhanced Sensitivity of a Nanogap-Interdigitated Electrode Array Impedimetric Biosensor. Langmuir 2011, 27, 13931–13939. [Google Scholar] [CrossRef]
  62. Schrattenecker, J.D.; Heer, R.; Hainberger, R.; Fafilek, G. Impedimetric IgG-Biosensor with In-Situ Generation of the Redox-Probe. Proceedings 2017, 1, 534. [Google Scholar] [CrossRef]
  63. Wang, S.X.; Acha, D.; Shah, A.J.; Hills, F.; Roitt, I.; Demosthenous, A.; Bayford, R.H. Detection of the Tau Protein in Human Serum by a Sensitive Four-Electrode Electrochemical Biosensor. Biosens. Bioelectron. 2017, 92, 482–488. [Google Scholar] [CrossRef]
  64. Ruankham, W.; Morales Frías, I.A.; Phopin, K.; Tantimongcolwat, T.; Bausells, J.; Zine, N.; Errachid, A. One-Step Impedimetric NT-ProBNP Aptasensor Targeting Cardiac Insufficiency in Artificial Saliva. Talanta 2023, 256, 124280. [Google Scholar] [CrossRef]
  65. Pal, A.; Biswas, S.; Kare, S.P.O.; Biswas, P.; Jana, S.K.; Das, S.; Chaudhury, K. Development of an Impedimetric Immunosensor for Machine Learning-Based Detection of Endometriosis: A Proof of Concept. Sens. Actuators B Chem. 2021, 346, 130460. [Google Scholar] [CrossRef]
  66. Bachour Junior, B.; Batistuti, M.R.; Pereira, A.S.; de Sousa Russo, E.M.; Mulato, M. Electrochemical Aptasensor for NS1 Detection: Towards a Fast Dengue Biosensor. Talanta 2021, 233, 122527. [Google Scholar] [CrossRef]
  67. Yağar, H.; Özcan, H.M.; Mehmet, O. A New Electrochemical Impedance Biosensor Based on Aromatic Thiol for Alpha-1 Antitrypsin Determination. Turk. J. Chem. 2021, 45, 104–118. [Google Scholar] [CrossRef]
  68. Ben Messaoud, N.; Barreiros dos Santos, M.; Trocado, V.; Nogueira-Silva, C.; Queirós, R. A Novel Label-Free Electrochemical Immunosensor for Detection of Surfactant Protein B in Amniotic Fluid. Talanta 2023, 251, 123744. [Google Scholar] [CrossRef]
  69. Díaz-Fernández, A.; Miranda-Castro, R.; de-los-Santos-Álvarez, N.; Lobo-Castañón, M.J.; Estrela, P. Impedimetric Aptamer-Based Glycan PSA Score for Discrimination of Prostate Cancer from Other Prostate Diseases. Biosens. Bioelectron. 2021, 175, 112872. [Google Scholar] [CrossRef]
  70. Nan, M.-N.; Bi, Y.; Xue, H.-L.; Long, H.-T.; Xue, S.-L.; Pu, L.-M.; Prusky, D. Modification Performance and Electrochemical Characteristics of Different Groups of Modified Aptamers Applied for Label-Free Electrochemical Impedimetric Sensors. Food Chem. 2021, 337, 127761. [Google Scholar] [CrossRef]
  71. Svigelj, R.; Zuliani, I.; Grazioli, C.; Dossi, N.; Toniolo, R. An Effective Label-Free Electrochemical Aptasensor Based on Gold Nanoparticles for Gluten Detection. Nanomaterials 2022, 12, 987. [Google Scholar] [CrossRef] [PubMed]
  72. Tertis, M.; Leva, P.I.; Bogdan, D.; Suciu, M.; Graur, F.; Cristea, C. Impedimetric Aptasensor for the Label-Free and Selective Detection of Interleukin-6 for Colorectal Cancer Screening. Biosens. Bioelectron. 2019, 137, 123–132. [Google Scholar] [CrossRef] [PubMed]
  73. Aydın, E.B.; Sezgintürk, M.K. An Impedimetric Immunosensor for Highly Sensitive Detection of IL-8 in Human Serum and Saliva Samples: A New Surface Modification Method by 6-Phosphonohexanoic Acid for Biosensing Applications. Anal. Biochem. 2018, 554, 44–52. [Google Scholar] [CrossRef] [PubMed]
  74. Aydın, E.B.; Sezgintürk, M.K. Ultrasensitive Detection of Interleukin 1α Using 3-Phosphonopropionic Acid Modified FTO Surface as an Effective Platform for Disposable Biosensor Fabrication. Bioelectrochemistry 2021, 138, 107698. [Google Scholar] [CrossRef] [PubMed]
  75. Gündoğdu, A.; Aydın, E.B.; Sezgintürk, M.K. A Novel Electrochemical Immunosensor Based on ITO Modified by Carboxyl-Ended Silane Agent for Ultrasensitive Detection of MAGE-1 in Human Serum. Anal. Biochem. 2017, 537, 84–92. [Google Scholar] [CrossRef]
  76. Yılmaz, N.; Aydın, E.B.; Sezgintürk, M.K. An Epoxysilane Modified Indium Tin Oxide Electrode for the Determination of PAK 2: Application in Human Serum Samples. Anal. Chim. Acta 2019, 1062, 68–77. [Google Scholar] [CrossRef]
  77. Demirbakan, B.; Sezgintürk, M.K. A Novel Electrochemical Immunosensor Based on Disposable ITO-PET Electrodes for Sensitive Detection of PAK 2 Antigen. J. Electroanal. Chem. 2019, 848, 113304. [Google Scholar] [CrossRef]
  78. Nur Altay, D.; Yagar, H.; Ozcan, H.M. A new ITO-based Aβ42 biosensor for early detection of Alzheimer’s disease. Bioelectrochemistry 2023, 153, 108501. [Google Scholar] [CrossRef]
  79. Yang, Y.-P.; Lu, Y.-L.; Kumar Gupta, A.; Lin, S.-P. Real-Time Impedimetric Detection of Cardiac Troponin I Using ITO-Coated Vertically Aligned Silicon Nanowires. Mater. Lett. 2022, 311, 131575. [Google Scholar] [CrossRef]
  80. Moro, G.; Ferrari, L.; Angelini, A.; Polo, F. An Impedimetric Biosensing Strategy Based on BicyclicPeptides as Bioreceptors for Monitoring H-UPA Cancer Biomarkers. Chemosensors 2023, 11, 234. [Google Scholar] [CrossRef]
  81. Senturk, H.; Eksin, E.; Işık, Ö.; İlaslan, Z.; Mısırlı, F.; Erdem, A. Impedimetric Aptasensor for Lysozyme Detection Based on Carbon Nanofibres Enriched Screen-Printed Electrodes. Electrochim. Acta 2021, 377, 138078. [Google Scholar] [CrossRef]
  82. Khan, R.; Pal, M.; Kuzikov, A.V.; Bulko, T.; Suprun, E.V.; Shumyantseva, V.V. Impedimetric Immunosensor for Detection of Cardiovascular Disorder Risk Biomarker. Mater. Sci. Eng. C 2016, 68, 52–58. [Google Scholar] [CrossRef]
  83. Kara, P.; de la Escosura-Muñiz, A.; Maltez-da Costa, M.; Guix, M.; Ozsoz, M.; Merkoçi, A. Aptamers Based Electrochemical Biosensor for Protein Detection Using Carbon Nanotubes Platforms. Biosens. Bioelectron. 2010, 26, 1715–1718. [Google Scholar] [CrossRef] [PubMed]
  84. Rohrbach, F.; Karadeniz, H.; Erdem, A.; Famulok, M.; Mayer, G. Label-Free Impedimetric Aptasensor for Lysozyme Detection Based on Carbon Nanotube-Modified Screen-Printed Electrodes. Anal. Biochem. 2012, 421, 454–459. [Google Scholar] [CrossRef] [PubMed]
  85. Azzouzi, S.; Ben Ali, M.; Bellagambi, F.; Elaissari, A.; Jaffrezic-Renault, N.; Errachid, A.; Zine, N. Spatially Hierarchical Nano-Architecture for Real Time Detection of Interleukin-8 Cancer Biomarker. Talanta 2022, 246, 123436. [Google Scholar] [CrossRef]
  86. Sonuç Karaboğa, M.N.; Sezgintürk, M.K. A Nano-Composite Based Regenerative Neuro Biosensor Sensitive to Parkinsonism-Associated Protein DJ-1/Park7 in Cerebrospinal Fluid and Saliva. Bioelectrochemistry 2021, 138, 107734. [Google Scholar] [CrossRef]
  87. Ghanavati, M.; Tadayon, F.; Bagheri, H. A Novel Label-Free Impedimetric Immunosensor for Sensitive Detection of Prostate Specific Antigen Using Au Nanoparticles/MWCNTs-Graphene Quantum Dots Nanocomposite. Microchem. J. 2020, 159, 105301. [Google Scholar] [CrossRef]
  88. Demirbakan, B.; Kemal Sezgintürk, M. An Impedimetric Biosensor System Based on Disposable Graphite Paper Electrodes: Detection of ST2 as a Potential Biomarker for Cardiovascular Disease in Human Serum. Anal. Chim. Acta 2021, 1144, 43–52. [Google Scholar] [CrossRef]
  89. Özcan, B.; Sezgintürk, M.K. Highly Sensitive and Single-Use Biosensing System Based on a GP Electrode for Analysis of Adiponectin, an Obesity Biomarker. ACS Biomater. Sci. Eng. 2021, 7, 3658–3668. [Google Scholar] [CrossRef]
  90. Demirbakan, B.; Sezgintürk, M.K. An Electrochemical Immunosensor Based on Graphite Paper Electrodes for the Sensitive Detection of Creatine Kinase in Actual Samples. J. Electroanal. Chem. 2022, 921, 116656. [Google Scholar] [CrossRef]
  91. Witt, S.; Rogien, A.; Werner, D.; Siegenthaler, J.; Lesiyon, R.; Kurien, N.; Rechenberg, R.; Baule, N.; Hardy, A.; Becker, M. Boron Doped Diamond Thin Films for the Electrochemical Detection of SARS-CoV-2 S1 Protein. Diam. Relat. Mater. 2021, 118, 108542. [Google Scholar] [CrossRef]
  92. Frias, I.A.M.; Zine, N.; Sigaud, M.; Lozano-Sanchez, P.; Caffio, M.; Errachid, A. Non-Covalent π–π Functionalized Gii-Sense® Graphene Foam for Interleukin 10 Impedimetric Detection. Biosens. Bioelectron. 2023, 222, 114954. [Google Scholar] [CrossRef] [PubMed]
  93. Filip, J.; Zavahir, S.; Klukova, L.; Tkac, J.; Kasak, P. Immobilization of Concanavalin A Lectin on a Reduced Graphene Oxide-Thionine Surface by Glutaraldehyde Crosslinking for the Construction of an Impedimetric Biosensor. J. Electroanal. Chem. 2017, 794, 156–163. [Google Scholar] [CrossRef]
  94. Sonuç Karaboga, M.N.; Sezgintürk, M.K. Analysis of Tau-441 Protein in Clinical Samples Using RGO/AuNP Nanocomposite-Supported Disposable Impedimetric Neuro-Biosensing Platform: Towards Alzheimer’s Disease Detection. Talanta 2020, 219, 121257. [Google Scholar] [CrossRef]
  95. Kazemi, S.H.; Ghodsi, E.; Abdollahi, S.; Nadri, S. Porous Graphene Oxide Nanostructure as an Excellent Scaffold for Label-Free Electrochemical Biosensor: Detection of Cardiac Troponin I. Mater. Sci. Eng. C 2016, 69, 447–452. [Google Scholar] [CrossRef]
  96. Jozghorbani, M.; Fathi, M.; Kazemi, S.H.; Alinejadian, N. Determination of Carcinoembryonic Antigen as a Tumor Marker Using a Novel Graphene-Based Label-Free Electrochemical Immunosensor. Anal. Biochem. 2021, 613, 114017. [Google Scholar] [CrossRef]
  97. Harmanci, D.; Onur Uygun, Z.; Koçak Sezgin, A.; Demirdöver, C.; Girgin Sagin, F.; Akdoğan, G. Rapid Fibroblast Cell Culture Characterization with Impedimetric Label-free Heat Shock Protein—47 Biosensor. Electroanalysis 2020, 32, 2310–2315. [Google Scholar] [CrossRef]
  98. Yagati, A.K.; Pyun, J.-C.; Min, J.; Cho, S. Label-Free and Direct Detection of C-Reactive Protein Using Reduced Graphene Oxide-Nanoparticle Hybrid Impedimetric Sensor. Bioelectrochemistry 2016, 107, 37–44. [Google Scholar] [CrossRef]
  99. Brodowski, M.; Kowalski, M.; Skwarecka, M.; Pałka, K.; Skowicki, M.; Kula, A.; Lipiński, T.; Dettlaff, A.; Ficek, M.; Ryl, J.; et al. Highly Selective Impedimetric Determination of Haemophilus Influenzae Protein D Using Maze-like Boron-Doped Carbon Nanowall Electrodes. Talanta 2021, 221, 121623. [Google Scholar] [CrossRef]
  100. Wang, C.; Li, J.; Kang, M.; Huang, X.; Liu, Y.; Zhou, N.; Zhang, Z. Nanodiamonds and Hydrogen-Substituted Graphdiyne Heteronanostructure for the Sensitive Impedimetric Aptasensing of Myocardial Infarction and Cardiac Troponin I. Anal. Chim. Acta 2021, 1141, 110–119. [Google Scholar] [CrossRef]
