Review

Current Status of Needles in the Optimization of Endoscopic Ultrasound-Guided Procedures

Department of Gastroenterology, Saitama Medical University International Medical Center, 1397-1, Yamane, Hidaka, Saitama 350-1298, Japan
*
Author to whom correspondence should be addressed.
Submission received: 21 June 2020 / Revised: 5 July 2020 / Accepted: 7 July 2020 / Published: 8 July 2020
(This article belongs to the Special Issue The Advanced Role of Diagnostic Endoscopic Ultrasonography)

Abstract

:
Endoscopic ultrasound (EUS) is among the most important tools for the evaluation of gastrointestinal tumors and affected areas around the gastrointestinal tract. It enables the acquisition of material from abnormal lesions via the gastrointestinal wall for tissue confirmation via endoscopic ultrasound-guided fine-needle aspiration (EUS-FNA). EUS-FNA has played a vital role in oncological care and has become the standard method for tissue sampling. The choice of needle type is an important factor determining tissue acquisition and has been evaluated by many researchers. New needles are introduced into the market almost every year, and opinions vary regarding proper needle selection. While there are diverse opinions but no definitive recommendations about the use of one particular device, fine-needle biopsy needles may provide detailed information on a tissue’s architecture based on greater sample yields. This permits additional analyses, including genetic sequencing and phenotyping, thereby enabling the provision of more personalized treatment plans. Furthermore, other EUS-guided procedures have been developed, including interventional EUS and through-the-needle devices. Given the continued attempts to improve the diagnostic ability and therapeutic techniques, we review in detail the available types of puncture needles to provide guidance on the selection of the appropriate needle types.

1. Introduction

Endoscopic ultrasound (EUS) is one of the important tools for evaluation of gastrointestinal tumors and affected areas around the gastrointestinal tract. It has great advantages over CT and other imaging modalities because it enables assessment of the echo structure in < 1 cm diameter lesions. In addition, endoscopic ultrasound-guided fine-needle aspiration (EUS-FNA) provides the capability to obtain material from abnormal lesions via the gastrointestinal wall for tissue analysis [1].
This safe and widely used tissue sampling procedure provides consistent results under ultrasound guidance [2,3,4], and it is thought to be more effective compared to CT-guided or ultrasound-guided biopsy of lesions [5]. EUS-FNA has been mainly used for pancreatic solid lesions [6,7], abdominal or mediastinal lymph nodes [8,9,10], and gastrointestinal subepithelial lesions (SELs) (Figure 1) [3,11].
More recently, its range of indications has expanded so as to include liver lesions [12], adrenal grand lesions [13], and biliary strictures [14,15]. For pancreatic cystic lesions, indications vary significantly between countries due to the risk of dissemination [16,17]. EUS-FNA enables the acquisition of histological evidence of cancer when chemotherapy is being considered, to distinguish benignancy or malignancy when deciding if surgery or follow-up is needed, and assessment of the degree of progression of malignant tumors when unexplained lymph-node swelling is detected. The sensitivity and specificity of EUS-FNA for detecting solid pancreatic lesions have been estimated to lie between 85–89% and 96–99%, respectively [18,19,20]. Although there may be complications such as pancreatitis and bleeding, EUS-FNA has been very safe, with < 1% of morbidity and mortality rates [21].
Several factors affect the outcome of this technique, such as the exact location and characteristics of the lesion, the skill level of the endoscopist, the used sampling technique, availability of rapid on-site evaluation (ROSE), and needle gauge or type [22]. Selection of the needle is considered to be an essential factor, which affects the quality of tissue acquisition and has consequently been evaluated in several studies. A diversity of needles has been in use over the past 25 years, supported by a continuous stream of design innovations. New needle designs are introduced almost annually, and views on the proper needle selection are wide-ranging (Figure 2), partly because they are used not only for tissue sampling but also for various interventional procedures.
This review focuses on past and current advances in design and status of needles and other devices that are used in EUS-guided procedures.

2. The History of Needle Development

EUS has made great progress over the past decades. In the late 1970s and 1980s, it was developed by the Olympus Corporation as an improvement to ultrasound imaging techniques of the pancreaticobiliary system. The initial EUS prototype was a 180-degree mechanical radial scanning instrument [23]. The first commercial EUS instrument, introduced in 1982, was a mechanical–radial type. The US transducer was located on the tip of the instrument and could be rotated by a motor within the endoscope handle. A 360-degree image was obtained perpendicular to the insertion shaft of the endoscope, which allowed for convenient, real-time mapping of the anatomy [24]. In the early 1990s, Pentax Medical, in cooperation with Hitachi, developed the first commercially available linear-array echoendoscope [23]. Linear echoendoscopes altered the landscape of EUS due to their ability to track a needle in real time, and across the image plane, into a targeted lesion. The electronic instruments also permitted the use of Doppler technology to assess vascular flow.
Around that time, the innovation of needle designs also began. Vilmann et al., in collaboration with Medi-Globe GmbH, created a special biopsy device, which was a major step leading to the clinical application of the biopsy method [25,26]. In 1992, these researchers reported the first case of using EUS-FNA to examine a pancreatic head lesion [25]. In 1993, they reported a further attempt, in which they used a newly developed steel needle with a Teflon sheath to examine upper-gastrointestinal tract lesions [27]. In 1996, Vilmann and Hancke reported the development of a new biopsy handle instrument (type Hancke/Vilmann) [28]. One year later, Binmoeller et al. first described an automated biopsy device for pancreatic lesions that could not be punctured with conventional aspiration needles [29]. The automatic spring-loaded biopsy needle allowed tissue sampling also of indurated pancreatic lesions. However, this instrument never attained success.
Thereafter, the range of indications for EUS-FNA expanded, and new clinical applications for biopsies other than conventional diagnostics were envisioned, including genetic diagnostics and anticancer drug-sensitivity assays. Such applications inevitably required the collection of an adequate tissue volume to achieve optimal diagnostic accuracy [30,31]. Therefore, the number of options for fine-needle biopsies (FNB) has increased in recent needle developments. FNB needles were especially designed to obtain core specimens with preserved tissue architectures. In 2002, Wiersema et al. reported initial experiences with EUS-FNB of perigastric organs [32].
The initial core-biopsy EUS needle used was the 19-G Tru-Cut needle (Quick-Core; Cook Medical, Bloomington, IN, USA). Although in some cases, high diagnostic yields were obtained [33], the production of this needle was terminated due to its limited flexibility and occurring adverse events (AEs). The Tru-Cut was replaced by the ProCore FNB needles (Cook Medical) (Figure 3), which is characterized by a cutting bevel (reverse for 19-, 22- and 25-G, and 20-G antegrade beveled side slot) at the needle tip [34,35,36].
Recently, new FNB needles have been released in the market, including the fork-tip needle (SharkCore; Medtronic, Newton, Mass and Covidien, Dublin, Ireland), which is characterized by two sharp tips on the opposite side of the lumen [37], and the franseen-type needle (Acquire; Boston Scientific, Marlborough, MA, USA) (Figure 4), which is characterized by three symmetric cutting tips [38]. Currently, a variety of puncture needles that improve lesion accessibility and puncture performance are commercially available (Table 1).

3. Needle Size

At present, the most frequently used needle sizes are 22 G and 25 G. A 19-G needle ensures that a sufficiently large sample is collected for histologic diagnostics and immunostaining in cases that are examined for autoimmune pancreatitis and malignant lymphoma [10,39]. However, the greater puncture resistance encountered by these larger-sized needles increases the difficulty of the procedure. In particular, transduodenal puncture is thought to be difficult to perform [40]. Additionally, the larger the caliber of the needle, the higher the risk of post-puncture bleeding [41]. A smaller needle is technically more easy to handle within the device [42]. Therefore, 22-G or 25-G needles are more commonly used for EUS-FNA. Several studies have compared 22-G vs. 25-G needles (Table 2). In a review of a number of randomized clinical trials, one study showed a tendency towards higher accuracy of the 25-G needle [43], whereas the others showed a similar diagnostic yield of malignancy [44,45,46,47,48,49]. Studies comparing FNA with 22-G and 25-G needles were also reviewed as five meta-analyses [41,50,51,52,53]. However, more recent meta-analyses showed different results. Facciorusso et al. analyzed seven trials with 689 patients and 732 lesions (295 sampled with 22-G, 309 with 25-G, and 128 with needles of both sizes) [51]. A non-significant superiority of 25-G needles in pooled sensitivity (risk ratio: 0.93, 0.91–0.95, versus 0.89, 0.85–0.94 of 22-G needles; p = 0.13) and no difference in specificity (1.00, 0.98–1.00 in both groups; p = 0.85) were observed. Sample adequacy was similar between the two devices (risk ratio: 1.03, 0.99–1.06; p = 0.15). Very few AEs were observed and did not impact patient outcomes. This study showed no significant superiority of either group regarding pooled sensitivity, specificity, and safety.
On the other hand, Xu et al. analyzed eleven trials with 837 patients (412 sampled with 22- G vs. 425 with 25-G needles) for diagnosis of solid pancreatic lesions [52]. Their outcomes revealed that 25-G needles have a higher sensitivity than 22-G needles when used on solid pancreatic lesions EUS-FNA (92% (95% CI, 89–95%) versus 88% (95% CI, 84–91%); p = 0.046), although there were no significant differences between the two groups in overall diagnostic specificity (p = 0.842).
In 2018, Guedes et al. analyzed four randomized trials with 462 patients (229 sampled with 22-G vs. 233 with 25-G needles) for the diagnosis of solid pancreatic lesions. They found that the diagnostic sensitivity was 93% for 25-G and 91% for 22-G needles. In this study, the specificity of the 25-G needle was 87%, and that of the 22-G needle was 83%, with no statistically significant difference between the two needle types (p = 0.497). Based on randomized trials, this meta-analysis did not show a significant difference between the 22 G and 25 G needles [53]. As mentioned above, a range of opinions persist and no definitive recommendations can be made in favor of using one particular device or the other, as there appears to be no strong superiority of one needle over the other.