  101. Zakaria, N.; Mohd, Y.; Chin, L.Y.; Jalil, M.N.; Zain, Z.M. Poly-Ortho-Phenylenediamine Modified Pt/Ir Microelectrode as Impedimetric Biosensor. Int. J. Electrochem. Sci. 2021, 16, 210736. [Google Scholar] [CrossRef]
  102. Hassan-Nixon, H.A.G.; Singh, N.; Cass, A.E.G. A Sensitive Impedimetric Immunosensor for the Detection of Interleukin-8 in Nasal Epithelial Lining Fluid of Asthma Patients. Biosens. Bioelectron. X 2022, 10, 100118. [Google Scholar] [CrossRef]
  103. Thangsunan, P.; Lal, N.; Tiede, C.; Moul, S.; Robinson, J.I.; Knowles, M.A.; Stockley, P.G.; Beales, P.A.; Tomlinson, D.C.; McPherson, M.J.; et al. Affimer-Based Impedimetric Biosensors for Fibroblast Growth Factor Receptor 3 (FGFR3): A Novel Tool for Detection and Surveillance of Recurrent Bladder Cancer. Sens. Actuators B Chem. 2021, 326, 128829. [Google Scholar] [CrossRef]
  104. Rushworth, J.V.; Ahmed, A.; Griffiths, H.H.; Pollock, N.M.; Hooper, N.M.; Millner, P.A. A Label-Free Electrical Impedimetric Biosensor for the Specific Detection of Alzheimer’s Amyloid-Beta Oligomers. Biosens. Bioelectron. 2014, 56, 83–90. [Google Scholar] [CrossRef] [PubMed]
  105. Shamsuddin, S.H.; Gibson, T.D.; Tomlinson, D.C.; McPherson, M.J.; Jayne, D.G.; Millner, P.A. Reagentless Affimer- and Antibody-Based Impedimetric Biosensors for CEA-Detection Using a Novel Non-Conducting Polymer. Biosens. Bioelectron. 2021, 178, 113013. [Google Scholar] [CrossRef]
  106. Aydın, E.B. Highly Sensitive Impedimetric Immunosensor for Determination of Interleukin 6 as a Cancer Biomarker by Using Conjugated Polymer Containing Epoxy Side Groups Modified Disposable ITO Electrode. Talanta 2020, 215, 120909. [Google Scholar] [CrossRef]
  107. María Dip Gandarilla, A.; Correa Glória, J.; Romaguera Barcelay, Y.; de Souza, R.F.B.; André Morais Mariuba, L.; Ricardo Brito, W. Application of Egg Yolk IgY on Carboxylated Polypyrrole Films for Impedimetric Detection of PfHRP2 Antigen. Bioelectrochemistry 2022, 148, 108273. [Google Scholar] [CrossRef]
  108. Yuan, G.; He, J.; Li, Y.; Xu, W.; Gao, L.; Yu, C. A Novel Ultrasensitive Electrochemical Immunosensor Based on Carboxy-Endcapped Conductive Polypyrrole for the Detection of Gypican-3 in Human Serum. Anal. Methods 2015, 7, 1745–1750. [Google Scholar] [CrossRef]
  109. Aydın, E.B.; Aydın, M.; Sezgintürk, M.K. Fabrication of Electrochemical Immunosensor Based on Acid-Substituted Poly(Pyrrole) Polymer Modified Disposable ITO Electrode for Sensitive Detection of CCR4 Cancer Biomarker in Human Serum. Talanta 2021, 222, 121487. [Google Scholar] [CrossRef]
  110. Aydın, E.B.; Aydın, M.; Sezgintürk, M.K. A Label-Free Electrochemical Biosensor for Highly Sensitive Detection of GM2A Based on Gold Nanoparticles/Conducting Amino-Functionalized Thiophene Polymer Layer. Sens. Actuators B Chem. 2023, 392, 134025. [Google Scholar] [CrossRef]
  111. Aydın, E.B.; Aydın, M.; Sezgintürk, M.K. A Novel Electrochemical Immunosensor Based on Acetylene Black/Epoxy-Substituted-Polypyrrole Polymer Composite for the Highly Sensitive and Selective Detection of Interleukin 6. Talanta 2021, 222, 121596. [Google Scholar] [CrossRef]
  112. Amouzadeh Tabrizi, M.; Shamsipur, M.; Mostafaie, A. A High Sensitive Label-Free Immunosensor for the Determination of Human Serum IgG Using Overoxidized Polypyrrole Decorated with Gold Nanoparticle Modified Electrode. Mater. Sci. Eng. C 2016, 59, 965–969. [Google Scholar] [CrossRef]
  113. Taheri, N.; Khoshsafar, H.; Ghanei, M.; Ghazvini, A.; Bagheri, H. Dual-Template Rectangular Nanotube Molecularly Imprinted Polypyrrole for Label-Free Impedimetric Sensing of AFP and CEA as Lung Cancer Biomarkers. Talanta 2022, 239, 123146. [Google Scholar] [CrossRef]
  114. Aydın, E.B.; Aydın, M.; Sezgintürk, M.K. New Impedimetric Sandwich Immunosensor for Ultrasensitive and Highly Specific Detection of Spike Receptor Binding Domain Protein of SARS-CoV-2. ACS Biomater. Sci. Eng. 2021, 7, 3874–3885. [Google Scholar] [CrossRef]
  115. Ruecha, N.; Shin, K.; Chailapakul, O.; Rodthongkum, N. Label-Free Paper-Based Electrochemical Impedance Immunosensor for Human Interferon Gamma Detection. Sens. Actuators B Chem. 2019, 279, 298–304. [Google Scholar] [CrossRef]
  116. Kim, M.; Iezzi, R.; Shim, B.S.; Martin, D.C. Impedimetric Biosensors for Detecting Vascular Endothelial Growth Factor (VEGF) Based on Poly(3,4-Ethylene Dioxythiophene) (PEDOT)/Gold Nanoparticle (Au NP) Composites. Front. Chem. 2019, 7, 234. [Google Scholar] [CrossRef]
  117. Xu, M.; Yadavalli, V.K. Flexible Biosensors for the Impedimetric Detection of Protein Targets Using Silk-Conductive Polymer Biocomposites. ACS Sens. 2019, 4, 1040–1047. [Google Scholar] [CrossRef]
  118. Aydın, E.B.; Aydın, M.; Sezgintürk, M.K. A novel electrochemical impedance immunosensor for the quantification of CYFRA 21-1 in human serum. Microchim. Acta 2023, 190, 235. [Google Scholar] [CrossRef]
  119. Karami, P.; Bagheri, H.; Johari-Ahar, M.; Khoshsafar, H.; Arduini, F.; Afkhami, A. Dual-Modality Impedimetric Immunosensor for Early Detection of Prostate-Specific Antigen and Myoglobin Markers Based on Antibody-Molecularly Imprinted Polymer. Talanta 2019, 202, 111–122. [Google Scholar] [CrossRef]
  120. Özcan, B.; Sezgintürk, M.K. A Novel and Disposable GP- Based Impedimetric Biosensor Using Electropolymerization Process with PGA for Highly Sensitive Determination of Leptin: Early Diagnosis of Childhood Obesity. Talanta 2021, 225, 121985. [Google Scholar] [CrossRef]
  121. Xia, N.; Huang, Y.; Cui, Z.; Liu, S.; Deng, D.; Liu, L.; Wang, J. Impedimetric Biosensor for Assay of Caspase-3 Activity and Evaluation of Cell Apoptosis Using Self-Assembled Biotin-Phenylalanine Network as Signal Enhancer. Sens. Actuators B Chem. 2020, 320, 128436. [Google Scholar] [CrossRef]
  122. Lobato, A.; Šubic, M.; Romih, T.; Žibret, L.; Metarapi, D.; Benčina, M.; Jerala, R.; Vidović, K.; Hočevar, S.B.; Tasić, N. Development of a Simple Gelatin-Based Sensing Platform for the Sensitive Label-Free Impedimetric Detection of SARS-CoV-2. Electrochim. Acta 2023, 463, 142823. [Google Scholar] [CrossRef]
  123. Ben Hassine, A.; Raouafi, N.; Moreira, F.T.C. Novel Electrochemical Molecularly Imprinted Polymer-Based Biosensor for Tau Protein Detection. Chemosensors 2021, 9, 238. [Google Scholar] [CrossRef]
  124. Yang, J.C.; Cho, C.H.; Choi, D.Y.; Park, J.P.; Park, J. Microcontact Surface Imprinting of Affinity Peptide for Electrochemical Impedimetric Detection of Neutrophil Gelatinase-Associated Lipocalin. Sens. Actuators B Chem. 2022, 364, 131916. [Google Scholar] [CrossRef]
  125. Tugce Yaman, Y.; Akbal Vural, O.; Bolat, G.; Abaci, S. Peptide Nanotubes/Self-Assembled Polydopamine Molecularly Imprinted Biochip for the Impedimetric Detection of Human Interleukin-6. Bioelectrochemistry 2022, 145, 108053. [Google Scholar] [CrossRef] [PubMed]
  126. Roushani, M.; Zalpour, N. Impedimetric Ultrasensitive Detection of Trypsin Based on Hybrid Aptamer-2DMIP Using a Glassy Carbon Electrode Modified by Nickel Oxide Nanoparticle. Microchem. J. 2022, 172, 106955. [Google Scholar] [CrossRef]
  127. Choi, D.Y.; Yang, J.C.; Hong, S.W.; Park, J. Molecularly Imprinted Polymer-Based Electrochemical Impedimetric Sensors on Screen-Printed Carbon Electrodes for the Detection of Trace Cytokine IL-1β. Biosens. Bioelectron. 2022, 204, 114073. [Google Scholar] [CrossRef] [PubMed]
  128. Tasić, N.; Cavalcante, L.; Deffune, E.; Góes, M.S.; Paixão, T.R.L.C.; Gonçalves, L.M. Probeless and Label-Free Impedimetric Biosensing of D-Dimer Using Gold Nanoparticles Conjugated with Dihexadecylphosphate on Screen-Printed Carbon Electrodes. Electrochim. Acta 2021, 397, 139244. [Google Scholar] [CrossRef]
  129. Erdem, A.; Congur, G. Dendrimer Enriched Single-Use Aptasensor for Impedimetric Detection of Activated Protein C. Colloids Surf. B Biointerfaces 2014, 117, 338–345. [Google Scholar] [CrossRef]
  130. Erdem, A.; Congur, G. Dendrimer Modified 8-Channel Screen-Printed Electrochemical Array System for Impedimetric Detection of Activated Protein C. Sens. Actuators B Chem. 2014, 196, 168–174. [Google Scholar] [CrossRef]
  131. Uygun, Z.O.; Şahin, Ç.; Yılmaz, M.; Akçay, Y.; Akdemir, A.; Sağın, F. Fullerene-PAMAM(G5) Composite Modified Impedimetric Biosensor to Detect Fetuin-A in Real Blood Samples. Anal. Biochem. 2018, 542, 11–15. [Google Scholar] [CrossRef]
  132. Zhang, C.; Zhang, D.; Ma, Z.; Han, H. Cascade Catalysis-Initiated Radical Polymerization Amplified Impedimetric Immunosensor for Ultrasensitive Detection of Carbohydrate Antigen 15-3. Biosens. Bioelectron. 2019, 137, 1–7. [Google Scholar] [CrossRef]
  133. Gupta, A.; Sharma, S.K.; Pachauri, V.; Ingebrandt, S.; Singh, S.; Sharma, A.L.; Deep, A. Sensitive Impedimetric Detection of Troponin I with Metal–Organic Framework Composite Electrode. RSC Adv. 2021, 11, 2167–2174. [Google Scholar] [CrossRef]
  134. Vijayaraghavan, P.; Wang, Y.-Y.; Palanisamy, S.; Lee, L.-Y.; Chen, Y.-K.; Tzou, S.-C.; Yuan, S.-S.F.; Wang, Y.-M. Hierarchical Ensembles of FeCo Metal-Organic Frameworks Reinforced Nickel Foam as an Impedimetric Sensor for Detection of IL-1RA in Human Samples. Chem. Eng. J. 2023, 458, 141444. [Google Scholar] [CrossRef]
  135. Han, C.; Xing, W.; Li, W.; Fang, X.; Zhao, J.; Ge, F.; Ding, W.; Qu, P.; Luo, Z.; Zhang, L. Aptamers Dimerization Inspired Biomimetic Clamp Assay towards Impedimetric SARS-CoV-2 Antigen Detection. Sens. Actuators B Chem. 2023, 380, 133387. [Google Scholar] [CrossRef]
  136. Li, Y.; Wang, C.; Li, Z.; Wang, M.; He, L.; Zhang, Z. Zirconium-Porphyrin Complex as Novel Nanocarrier for Label-Free Impedimetric Biosensing Neuron-Specific Enolase. Sens. Actuators B Chem. 2020, 314, 128090. [Google Scholar] [CrossRef]
  137. Centane, S.; Mgidlana, S.; Openda, Y.; Nyokong, T. Electrochemical Detection of Human Epidermal Growth Factor Receptor 2 Using an Aptamer on Cobalt Phthalocyanines—Cerium Oxide Nanoparticle Conjugate. Bioelectrochemistry 2022, 146, 108146. [Google Scholar] [CrossRef]
  138. Centane, S.; Mgidlana, S.; Openda, Y.I.; Ndebele, N.; Nyokong, T. Effect of Symmetry and Substituents of Cobalt Based Phthalocyanines in Aptasensor Design for the Electrochemical Impedimetric Detection of the Human Epidermal Growth Factor Receptor 2. J. Electroanal. Chem. 2023, 941, 117524. [Google Scholar] [CrossRef]
  139. Erkmen, C.; Tiğ, G.A.; Uslu, B. First Label-Free Impedimetric Aptasensor Based on Au NPs/TiO2 NPs for the Determination of Leptin. Sens. Actuators B Chem. 2022, 358, 131420. [Google Scholar] [CrossRef]
  140. Derkus, B.; Cebesoy Emregul, K.; Mazi, H.; Emregul, E.; Yumak, T.; Sinag, A. Protein A Immunosensor for the Detection of Immunoglobulin G by Impedance Spectroscopy. Bioprocess Biosyst. Eng. 2013, 37, 965–976. [Google Scholar] [CrossRef]
  141. Arkusz, K.; Paradowska, E. Impedimetric Detection of Femtomolar Levels of Interleukin 6, Interleukin 8, and Tumor Necrosis Factor Alpha Based on Thermally Modified Nanotubular Titanium Dioxide Arrays. Nanomaterials 2020, 10, 2399. [Google Scholar] [CrossRef]
  142. Eveness, J.; Cao, L.; Kiely, J.; Luxton, R. Equivalent Circuit Model of a Non-Faradaic Impedimetric ZnO Nano-Crystal Biosensor. J. Electroanal. Chem. 2022, 906, 116003. [Google Scholar] [CrossRef]
  143. Nunez, F.A.; Castro, A.C.H.; Daher, I.P.; Cunha-Neto, E.; Kalil, J.; Boscardin, S.B.; Lanfredi, A.J.C.; de Oliveira, V.L.; Alves, W.A. ZnO-Based Electrochemical Immunosensor to Assess Vaccine-Induced Antibody-Mediated Immunity against Wild-Type and Gamma SARS-CoV-2 Strains. Biosensors 2023, 13, 371. [Google Scholar] [CrossRef] [PubMed]
  144. Hussein, H.A.; Kandeil, A.; Gomaa, M.R.; Hassan, R.Y.A. Double-Antibody-Based Nano-Biosensing System for the Onsite Monitoring of SARS-CoV-2 Variants. Microsyst. Nanoeng. 2023, 9, 105. [Google Scholar] [CrossRef]
  145. Sandil, D.; Srivastava, S.; Khatri, R.; Sharma, K.; Puri, N.K. Synthesis and Fabrication of 2D Tungsten Trioxide Nanosheets Based Platform for Impedimetric Sensing of Cardiac Biomarker. Sens. Bio-Sens. Res. 2021, 32, 100423. [Google Scholar] [CrossRef]
  146. Caballero, D.; Martinez, E.; Bausells, J.; Errachid, A.; Samitier, J. Impedimetric Immunosensor for Human Serum Albumin Detection on a Direct Aldehyde-Functionalized Silicon Nitride Surface. Anal. Chim. Acta 2012, 720, 43–48. [Google Scholar] [CrossRef] [PubMed]
  147. Chuang, Y.-H.; Chang, Y.-T.; Liu, K.-L.; Chang, H.-Y.; Yew, T.-R. Electrical Impedimetric Biosensors for Liver Function Detection. Biosens. Bioelectron. 2011, 28, 368–372. [Google Scholar] [CrossRef] [PubMed]
  148. Low, Y.K.; Chan, J.; Soraya, G.V.; Buffet, C.; Abeyrathne, C.D.; Huynh, D.H.; Skafidas, E.; Kwan, P.; Rogerson, S.J. Development of an Ultrasensitive Impedimetric Immunosensor Platform for Detection of Plasmodium Lactate Dehydrogenase. Sensors 2019, 19, 2446. [Google Scholar] [CrossRef] [PubMed]
  149. Tanak, A.S.; Jagannath, B.; Tamrakar, Y.; Muthukumar, S.; Prasad, S. Non-Faradaic Electrochemical Impedimetric Profiling of Procalcitonin and C-Reactive Protein as a Dual Marker Biosensor for Early Sepsis Detection. Anal. Chim. Acta X 2019, 3, 100029. [Google Scholar] [CrossRef] [PubMed]
  150. Cao, L.; Kiely, J.; Piano, M.; Luxton, R. A Copper Oxide/Zinc Oxide Composite Nano-Surface for Use in a Biosensor. Materials 2019, 12, 1126. [Google Scholar] [CrossRef] [PubMed]
  151. Cao, L.; Kiely, J.; Piano, M.; Luxton, R. Nanoparticle-Based 3D Membrane for Impedimetric Biosensor Applications. Bioelectrochemistry 2020, 136, 107593. [Google Scholar] [CrossRef] [PubMed]
  152. Januarie, K.C.; Oranzie, M.; Feleni, U.; Iwuoha, E. Quantum Dot Amplified Impedimetric Aptasensor for Interferon-Gamma. Electrochim. Acta 2023, 463, 142825. [Google Scholar] [CrossRef]
  153. Ribeiro, J.F.F.; Melo, J.R.S.; de L. Santos, C.; Chaves, C.R.; Cabral Filho, P.E.; Pereira, G.; Santos, B.S.; Pereira, G.A.L.; Rosa, D.S.; Ribeiro, R.T.; et al. Sensitive Zika Biomarker Detection Assisted by Quantum Dot-Modified Electrochemical Immunosensing Platform. Colloids Surf. B Biointerfaces 2023, 221, 112984. [Google Scholar] [CrossRef]
  154. Sri, S.; Chauhan, D.; Lakshmi, G.B.V.S.; Thakar, A.; Solanki, P.R. MoS2 Nanoflower Based Electrochemical Biosensor for TNF Alpha Detection in Cancer Patients. Electrochim. Acta 2022, 405, 139736. [Google Scholar] [CrossRef]
  155. Mi, X.; Li, H.; Tu, Y. An Aptamer Biosensing Strategy for Label-Free Assay of Dual Acute Myocardial Infarction Biomarkers Built upon AuNPs/Ti3C2-MXenes. Chemosensors 2023, 11, 157. [Google Scholar] [CrossRef]
  156. Dalila, R.N.; Arshad, M.K.M.; Gopinath, S.C.B.; Ibau, C.; Nuzaihan, M.N.M.; Fathil, M.F.M.; Azmi, U.Z.M.; Anbu, P. Faradaic Electrochemical Impedimetric Analysis on MoS2/Au-NPs Decorated Surface for C-Reactive Protein Detection. J. Taiwan Inst. Chem. Eng. 2022, 138, 104450. [Google Scholar] [CrossRef]
  157. Bertok, T.; Lorencova, L.; Hroncekova, S.; Gajdosova, V.; Jane, E.; Hires, M.; Kasak, P.; Kaman, O.; Sokol, R.; Bella, V.; et al. Advanced Impedimetric Biosensor Configuration and Assay Protocol for Glycoprofiling of a Prostate Oncomarker Using Au Nanoshells with a Magnetic Core. Biosens. Bioelectron. 2019, 131, 24–29. [Google Scholar] [CrossRef]
  158. Adesina, A.; Adeniyi, O.; Mashazi, P. Nanomagnet Bioconjugates with Anti-CRP Polyclonal Antibodies as Nanobioprobes for Enhanced Impedimetric Detection of CRP. Electroanalysis 2022, 35, e202200059. [Google Scholar] [CrossRef]
  159. Prajapati, V.; Shrivastav, P.S. Bioreceptors for Antigen–Antibody Interactions. In Biosensors Nanotechnology; Inamuddin, Altalhi, T., Eds.; John Wiley & Sons, Inc.: Hoboken, NJ, USA, 2023; pp. 371–394. [Google Scholar] [CrossRef]
  160. Balayan, S.; Chauhan, N.; Rosario, W.; Jain, U. Biosensor Development for C-Reactive Protein Detection: A Review. Appl. Surf. Sci. Adv. 2022, 12, 100343. [Google Scholar] [CrossRef]
  161. Chen, Z.-B.; Jin, H.-H.; Yang, Z.-G.; He, D.-P. Recent Advances on Bioreceptors and Metal Nanomaterials-Based Electrochemical Impedance Spectroscopy Biosensors. Rare Met. 2022, 42, 1098–1117. [Google Scholar] [CrossRef]
  162. Gupta, U.; Gupta, V.; Arun, R.K.; Chanda, N. Recent Advances in Enzymatic Biosensors for Point-of-care Detection of Biomolecules. Biotechnol. Bioeng. 2022, 119, 3393–3407. [Google Scholar] [CrossRef] [PubMed]
  163. Dkhar, D.S.; Kumari, R.; Mahapatra, S.; Divya; Kumar, R.; Tripathi, T.; Chandra, P. Antibody-Receptor Bioengineering and Its Implications in Designing Bioelectronic Devices. Int. J. Biol. Macromol. 2022, 218, 225–242. [Google Scholar] [CrossRef]
  164. Corrigan, D.K. Impedance Biosensors. In Fundamentals of Sensor Technology; Woodhead Publishing: Sawston, UK, 2023; pp. 239–264. [Google Scholar] [CrossRef]
  165. Balayan, S.; Chauhan, N.; Chandra, R.; Jain, U. Molecular Imprinting Based Electrochemical Biosensor for Identification of Serum Amyloid A (SAA), a Neonatal Sepsis Biomarker. Int. J. Biol. Macromol. 2022, 195, 589–597. [Google Scholar] [CrossRef]
  166. Weaver, S.; Mohammadi, M.H.; Nakatsuka, N. Aptamer-Functionalized Capacitive Biosensors. Biosens. Bioelectron. 2023, 224, 115014. [Google Scholar] [CrossRef]
  167. Lee, J.; Kane, B.J.; Khanwalker, M.; Sode, K. Development of an Electrochemical Impedance Spectroscopy Based Biosensor for Detection of Ubiquitin C-Terminal Hydrolase L1. Biosens. Bioelectron. 2022, 208, 114232. [Google Scholar] [CrossRef] [PubMed]
  168. Choudhury, S.; Nautiyal, R.; Thakkar, D.K.; Betty, C.A. Thickness Dependence of Nanocrystalline Tin Oxide Thin Films in Capacitive Biosensor Characterization. J. Electroanal. Chem. 2020, 877, 114742. [Google Scholar] [CrossRef]
  169. Chen, H.-J.; Chen, R.L.C.; Hsieh, B.-C.; Hsiao, H.-Y.; Kung, Y.; Hou, Y.-T.; Cheng, T.-J. Label-Free and Reagentless Capacitive Aptasensor for Thrombin. Biosens. Bioelectron. 2019, 131, 53–59. [Google Scholar] [CrossRef] [PubMed]
  170. Shanbhag, M.M.; Manasa, G.; Mascarenhas, R.J.; Mondal, K.; Shetti, N.P. Fundamentals of Bio-Electrochemical Sensing. Chem. Eng. J. Adv. 2023, 16, 100516. [Google Scholar] [CrossRef]
  171. Nano, A.; Furst, A.L.; Hill, M.G.; Barton, J.K. DNA Electrochemistry: Charge-Transport Pathways through DNA Films on Gold. J. Am. Chem. Soc. 2021, 143, 11631–11640. [Google Scholar] [CrossRef] [PubMed]
  172. Yuan, L.; Liu, L. Peptide-Based Electrochemical Biosensing. Sens. Actuators B Chem. 2021, 344, 130232. [Google Scholar] [CrossRef]
  173. Brothers, M.C.; Moore, D.; St. Lawrence, M.; Harris, J.; Joseph, R.M.; Ratcliff, E.; Ruiz, O.N.; Glavin, N.; Kim, S.S. Impact of Self-Assembled Monolayer Design and Electrochemical Factors on Impedance-Based Biosensing. Sensors 2020, 20, 2246. [Google Scholar] [CrossRef]
  174. Xu, X.; Makaraviciute, A.; Kumar, S.; Wen, C.; Sjödin, M.; Abdurakhmanov, E.; Danielson, U.H.; Nyholm, L.; Zhang, Z. Structural Changes of Mercaptohexanol Self-Assembled Monolayers on Gold and Their Influence on Impedimetric Aptamer Sensors. Anal. Chem. 2019, 91, 14697–14704. [Google Scholar] [CrossRef]
  175. Nakatsuka, N.; Faillétaz, A.; Eggemann, D.; Forró, C.; Vörös, J.; Momotenko, D. Aptamer Conformational Change Enables Serotonin Biosensing with Nanopipettes. Anal. Chem. 2021, 93, 4033–4041. [Google Scholar] [CrossRef]
  176. Raposo, M.; Soreto, S.; Moreirinha, C.; Gomes, M.T.S.R.; Botelho, M.J.; Costa, L.C.; Rudnitskaya, A. Notice of Removal: Impedimetric Electronic Tongue for the Detection of Marine Toxins. In Proceedings of the 2022 IEEE International Symposium on Olfaction and Electronic Nose (ISOEN), Aveiro, Portugal, 29 May–1 June 2022. [Google Scholar] [CrossRef]
  177. Kumar, D.; Prasad, M.; Mohan, H. Biological Recognition Elements. In Electrochemical Sensors; Woodhead Publishing: Sawston, UK, 2022; pp. 213–239. [Google Scholar] [CrossRef]
  178. Costa, S.P.; Nogueira, C.L.; Cunha, A.P.; Lisac, A.; Carvalho, C.M. Potential of Bacteriophage Proteins as Recognition Molecules for Pathogen Detection. Crit. Rev. Biotechnol. 2022, 43, 787–804. [Google Scholar] [CrossRef]
  179. Di Mauro, V.; Lauta, F.C.; Modica, J.; Appleton, S.L.; De Franciscis, V.; Catalucci, D. Diagnostic and Therapeutic Aptamers: A Promising Pathway to Improved Cardiovascular Disease Management. JACC Basic Transl. Sci. 2023, in press. [Google Scholar] [CrossRef]
  180. Debie, P.; Lafont, C.; Defrise, M.; Hansen, I.; van Willigen, D.M.; van Leeuwen, F.W.B.; Gijsbers, R.; D’Huyvetter, M.; Devoogdt, N.; Lahoutte, T.; et al. Size and Affinity Kinetics of Nanobodies Influence Targeting and Penetration of Solid Tumours. J. Control. Release 2020, 317, 34–42. [Google Scholar] [CrossRef] [PubMed]