4. EUS-FNB

Several studies about diagnostic ability of FNB needles have been published. Early published results on the performance of ProCore FNB needles demonstrated high diagnostic accuracy [35,54,55]. Larghi et al. reported the first prospective, multicenter study in 61 patients in whom only a single needle pass with a 22-G FNB needle was performed [54]. The procedure was successfully performed in all but one patient, despite the fact that in 57% of cases, the puncture was carried out transduodenally. The rates of sensitivity, specificity, positive predictive value, negative predictive value, and accuracy for the histologic diagnosis of a pancreatic mass were 87.5, 100, 100, 41.7, and 88.5%, respectively. Two studies have compared the 20-G antegrade beveled biopsy needle with the 22-G and 25-G reverse beveled needles. According to these studies, the 20-G antegrade beveled biopsy needle is better than the reverse beveled needles when retrieving samples for histological analysis [56,57]. Moreover, a recent meta-analysis confirmed the superiority of the 20-G antegrade beveled biopsy needle compared with reverse beveled needles [58].
A large and early retrospective multicenter study of EUS-FNB obtained by fork-tip FNB needle on different solid lesions (pancreatic, SEL, and lymph-node) demonstrated an excellent 88% overall pathologic diagnostic yield based on a median of only two passes [59]. Overall, histological diagnostic ability and thus pathologic yield for the different lesion subtypes were as follows: pancreatic lesions 86%, SELs 87%, lymph nodes 93%. Needle size did not affect pathologic diagnostic yield, with both 25-G and 22-G needles delivering high yields of 86% and 89%, respectively. The 25-G fork-tip FNB needle demonstrated high diagnostic accuracy (93%) in a recent multicenter prospective study [60].
A retrospective multicenter study of 200 patients undergoing EUS-FNB of solid lesions (pancreas, SEL, lymph node, and other) using franseen needles also showed a high rate of tissue adequacy and tissue core [61]. The tissue obtained by these needles was adequate for evaluation and diagnosis by ROSE in 98.5% of cases. In 131/145 (90%) of cases, a core of tissue was obtained.
Further studies comparing outcomes based on fork-tip versus franseen needles could determine which needle design is superior in terms of diagnostic yield and safety. For pancreatic lesions, one comparative was reported [62]. Sampling was performed twice in 50 patients using the two needle types in randomized order. The results showed comparable diagnostic yields (96% vs. 92%, p = 0.32) and diagnostic adequacy with ROSE (94% vs. 96%, p = 0.32) of the franseen and fork-tip needles, respectively. In addition, no statistically significant differences were found in terms of tissue quality or quantity, total tissue obtained, total tumor tissue, and the desmoplastic fibrosis yield obtained by the two needles. Another randomized controlled trial also demonstrated no significant difference in the diagnostic accuracy, 64 out of 75 (85.3%) versus 68 out of 75 (90.7%), (p = 0.45) between franseen and fork-tip needles [63]. Recently, a meta-analysis comparing the two needles for EUS FNB of solid mass lesions was published [64]. The analysis featured a total of 21 studies with 1632 patients. The pooled diagnostic yield with fork-tip needle was 92.8% (95% CI 85.3–96.6, I2 = 73.1), whereas the pooled diagnostic yield using the franseen needle was 92.7% (95% CI 86.4–96.2, I2 = 88.4), demonstrating no statistical difference between the needles (p = 0.98). The overall rate of AEs for both needles combined was 4.2% (95% CI 2.8–6.4, I2 = 0.0); for the fork-tip and franseen needles, the rates were 3.7% (95% CI 2.3–6.0, I2 = 0), and 6.2% (95% CI 2.6–14.1, I2 = 0.0), respectively (p = 0.31, not significant). Of note, the AE rates of EUS-FNB were slightly higher than what would be anticipated, given that EUS-FNA is generally considered to be very safe. It is theoretically possible that the novel design of these FNB needle tips may allow for more effective capture of the tissue prior to shearing it off, but that this may induce more localized trauma, resulting in more frequent AEs.
Thus far, no consensus has been reached regarding optimally effective and safe needle types. Therefore, a need exists to conduct larger, multicenter, randomized trials to evaluate subtle advantages and disadvantages among new FNB needles.
ROSE on EUS-FNB is performed using the touch imprint cytology. The acquired material from the EUS-FNB was placed on a smear and examined for the presence of a solid component. The worm-like core tissue was then separated from the bloody material and was transferred to a clean slide and gently pushed and rubbed down with another slide. The remaining solid tissue from the first pass (if available) was placed in formalin for histological evaluation [65]. A large international randomized trial investigating the usefulness of ROSE coupled with EUS-FNB vs. EUS-FNB alone is ongoing [66].

5. EUS-FNB vs. EUS-FNA

Several randomized controlled trials have compared diagnostic yields of FNA and FNB. Alatawi et al. performed a randomized controlled trial to compare the outcomes of the 22 G Procore FNB-versus the 22-G FNA needle in patients with solid pancreatic lesions [67]. They included 100 patients who were equally allocated to FNA and FNB groups. Sampling outcomes showed a significantly lower adequate sample rate in the FNA than in the FNB group (90% vs. 100%, p = 0.02). Additionally, this study showed that the use of Procore FNB needles can attain a high diagnostic performance after two needle passes, and that additional passes do not significantly improve the diagnostic yield. FNA needles may achieve the same efficacy for the cytological diagnosis of pancreatic cancer, but this requires a higher number of needle passes. The researchers concluded that the samples obtained with 22-G Procore FNB needles were of significantly higher histological quality and required fewer needle passes for diagnosis. Another randomized controlled trial was conducted by Kim et al. [68]. Their study featured a comparison of the 22-G Procore FNB- and the 22-G FNA needles in patients with SELs. Compared to the EUS-FNA group, the EUS-FNB group required a significantly lower number of needle passes to obtain macroscopically optimal core samples (median: 4 vs. 2, p = 0.025). Moreover, compared to EUS-FNA needles, EUS-FNB needles produced higher yields of macroscopically and histologically optimal core samples after three needle passes (30% vs. 92%, p = 0.006, and 20% vs. 75%, p = 0.010 respectively). Finally, a higher diagnostic accuracy was obtained with EUS-FNB than with EUS-FNA needles (75% vs. 20%, respectively; p = 0.010). The authors concluded that EUS-FNB has a superior ability to obtain histological core samples and also a higher diagnostic accuracy than EUS-FNA for SELs.
A recent systematic review and meta-analysis compared the diagnostic yields of FNA and FNB in patients with lesions of the pancreas, lymph nodes, or solid GI lesions, while specifically evaluating the diagnostic value of ROSE [69]. Fifteen studies (n = 1024) were included in the analysis. No significant difference was observed in diagnostic adequacy (defined as the ability to procure cytological and/or histological samples adequate for interpretation) (Relative risk (RR): 0.98, CI: 0.91-1.06, I2 = 51%). In the absence of ROSE, FNB showed a trend toward better diagnostic adequacy, which, however, fell short of statistical significance (p = 0.06). Only in patients with solid pancreatic lesions, no difference in diagnostic adequacy was observed (RR: 0.96, CI: 0.86–1.09, I2 = 66%), but, in the absence of ROSE, FNB was associated with a higher diagnostic adequacy (p = 0.02). No significant differences were found in diagnostic accuracy (RR: 0.99, CI: 0.95–1.03, I2 = 27%) and procurement of optimal-quality core histological samples (RR: 0.97, CI: 0.89–1.05, I2 = 9.6%). However, FNB enabled the diagnosis with fewer passes (Standardized mean: 0.93, CI: 0.45–1.42, I2 = 84%) (Table 3).
Given the above results, FNB without ROSE could replace FNA with ROSE without loss of diagnostic accuracy. So, in medical centers where ROSE has not been applicable, FNB may still be an effective option.