  181. Plach, M.; Schubert, T. Biophysical Characterization of Aptamer-Target Interactions. In Aptamers in Biotechnology; Advances in Biochemical Engineering/Biotechnology; Springer: Cham, Switzerland, 2020; pp. 1–15. [Google Scholar] [CrossRef]
  182. Mollarasouli, F.; Kurbanoglu, S.; Ozkan, S.A. The Role of Electrochemical Immunosensors in Clinical Analysis. Biosensors 2019, 9, 86. [Google Scholar] [CrossRef] [PubMed]
  183. Jha, S.K.; Soni, A.; Raj, R.; Bala, S.; Sharma, K.; Panwar, S.; Singh, H. Functionalization, Immobilization and Stabilization of Biomolecules in Microfluidic Devices. In Immobilization Strategies; Gels Horizons: From Science to Smart Materials; Springer: Singapore, 2020; pp. 509–533. [Google Scholar] [CrossRef]
  184. Cho, I.-H.; Kim, D.H.; Park, S. Electrochemical Biosensors: Perspective on Functional Nanomaterials for on-Site Analysis. Biomater. Res. 2020, 24, 6. [Google Scholar] [CrossRef] [PubMed]
  185. Yan, X.; Tang, J.; Tanner, D.; Ulstrup, J.; Xiao, X. Direct Electrochemical Enzyme Electron Transfer on Electrodes Modified by Self-Assembled Molecular Monolayers. Catalysts 2020, 10, 1458. [Google Scholar] [CrossRef]
  186. Patel, M.; Agrawal, M.; Srivastava, A. Signal Amplification Strategies in Electrochemical Biosensors via Antibody Immobilization and Nanomaterial-Based Transducers. Mater. Adv. 2022, 3, 8864–8885. [Google Scholar] [CrossRef]
  187. Sun, H.; Li, D.; Yue, X.; Hong, R.; Yang, W.; Liu, C.; Xu, H.; Lu, J.; Dong, L.; Wang, G.; et al. A Review of Transition Metal Dichalcogenides-Based Biosensors. Front. Bioeng. Biotechnol. 2022, 10, 941135. [Google Scholar] [CrossRef]
  188. Puumala, L.S.; Grist, S.M.; Morales, J.M.; Bickford, J.R.; Chrostowski, L.; Shekhar, S.; Cheung, K.C. Biofunctionalization of Multiplexed Silicon Photonic Biosensors. Biosensors 2023, 13, 53. [Google Scholar] [CrossRef]
  189. Zhu, L.; Chang, Y.; Li, Y.; Qiao, M.; Liu, L. Biosensors Based on the Binding Events of Nitrilotriacetic Acid–Metal Complexes. Biosensors 2023, 13, 507. [Google Scholar] [CrossRef]
  190. Cajigas, S.; Orozco, J. Nanobioconjugates for Signal Amplification in Electrochemical Biosensing. Molecules 2020, 25, 3542. [Google Scholar] [CrossRef]
  191. Maziz, A.; Özgür, E.; Bergaud, C.; Uzun, L. Progress in Conducting Polymers for Biointerfacing and Biorecognition Applications. Sens. Actuators Rep. 2021, 3, 100035. [Google Scholar] [CrossRef]
  192. Lakard, B. Electrochemical Biosensors Based on Conducting Polymers: A Review. Appl. Sci. 2020, 10, 6614. [Google Scholar] [CrossRef]
  193. Lamaoui, A.; García-Guzmán, J.J.; Amine, A.; Palacios-Santander, J.M.; Cubillana-Aguilera, L. Synthesis Techniques of Molecularly Imprinted Polymer Composites. In Molecularly Imprinted Polymer Composites; Woodhead Publishing: Sawston, UK, 2021; pp. 49–91. [Google Scholar] [CrossRef]
  194. Baio, J.E.; Graham, D.J.; Castner, D.G. Surface Analysis Tools for Characterizing Biological Materials. Chem. Soc. Rev. 2020, 49, 3278–3296. [Google Scholar] [CrossRef]
  195. Kaur, S.; Kaur, I. Self-assembly of P-Aminothiophenol on Gold Surface: Application for Impedimetric and Potentiometric Sensing of Cobalt (II) Ions—A Comparative Study. Electroanalysis 2019, 31, 2507–2517. [Google Scholar] [CrossRef]
  196. Ghorbanizamani, F.; Yagci, Y.; Moulahoum, H.; Timur, S. Self-Assembled Monolayer–Based Nanoscaled Surfaces. In Biophysics at the Nanoscale; Academic Press: Cambridge, MA, USA, 2024; pp. 1–25. [Google Scholar] [CrossRef]
  197. Miller, N.L.; Clark, T.; Raman, R.; Sasisekharan, R. Learned Features of Antibody-Antigen Binding Affinity. Front. Mol. Biosci. 2023, 10, 1112738. [Google Scholar] [CrossRef] [PubMed]
  198. Van Regenmortel, M.H.V. Specificity, Polyspecificity and Heterospecificity of Antibody-Antigen Recognition. In HIV/AIDS: Immunochemistry, Reductionism and Vaccine Design; Springer: Cham, Switzerland, 2019; pp. 39–56. [Google Scholar] [CrossRef]
  199. Kapingidza, A.B.; Kowal, K.; Chruszcz, M. Antigen–Antibody Complexes. In Vertebrate and Invertebrate Respiratory Proteins, Lipoproteins and Other Body Fluid Proteins; Springer: Cham, Switzerland, 2020; pp. 465–497. [Google Scholar] [CrossRef]
  200. Shoute, L.C.T.; Abdelrasoul, G.N.; Ma, Y.; Duarte, P.A.; Edwards, C.; Zhuo, R.; Zeng, J.; Feng, Y.; Charlton, C.L.; Kanji, J.N.; et al. Label-Free Impedimetric Immunosensor for Point-of-Care Detection of COVID-19 Antibodies. Microsyst. Nanoeng. 2023, 9, 3. [Google Scholar] [CrossRef] [PubMed]
  201. Lorenzen, A.L.; dos Santos, A.M.; dos Santos, L.P.; da Silva Pinto, L.; Conceição, F.R.; Wolfart, F. PEDOT-AuNPs-Based Impedimetric Immunosensor for the Detection of SARS-CoV-2 Antibodies. Electrochim. Acta 2022, 404, 139757. [Google Scholar] [CrossRef]
  202. Trier, N.; Hansen, P.; Houen, G. Peptides, Antibodies, Peptide Antibodies and More. Int. J. Mol. Sci. 2019, 20, 6289. [Google Scholar] [CrossRef]
  203. Yin, S.; Niu, L.; Liu, Y. Recent Progress on Techniques in the Detection of Aflatoxin B1 in Edible Oil: A Mini Review. Molecules 2022, 27, 6141. [Google Scholar] [CrossRef]
  204. Xia, Y.; Chen, H.; Liu, R.; Shi, F.; Ren, C.; Li, J.; Zhao, J.; Chen, X.; Yang, Z. Mesoporous SiO2 Sphere-Based Electrochemical Impedance Immunosensor for Ultrasensitive Detection of Bovine Interferon-γ. J. Anal. Test. 2023, 7, 295–303. [Google Scholar] [CrossRef]
  205. Pang, S.; Yu, H.; Zhang, Y.; Jiao, Y.; Zheng, Z.; Wang, M.; Zhang, H.; Liu, A. Bioscreening Specific Peptide-Expressing Phage and Its Application in Sensitive Dual-Mode Immunoassay of SARS-CoV-2 Spike Antigen. Talanta 2024, 266, 125093. [Google Scholar] [CrossRef]
  206. Devi, S.; Prakash, J.; Tsujimura, S. Graphene Oxide-Based Nanomaterials for Electrochemical Bio/Immune Sensing and Its Advancements in Health Care Applications: A Review. Hybrid Adv. 2024, 5, 100123. [Google Scholar] [CrossRef]
  207. Khorshed, A.A.; Savchenko, O.; Liu, F.; Lian, S.; Zeng, J.; Ren, S.; Gu, J.; Jha, N.; Zhong, Y.; Wang, J.; et al. Development of an Impedance-Based Biosensor for Determination of IgG Galactosylation Levels. Biosens. Bioelectron. 2024, 245, 115793. [Google Scholar] [CrossRef]
  208. Sharafeldin, M.; McCaffrey, K.; Rusling, J.F. Influence of Antibody Immobilization Strategy on Carbon Electrode Immunoarrays. Analyst 2019, 144, 5108–5116. [Google Scholar] [CrossRef]
  209. Özcan, B.; Sezgintürk, M.K. Fabrication of an Ultrasensitive and Single-Use Graphite Paper Based Immunosensor for Neuropeptide Y Detection: A Promising Biosensing System for Early Detection of Childhood Obesity. Mater. Today Commun. 2022, 33, 104797. [Google Scholar] [CrossRef]
  210. Santos, A.; de Souza Brandão, A.M.; Hryniewicz, B.M.; Abreu, H.; Bach-Toledo, L.; da Silva, S.S.; Deller, A.E.; Rogerio, V.Z.; Rodrigues, D.B.; Hiraiwa, P.M.; et al. COVID-19 Impedimetric Biosensor Based on Polypyrrole Nanotubes, Nickel Hydroxide and VHH Antibody Fragment: Specific, Sensitive, and Rapid Viral Detection in Saliva Samples. Mater. Today Chem. 2023, 30, 101597. [Google Scholar] [CrossRef]
  211. Adesina, A.; Adeniyi, O.; Mashazi, P. Online Regeneration of a Piezoelectric and Impedimetric Immunosensor for the Detection of C-Reactive Protein on the Oriented Antibody Gold Surface. Anal. Methods 2023, 15, 1157–1167. [Google Scholar] [CrossRef] [PubMed]
  212. Karuppaiah, G.; Lee, M.; Bhansali, S.; Manickam, P. Electrochemical Sensors for Cortisol Detection: Principles, Designs, Fabrication, and Characterisation. Biosens. Bioelectron. 2023, 239, 115600. [Google Scholar] [CrossRef] [PubMed]
  213. Vidic, J.; Manzano, M. Electrochemical Biosensors for Rapid Pathogen Detection. Curr. Opin. Electrochem. 2021, 29, 100750. [Google Scholar] [CrossRef]
  214. Upasham, S.; Banga, I.K.; Jagannath, B.; Paul, A.; Lin, K.-C.; Muthukumar, S.; Prasad, S. Electrochemical Impedimetric Biosensors, Featuring the Use of Room Temperature Ionic Liquids (RTILs): Special Focus on Non-Faradaic Sensing. Biosens. Bioelectron. 2021, 177, 112940. [Google Scholar] [CrossRef]
  215. Zhang, X.; Qi, J.; Zhang, Q.; Xue, Y.; Meng, F.; Zhang, J.; Liu, Y.; Yang, G.; Wu, C. A Novel Sandwich Impedimetric Immunosensor for Detection of Apolipoprotein-A1 Based on the Gold Nanoparticle–Hybridized Mercapto-β-Cyclodextrin-Pb(II) Metal–Organic Framework. Microchim. Acta 2022, 190, 33. [Google Scholar] [CrossRef]
  216. Dahiya, T.; Ravina; Mann, A.; Mohan, H.; Sharma, M.; Pundir, C.S.; Rana, J.S. Impedimetric Immunosensor Based on Chitosan-Modified Gold Wire with Au@rGO Nanocomposite for the Detection of Brain Natriuretic Peptide (BNP). J. Mater. Sci. 2023, 58, 4739–4752. [Google Scholar] [CrossRef]
  217. Khan, S.; Hussain, A.; Fahimi, H.; Aliakbari, F.; Haj Bloukh, S.; Edis, Z.; Mahdi Nejadi Babadaei, M.; Izadi, Z.; Shiri Varnamkhasti, B.; Jahanshahi, F.; et al. A Review on the Therapeutic Applications of Aptamers and Aptamer-Conjugated Nanoparticles in Cancer, Inflammatory and Viral Diseases. Arab. J. Chem. 2022, 15, 103626. [Google Scholar] [CrossRef]
  218. Douaki, A.; Garoli, D.; Inam, A.K.M.S.; Angeli, M.A.C.; Cantarella, G.; Rocchia, W.; Wang, J.; Petti, L.; Lugli, P. Smart Approach for the Design of Highly Selective Aptamer-Based Biosensors. Biosensors 2022, 12, 574. [Google Scholar] [CrossRef]
  219. Sousa, D.A.; Carneiro, M.; Ferreira, D.; Moreira, F.T.C.; Sales, M.G.F.; Rodrigues, L.R. Recent Advances in the Selection of Cancer-Specific Aptamers for the Development of Biosensors. Curr. Med. Chem. 2022, 29, 5850–5880. [Google Scholar] [CrossRef]
  220. Preuß, J.-A.; Reich, P.; Bahner, N.; Bahnemann, J. Impedimetric Aptamer-Based Biosensors: Applications. In Aptamers in Biotechnology; Advances in Biochemical Engineering/Biotechnology; Springer: Cham, Switzerland, 2020; pp. 43–91. [Google Scholar] [CrossRef]
  221. Ramanathan, S.; Gopinath, S.C.B.; Ismail, Z.H.; Md Arshad, M.K.; Poopalan, P. Aptasensing Nucleocapsid Protein on Nanodiamond Assembled Gold Interdigitated Electrodes for Impedimetric SARS-CoV-2 Infectious Disease Assessment. Biosens. Bioelectron. 2022, 197, 113735. [Google Scholar] [CrossRef] [PubMed]
  222. Paimard, G.; Shahlaei, M.; Moradipour, P.; Karamali, V.; Arkan, E. Impedimetric Aptamer Based Determination of the Tumor Marker MUC1 by Using Electrospun Core-Shell Nanofibers. Microchim. Acta 2019, 187, 5. [Google Scholar] [CrossRef] [PubMed]
  223. Ahangarzadeh, S.; Kanafi, M.M.; Hosseinzadeh, S.; Mokhtarzadeh, A.; Barati, M.; Ranjbari, J.; Tayebi, L. Bicyclic Peptides: Types, Synthesis and Applications. Drug Discov. Today 2019, 24, 1311–1319. [Google Scholar] [CrossRef] [PubMed]
  224. Jaroszewicz, W.; Morcinek-Orłowska, J.; Pierzynowska, K.; Gaffke, L.; Węgrzyn, G. Phage Display and Other Peptide Display Technologies. FEMS Microbiol. Rev. 2021, 46, fuab052. [Google Scholar] [CrossRef] [PubMed]
  225. Liu, G.; Li, Y.; Liu, M.; Cheng, J.; Yang, S.; Gao, F.; Liu, L. Overview on Peptide-Based Electrochemical Biosensors. Int. J. Electrochem. Sci. 2023, 18, 100395. [Google Scholar] [CrossRef]
  226. Negahdary, M.; Angnes, L. Electrochemical Nanobiosensors Equipped with Peptides: A Review. Microchim. Acta 2022, 189, 94. [Google Scholar] [CrossRef] [PubMed]
  227. Zhao, J.-G.; Cao, J.; Wang, W.-Z. Peptide-Based Electrochemical Biosensors and Their Applications in Disease Detection. J. Anal. Test. 2022, 6, 193–203. [Google Scholar] [CrossRef]
  228. Zimina, T.; Sitkov, N.; Karasev, V.; Skorik, Y.; Kolobov, A.; Kolobov, A.; Bunenkov, N.; Luchinin, V. Design of Peptide Ligand for Lactoferrin and Study of Its Binding Specificity. Chemosensors 2023, 11, 162. [Google Scholar] [CrossRef]
  229. Soto, D.; Orozco, J. Peptide-Based Simple Detection of SARS-CoV-2 with Electrochemical Readout. Anal. Chim. Acta 2022, 1205, 339739. [Google Scholar] [CrossRef]
  230. Cho, C.H.; Kim, J.H.; Song, D.-K.; Park, T.J.; Park, J.P. An Affinity Peptide-Incorporated Electrochemical Biosensor for the Detection of Neutrophil Gelatinase-Associated Lipocalin. Biosens. Bioelectron. 2019, 142, 111482. [Google Scholar] [CrossRef]
  231. Matsubara, T.; Ujie, M.; Yamamoto, T.; Einaga, Y.; Daidoji, T.; Nakaya, T.; Sato, T. Avian Influenza Virus Detection by Optimized Peptide Termination on a Boron-Doped Diamond Electrode. ACS Sens. 2020, 5, 431–439. [Google Scholar] [CrossRef]
  232. Bucur, B.; Purcarea, C.; Andreescu, S.; Vasilescu, A. Addressing the Selectivity of Enzyme Biosensors: Solutions and Perspectives. Sensors 2021, 21, 3038. [Google Scholar] [CrossRef] [PubMed]
  233. Bollella, P.; Katz, E. Enzyme-Based Biosensors: Tackling Electron Transfer Issues. Sensors 2020, 20, 3517. [Google Scholar] [CrossRef]
  234. Nguyen, H.H.; Lee, S.H.; Lee, U.J.; Fermin, C.D.; Kim, M. Immobilized Enzymes in Biosensor Applications. Materials 2019, 12, 121. [Google Scholar] [CrossRef]
  235. Chan, D.; Barsan, M.M.; Korpan, Y.; Brett, C.M.A. L-Lactate Selective Impedimetric Bienzymatic Biosensor Based on Lactate Dehydrogenase and Pyruvate Oxidase. Electrochim. Acta 2017, 231, 209–215. [Google Scholar] [CrossRef]
  236. Hu, X.; Liao, M.; Gong, H.; Zhang, L.; Cox, H.; Waigh, T.A.; Lu, J.R. Recent Advances in Short Peptide Self-Assembly: From Rational Design to Novel Applications. Curr. Opin. Colloid Interface Sci. 2020, 45, 1–13. [Google Scholar] [CrossRef]
  237. Tse Sum Bui, B.; Mier, A.; Haupt, K. Molecularly Imprinted Polymers as Synthetic Antibodies for Protein Recognition: The Next Generation. Small 2023, 19, 2206453. [Google Scholar] [CrossRef]
  238. Cordell, P.; Carrington, G.; Curd, A.; Parker, F.; Tomlinson, D.; Peckham, M. Affimers and Nanobodies as Molecular Probes and Their Applications in Imaging. J. Cell Sci. 2022, 135, jcs259168. [Google Scholar] [CrossRef]
  239. Shamsuddin, S.H.; Jayne, D.G.; Tomlinson, D.C.; McPherson, M.J.; Millner, P.A. Selection and Characterisation of Affimers Specific for CEA Recognition. Sci. Rep. 2021, 11, 744. [Google Scholar] [CrossRef] [PubMed]
  240. Jolugbo, P.; Willot, T.; Hsiang, W.; Maisey, T.; O’Callaghan, D.W.; Green, M.; Jayne, D.M. Ibrahim Khot Fluorescent Imaging Using Novel Conjugated Polymeric Nanoparticles-Affimer Probes in Complex In Vitro Models of Colorectal Cancer. Nanoscale 2023, 15, 12476–12480. [Google Scholar] [CrossRef] [PubMed]
  241. Kamara, S.; Guo, Y.; Mao, S.; Ye, X.; Li, Q.; Zheng, M.; Zhu, J.; Zhang, J.; Du, W.; Chen, J.; et al. Novel EBV LMP1 C-Terminal Domain Binding Affibody Molecules as Potential Agents for In Vivo Molecular Imaging Diagnosis of Nasopharyngeal Carcinoma. Appl. Microbiol. Biotechnol. 2021, 105, 7283–7293. [Google Scholar] [CrossRef] [PubMed]
  242. Gad, S.; Ayakar, S. Protein Scaffolds: A Tool for Multi-Enzyme Assembly. Biotechnol. Rep. 2021, 32, e00670. [Google Scholar] [CrossRef] [PubMed]
  243. Friedman, M.; Nordberg, E.; Höidén-Guthenberg, I.; Brismar, H.; Adams, G.P.; Nilsson, F.Y.; Carlsson, J.; Ståhl, S. Phage Display Selection of Affibody Molecules with Specific Binding to the Extracellular Domain of the Epidermal Growth Factor Receptor. Protein Eng. Des. Sel. 2007, 20, 189–199. [Google Scholar] [CrossRef]
  244. Ravalli, A.; da Rocha, C.G.; Yamanaka, H.; Marrazza, G. A Label-Free Electrochemical Affisensor for Cancer Marker Detection: The Case of HER2. Bioelectrochemistry 2015, 106, 268–275. [Google Scholar] [CrossRef] [PubMed]
  245. Ahmad, M.I.; Amorim, C.G.; Fawzi, L.; Montenegro, M.C.B.S.M. Nanobody-Based Immunodiagnostics: A Systematic Review of Nanobody Integration in Diagnostics and Deep Insight into Electrochemical Immunoassays. Microchem. J. 2024, 196, 109628. [Google Scholar] [CrossRef]
  246. Salvador, J.-P.; Vilaplana, L.; Marco, M.-P. Nanobody: Outstanding Features for Diagnostic and Therapeutic Applications. Anal. Bioanal. Chem. 2019, 411, 1703–1713. [Google Scholar] [CrossRef] [PubMed]
  247. Sánchez-Salcedo, R.; Miranda-Castro, R.; de-Los-Santos-Álvarez, N.; Lobo-Castañón, M.J.; Corrigan, D.K. Comparing Nanobody and Aptamer-Based Capacitive Sensing for Detection of Interleukin-6 (IL-6) at Physiologically Relevant Levels. Anal. Bioanal. Chem. 2023, 415, 7035–7045. [Google Scholar] [CrossRef] [PubMed]
  248. Cruz-Pacheco, A.F.; Monsalve, Y.; Serrano-Rivero, Y.; Salazar-Uribe, J.; Moreno, E.; Orozco, J. Engineered Synthetic Nanobody-Based Biosensors for Electrochemical Detection of Epidermal Growth Factor Receptor. Chem. Eng. J. 2023, 465, 142941. [Google Scholar] [CrossRef]
  249. Paratore, F.; Bacheva, V.; Bercovici, M.; Kaigala, G.V. Reconfigurable Microfluidics. Nat. Rev. Chem. 2021, 6, 70–80. [Google Scholar] [CrossRef] [PubMed]
  250. Sharma, B.; Sharma, A. Microfluidics: Recent Advances Toward Lab-on-Chip Applications in Bioanalysis. Adv. Eng. Mater. 2021, 24, 2100738. [Google Scholar] [CrossRef]
  251. Hosseini, H.R.; Nikookar, H.; Yesiloz, G.; Naseh, M.; Mohammadi, M. An Overview on Micropumps, Micromixers, and Their Applications in Bioprocess. In Advances in Bioenergy and Microfluidic Applications; Elsevier: Amsterdam, The Netherlands, 2021; pp. 365–386. [Google Scholar] [CrossRef]
  252. Mohith, S.; Karanth, P.N.; Kulkarni, S.M. Recent Trends in Mechanical Micropumps and Their Applications: A Review. Mechatronics 2019, 60, 34–55. [Google Scholar] [CrossRef]
  253. Nigam, H.L.; Nambi Krishnan, J. Advances in Micropumps for Microfluidic Systems; Advances in Mechatronics and Mechanical Engineering (AMME) Book Series; IGI Global: Hershey, PA, USA, 2023; pp. 51–74. [Google Scholar] [CrossRef]
  254. Buttkewitz, M.A.; Heuer, C.P.; Bahnemann, J. Sensor Integration into Microfluidic Systems: Trends and Challenges. Curr. Opin. Biotechnol. 2023, 83, 102978. [Google Scholar] [CrossRef]
  255. Lai, X.; Yang, M.; Wu, H.; Li, D. Modular Microfluidics: Current Status and Future Prospects. Micromachines 2022, 13, 1363. [Google Scholar] [CrossRef]
  256. Song, K.; Li, G.; Zu, X.; Du, Z.; Liu, L.; Hu, Z. The Fabrication and Application Mechanism of Microfluidic Systems for High Throughput Biomedical Screening: A Review. Micromachines 2020, 11, 297. [Google Scholar] [CrossRef]
  257. Choi, S.; Goryll, M.; Sin, L.Y.M.; Wong, P.K.; Chae, J. Microfluidic-Based Biosensors toward Point-of-Care Detection of Nucleic Acids and Proteins. Microfluid. Nanofluidics 2010, 10, 231–247. [Google Scholar] [CrossRef]
  258. Abbas, N.; Song, S.; Chang, M.-S.; Chun, M.-S. Point-of-Care Diagnostic Devices for Detection of Escherichia coli O157:H7 Using Microfluidic Systems: A Focused Review. Biosensors 2023, 13, 741. [Google Scholar] [CrossRef]
  259. Tavakoli, H.; Mohammadi, S.; Li, X.; Fu, G.; Li, X. Microfluidic Platforms Integrated with Nano-Sensors for Point-of-Care Bioanalysis. TrAC Trends Anal. Chem. 2022, 157, 116806. [Google Scholar] [CrossRef]
  260. Dhanapala, L.; Krause, C.E.; Jones, A.L.; Rusling, J.F. Printed Electrodes in Microfluidic Arrays for Cancer Biomarker Protein Detection. Biosensors 2020, 10, 115. [Google Scholar] [CrossRef]
  261. Kim, H.J.; Choi, W.; Kim, J.; Choi, J.; Choi, N.; Hwang, K.S. Highly Sensitive Three-Dimensional Interdigitated Microelectrode Biosensors Embedded with Porosity Tunable Hydrogel for Detecting Proteins. Sens. Actuators B Chem. 2020, 302, 127190. [Google Scholar] [CrossRef]
  262. Garg, M.; Christensen, M.G.; Iles, A.; Sharma, A.L.; Singh, S.; Pamme, N. Microfluidic-Based Electrochemical Immunosensing of Ferritin. Biosensors 2020, 10, 91. [Google Scholar] [CrossRef] [PubMed]
  263. Devi, K.S.S.; Krishnan, U.M. Microfluidic Electrochemical Immunosensor for the Determination of Cystatin C in Human Serum. Microchim. Acta 2020, 187, 585. [Google Scholar] [CrossRef] [PubMed]
  264. Lakey, A.; Ali, Z.; Scott, S.M.; Chebil, S.; Korri-Youssoufi, H.; Hunor, S.; Ohlander, A.; Kuphal, M.; Marti, J.S. Impedimetric Array in Polymer Microfluidic Cartridge for Low Cost Point-of-Care Diagnostics. Biosens. Bioelectron. 2019, 129, 147–154. [Google Scholar] [CrossRef] [PubMed]
  265. Quoc, T.V.; Ngoc, V.N.; Bui, T.T.; Jen, C.-P.; Duc, T.C. High-Frequency Interdigitated Array Electrode-Based Capacitive Biosensor for Protein Detection. BioChip J. 2019, 13, 403–415. [Google Scholar] [CrossRef]
  266. Alsabbagh, K.; Hornung, T.; Voigt, A.; Sadir, S.; Rajabi, T.; Länge, K. Microfluidic Impedance Biosensor Chips Using Sensing Layers Based on DNA-Based Self-Assembled Monolayers for Label-Free Detection of Proteins. Biosensors 2021, 11, 80. [Google Scholar] [CrossRef]