6. Interventional EUS

EUS is clinically useful not only as a diagnostic tool but also during interventions. Many interventional EUS procedures have replaced conventional therapeutic approaches in the treatment of various gastrointestinal diseases. In fact, this procedure has emerged as the treatment of choice for management of pancreatic fluid collections (PFC) and drainage of mediastinal and intra-abdominal abscesses. Similarly, EUS-guided biliary drainage (EUS-BD) is applied in patients with failed cannulations and inaccessible papillae, as well as in those with duodenal/gastric obstructions or surgically altered anatomies. In addition, EUS-guided gall bladder drainage may gain prominence as a useful procedure in patients with acute cholecystitis who are unfit for surgery [70]. While interventional EUS has clinical benefits in these patients and demonstrates high technical and clinical success rates in high-volume centers, its use is associated with various AEs, including stent migration. If stent migration does occur, it is occasionally fatal [71]. In this sense, interventional EUS can be a high-risk procedure, and only endoscopists skilled in both ERCP and EUS should be permitted to perform it. Clearly, there is a practical need for the establishment of techniques and devices that are more feasible and safer.
A 19 G or 22 G FNA needle or needle knife is used to puncture the target when interventional EUS is performed. To enable passage of 0.035- or 0.025-inch guidewires through a needle, a 19-gauge needle should be selected. The SonoTip Pro Control 19 G needle (Medi-Globe GmbH, Rosenheim, Germany) has been reported as a suitable needle to perform interventional EUS (Figure 5). Since the needle is highly flexible, it is easy to adjust the insertion angle of the needle. In addition, the cutting surface of this FNA needle is 5 mm long and is considered to be extremely sharp, so it is easy to operate the guidewire without kinking. Therefore, the needle is considered to be appropriate for use in interventional EUS [72]. However, when attempting to puncture an intrahepatic bile duct at a distant position, the highly flexible needle may not move straight and may deviate from its target. In such cases, it may be necessary to select another needle.
Interventional EUS continues to evolve as new devices dedicated to this technique are being developed. One of the greatest recent impacts in this field has been the emergence of a new stent delivery system integrated with an electrocautery needle. In 2014, Teoh et al. reported the testing of a new device dedicated to interventional EUS, which creates a fistula by electrocautery followed by placement of a fully covered metal stent in one procedure (Hot-Axios; Boston Scientific, Marlborough, MA, USA) (Figure 6) [73]. This device is a solution for endoscopists to tackle different complex tasks, i.e., to safely dilate the puncture tract, and to steadily and precisely deploy a stent in many EUS-guided interventions. The device has also prompted further refinement of procedures and the introduction of new techniques [74]. Since its introduction into the market, many reports on interventional EUS have been published based on this device. Its most common application is drainage for PFC, including the treatment of walled-off necrosis. Walter et al. reported the results of a prospective cohort study of 61 cases with PFC in which this lumen-apposing metal stent was used. The results of this study confirmed the stent’s feasibility and good clinical outcomes with low AE rates [75]. Several retrospective studies with larger numbers of cases have already shown favorable technical and clinical outcomes. Nevertheless, further evaluation is still needed of the relative efficacy and the cost–benefit ratio of this technique through comparisons with established techniques such as double pig-tail stent deployment. Future improvements might include the development of a dedicated device for EUS-BD, similar to those such as Hot-Axios that is optimized to achieve puncture and stent placement in one step, to prevent migration.

7. The Use of Through-the-Needle Imaging

Confocal laser endomicroscopy (CLE) is a novel imaging method that allows microscopy of the gastrointestinal mucosal epithelium during endoscopy, enabling real-time optical biopsies [76]. Needle-based confocal laser endomicroscopy (nCLE) is a promising endoscopic diagnostic technique that can be delivered through a 19-G needle into lesions and provides real-time, in vivo, 500–1000x magnified microscopic tissue imaging during EUS [77]. Fluorescein is usually administered immediately before imaging, and optimum images are obtained between 30 s to 8 min after injection but can be interpreted for as long as 60 min [78]. Recent studies have revealed the diagnostic value of EUS-guided nCLE for various lesions.
According to several studies, a classification of nCLE patterns could be established, facilitating the diagnosis of pancreatic cystic lesions [79]. A recent prospective multicenter study of 206 patients with pancreatic cystic lesions showed thay sensitivities and specificities of nCLE for the diagnosis of serous cystadenomas, mucinous cystic neoplasms, and premalignant cystic lesions were all 0.95 or higher; the area under the receiver operating characteristic curve was significantly larger for nCLE than for cyst fluid carcinoembryonic antigen or EUS [80]. A number of studies also evaluated the feasibility and safety of nCLE for the assessment of solid organs (i.e., solid pancreatic lesions, lymph nodes, and SELs) [81,82,83]. Although in vivo real-time imaging using nCLE is highly concordant with pathology and may replace ROSE, it is important to keep in mind that EUS-FNA alone has a sensitivity around 90% for malignancy. Therefore, it is questionable whether the adjunct of nCLE can improve the diagnostic value for solid mass any further.
Although nCLE may be used routinely for diagnosing pancreatic cysts in the future, there are several limitations to this technology. During real-time nCLE diagnosis, blinding the endoscopist to EUS features of the tumor was impossible, which inevitably affected the diagnosis of real-time nCLE [77]. Additionally, heterogenicity in endoscopist’s experience, inter-observer variability, and reproducibility of nCLE image should be taken into account [84]. To justify routine use of nCLE, advanced training and validation are needed.
Another through-the-needle imaging technique is EUS-guided through-the-needle biopsy (EUS–TTNB). The instrument has a sheath diameter of 0.8 mm and a jaw-opening width of 4.3 mm. It can be introduced through a 19-G FNA needle, after puncturing a pancreatic cystic lesion. Consequently, it enables the collection of histological specimens from the pancreatic cyst wall. Interobserver agreement between expert pathologists in the evaluation of TTNB samples from pancreatic cystic lesions has demonstrated by Larghi et al. [85]. It was close to perfection for all evaluated parameters, except definitive diagnosis. When mucinous cystic lesions were compared versus all other diagnoses, the agreement became substantial, thus indicating that TTNB specimens can provide important information for pancreatic cystic lesions. Recently, a multicenter retrospective study of this intervention was published, which featured 28 patients. The technical and clinical success rates of this technique were found to be 85.7% and 71.4%, respectively. Technical failure seems to be caused by loss of flexibility of the echoendoscope when forceps are inserted. Even though no severe or fatal AEs were observed, a rate of 10.7% is notable and should be interpreted with caution because of the limited number of patients. One patient was hospitalized due to nonspecific abdominal pain and was discharged the following day after the symptoms resolved. Two patients suffered acute pancreatitis in mild forms and hospitalized for 4 and 6 days [86]. Larger retrospective studies have demonstrated a high diagnostic yield. Crinò et al. evaluated 61 patients, and the diagnostic reliability of TTNB sampling compared with surgical histology was 90% [87]. Barresi et al. evaluated 56 patients, and overall diagnostic yield by combining cytological and histological samples was 47/56 (83.9%, 95% CI, 72–92%) [88]. Yang et al. performed a prospective study of 114 patients and compared EUS-FNA with TTNB. In their study, samples collected by EUS-FNA were analyzed by cytology, and samples collected by TTNB were analyzed by histology. Tissue acquisition yield was significantly higher with TTNB (95 of 114, 83.3%) than FNA (43 of 114, 37.7%) (p < 0.001). Findings from samples collected by TTNB were 100% concordant with findings from surgical specimens (14 of 14), whereas only 3 of 14 findings from analysis of samples collected by FNA were in agreement with findings from surgical specimens (21.4%) (p < 0.001) [89]. Moreover, three meta-analyses comparing EUS-FNA with TTNB were published. The pooled diagnostic accuracy rate, sensitivity, and specificity of TTNB were 78.8%, 82.2%, and 96.8%, respectively. The AEs rates were 6.1–8.6% [90,91,92].
The possibility of obtaining histological specimens from the cyst wall when using this novel microbiopsy forceps could lead to drastic changes in the management algorithm of pancreatic cysts if it can be performed safely. If biopsy specimens extracted through this procedure contain sufficient material for supplementary immunohistochemical (IHC) analysis, this will enable specialists to reach more accurate histological diagnoses. In cases of intraductal papillary mucinous neoplasm (IPMN), histological subtyping will become a possibility. When compared to gastric and oncocytic subtypes, pancreatobiliary and intestinal subtypes are associated with progression to high-grade dysplasia and invasive carcinoma [93]. Previous surgical series have shown that the clinical behavior of invasive carcinomas derived from pancreatobiliary type IPMN has a significantly poorer prognosis than carcinomas of the intestinal subtype [94,95]. In addition, procurement of cellular material from the cysts enables next-generation sequencing (NGS) and testing for possible gene mutations. When combined with microbiopsy and IHC, NGS could potentially increase diagnostic accuracy and lead to a personalized approach when deciding on a treatment strategy. Prospective multicenter studies are needed to determine the feasibility of its more widespread use in clinical practice.

8. Conclusions

This article has summarized the historical advances in the design and performance of needles used for EUS-guided procedures. Since the early 1990s, EUS-FNA has played a clear and vital role in oncological care and has become the standard procedure for sampling tissues with high diagnostic accuracy. Although there are diverse opinions and no definitive recommendations regarding the preferred use of one particular device, FNB may provide further information on tissue architecture and provides a greater sample yield, which allows for further analyses, such as genetic sequencing and phenotyping. This again may enable more personalized treatment strategies.
Moreover, other EUS-guided procedures such as interventional EUS and through-the-needle devices have been developed. As mentioned above, there are continuing attempts to improve the diagnostic ability and therapeutic technique, which will benefit from a better understanding and the selection of the most appropriate types of puncture needles.