  267. Wu, C.-C.; Singh, K.; Ye, J.-X.; Chuang, Y.-S.; Wen, H. A Microfluidic Chip Integrating Electrochemical Impedimetric Immunosensors Constructed by Top-Bottom Opposite Electrodes for Rapid Detection of Peanut Allergen-Ara H 1. Sens. Actuators B Chem. 2023, 396, 134637. [Google Scholar] [CrossRef]
  268. Bhardwaj, T.; Dalal, P.; Rathore, A.S.; Jha, S.K. An Aptamer Based Microfluidic Chip for Impedimetric Detection of Ranibizumab in a Bioreactor. Sens. Actuators B Chem. 2020, 312, 127941. [Google Scholar] [CrossRef]
  269. Torul, H.; Arslan, Z.Ç.; Tezcan, T.; Kayiş, E.Ç.; Çalımcı, M.; Gumustas, A.; Yildirim, E.; Külah, H.; Tamer, U. Microfluidic-Based Blood Immunoassays. J. Pharm. Biomed. Anal. 2023, 228, 115313. [Google Scholar] [CrossRef] [PubMed]
  270. Shahbazi, F.; Jabbari, M.; Esfahani, M.N.; Keshmiri, A. Microfluidic-Integrated Biosensors. In Applied Complex Flow; Emerging Trends in Mechatronics; Springer: Singapore, 2023; pp. 21–42. [Google Scholar] [CrossRef]
  271. Muhsin, S.A.; Abdullah, A.; Kobashigawa, E.; Al-Amidie, M.; Russell, S.; Zhang, M.Z.; Zhang, S.; Almasri, M. A Microfluidic Biosensor for the Diagnosis of Chronic Wasting Disease. Microsyst. Nanoeng. 2023, 9, 104. [Google Scholar] [CrossRef] [PubMed]
  272. Adam, H.; Gopinath, S.C.B.; Md Arshad, M.K.; Adam, T.; Hashim, U.; Sauli, Z.; Fakhri, M.A.; Subramaniam, S.; Chen, Y.; Sasidharan, S.; et al. Integration of Microfluidic Channel on Electrochemical-Based Nanobiosensors for Monoplex and Multiplex Analyses: An Overview. J. Taiwan Inst. Chem. Eng. 2023, 146, 104814. [Google Scholar] [CrossRef]
  273. Schmidt-Speicher, L.M.; Länge, K. Microfluidic Integration for Electrochemical Biosensor Applications. Curr. Opin. Electrochem. 2021, 29, 100755. [Google Scholar] [CrossRef]
  274. Zhang, H.; Miller, B.L. Immunosensor-Based Label-Free and Multiplex Detection of Influenza Viruses: State of the Art. Biosens. Bioelectron. 2019, 141, 111476. [Google Scholar] [CrossRef]
  275. Khodayari Bavil, A.; Sticker, D.; Rothbauer, M.; Ertl, P.; Kim, J. A Microfluidic Microparticle-Labeled Impedance Sensor Array for Enhancing Immunoassay Sensitivity. Analyst 2021, 146, 3289–3298. [Google Scholar] [CrossRef]
Figure 1. A general scheme for the development of a hybrid impedimetric biosensor for protein detection. 1—selecting the design and material of electrodes; 2—electrode modification and functionalization; 3—selection and immobilization of biorecognition element; 4—implementation of the interface with measurement circuit and sample analysis system.
Figure 1. A general scheme for the development of a hybrid impedimetric biosensor for protein detection. 1—selecting the design and material of electrodes; 2—electrode modification and functionalization; 3—selection and immobilization of biorecognition element; 4—implementation of the interface with measurement circuit and sample analysis system.
Micromachines 15 00181 g001
Figure 2. Typical Nyquist diagram (a) with corresponding equivalent electrical circuit (b).
Figure 2. Typical Nyquist diagram (a) with corresponding equivalent electrical circuit (b).
Micromachines 15 00181 g002
Figure 3. Schematic example of a Bode plot for impedance module, |Z|, and phase shift versus frequency, f.
Figure 3. Schematic example of a Bode plot for impedance module, |Z|, and phase shift versus frequency, f.
Micromachines 15 00181 g003
Figure 4. Schematic presentation of immunosensor preparation process for SPB detection. Reprinted from [68] with permission of Elsevier provided by Copyright Clearance Center.
Figure 4. Schematic presentation of immunosensor preparation process for SPB detection. Reprinted from [68] with permission of Elsevier provided by Copyright Clearance Center.
Micromachines 15 00181 g004
Figure 5. Scheme of an immunosensor for PAK 2 based on an ITO electrode. Reprinted from [76] with permission of Elsevier provided by Copyright Clearance Center.
Figure 5. Scheme of an immunosensor for PAK 2 based on an ITO electrode. Reprinted from [76] with permission of Elsevier provided by Copyright Clearance Center.
Micromachines 15 00181 g005
Figure 6. ST2 immunosensor immobilization scheme. Reprinted from [88] with permission of Elsevier provided by Copyright Clearance Center.
Figure 6. ST2 immunosensor immobilization scheme. Reprinted from [88] with permission of Elsevier provided by Copyright Clearance Center.
Micromachines 15 00181 g006
Figure 7. SEM micro-images of the bare B:CNW surface. Reprinted from [99] with permission of Elsevier provided by Copyright Clearance Center.
Figure 7. SEM micro-images of the bare B:CNW surface. Reprinted from [99] with permission of Elsevier provided by Copyright Clearance Center.
Micromachines 15 00181 g007
Figure 8. SEM image with low and high magnification (a,b), TEM (c), and high-resolution TEM (d) images of HsGDY@NDs. Reprinted from [100] with permission of Elsevier provided by Copyright Clearance Center.
Figure 8. SEM image with low and high magnification (a,b), TEM (c), and high-resolution TEM (d) images of HsGDY@NDs. Reprinted from [100] with permission of Elsevier provided by Copyright Clearance Center.
Micromachines 15 00181 g008
Figure 9. Schematic presentation of manufacturing steps of a biosensor with a Au electrode: (A)—Au electrode without functionalization; (B)—adsorption of cysteamine on the gold surface; (C) covalent binding of carboxybetaine monomer (CBMA) in the presence of EDC/NHS for activation; (D)—photopolymerization of the CBMA monomer in solution led to the formation of a zwitterionic polymer pCBMA with a CBMA-receptive interface for covalent binding of the IL-8 polyclonal antibody (Ab) in the presence of EDC/NHS (E) to capture the target analyte (F). Reprinted from [102] with permission of Elsevier provided by Copyright Clearance Center.
Figure 9. Schematic presentation of manufacturing steps of a biosensor with a Au electrode: (A)—Au electrode without functionalization; (B)—adsorption of cysteamine on the gold surface; (C) covalent binding of carboxybetaine monomer (CBMA) in the presence of EDC/NHS for activation; (D)—photopolymerization of the CBMA monomer in solution led to the formation of a zwitterionic polymer pCBMA with a CBMA-receptive interface for covalent binding of the IL-8 polyclonal antibody (Ab) in the presence of EDC/NHS (E) to capture the target analyte (F). Reprinted from [102] with permission of Elsevier provided by Copyright Clearance Center.
Micromachines 15 00181 g009
Figure 10. FE-SEM images of electropolymerized coatings of FTO/PPy (AC), FTO/PPy-MO NIP (DF), and FTO/PPy-MO DMIP electrodes (GI). The DMIP electrode was prepared by imprinting 100 ng/mL of each biomarker AFP and CEA. Reprinted from [113] with permission of Elsevier provided by Copyright Clearance Center.
Figure 10. FE-SEM images of electropolymerized coatings of FTO/PPy (AC), FTO/PPy-MO NIP (DF), and FTO/PPy-MO DMIP electrodes (GI). The DMIP electrode was prepared by imprinting 100 ng/mL of each biomarker AFP and CEA. Reprinted from [113] with permission of Elsevier provided by Copyright Clearance Center.
Micromachines 15 00181 g010
Figure 11. SEM images of electrode modifications AuE (A), AuE/4-ATP/fullerene (B), and AuE/4-ATP/fullerene/PAMAM (C). Reprinted from [131] with permission of Elsevier provided by Copyright Clearance Center.
Figure 11. SEM images of electrode modifications AuE (A), AuE/4-ATP/fullerene (B), and AuE/4-ATP/fullerene/PAMAM (C). Reprinted from [131] with permission of Elsevier provided by Copyright Clearance Center.
Micromachines 15 00181 g011
Figure 12. The original nickel foam and the resulting FeCo-MOF/NF: (a) SEM images of bare porous NF, (be) FeCo-MOF grown on the NF skeleton (FeCo-MOF/NF) after in situ hydrothermal reactions using precursors, ligands, and Ni foam skeleton at different magnifications. The marks in (f) are the average thickness of FeCo-MOF nanosheets grown on porous NF substrates. Reprinted from [134] with permission of Elsevier provided by Copyright Clearance Center.
Figure 12. The original nickel foam and the resulting FeCo-MOF/NF: (a) SEM images of bare porous NF, (be) FeCo-MOF grown on the NF skeleton (FeCo-MOF/NF) after in situ hydrothermal reactions using precursors, ligands, and Ni foam skeleton at different magnifications. The marks in (f) are the average thickness of FeCo-MOF nanosheets grown on porous NF substrates. Reprinted from [134] with permission of Elsevier provided by Copyright Clearance Center.
Micromachines 15 00181 g012
Figure 13. SEM images of top view (A,B) and cross-view (C,D) of TNT (A,C) and TNT annealed in argon (B,D) at 550 °C for 2 h. Reprinted from [141], license CC BY-NC-ND 4.0.
Figure 13. SEM images of top view (A,B) and cross-view (C,D) of TNT (A,C) and TNT annealed in argon (B,D) at 550 °C for 2 h. Reprinted from [141], license CC BY-NC-ND 4.0.
Micromachines 15 00181 g013
Figure 14. Non-faradaic impedimetric biosensor monitoring of caspase-9 in mammalian cell culture: (a) SEM image of a film of ZnO/CuO nanoparticles; (b) illustration of a cross-section of the biosensor surface and biorecognition interactions between anti-cas9 and caspase-9. Reprinted from [51], license CC BY-NC-ND 4.0.
Figure 14. Non-faradaic impedimetric biosensor monitoring of caspase-9 in mammalian cell culture: (a) SEM image of a film of ZnO/CuO nanoparticles; (b) illustration of a cross-section of the biosensor surface and biorecognition interactions between anti-cas9 and caspase-9. Reprinted from [51], license CC BY-NC-ND 4.0.
Micromachines 15 00181 g014
Figure 15. Schematic representation of the stages of immunosensor formation with QD for detection of EP-ZIKV. Reprinted from [153] with permission of Elsevier provided by Copyright Clearance Center.
Figure 15. Schematic representation of the stages of immunosensor formation with QD for detection of EP-ZIKV. Reprinted from [153] with permission of Elsevier provided by Copyright Clearance Center.
Micromachines 15 00181 g015
Figure 16. SEM electron microscopy images of MoS2 nanoflowers: (a) flower like morphology; (b) single bud with a more magnified petals. Reprinted from [154] with permission of Elsevier provided by Copyright Clearance Center.
Figure 16. SEM electron microscopy images of MoS2 nanoflowers: (a) flower like morphology; (b) single bud with a more magnified petals. Reprinted from [154] with permission of Elsevier provided by Copyright Clearance Center.
Micromachines 15 00181 g016
Figure 17. Modification stages of Au electrode using aryldiazonium (CB) carboxybetaine derivative and lectin (SNA-I). Reprinted from [157] with permission of Elsevier provided by Copyright Clearance Center.
Figure 17. Modification stages of Au electrode using aryldiazonium (CB) carboxybetaine derivative and lectin (SNA-I). Reprinted from [157] with permission of Elsevier provided by Copyright Clearance Center.
Micromachines 15 00181 g017
Figure 18. COVID-19 impedimetric biosensor based on polypyrrole nanotubes, nickel hydroxide, and VHH antibody fragment. Reprinted from [210] with permission of Elsevier provided by Copyright Clearance Center.
Figure 18. COVID-19 impedimetric biosensor based on polypyrrole nanotubes, nickel hydroxide, and VHH antibody fragment. Reprinted from [210] with permission of Elsevier provided by Copyright Clearance Center.
Micromachines 15 00181 g018
Figure 19. Preparation of L-cysteine-SnTeSe QD aptasensor. Reprinted from [152] with permission of Elsevier provided by Copyright Clearance Center.
Figure 19. Preparation of L-cysteine-SnTeSe QD aptasensor. Reprinted from [152] with permission of Elsevier provided by Copyright Clearance Center.
Micromachines 15 00181 g019
Figure 20. Analytical performance for the detection of NGAL: a commercially available ELISA NGAL detection kit (a) compared to sensor system from [230] (b). Statistical significance between groups was determined by ANOVA test, where p value is * p < 0.02 or ** p < 0.256 in (a), * p < 0.017 or ** p < 0.246 in (b), respectively. Mean values of each group are indicated with bold lines. Reprinted from [230] with permission of Elsevier provided by Copyright Clearance Center.
Figure 20. Analytical performance for the detection of NGAL: a commercially available ELISA NGAL detection kit (a) compared to sensor system from [230] (b). Statistical significance between groups was determined by ANOVA test, where p value is * p < 0.02 or ** p < 0.256 in (a), * p < 0.017 or ** p < 0.246 in (b), respectively. Mean values of each group are indicated with bold lines. Reprinted from [230] with permission of Elsevier provided by Copyright Clearance Center.
Micromachines 15 00181 g020
Figure 21. (a) Fabrication process of the (NGAL-BP1)-modified PDMS mold with hexagonally patterned pores for peptide surface imprinting via incubation in a peptide solution with a specific concentration. (b) Atomic force microscopy (AFM) image of the polystyrene (PS) colloidal monolayer and (c,d) scanning electron microscopy (SEM) images of oxidized PDMS molds with patterned concave pores after NGAL-BP1 adsorption during 2.5 h incubation in 9.25 and 18.51 μg/mL peptide solutions. All scale bars are 1 µm. Reprinted from [124] with permission of Elsevier provided by Copyright Clearance Center.
Figure 21. (a) Fabrication process of the (NGAL-BP1)-modified PDMS mold with hexagonally patterned pores for peptide surface imprinting via incubation in a peptide solution with a specific concentration. (b) Atomic force microscopy (AFM) image of the polystyrene (PS) colloidal monolayer and (c,d) scanning electron microscopy (SEM) images of oxidized PDMS molds with patterned concave pores after NGAL-BP1 adsorption during 2.5 h incubation in 9.25 and 18.51 μg/mL peptide solutions. All scale bars are 1 µm. Reprinted from [124] with permission of Elsevier provided by Copyright Clearance Center.
Micromachines 15 00181 g021
Figure 22. Molecularly imprinted polymer-based electrochemical impedimetric sensor for detection of trace cytokine IL-1β. Reprinted from [127] with permission of Elsevier provided by Copyright Clearance Center.
Figure 22. Molecularly imprinted polymer-based electrochemical impedimetric sensor for detection of trace cytokine IL-1β. Reprinted from [127] with permission of Elsevier provided by Copyright Clearance Center.
Micromachines 15 00181 g022
Figure 23. Schematic drawing of conceptual basis of IL-6-recognizing biosensor. Gold electrodes are modified with a nanobody or an aptamer specifically recognizing IL-6. There is a change in the non-faradaic impedance spectra when these surfaces are challenged in serum samples containing IL-6, which can be observed in both impedance (a) and capacitance planes (b). Reprinted from [247], license CC BY-NC-ND 4.0.
Figure 23. Schematic drawing of conceptual basis of IL-6-recognizing biosensor. Gold electrodes are modified with a nanobody or an aptamer specifically recognizing IL-6. There is a change in the non-faradaic impedance spectra when these surfaces are challenged in serum samples containing IL-6, which can be observed in both impedance (a) and capacitance planes (b). Reprinted from [247], license CC BY-NC-ND 4.0.
Micromachines 15 00181 g023
Figure 24. The schematic illustration of the immunosensor preparation based on gold nanostructure (AuNS) electrodeposition followed by 3-mercaptopropionic acid (MPA) modification, antibody immobilization, and BSA blocking (A). The design and assembly of a microfluidic device integrating top–bottom opposite electrodes (B). The device was connected to a syringe pump for controlling the flow rate and an electrochemical apparatus for EIS measurement (C). Reprinted from [267] with permission of Elsevier provided by Copyright Clearance Center.
Figure 24. The schematic illustration of the immunosensor preparation based on gold nanostructure (AuNS) electrodeposition followed by 3-mercaptopropionic acid (MPA) modification, antibody immobilization, and BSA blocking (A). The design and assembly of a microfluidic device integrating top–bottom opposite electrodes (B). The device was connected to a syringe pump for controlling the flow rate and an electrochemical apparatus for EIS measurement (C). Reprinted from [267] with permission of Elsevier provided by Copyright Clearance Center.
Micromachines 15 00181 g024
Figure 25. A 3D view of the biosensor CHD for diagnosis. Reprinted from [271], license CC BY-NC-ND 4.0.
Figure 25. A 3D view of the biosensor CHD for diagnosis. Reprinted from [271], license CC BY-NC-ND 4.0.
Micromachines 15 00181 g025
Table 1. Examples of impedimetric biosensors based on metal electrodes for protein detection.
Table 1. Examples of impedimetric biosensors based on metal electrodes for protein detection.
ElectrodesBiorecognition
Element
Target ProteinMatrixSensitivityRefs.
Au-IDWμEIgG-Fc fragmentIgM-RFHuman serum10–200 IU/mL range; LOD = 0.22 IU/mL[57]
Auanti-NS1 antibodyNS1Neat serum0.01–2.00 µg/mL range; LOD = 3 ng/mL[58]
Auanti-N antibodyN protein of SARS-CoV-2Human saliva4.4 ng/mL–4.4 pg/mL range; LOD = 0.362 ng/mL[59]
Auanti-human CRP antibodyC-reactive proteinHuman serum0.5–50 nM range; LOD = 176 pM[60]
Au-IDEs; Au-Colloidal Au NPsantigen tTGαtTG-AbsSerum30 pM–30 nM range[61]
AuIgG-antibodyIgGPBS buffer0.9–50 mg/L range[62]
Auanti-tau antibodyTauHuman serum10−14–10−7 M range; LOD = 0.03 pM [63]
AuaptamerN-terminal pro b-type natriuretic peptide Artificial human saliva5.0 × 10−3–1.0 pg/mL range; LOD = 5.0 × 10−3 pg/mL[64]
Auanti-A1BG antibodyA1BGHuman serum1–300 ng/mL range; LOF = 1 ng/mL[65]
AuDNA aptamersNS1Human serum0.025 ng/mL[66]
Auanti-A1AT antibodyA1ATArtificial serum100–600 μg/mL range; LOD = 11.9 μg/mL[67]
Auanti-SPB antibodySPBAmniotic fluid2–2000 ng/m L range; LOD = 0.1 ng/mL[68]
Au/Au NPsaptamers anti-PSA and PSAG-1PSAHuman serum0.26–62.5 ng/mL (PSAG-1), 0.64–62.5 ng/mL (anti-PSA) range[69]
Carbon; Au NPsaptamer Gli4-TPWG-GliadinBuffer0.1–1 mg/L of gliadin range; LOD = 0.05 mg/L[71]
Glassy carbon/Au NPsaptamerIL-6Human serum5 pg/mL–100 ng/mL range; LOD = 1.6 pg/mL[72]
Table 2. Examples of impedimetric biosensors based on conducting oxides for protein detection.
Table 2. Examples of impedimetric biosensors based on conducting oxides for protein detection.
ElectrodesBiorecognition
Element
Target ProteinMatrixSensitivityRefs.
ITOanti-IL-8 antibodyIL-8Human serum and saliva0.02–3 pg/mL range; LOD = 6 fg/mL[73]
FTOanti-IL-1α antibodyIL-1αHuman serum and saliva0.02–2 pg/mL range; LOD = 6 fg/mL[74]
ITOanti-MAGE-1 antibodyMAGE-1Human serum4–200 fg/mL; LOD = 1.30 fg/mL[75]
ITOanti-PAK 2 antibodyPAK 2Human serum0.005–0.075 pg/mL range; LOD = 1.5 fg/mL[76]
ITOanti-PAK2 antibodyPAK-2Artificial and real human serum0.05–2.5 pg/mL range; LOD = 0.0252 pg/mL[77]
ITOanti-Aβ42 antibody42Human serum1–100 pg/mL range; LOD = 0.37 pg/mL[78]
VASiNWs/ITOantibody (cTnIAb)cTnIHuman blood sera1.76011–17,601.1 ng/mL range; LOD = 0.53 ng/mL[79]
Table 3. Examples of impedimetric biosensors based on carbon materials for protein detection.
Table 3. Examples of impedimetric biosensors based on carbon materials for protein detection.
ElectrodesBiorecognition
Element
Target ProteinMatrixSensitivityRefs.
Carbon (SPCE)bicyclic peptidesh-uPABuffer0.01–1 ng/mL range; LOD = 9 ng/mL[80]
Carbon nanofibersaptamer-LYSLysozyme (Lys)Fetal bovine serumLOD = 0.36 μg/mL[81]
MWCNTsanti-Mb-IgG antibodyMbHuman serum0.1–90 ng/mL range; LOD = 0.08 ng/mL[82]
MWCNTsthrombin aptamerα-thrombin5 mM [Fe(CN)6]3−/4− in PBS0.39–1.95 nM range; LOD = 105 pM[83]
Carbon nanotubeanti-lysozyme DNA aptamerLYS2.5 mM [Fe(CN)6]3−/4− in KCl12–400 µg/mL range; LOD = 12.09 µg/mL[84]
MWCNTanti-IL-8 antibodyIL-8PBS1–1000 pg/mL range; LOD = 0.1 pg/mL[85]
MWCNT/AuNPsanti-DJ-1antibodyDJ-1Cerebrospinal fluid and saliva4.7–4700 fg/mL range; LOD = 0.5 fg/mL[86]
AuNPs@MWCNTs-GQDsanti-PSA antibodyPSA[Fe(CN)6]3−/4− in KCl1–10,000 pg/mL range; LOD = 0.48 pg/mL[87]
Glassy carbon; boron-doped diamondanti-N antibodyN protein of SARS-CoV-2Human saliva4.4 ng/mL–4.4 pg/mL range; LOD = 0.227 ng/mL (glassy carbon) and 0.334 ng/mL (boron-doped diamond)[59]
C60-modified graphite paperanti-ST2 antibodyST2Human serum0.1–100 fg/mL range; LOD = 0.124 fg/mL[88]
GPanti-adiponectin antibodyAdiponectinHuman serum0.05–25 pg/mL range; LOD = 0.0033 pg/mL[89]
GP/Au NPsanti-CK antibodyCreatine kinase (CK)Human blood0.1–50 pg/mL range; LOD = 0.045 pg/mL[90]
Boron-doped diamondanti-S1 antibodySubunit S1 of SARS-CoV-2Complex matrix1 fg/mL[91]
Gii-sense® Graphene Foamanti-IL-10 antibodyIL-10Artificial saliva10–100 fg/mL range; LOD = 7.89 fg/mL[92]
ErGOlectin concanavalin AGlycoprotein
invertase (INV)
PBS10−14–10−8 mol range[93]
rGO/AuNPsanti-Tau antibodyTau-441Serum and cerebrospinal fluid (CSF)1–500 pg/mL range; LOD = 0.091 pg/mL[94]
Porous graphene oxide (PrGO)anti-cTnI antibodycTnIHuman blood0.1–10 ng/mL range; LOD = 0.07 ng/mL [95]
rGOanti-CEA antibodyCEAHuman blood serum0.1–5 ng/mL range; LOD = 0.05 ng/mL[96]
AuNP electrode/GOanti -HSP-47 antibodyHSP-47Cell lysates10–160 pg/mL range; LOD = 9.47 pg/mL[97]
rGO-NPanti-CRP antibodyCRPPBS, Human serum1–1000 ng/mL range; LOD = 0.06 ng/mL (in PBS) and 0.08 ng/mL (in human serum)[98]
Maze-like boron-doped carbon nanowallanti-protein D antibodiesProtein DPBS3.37 × 10−11–3.37 × 10−3 µg/mL range; LOD = 2.39 × 102 fg/mL[99]
HsGDY@NDsaptamerscTnI; MyoPBSLOD = 6.29 fg/mL (cTnI); 9.04 fg/mL (Myo)[100]
Table 4. Examples of impedimetric biosensors based on polymeric materials for protein detection.