Author Contributions

A.F. and S.R. designed the report; M.M., Y.T., T.O., M.S., T.N., and H.K. contributed to analysis and interpretation of previous reports and assisted in the preparation of the manuscript; A.F. and S.R. organized the report; and A.F. wrote the paper. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Irisawa, A.; Yamao, K. Curved linear array EUS technique in the pancreas and biliary tree: Focusing on the stations. Gastrointest. Endosc. 2009, 69, S84–S89. [Google Scholar] [CrossRef] [PubMed]
  2. Haba, S.; Yamao, K.; Bhatia, V.; Mizuno, N.; Hara, K.; Hijioka, S.; Imaoka, H.; Niwa, Y.; Tajika, M.; Kondo, S.; et al. Diagnostic ability and factors affecting accuracy of endoscopic ultrasound-guided fine needle aspiration for pancreatic solid lesions: Japanese large single center experience. J. Gastroenterol. 2013, 48, 973–981. [Google Scholar] [CrossRef] [PubMed]
  3. Mekky, M.A.; Yamao, K.; Sawaki, A.; Mizuno, N.; Hara, K.; Nafeh, M.A.; Osman, A.M.; Koshikawa, T.; Yatabe, Y.; Bhatia, V. Diagnostic utility of EUS-guided FNA in patients with gastric submucosal tumors. Gastrointest. Endosc. 2010, 71, 913–919. [Google Scholar] [CrossRef] [PubMed]
  4. Yasuda, I.; Tsurumi, H.; Omar, S.; Iwashita, T.; Kojima, Y.; Yamada, T.; Sawada, M.; Takami, T.; Moriwaki, H.; Soehendra, N. Endoscopic ultrasound-guided fine-needle aspiration biopsy for lymphadenopathy of unknown origin. Endoscopy 2006, 38, 919–924. [Google Scholar] [CrossRef]
  5. Horwhat, J.D.; Paulson, E.K.; McGrath, K.; Branch, M.S.; Baillie, J.; Tyler, D.; Pappas, T.; Enns, R.; Robuck, G.; Stiffler, H.; et al. A randomized comparison of EUS-guided FNA versus CT or US-guided FNA for the evaluation of pancreatic mass lesions. Gastrointest. Endosc. 2006, 63, 966–975. [Google Scholar] [CrossRef] [PubMed]
  6. Khalid, A.; Nodit, L.; Zahid, M.; Bauer, K.; Brody, D.; Finkelstein, S.D.; McGrath, K.M. Endoscopic ultrasound fine needle aspirate DNA analysis to differentiate malignant and benign pancreatic masses. Am. J. Gastroenterol. 2006, 101, 2493–2500. [Google Scholar] [CrossRef]
  7. Eloubeidi, M.A.; Chen, V.K.; Eltoum, I.A.; Jhala, D.; Chhieng, D.C.; Jhala, N.; Vickers, S.M.; Wilcox, C.M. Endoscopic ultrasound-guided fine needle aspiration biopsy of patients with suspected pancreatic cancer: Diagnostic accuracy and acute and 30-day complications. Am. J. Gastroenterol. 2003, 98, 2663–2668. [Google Scholar] [CrossRef]
  8. Catalano, M.F.; Rosenblatt, M.L.; Chak, A.; Sivak, M.V., Jr.; Scheiman, J.; Gress, F. Endoscopic ultrasound-guided fine needle aspiration in the diagnosis of mediastinal masses of unknown origin. Am. J. Gastroenterol. 2002, 97, 2559–2565. [Google Scholar] [CrossRef]
  9. Tournoy, K.G.; Praet, M.M.; Van Maele, G.; Van Meerbeeck, J.P. Esophageal endoscopic ultrasound with fine-needle aspiration with an on-site cytopathologist: High accuracy for the diagnosis of mediastinal lymphadenopathy. Chest 2005, 128, 3004–3009. [Google Scholar] [CrossRef]
  10. Yasuda, I.; Goto, N.; Tsurumi, H.; Nakashima, M.; Doi, S.; Iwashita, T.; Kanemura, N.; Kasahara, S.; Adachi, S.; Hara, T.; et al. Endoscopic ultrasound-guided fine needle aspiration biopsy for diagnosis of lymphoproliferative disorders: Feasibility of immunohistological, flow cytometric, and cytogenetic assessments. Am. J. Gastroenterol. 2012, 107, 397–404. [Google Scholar] [CrossRef]
  11. Philipper, M.; Hollerbach, S.; Gabbert, H.E.; Heikaus, S.; Bocking, A.; Pomjanski, N.; Neuhaus, H.; Frieling, T.; Schumacher, B. Prospective comparison of endoscopic ultrasound-guided fine-needle aspiration and surgical histology in upper gastrointestinal submucosal tumors. Endoscopy 2010, 42, 300–305. [Google Scholar] [CrossRef] [PubMed]
  12. Singh, P.; Mukhopadhyay, P.; Bhatt, B.; Patel, T.; Kiss, A.; Gupta, R.; Bhat, S.; Erickson, R.A. Endoscopic ultrasound versus CT scan for detection of the metastases to the liver: Results of a prospective comparative study. J. Clin. Gastroenterol. 2009, 43, 367–373. [Google Scholar] [CrossRef] [PubMed]
  13. DeWitt, J.; Alsatie, M.; LeBlanc, J.; McHenry, L.; Sherman, S. Endoscopic ultrasound-guided fine-needle aspiration of left adrenal gland masses. Endoscopy 2007, 39, 65–71. [Google Scholar] [CrossRef] [PubMed]
  14. Lee, J.H.; Salem, R.; Aslanian, H.; Chacho, M.; Topazian, M. Endoscopic ultrasound and fine-needle aspiration of unexplained bile duct strictures. Am. J. Gastroenterol. 2004, 99, 1069–1073. [Google Scholar] [CrossRef]
  15. Eloubeidi, M.A.; Chen, V.K.; Jhala, N.C.; Eltoum, I.E.; Jhala, D.; Chhieng, D.C.; Syed, S.A.; Vickers, S.M.; Mel Wilcox, C. Endoscopic ultrasound-guided fine needle aspiration biopsy of suspected cholangiocarcinoma. Clin. Gastroenterol. Hepatol. 2004, 2, 209–213. [Google Scholar] [CrossRef]
  16. Tanaka, M.; Fernandez-Del Castillo, C.; Kamisawa, T.; Jang, J.Y.; Levy, P.; Ohtsuka, T.; Salvia, R.; Shimizu, Y.; Tada, M.; Wolfgang, C.L. Revisions of international consensus Fukuoka guidelines for the management of IPMN of the pancreas. Pancreatology 2017, 17, 738–753. [Google Scholar] [CrossRef]
  17. Hirooka, Y.; Goto, H.; Itoh, A.; Hashimoto, S.; Niwa, K.; Ishikawa, H.; Okada, N.; Itoh, T.; Kawashima, H. Case of intraductal papillary mucinous tumor in which endosonography-guided fine-needle aspiration biopsy caused dissemination. J. Gastroenterol. Hepatol. 2003, 18, 1323–1324. [Google Scholar] [CrossRef]
  18. Hebert-Magee, S.; Bae, S.; Varadarajulu, S.; Ramesh, J.; Frost, A.R.; Eloubeidi, M.A.; Eltoum, I.A. The presence of a cytopathologist increases the diagnostic accuracy of endoscopic ultrasound-guided fine needle aspiration cytology for pancreatic adenocarcinoma: A meta-analysis. Cytopathology 2013, 24, 159–171. [Google Scholar] [CrossRef] [Green Version]
  19. Hewitt, M.J.; McPhail, M.J.; Possamai, L.; Dhar, A.; Vlavianos, P.; Monahan, K.J. EUS-guided FNA for diagnosis of solid pancreatic neoplasms: A meta-analysis. Gastrointest. Endosc. 2012, 75, 319–331. [Google Scholar] [CrossRef]
  20. Puli, S.R.; Bechtold, M.L.; Buxbaum, J.L.; Eloubeidi, M.A. How good is endoscopic ultrasound-guided fine-needle aspiration in diagnosing the correct etiology for a solid pancreatic mass? A meta-analysis and systematic review. Pancreas 2013, 42, 20–26. [Google Scholar] [CrossRef]
  21. Wang, K.X.; Ben, Q.W.; Jin, Z.D.; Du, Y.Q.; Zou, D.W.; Liao, Z.; Li, Z.S. Assessment of morbidity and mortality associated with EUS-guided FNA: A systematic review. Gastrointest. Endosc. 2011, 73, 283–290. [Google Scholar] [CrossRef] [PubMed]
  22. Wani, S.; Muthusamy, V.R.; Komanduri, S. EUS-guided tissue acquisition: An evidence-based approach (with videos). Gastrointest. Endosc. 2014, 80, 939–959. [Google Scholar] [CrossRef] [PubMed]
  23. Gress, F.G. The Early History of Interventional Endoscopic Ultrasound. Gastrointest. Endosc. Clin. N. Am. 2017, 27, 547–550. [Google Scholar] [CrossRef] [PubMed]
  24. Strohm, W.D.; Phillip, J.; Hagenmuller, F.; Classen, M. Ultrasonic tomography by means of an ultrasonic fiberendoscope. Endoscopy 1980, 12, 241–244. [Google Scholar] [CrossRef] [PubMed]
  25. Vilmann, P.; Jacobsen, G.K.; Henriksen, F.W.; Hancke, S. Endoscopic ultrasonography with guided fine needle aspiration biopsy in pancreatic disease. Gastrointest. Endosc. 1992, 38, 172–173. [Google Scholar] [CrossRef]
  26. Vilmann, P.; Saftoiu, A. Endoscopic ultrasound-guided fine needle aspiration biopsy: Equipment and technique. J. Gastroenterol. Hepatol. 2006, 21, 1646–1655. [Google Scholar] [CrossRef]
  27. Vilmann, P.; Hancke, S.; Henriksen, F.W.; Jacobsen, G.K. Endosonographically-guided fine needle aspiration biopsy of malignant lesions in the upper gastrointestinal tract. Endoscopy 1993, 25, 523–527. [Google Scholar] [CrossRef]
  28. Vilmann, P.; Hancke, S. A new biopsy handle instrument for endoscopic ultrasound-guided fine-needle aspiration biopsy. Gastrointest. Endosc. 1996, 43, 238–242. [Google Scholar] [CrossRef]
  29. Binmoeller, K.F.; Jabusch, H.C.; Seifert, H.; Soehendra, N. Endosonography-guided fine-needle biopsy of indurated pancreatic lesions using an automated biopsy device. Endoscopy 1997, 29, 384–388. [Google Scholar] [CrossRef]
  30. Fujita, A.; Ryozawa, S.; Kobayashi, M.; Araki, R.; Nagata, K.; Minami, K.; Tanisaka, Y.; Kobatake, T.; Mizuide, M. Diagnostic ability of a 22G Franseen needle in endoscopic ultrasound-guided fine needle aspiration of subepithelial lesions. Mol. Clin. Oncol. 2018, 9, 527–531. [Google Scholar] [CrossRef]
  31. Fujita, A.; Ryozawa, S.; Mizuide, M.; Araki, R.; Nagata, K.; Tanisaka, Y.; Harada, M.; Ogawa, T.; Tashima, T.; Nonaka, K. Does endoscopic ultrasound-guided fine needle biopsy using a Franseen needle really offer high diagnostic accuracy? A propensity-matched analysis. Endosc. Int. Open 2019, 7, e1327–e1332. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  32. Wiersema, M.J.; Levy, M.J.; Harewood, G.C.; Vazquez-Sequeiros, E.; Jondal, M.L.; Wiersema, L.M. Initial experience with EUS-guided trucut needle biopsies of perigastric organs. Gastrointest. Endosc. 2002, 56, 275–278. [Google Scholar] [CrossRef]
  33. Gleeson, F.C.; Clayton, A.C.; Zhang, L.; Clain, J.E.; Gores, G.J.; Rajan, E.; Smyrk, T.C.; Topazian, M.D.; Wang, K.K.; Wiersema, M.J.; et al. Adequacy of endoscopic ultrasound core needle biopsy specimen of nonmalignant hepatic parenchymal disease. Clin. Gastroenterol. Hepatol. 2008, 6, 1437–1440. [Google Scholar] [CrossRef] [PubMed]
  34. Fabbri, C.; Fornelli, A.; Fuccio, L.; Giovanelli, S.; Tarantino, I.; Antonini, F.; Liotta, R.; Frazzoni, L.; Gusella, P.; La Marca, M.; et al. High diagnostic adequacy and accuracy of the new 20G procore needle for EUS-guided tissue acquisition: Results of a large multicentre retrospective study. Endosc. Ultrasound 2019, 8, 261–268. [Google Scholar] [CrossRef]
  35. Iglesias-Garcia, J.; Poley, J.W.; Larghi, A.; Giovannini, M.; Petrone, M.C.; Abdulkader, I.; Monges, G.; Costamagna, G.; Arcidiacono, P.; Biermann, K.; et al. Feasibility and yield of a new EUS histology needle: Results from a multicenter, pooled, cohort study. Gastrointest. Endosc. 2011, 73, 1189–1196. [Google Scholar] [CrossRef]
  36. DeWitt, J.; Cho, C.M.; Lin, J.; Al-Haddad, M.; Canto, M.I.; Salamone, A.; Hruban, R.H.; Messallam, A.A.; Khashab, M.A. Comparison of EUS-guided tissue acquisition using two different 19-gauge core biopsy needles: A multicenter, prospective, randomized, and blinded study. Endosc. Int. Open 2015, 3, e471–e478. [Google Scholar] [CrossRef] [Green Version]
  37. Adler, D.G.; Witt, B.; Chadwick, B.; Wells, J.; Taylor, L.J.; Dimaio, C.; Schmidt, R. Pathologic evaluation of a new endoscopic ultrasound needle designed to obtain core tissue samples: A pilot study. Endosc. Ultrasound 2016, 5, 178–183. [Google Scholar] [CrossRef] [Green Version]
  38. Bang, J.Y.; Hebert-Magee, S.; Hasan, M.K.; Navaneethan, U.; Hawes, R.; Varadarajulu, S. Endoscopic ultrasonography-guided biopsy using a Franseen needle design: Initial assessment. Dig. Endosc. Off. J. Jpn. Gastroenterol. Endosc. Soc. 2017, 29, 338–346. [Google Scholar] [CrossRef]
  39. Iwashita, T.; Yasuda, I.; Doi, S.; Ando, N.; Nakashima, M.; Adachi, S.; Hirose, Y.; Mukai, T.; Iwata, K.; Tomita, E.; et al. Use of samples from endoscopic ultrasound-guided 19-gauge fine-needle aspiration in diagnosis of autoimmune pancreatitis. Clin. Gastroenterol. Hepatol. 2012, 10, 316–322. [Google Scholar] [CrossRef]
  40. Vilmann, P.; Seicean, A.; Saftoiu, A. Tips to overcome technical challenges in EUS-guided tissue acquisition. Gastrointest. Endosc. Clin. N. Am. 2014, 24, 109–124. [Google Scholar] [CrossRef]
  41. Affolter, K.E.; Schmidt, R.L.; Matynia, A.P.; Adler, D.G.; Factor, R.E. Needle size has only a limited effect on outcomes in EUS-guided fine needle aspiration: A systematic review and meta-analysis. Dig. Dis. Sci. 2013, 58, 1026–1034. [Google Scholar] [CrossRef] [PubMed]
  42. Tanisaka, Y.; Ryozawa, S.; Kobayashi, M.; Harada, M.; Kobatake, T.; Omiya, K.; Iwano, H.; Arai, S.; Nonaka, K.; Mashimo, Y. Usefulness of endoscopic ultrasound-guided fine needle aspiration for lymphadenopathy. Oncol. Lett. 2018, 15, 4759–4766. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  43. Carrara, S.; Anderloni, A.; Jovani, M.; Di Tommaso, L.; Rahal, D.; Hassan, C.; Ridola, L.; Federico, D.; Loriga, A.; Repici, A. A prospective randomized study comparing 25-G and 22-G needles of a new platform for endoscopic ultrasound-guided fine needle aspiration of solid masses. Dig. Liver Dis. 2016, 48, 49–54. [Google Scholar] [CrossRef] [PubMed]
  44. Lee, J.H.; Stewart, J.; Ross, W.A.; Anandasabapathy, S.; Xiao, L.; Staerkel, G. Blinded prospective comparison of the performance of 22-gauge and 25-gauge needles in endoscopic ultrasound-guided fine needle aspiration of the pancreas and peri-pancreatic lesions. Dig. Dis. Sci. 2009, 54, 2274–2281. [Google Scholar] [CrossRef]
  45. Siddiqui, U.D.; Rossi, F.; Rosenthal, L.S.; Padda, M.S.; Murali-Dharan, V.; Aslanian, H.R. EUS-guided FNA of solid pancreatic masses: A prospective, randomized trial comparing 22-gauge and 25-gauge needles. Gastrointest. Endosc. 2009, 70, 1093–1097. [Google Scholar] [CrossRef]
  46. Camellini, L.; Carlinfante, G.; Azzolini, F.; Iori, V.; Cavina, M.; Sereni, G.; Decembrino, F.; Gallo, C.; Tamagnini, I.; Valli, R.; et al. A randomized clinical trial comparing 22G and 25G needles in endoscopic ultrasound-guided fine-needle aspiration of solid lesions. Endoscopy 2011, 43, 709–715. [Google Scholar] [CrossRef]
  47. Fabbri, C.; Polifemo, A.M.; Luigiano, C.; Cennamo, V.; Baccarini, P.; Collina, G.; Fornelli, A.; Macchia, S.; Zanini, N.; Jovine, E.; et al. Endoscopic ultrasound-guided fine needle aspiration with 22- and 25-gauge needles in solid pancreatic masses: A prospective comparative study with randomisation of needle sequence. Dig. Liver Dis. 2011, 43, 647–652. [Google Scholar] [CrossRef]
  48. Lee, J.K.; Lee, K.T.; Choi, E.R.; Jang, T.H.; Jang, K.T.; Lee, J.K.; Lee, K.H. A prospective, randomized trial comparing 25-gauge and 22-gauge needles for endoscopic ultrasound-guided fine needle aspiration of pancreatic masses. Scand. J. Gastroenterol. 2013, 48, 752–757. [Google Scholar] [CrossRef]
  49. Vilmann, P.; Saftoiu, A.; Hollerbach, S.; Skov, B.G.; Linnemann, D.; Popescu, C.F.; Wellmann, A.; Gorunescu, F.; Clementsen, P.; Freund, U.; et al. Multicenter randomized controlled trial comparing the performance of 22 gauge versus 25 gauge EUS-FNA needles in solid masses. Scand. J. Gastroenterol. 2013, 48, 877–883. [Google Scholar] [CrossRef]
  50. Madhoun, M.F.; Wani, S.B.; Rastogi, A.; Early, D.; Gaddam, S.; Tierney, W.M.; Maple, J.T. The diagnostic accuracy of 22-gauge and 25-gauge needles in endoscopic ultrasound-guided fine needle aspiration of solid pancreatic lesions: A meta-analysis. Endoscopy 2013, 45, 86–92. [Google Scholar] [CrossRef]
  51. Facciorusso, A.; Stasi, E.; Di Maso, M.; Serviddio, G.; Ali Hussein, M.S.; Muscatiello, N. Endoscopic ultrasound-guided fine needle aspiration of pancreatic lesions with 22 versus 25 Gauge needles: A meta-analysis. United Eur. Gastroenterol. J. 2017, 5, 846–853. [Google Scholar] [CrossRef] [Green Version]
  52. Xu, M.M.; Jia, H.Y.; Yan, L.L.; Li, S.S.; Zheng, Y. Comparison of two different size needles in endoscopic ultrasound-guided fine-needle aspiration for diagnosing solid pancreatic lesions: A meta-analysis of prospective controlled trials. Medicine (Baltim.) 2017, 96, e5802. [Google Scholar] [CrossRef] [PubMed]
  53. Guedes, H.G.; Moura, D.T.H.; Duarte, R.B.; Cordero, M.A.C.; Santos, M.; Cheng, S.; Matuguma, S.E.; Chaves, D.M.; Bernardo, W.M.; Moura, E.G.H. A comparison of the efficiency of 22G versus 25G needles in EUS-FNA for solid pancreatic mass assessment: A systematic review and meta-analysis. Clinics (Sao Paulo, Braz.) 2018, 73, e261. [Google Scholar] [CrossRef] [PubMed]
  54. Larghi, A.; Iglesias-Garcia, J.; Poley, J.W.; Monges, G.; Petrone, M.C.; Rindi, G.; Abdulkader, I.; Arcidiacono, P.G.; Costamagna, G.; Biermann, K.; et al. Feasibility and yield of a novel 22-gauge histology EUS needle in patients with pancreatic masses: A multicenter prospective cohort study. Surg. Endosc. 2013, 27, 3733–3738. [Google Scholar] [CrossRef] [PubMed]
  55. Iwashita, T.; Nakai, Y.; Samarasena, J.B.; Park, D.H.; Zhang, Z.; Gu, M.; Lee, J.G.; Chang, K.J. High single-pass diagnostic yield of a new 25-gauge core biopsy needle for EUS-guided FNA biopsy in solid pancreatic lesions. Gastrointest. Endosc. 2013, 77, 909–915. [Google Scholar] [CrossRef] [PubMed]
  56. Armellini, E.; Manfrin, E.; Trisolini, E.; Andorno, S.; Ballarè, M.; Bernardoni, L.; Boldorini, R.L.; Gabbrielli, A.; Frulloni, L.; Larghi, A.; et al. Histologic retrieval rate of a newly designed side-bevelled 20G needle for EUS-guided tissue acquisition of solid pancreatic lesions. United Eur. Gastroenterol. J. 2019, 7, 96–104. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  57. Cho, E.; Park, C.H.; Kim, T.H.; Cho, C.M.; Seo, D.W.; Kim, J.; Choi, J.H.; Moon, S.H. A prospective, randomized, multicenter clinical trial comparing 25-gauge and 20-gauge biopsy needles for endoscopic ultrasound-guided sampling of solid pancreatic lesions. Surg. Endosc. 2020, 34, 1310–1317. [Google Scholar] [CrossRef]
  58. Van Riet, P.; Erler, N.; Bruno, M.; Cahen, D. Comparison of fine needle aspiration and fine needle biopsy devices for endoscopic ultrasound-guided sampling of solid lesions: A systemic review and meta-analysis. Endoscopy 2020. [Google Scholar] [CrossRef]
  59. DiMaio, C.J.; Kolb, J.M.; Benias, P.C.; Shah, H.; Shah, S.; Haluszka, O.; Maranki, J.; Sharzehi, K.; Lam, E.; Gordon, S.R.; et al. Initial experience with a novel EUS-guided core biopsy needle (SharkCore): Results of a large North American multicenter study. Endosc. Int. Open 2016, 4, E974–E979. [Google Scholar] [CrossRef] [Green Version]
  60. Di Leo, M.; Crinò, S.F.; Bernardoni, L.; Rahal, D.; Auriemma, F.; Correale, L.; Donato, G.; Massidda, M.; Anderloni, A.; Manfrin, E.; et al. EUS-guided core biopsies of pancreatic solid masses using a new fork-tip needle: A multicenter prospective study. Dig. Liver Dis. 2019, 51, 1275–1280. [Google Scholar] [CrossRef]
  61. Adler, D.G.; Muthusamy, V.R.; Ehrlich, D.S.; Parasher, G.; Thosani, N.C.; Chen, A.; Buscaglia, J.M.; Appannagari, A.; Quintero, E.; Aslanian, H.; et al. A multicenter evaluation of a new EUS core biopsy needle: Experience in 200 patients. Endosc. Ultrasound 2019, 8, 99–104. [Google Scholar] [CrossRef] [Green Version]
  62. Bang, J.Y.; Hebert-Magee, S.; Navaneethan, U.; Hasan, M.K.; Hawes, R.; Varadarajulu, S. Randomized trial comparing the Franseen and Fork-tip needles for EUS-guided fine-needle biopsy sampling of solid pancreatic mass lesions. Gastrointest. Endosc. 2018, 87, 1432–1438. [Google Scholar] [CrossRef] [PubMed]
  63. Ashat, M.; Klair, J.S.; Rooney, S.L.; Vishal, S.J.; Jensen, C.; Murali, A.R.; Sahar, N.; El-Abiad, R.; Gerke, H. Randomized controlled trial comparing the Franseen needle with a Fork-tip needle for EUS-guided fine-needle biopsy. Gastrointest. Endosc. 2020. [Google Scholar] [CrossRef] [PubMed]
  64. Mohan, B.P.; Shakhatreh, M.; Garg, R.; Asokkumar, R.; Jayaraj, M.; Ponnada, S.; Navaneethan, U.; Adler, D.G. Comparison of Franseen and fork-tip needles for EUS-guided fine-needle biopsy of solid mass lesions: A systematic review and meta-analysis. Endosc. Ultrasound 2019, 8, 382–391. [Google Scholar] [CrossRef] [PubMed]
  65. Crinò, S.F.; Larghi, A.; Bernardoni, L.; Parisi, A.; Frulloni, L.; Gabbrielli, A.; Parcesepe, P.; Scarpa, A.; Manfrin, E. Touch imprint cytology on endoscopic ultrasound fine-needle biopsy provides comparable sample quality and diagnostic yield to standard endoscopic ultrasound fine-needle aspiration specimens in the evaluation of solid pancreatic lesions. Cytopathology 2019, 30, 179–186. [Google Scholar] [CrossRef] [PubMed]
  66. Crinò, S.F.; Manfrin, E.; Scarpa, A.; Baldaque-Silva, F.; Carrara, S.; De Nucci, G.; Di Mitri, R.; Ginés, A.; Iglesias-Garcia, J.; Itoi, T.; et al. EUS-FNB with or without on-site evaluation for the diagnosis of solid pancreatic lesions (FROSENOR): Protocol for a multicenter randomized non-inferiority trial. Dig. Liver Dis. 2019, 51, 901–906. [Google Scholar] [CrossRef]
  67. Alatawi, A.; Beuvon, F.; Grabar, S.; Leblanc, S.; Chaussade, S.; Terris, B.; Barret, M.; Prat, F. Comparison of 22G reverse-beveled versus standard needle for endoscopic ultrasound-guided sampling of solid pancreatic lesions. United Eur. Gastroenterol. J. 2015, 3, 343–352. [Google Scholar] [CrossRef]
  68. Kim, G.H.; Cho, Y.K.; Kim, E.Y.; Kim, H.K.; Cho, J.W.; Lee, T.H.; Moon, J.S. Comparison of 22-gauge aspiration needle with 22-gauge biopsy needle in endoscopic ultrasonography-guided subepithelial tumor sampling. Scand. J. Gastroenterol. 2014, 49, 347–354. [Google Scholar] [CrossRef]
  69. Khan, M.A.; Grimm, I.S.; Ali, B.; Nollan, R.; Tombazzi, C.; Ismail, M.K.; Baron, T.H. A meta-analysis of endoscopic ultrasound-fine-needle aspiration compared to endoscopic ultrasound-fine-needle biopsy: Diagnostic yield and the value of onsite cytopathological assessment. Endosc. Int. Open 2017, 5, E363–E375. [Google Scholar] [CrossRef] [Green Version]
  70. Sharma, V.; Rana, S.S.; Bhasin, D.K. Endoscopic ultrasound guided interventional procedures. World J. Gastrointest. Endosc. 2015, 7, 628–642. [Google Scholar] [CrossRef] [Green Version]
  71. Martins, F.P.; Rossini, L.G.; Ferrari, A.P. Migration of a covered metallic stent following endoscopic ultrasound-guided hepaticogastrostomy: Fatal complication. Endoscopy 2010, 42 (Suppl. 2), E126–E127. [Google Scholar] [CrossRef] [Green Version]
  72. Ogura, T.; Higuchi, K. Technical tips of endoscopic ultrasound-guided choledochoduodenostomy. World J. Gastroenterol. 2015, 21, 820–828. [Google Scholar] [CrossRef]
  73. Teoh, A.Y.; Binmoeller, K.F.; Lau, J.Y. Single-step EUS-guided puncture and delivery of a lumen-apposing stent for gallbladder drainage using a novel cautery-tipped stent delivery system. Gastrointest. Endosc. 2014, 80, 1171. [Google Scholar] [CrossRef] [PubMed]
  74. Ryozawa, S.; Fujita, N.; Irisawa, A.; Hirooka, Y.; Mine, T. Current status of interventional endoscopic ultrasound. Dig. Endosc. Offi. J. Jpn. Gastroenterol. Endosc. Soc. 2017, 29, 559–566. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  75. Walter, D.; Will, U.; Sanchez-Yague, A.; Brenke, D.; Hampe, J.; Wollny, H.; López-Jamar, J.M.; Jechart, G.; Vilmann, P.; Gornals, J.B.; et al. A novel lumen-apposing metal stent for endoscopic ultrasound-guided drainage of pancreatic fluid collections: A prospective cohort study. Endoscopy 2015, 47, 63–67. [Google Scholar] [CrossRef] [Green Version]
  76. Kiesslich, R.; Burg, J.; Vieth, M.; Gnaendiger, J.; Enders, M.; Delaney, P.; Polglase, A.; McLaren, W.; Janell, D.; Thomas, S.; et al. Confocal laser endoscopy for diagnosing intraepithelial neoplasias and colorectal cancer in vivo. Gastroenterology 2004, 127, 706–713. [Google Scholar] [CrossRef] [PubMed]
  77. Zhang, M.M.; Zhong, N.; Gu, X.; Wang, X.; Zuo, X.L.; Ji, R.; Li, C.Q.; Li, L.X.; Li, Z.; Yu, Y.B.; et al. In vivo real-time diagnosis of endoscopic ultrasound-guided needle-based confocal laser endomicroscopy in gastric subepithelial lesions. J. Gastroenterol. Hepatol. 2020, 35, 446–452. [Google Scholar] [CrossRef] [PubMed]
  78. Becker, V.; von Delius, S.; Bajbouj, M.; Karagianni, A.; Schmid, R.M.; Meining, A. Intravenous application of fluorescein for confocal laser scanning microscopy: Evaluation of contrast dynamics and image quality with increasing injection-to-imaging time. Gastrointest. Endosc. 2008, 68, 319–323. [Google Scholar] [CrossRef] [PubMed]
  79. Krishna, S.G.; Brugge, W.R.; Dewitt, J.M.; Kongkam, P.; Napoleon, B.; Robles-Medranda, C.; Tan, D.; El-Dika, S.; McCarthy, S.; Walker, J.; et al. Needle-based confocal laser endomicroscopy for the diagnosis of pancreatic cystic lesions: An international external interobserver and intraobserver study (with videos). Gastrointest. Endosc. 2017, 86, 644–654.e642. [Google Scholar] [CrossRef] [PubMed]
  80. Napoleon, B.; Palazzo, M.; Lemaistre, A.I.; Caillol, F.; Palazzo, L.; Aubert, A.; Buscail, L.; Maire, F.; Morellon, B.M.; Pujol, B.; et al. Needle-based confocal laser endomicroscopy of pancreatic cystic lesions: A prospective multicenter validation study in patients with definite diagnosis. Endoscopy 2019, 51, 825–835. [Google Scholar] [CrossRef]
  81. Karstensen, J.G.; Cartana, T.; Constantinescu, C.; Dumitrascu, S.; Kovacevic, B.; Klausen, P.; Hassan, H.; Klausen, T.W.; Bertani, H.; Bhutani, M.S.; et al. Endoscopic ultrasound guided needle-based confocal laser endomicroscopy in solid pancreatic masses—A prospective validation study. Endosc. Int. Open 2018, 6, E78–E85. [Google Scholar] [CrossRef] [Green Version]
  82. Benias, P.C.; D’Souza, L.S.; Papafragkakis, H.; Kim, J.; Harshan, M.; Theise, N.D.; Carr-Locke, D.L. Needle-based confocal endomicroscopy for evaluation of malignant lymph nodes—A feasibility study. Endoscopy 2016, 48, 923–928. [Google Scholar] [CrossRef] [PubMed]
  83. Terada, T.; Tahara, T.; Hashimoto, S.; Horiguchi, N.; Funasaka, K.; Nagasaka, M.; Nakagawa, Y.; Shibata, T.; Yoshioka, K.; Tsukamoto, T.; et al. Endoscopic ultrasound-guided needle-based confocal laser endomicroscopy in gastrointestinal subepithelial lesions: Feasibility study. Dig. Endosc. Off. J. Jpn. Gastroenterol. Endosc. Soc. 2020, 32, 574–584. [Google Scholar] [CrossRef]
  84. Bhutani, M.S.; Koduru, P.; Joshi, V.; Karstensen, J.G.; Saftoiu, A.; Vilmann, P.; Giovannini, M. EUS-Guided Needle-Based Confocal Laser Endomicroscopy: A Novel Technique With Emerging Applications. Gastroenterol. Hepatol. (NY) 2015, 11, 235–240. [Google Scholar]
  85. Larghi, A.; Manfrin, E.; Fabbri, C.; Crinò, S.F.; Correale, L.; Chiarello, G.; Barresi, L.; Van Velthuysen, M.L.; Poley, J.W.; Rahal, D.; et al. Interobserver agreement among expert pathologists on through-the-needle microforceps biopsy samples for evaluation of pancreatic cystic lesions. Gastrointest. Endosc. 2019, 90, 784–792. [Google Scholar] [CrossRef] [PubMed]
  86. Kovacevic, B.; Karstensen, J.G.; Havre, R.F.; Pham, K.D.; Giovannini, M.; Dabizzi, E.; Arcidiacono, P.; Santo, E.; Sequeiros, E.V.; Klausen, P.; et al. Initial experience with EUS-guided microbiopsy forceps in diagnosing pancreatic cystic lesions: A multicenter feasibility study (with video). Endosc. Ultrasound 2018, 7, 383–388. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  87. Crinò, S.F.; Bernardoni, L.; Brozzi, L.; Barresi, L.; Malleo, G.; Salvia, R.; Frulloni, L.; Sina, S.; Parisi, A.; Remo, A.; et al. Association between macroscopically visible tissue samples and diagnostic accuracy of EUS-guided through-the-needle microforceps biopsy sampling of pancreatic cystic lesions. Gastrointest. Endosc. 2019, 90, 933–943. [Google Scholar] [CrossRef]
  88. Barresi, L.; Crinò, S.F.; Fabbri, C.; Attili, F.; Poley, J.W.; Carrara, S.; Tarantino, I.; Bernardoni, L.; Giovanelli, S.; Di Leo, M.; et al. Endoscopic ultrasound-through-the-needle biopsy in pancreatic cystic lesions: A multicenter study. Dig. Endosc. Off. J. Jpn. Gastroenterol. Endosc. Soc. 2018, 30, 760–770. [Google Scholar] [CrossRef]
  89. Yang, D.; Trindade, A.J.; Yachimski, P.; Benias, P.; Nieto, J.; Manvar, A.; Ho, S.; Esnakula, A.; Gamboa, A.; Sethi, A.; et al. Histologic Analysis of Endoscopic Ultrasound-Guided Through the Needle Microforceps Biopsies Accurately Identifies Mucinous Pancreas Cysts. Clin. Gastroenterol. Hepatol. 2019, 17, 1587–1596. [Google Scholar] [CrossRef]
  90. Facciorusso, A.; Del Prete, V.; Antonino, M.; Buccino, V.R.; Wani, S. Diagnostic yield of EUS-guided through-the-needle biopsy in pancreatic cysts: A meta-analysis. Gastrointest. Endosc. 2020, 92, 1–8.e3. [Google Scholar] [CrossRef]
  91. Tacelli, M.; Celsa, C.; Magro, B.; Barchiesi, M.; Barresi, L.; Capurso, G.; Arcidiacono, P.G.; Cammà, C.; Crinò, S.F. Diagnostic performance of endoscopic ultrasound through-the-needle microforceps biopsy of pancreatic cystic lesions: Systematic review with meta-analysis. Dig. Endosc. Off. J. Jpn. Gastroenterol. Endosc. Soc. 2020. [Google Scholar] [CrossRef] [Green Version]
  92. Westerveld, D.R.; Ponniah, S.A.; Draganov, P.V.; Yang, D. Diagnostic yield of EUS-guided through-the-needle microforceps biopsy versus EUS-FNA of pancreatic cystic lesions: A systematic review and meta-analysis. Endosc. Int. Open 2020, 8, E656–E667. [Google Scholar] [CrossRef] [Green Version]
  93. Adsay, N.V.; Merati, K.; Basturk, O.; Iacobuzio-Donahue, C.; Levi, E.; Cheng, J.D.; Sarkar, F.H.; Hruban, R.H.; Klimstra, D.S. Pathologically and biologically distinct types of epithelium in intraductal papillary mucinous neoplasms: Delineation of an "intestinal" pathway of carcinogenesis in the pancreas. Am. J. Surg. Pathol. 2004, 28, 839–848. [Google Scholar] [CrossRef] [PubMed]
  94. Kim, J.; Jang, K.T.; Mo Park, S.; Lim, S.W.; Kim, J.H.; Lee, K.H.; Lee, J.K.; Heo, J.S.; Choi, S.H.; Choi, D.W.; et al. Prognostic relevance of pathologic subtypes and minimal invasion in intraductal papillary mucinous neoplasms of the pancreas. Tumour Biol. 2011, 32, 535–542. [Google Scholar] [CrossRef] [PubMed]
  95. Sadakari, Y.; Ohuchida, K.; Nakata, K.; Ohtsuka, T.; Aishima, S.; Takahata, S.; Nakamura, M.; Mizumoto, K.; Tanaka, M. Invasive carcinoma derived from the nonintestinal type intraductal papillary mucinous neoplasm of the pancreas has a poorer prognosis than that derived from the intestinal type. Surgery 2010, 147, 812–817. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Endoscopic ultrasound-guided fine-needle aspiration (EUS-FNA) of a gastric subepithelial lesion.
Figure 1. Endoscopic ultrasound-guided fine-needle aspiration (EUS-FNA) of a gastric subepithelial lesion.
Diagnostics 10 00463 g001
Figure 2. The Ez shot3 (Olympus Medical Systems, Tokyo, Japan) with a Menghini shape of the needle tip (Image courtesy of Olympus Medical Systems).
Figure 2. The Ez shot3 (Olympus Medical Systems, Tokyo, Japan) with a Menghini shape of the needle tip (Image courtesy of Olympus Medical Systems).
Diagnostics 10 00463 g002
Figure 3. The 20G ProCore needle with an antegrade beveled side slot at the needle tip (Image courtesy of Cook Medical).
Figure 3. The 20G ProCore needle with an antegrade beveled side slot at the needle tip (Image courtesy of Cook Medical).
Diagnostics 10 00463 g003
Figure 4. The franseen needle with three symmetric cutting tips (Image courtesy of Boston Scientific).
Figure 4. The franseen needle with three symmetric cutting tips (Image courtesy of Boston Scientific).
Diagnostics 10 00463 g004
Figure 5. The 19 G Sono Tip Pro Control (Medi-Globe GmbH) with a cut-surface length of 5 mm.
Figure 5. The 19 G Sono Tip Pro Control (Medi-Globe GmbH) with a cut-surface length of 5 mm.
Diagnostics 10 00463 g005
Figure 6. Hot-Axios stent delivery system with electrocautery at the distal tip of the delivery catheter (Image courtesy of Boston Scientific).
Figure 6. Hot-Axios stent delivery system with electrocautery at the distal tip of the delivery catheter (Image courtesy of Boston Scientific).
Diagnostics 10 00463 g006
Table 1. List of needles used for endoscopic ultrasound-guided procedures.
Table 1. List of needles used for endoscopic ultrasound-guided procedures.
Release YearNeedleSizeMaterial of Needle
2000Echo Tip Ultra (Cook Medical)19, 22, 25Stainless
2001NA-11J-KB (Olympus Medical Systems)22Stainless
2003EZ Shot (Olympus Medical Systems)22Stainless
2004Quick-Core (Cook Medical)19Stainless
2011Expect (Boston Scientific)19, 22, 25Cobalt-chromium
2011EZ Shot 2(Olympus Medical Systems)19, 22, 25Stainless
2012Echo Tip Procore (Cook Medical)19, 22, 25Stainless
2012SONO tip Pro Control (Medi-Globe GmbH)19, 22, 25Stainless
2012Expect 19 G Flex Needle (Boston Scientific)19Nitinol
2013EUS Sonopsy CY (HAKKO)21Stainless
2016EZ Shot 3 Plus (Olympus Medical Systems)19, 22Nitinol
2016Echo Tip Procore 20 G (Cook Medical)20Stainless
2016SharkCore (Medtronic)19, 22, 25Stainless
2016Acquire (Boston)22, 25Cobalt-chromium
2017Acquire 19 G Flex Needle (Boston Scientific)19Nitinol
2018EZ Shot 3 Plus (Olympus Medical Systems)25Stainless
2020SONO tip TopGain (Medi-Globe GmbH)19, 22, 25Stainless
Table 2. Published comparative studies regarding 22 G versus 25 G FNA-needles.
Table 2. Published comparative studies regarding 22 G versus 25 G FNA-needles.
ReferenceStudy DesignCases (n)LesionRoseDiagnostic Ability (25G vs. 22G)
Lee [44] 2009RCT12Mainly pancreasYesn.s
Siddiqui [45] 2009RCT131PancreasYesAccuracy; 95.5% vs. 87.5%; p = 0.18; n.s
Camellini [46] 2011RCT127Mainly pancreasYesAccuracy; 78.1% vs. 77.8%; n.s
Fabbri [47] 2011RCT50PancreasYesAccuracy; 94% vs. 86%; n.s
Lee [48] 2013RCT188PancreasNoAccuracy; 88.3% vs. 89.4%; p = 0.82; n.s
Vilmann [49] 2013RCT135Pancreas and LNNoSensitivity; 94.1% vs. 94.1%; n.s
Carrara [43] 2016RCT144Mainly pancreasYesAccuracy; 81% vs. 68%; p = 0.09; n.s
Affolter [41] 2013MA (11 studies)1452Mainly pancreas5/11Sensitivity; 91% vs. 78%; p = 0.97; n.s
Madhoun [50] 2013MA (8 studies)1292Pancreas5/8Sensitivity; 93% vs. 85%; p = 0.0003
Facciorusso [51] 2017MA (7 studies)732Pancreas5/7Sensitivity; 93% vs. 89%; p = 0.13; n.s
Xu [52] 2017MA (11 studies)837Pancreas6/11Sensitivity; 92% vs. 88%; p = 0.046
Guedes [53] 2018MA (4 studies)462Pancreas2/4Sensitivity; 93% vs. 91%; p = 0.497; n.s
RCT, randomized controlled trial; MA, meta-analysis; n.s, not significant.
Table 3. Published comparative studies regarding EUS-FNA versus EUS-FNB.
Table 3. Published comparative studies regarding EUS-FNA versus EUS-FNB.
ReferenceStudy DesignCases (n)LesionDiagnostic Ability (FNA vs. FNB)
Alatawi [67] 2015RCT100PancreasAdequate sample rate; 90% vs. 100%; p = 0.0003
Kim [68] 2014RCT22SELsAccuracy; 20% vs. 75%; p = 0.010
Khan [69] 2017MA (15 studies)1024pancreas, LN and
solid GI lesion
FNB showed better diagnostic adequacy
without ROSE; p = 0.02
RCT, randomized controlled trial; MA, meta-analysis.

Share and Cite

MDPI and ACS Style

Fujita, A.; Ryozawa, S.; Tanisaka, Y.; Ogawa, T.; Suzuki, M.; Noguchi, T.; Katsuda, H.; Mizuide, M. Current Status of Needles in the Optimization of Endoscopic Ultrasound-Guided Procedures. Diagnostics 2020, 10, 463. https://0-doi-org.brum.beds.ac.uk/10.3390/diagnostics10070463

AMA Style

Fujita A, Ryozawa S, Tanisaka Y, Ogawa T, Suzuki M, Noguchi T, Katsuda H, Mizuide M. Current Status of Needles in the Optimization of Endoscopic Ultrasound-Guided Procedures. Diagnostics. 2020; 10(7):463. https://0-doi-org.brum.beds.ac.uk/10.3390/diagnostics10070463

Chicago/Turabian Style

Fujita, Akashi, Shomei Ryozawa, Yuki Tanisaka, Tomoya Ogawa, Masahiro Suzuki, Tatsuya Noguchi, Hiromune Katsuda, and Masafumi Mizuide. 2020. "Current Status of Needles in the Optimization of Endoscopic Ultrasound-Guided Procedures" Diagnostics 10, no. 7: 463. https://0-doi-org.brum.beds.ac.uk/10.3390/diagnostics10070463