Table 4. Examples of impedimetric biosensors based on polymeric materials for protein detection.
ElectrodesBiorecognition ElementTarget ProteinMatrixSensitivityRefs.
Poly ortho-phenylenediaminemAβ antibodyAβ40PBS1 pg/mL[101]
Polycarboxybetaine methacrylate (pCBMA)anti-IL-8 antibodyIL-8Nasal epithelial lining fluid55 fM–55 nM range; LOD = 10 fM[102]
Polytyramineanti-FGFR3-14 and FGFR3-21 affimer proteinsFGFR3PBS and synthetic urinesub-pM–nM range[103]
Copolymer of polytyr-amine/3-(4-hydroxyphenyl) propionic acid (POPA)fragment of the cellular prion protein (PrPC, residues 95–110)AβOPBS10−12–10−6 M range; LOD~0.5 pM; (Aβ peptide)[104]
PolyoctopamineCEA affimer; anti-CEA antibodyCEASpiked human serum1–100 fM (affimer) and 1 fM–100 nM (antibody) range; LOD = 11.76 fM[105]
Polypyrrole polymerIL-6 receptorIL-6Human serum0.02–16 pg/mL range; LOD = 6.0 fg/mL[106]
Poly-pyrrole-pyrrole 3 carboxylic acidantibodies (Ab-PfHRP2)PfHRP2[Fe(CN)6]4−/3−+ KCl solution100–1000 ng/mL range; LOD = 27.47 ng/mL[107]
Carboxy-endcapped conductive polypyrroleanti-GPC3 antibodygypican-3 (GPC3)Human serum0.9 pg/mL to 9 ng/mL range; LOD = 0.3 pg/mL[108]
Acid-substituted poly(pyrrole)anti-CC chemokine receptor 4 (CCR4) antibodiesCCR4Human serum0.02–8 pg/mL range; LOD = 6.4 fg/mL[109]
AuNPs/thiophene polymer P(ThiAmn)anti-GM2A antibodiesGM2AHuman serum0.0185–111 pg/mL range; LOD = 5.8 fg/mL[110]
acetylene black (AB)/epoxy-substituted-poly(pyrrole) polymeranti-IL-6 antibodiesIL-6Spiked human serum0.01–50 pg/mL range; LOD = 3.2 fg/mL[111]
Overoxidized polypyrrole decorated with gold nanoparticleanti-HigG antibodyHigGHuman serum0.5–125.0 ng/mL range; LOD = 0.02 ng/mL[112]
Methyl orange-doped polypyrroleDMIPCEA; AFPHuman serum5–104 pg/mL (CEA), 10–104 pg/mL (AFP) range; LOD = 1.6 pg/mL (CEA), 1.6 pg/mL (AFP)[113]
Acetylene black/thiophene polymeranti-RBD antibodyspike receptor (RBD) COVID-19Nasal secretions0.0012–120 pg/mL range; LOD = 0.58 fg/mL[114]
Polyaniline (PANI)IFN-γ antibodyIFN-γHuman serum5–1000 pg/mL range; LOD = 3.4 pg/mL[115]
(PEDOT)/Au NP compositesanti-VEGF antibodyVEGF[Fe(CN)6]4−/3−+ KCl solution1–20 pg/mL range; LOD = 0.5 pg/mL[116]
Silk protein with PEDOT:PSSanti-VEGF165 antibodiesVEGFHuman serum1 pg/mL–1 μg/mL range; LOD = 1.03 pg/mL[117]
Carbon black C45/polythiophene polymeranti-CYFRA 21-1 antibodyCYFRA 21-1Human serum0.03–90 pg/mL range; LOD = 4.7 fg/mL[118]
Acrylamide (and Fe3O4 NP decorated with MWCNT-GO composite)MIP + anti-PSA antibodyPSA; MyoHuman serum and urine0.01–100 ng/mL (PSA), 1–20,000 ng/mL (Myo) range; LOD = 5.4 pg/mL (PSA), 0.83 ng/mL (Myo)[119]
Polyglutamic acid (PGA)anti-leptin antibodyleptin[Fe(CN)6]4−/3−+ PBS solution0.2–20 pg/mL range; LOD = 0.00813 pg/mL[120]
Biotinylated phenylalanine nanoparticleAsp-Glu-Val-Asp (DEVD)-containing peptidecaspase-3[Fe(CN)6]4−/3− solution26 pM–125 pg/mL range [121]
Gelatinanti-SARS-CoV-2 Spike glycoprotein S1 antibodySARS-CoV-2 spike glycoprotein S1PBS and nasopharyngeal fluid0.001–10 µg/mL range; LOD = 169 pg/mL (in PBS), 90 pg/mL (in nasopharyngeal fluid)[122]
3-Aminophenol polymerMIPTau proteinPBS2.18 pM–2.18 nM range; LOD = 0.02 pM[123]
Binding peptide (BP1)-imprinted filmMIPneutrophil gelatinase-associated lipocalinPBSLOD = 0.07 μg/mL[124]
PolydopamineMIPIL-6[Fe(CN)6]4−/3−+ PBS solution1–200 pg/mL range; LOD = 0.25 pg/mL[125]
NiO NP/ePDAMIPTrp[Fe(CN)6]3−/4− solution1–90 pg/mL range; LOD = 0.75 pg/mL[126]
Poly(chromotrope 2R) (C2R)MIPIL-1βPBS solutionLOD = 0.23 pg/mL[127]
Au NPs conjugated with DihexadecylphosphatemAb-DD antibodyD-dimer (DD)PBSLOD = 8.92 ng/mL[128]
PAMAM generation 2DNA aptamerAPC[Fe(CN)6]4−/3−+ KCl solution; fetal bovine serum 0.74–7.5 range; µg/mL LOD = 0.74 µg/mL (in buffer), 2.03 µg/mL (in serum)[129]
PAMAM generation 2DNA aptamerAPC[Fe(CN)6]4−/3−+ KCl solution; fetal bovine serum1–2.5 µg/mL; LOD = 1.81 µg/mL (in buffer), 0.02 µg/mL (in diluted fetal bovine serum)[130]
Fullerene–PAMAM(G5) compositeanti-Fetuin-A antibodyHFAReal blood samples1.66–134 ng/mL range; LOD = 0.48 ng/mL[131]
Table 5. Examples of impedimetric biosensors based on metal complexes of porphyrins, phthalocyanines, and metal–organic frameworks for protein detection.
Table 5. Examples of impedimetric biosensors based on metal complexes of porphyrins, phthalocyanines, and metal–organic frameworks for protein detection.
ElectrodesBiorecognition
Element
Target ProteinMatrixSensitivityRefs.
Cu-MOFAb2-CA15-3 antibody (and Ab1 on Au NPs)CA15-3[Fe(CN)6]4−/3−+ KCl solution 10 μU/mL–10 mU/mL and 10 mU/mL–100 U/mL ranges; LOD = 5.06 μU/mL for CA15-3[132]
Cu3(BTC)2 (copper-MOF)/PANI compositeanti-cTnI antibodiescTnI[Fe(CN)6]4−/3−+ PBS solution1–400 ng/mL range; LOD = 0.8 ng/mL[133]
FeCo-MOF/nickel foamIL-1RA antibodiesIL-1RAPBS and human serum10 fg/mL–10 ng/mL range; LOD = 7.30 fg/mL (in buffer), 7.22 fg/mL (in serum)[134]
Au@UiO-66-NH2DNA biomimetic clamp 2 nucleocapsid protein (SARS-CoV-2 antigen)Saliva and serumLOD = 0.31 pg/mL[135]
Zr(III) tetra(4-aminophenyl) porphyrin (TAPP) complexantibody of NSENSE[Fe(CN)6]4−/3−+ PBS solution10.0 fg/m–2.0 ng/mL range; LOD = 7.1 fg/mL[136]
Co phthalocyanines–Ce oxide NP conjugateHB5 aptamerHER2Human serum1–10 ng/mL range; LOD = 0.2 ng/mL[137]
Cobalt-based phthalocyanine/graphene quantum dotsHB5 aptamerHER2Human serumLOD = 0.0027 ng/mL[138]
Table 6. Examples of impedimetric biosensors based on metal oxides and dielectrics for protein detection.
Table 6. Examples of impedimetric biosensors based on metal oxides and dielectrics for protein detection.
ElectrodesBiorecognition ElementTarget ProteinMatrixSensitivityRefs.
Au NPs/TiO2 NPsDNA aptamerleptinHuman serum1.0–100.0 pg/mL and 100.0–1000.0 pg/mL ranges; LOD = 0.312 pg/mL[139]
Polymer/TiO2 NPsprotein AIgG[Fe(CN)6]4−/3−+ KCl solution0.0062–500 µg/mL range; LOD = 0.57 ng/mL[140]
Titanium dioxide nanotube (TNT) arrayanti-IL-6, anti-IL-8, and anti-TNFα ABIL-6, IL-8, TNFαPBS5–2500 pg/mL range; LOD = 5 pg/mL[141]
AuNPs/WO3/CNTsSARS-CoV-2 antibodies SARS-CoV-2-S protein[Fe(CN)6]4−/3− in buffer solution0.125–16.0 pg/mL ranges; LOD = 1.8 pg/mL[144]
Tungsten trioxide nanosheets (WO3 NS)anti-cTnI antibodiescTnI[Fe(CN)6]4−/3−+ PBS solution0.1–100 ng/mL range[145]
ZnO/CuO composite nanoparticleanti-cas9 antibodycaspase-9Buffer solution0.1–1 U/mL range; LOD = 0.07 U/mL[51]
Silicon nitrideanti-HSA antibodiesHSAPBS10−13–10−7 M range; LOD = 10−14 M[146]
Glassanti-HSA antibodiesHSAPBSLOD = 2 × 10−4 mg/mL[147]
SiO2anti-pLDH AntibodypLDHSalivaLOD = 250 pg/mL[148]
ZnO thin filmantibodiesPCT, CRPHuman serum and blood0.01–10 ng/mL (PCT), 0.01–20 µg/mL (CRP), ranges; LOD = 0.10 ng/mL (PCT); 0.10 µg/mL (CRP)[149]
ZnO–CuO compositeanti-CRP antibodies.CRPPBS1–10 ng/mL range[150]
Nano-ZnO/CuO membranesanti-CRP antibodiesCRPPBS0.1–15 ng/mL range; LOD = 16 pg/mL[151]
Table 7. Examples of impedimetric biosensors based on other materials for protein detection.
Table 7. Examples of impedimetric biosensors based on other materials for protein detection.
ElectrodesBiorecognition ElementTarget ProteinMatrixSensitivityRefs.
L-cysteine-SnTeSe QDinterferon-gamma aptamerinterferon-gammaSerum samples10–55 pg/mL range; LOD = 0.151 pg/mL[152]
Carboxylated CdTe QDsanti-EP DIII ZIKV antibodiesEP-ZIKV[Fe(CN)6]3−/4− + bufferLOD = 0.1 ng/mL[153]
MoS2 nfanti-TNF-α antibodiesTNF-αSerum samples1–200 pg/mL range; LOD = 0.202 pg/mL[154]
AuNPs/Ti3C2-MXenesaptamerscTnI, Myo[Fe(CN)6]3−/4− + buffer0.24 fg/mL–24 ng/mL (cTnI), 1–72 ng/mL (Myo) ranges; LOD = 0.14 fg/mL (cTnI), 0.2 ng/mL (Myo)[155]
Au-NPs decorated MoS2 nanosheetC-RP antibodyC-RPHuman serum1 fg/mL–1 μg/mL range; LOD = 0.01 fg/mL[156]
MP@silica@Au/CBanti-fPSA antibody (and lectin (SNA-I) for Au-electrode/CB)PSASpiked human serum0.01–1 pg/mL range; LOD = 1.2 fM[157]
Magnetic nanoparticlesanti-CRP polyclonal antibody (and anti-CRP monoclonal antibody on electrode)CRPPBS10–200 ng/mL range; LOD = 0.34 ng/mL[158]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Sitkov, N.; Ryabko, A.; Moshnikov, V.; Aleshin, A.; Kaplun, D.; Zimina, T. Hybrid Impedimetric Biosensors for Express Protein Markers Detection. Micromachines 2024, 15, 181. https://0-doi-org.brum.beds.ac.uk/10.3390/mi15020181

AMA Style

Sitkov N, Ryabko A, Moshnikov V, Aleshin A, Kaplun D, Zimina T. Hybrid Impedimetric Biosensors for Express Protein Markers Detection. Micromachines. 2024; 15(2):181. https://0-doi-org.brum.beds.ac.uk/10.3390/mi15020181

Chicago/Turabian Style

Sitkov, Nikita, Andrey Ryabko, Vyacheslav Moshnikov, Andrey Aleshin, Dmitry Kaplun, and Tatiana Zimina. 2024. "Hybrid Impedimetric Biosensors for Express Protein Markers Detection" Micromachines 15, no. 2: 181. https://0-doi-org.brum.beds.ac.uk/10.3390/mi15020181

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics