Review

Spinal Implant Osseointegration and the Role of 3D Printing: An Analysis and Review of the Literature

1
Department of Orthopaedic Surgery, University of Connecticut, Farmington, CT 06032, USA
2
Bone and Joint Institute, Hartford Hospital, Hartford, CT 06106, USA
*
Author to whom correspondence should be addressed.
Submission received: 18 January 2022 / Revised: 12 February 2022 / Accepted: 4 March 2022 / Published: 6 March 2022

Abstract

:
The use of interbody implants for spinal fusion has been steadily increasing to avoid the risks of complications and donor site morbidity when using autologous bone. Understanding the pros and cons of various implant designs can assist the surgeon in choosing the ideal interbody for each individual patient. The goal of these interbody cages is to promote a surface area for bony ingrowth while having the biomechanical properties to support the axial skeleton. Currently, the majority of interbody implants consists of metal or polyether ether ketone (PEEK) cages with bone graft incorporated inside. Titanium alloy implants have been commonly used, however, the large difference in modulus of elasticity from bone has inherent issues. PEEK implants have a desirable surface area with the benefit of a modulus of elasticity closer to that of bone. Unfortunately, clinically, these devices have had increased risk of subsidence. More recently, 3D printed implants have come into the market, providing mechanical stability with increased surface design for bony ingrowth. While clinical outcomes studies are limited, early results have demonstrated more reliable and quicker fusion rates using 3D custom interbody devices. In this review, we discuss the biology of osseointegration, the use of surface coated implants, as well as the potential benefits of using 3D printed interbodies.

1. Introduction

Musculoskeletal conditions are among the most disabling and costly conditions experienced by Americans [1]. Spinal fusions have evolved into a viable treatment modality to treat chronic back pain and restore patients’ quality of life [2,3,4]. While autologous bone grafts are generally regarded as the standard augment for spinal fusion surgeries due to their osteogenic capabilities, complications and morbidity to the donor site have given rise to the use of substitutes and spinal implants [5,6]. As the baby-boomer generation continues to age and comprise a disproportionate amount of the musculoskeletal complaints, there is increased need for surgical technologies that utilize the intrinsic regenerative capacity of mineralized tissues to provide more permanent solutions for spinal pathologies [7].
Osseointegration refers to the direct integration of bone to metal resulting in structural and functional integration between the living bone and implant surface [8]. Advances in osseointegration have stemmed largely from the dental implant field, which has demonstrated that achieving successful bone implantation requires a strong and direct interaction between bone and implant surface [9,10,11,12,13]. Early literature on osseointegration was problematic, with reports of formation of fibrous capsules around metallic or polymeric implants leading to early failure [14,15,16,17,18,19,20,21]. There have also been reported cases in which significant osteolysis developed, despite indetectable traces of wear debris [21,22,23]. More recently, these issues have been minimized by controlling specific implant properties like surface roughness and nanostructures to promote bone apposition directly onto implant surfaces [24,25,26,27,28,29].
There are a variety of clinically available cages for spinal fusion that differ in shape, texture, and chemical composition. The purpose of this review is to (1) evaluate the key biological processes that occur around implants, (2) discuss the role that surface structure plays on osseointegration, and (3) discuss current literature on custom 3D printed cages and their impact on fusion rates.

2. The Biology of Osseointegration

Osseointegration begins with absorption of water molecules, proteins, and lipids to the implant surface [30,31]. The specific host response to the implant surface is dependent upon protein properties [31]. Proteins such as fibronectin and vitronectin tend to initiate inflammation through the attachment of platelets [31,32,33]. These platelet attachments have shown to lead to fibrin clot formation that contributes to a meshwork architecture, which facilitates cell migration toward the implant surface [7,34]. Neutrophils and macrophages utilize this meshwork to rid the area of pathogens and necrotic tissue [35,36]. Mesenchymal stem cells (MSCs) arriving at the implant surface are exposed to these inflammatory cytokines and influence subsequent differentiation into osteoblasts, chondrocytes, and fibroblasts [37,38]. Gittens et al. demonstrated that these MSCs tend to form bone and soft tissue in this environment, although the characteristics of the implant are thought to influence this process [7].
Upon fibrin meshwork formation, bone can form on the bone surface surrounding the implant (a process termed distance osteogenesis) and on the implant surface (contact osteogenesis) [39]. Osteoblasts encountering these two surfaces may reproduce for a few generations or lay down proteins to form the lamina limitans [39]. Often referred to as the cement line [39,40,41], the lamina limitans contains a protein profile which further induces osteoblast migration and maturation.
Bony remodeling then occurs, a process in which osteoclasts resorb the newly formed bone to amend microcracks and prepare the surface for new bone formation (Figure 1) [42]. This involves creating resorption lacunae with nanotopography thought to signal osteoblasts that an appropriate surface has been established for new bone formation [7]. If the surface properties are insufficient, migrating cells can form fibrous tissue between implant and bone, resulting in degradation of surrounding bone and implant loosening [24].

3. Implant Composition

Patient-specific conditions such as old age, poor bone quality, and smoking can threaten the outcome of successful fusion [40]. The goal of implant design is thus to minimize the effects of patient variables and improve implant-bone osseointegration by being cognizant of implant chemical composition, as well as the microscale (1 dimension <100 μm), submicroscale (1 dimension <1 μm), and nanoscale (1 dimension <100 nm) surface characteristics [40]. It is important to acknowledge that the region of the instrumented spine may have relevance for implant selection. Higher reported rates of subsidence and non-union in lumbar fusions compared to other parts of the spine underscores the need for careful implant development and selection when instrumentation of these levels is being planned.

3.1. Titanium and Titanium Alloys

Titanium and titanium alloy implants are widely used due to their suitable weight-to-strength ratio and good biological performance (Figure 2). Brånemark et al. demonstrated that the surface of titanium allows for osseointegration [9]. Upon exposure to air, titanium forms a thin oxide layer that inhibits further implant corrosion; the titanium oxide film also restricts the release of ionic and molecular titanium species, protecting the biological surroundings from the highly reactive metal [43]. It is suggested that the oxide later provides titanium’s good biological performance by mimicking the ceramic properties of hydroxyapatite in bone [25]. Titanium is relatively agreeable to the spine because of its biocompatibility, robust repassivation, resistance to corrosion, and low density [44].
The limitation of titanium alloys revolves around the mismatch in the elastic modulus between titanium (100 GPa) and bone (10–30 GPa) [10]. This mechanical difference can lead to stress shielding around the implant, local inflammation, bone atrophy, subsidence, and implant failure [10]. Moreover, assessing successful fusion radiographically may be difficult due to titanium’s high radiodensity [45].

3.2. Polyether Ether Ketone (PEEK)

Interbody cages are often made of polyether ether ketone (Figure 3), which is an inert semicrystalline polyaromatic linear polymer. PEEK is an inexpensive, radiolucent material that has a modulus of elasticity that approximates that of cortical bone [16]. Its comparable modulus of elasticity to bone has contributed to comparable arthrodesis rates between PEEK cages and autografts [16]. Unlike titanium, the radiolucent composition of these cages allows monitoring of bone growth on post-operative serial radiographs. In contrast to roughened titanium alloy surfaces that promote osteogenesis, PEEK surfaces result in the formation of fibrous tissue [46]. As a result, PEEK cages are often packed with bone graft to achieve vertebral fusion.
Prospective clinical studies comparing PEEK to autograft have suggested equivalent results in regard to patient-reported outcomes [47]. However, surgical outcomes have suggested higher rates of PEEK cage subsidence, ranging from 32% to 38% of cases [48,49]. Suggested theories for this include over-distraction, overly aggressive endplate preparation, or normal fusion processes [48].
Manipulation of the quantity and direction of fiber element to achieve desired material properties provides PEEK cages with the potential to minimize stress shielding compared to solid titanium implants [50]. Pelletier et al. performed a study comparing PEEK and titanium anterior lumbar interbody fusion implants in sheep, demonstrating no difference in initial biomechanics, mechanical properties, or fusion rates when similar amounts of bone graft were used [50]. There are varying reports in the literature of the interaction between PEEK and osteoblasts [51,52]. Olivares-Navarrete et al. found that osteoblasts differentiate to a lesser phenotypic degree on PEEK versus titanium surfaces, suggesting that PEEK cannot support osteogenic tissue as well as titanium [51]. However, Sagomonyants et al. demonstrated that PEEK and roughened titanium have comparable in vitro bone forming capacity [52].
Unlike comparisons between allograft/autograft and PEEK cages suggesting worse subsidence rates with PEEK, comparisons with titanium compounds have had different results [53]. A randomized controlled trial by Chen et al. comparing PEEK to titanium cages in the cervical spine showed subsidence rates of 34.5% in titanium compared to 5.4% in PEEK [53]. Several subsequent studies demonstrated that PEEK preserves intervertebral heights and Cobb angles more effectively than their metallic counterparts [54,55].

3.3. Surface-Coated Cages

The development of surface-coated interbody cages arose from the abovementioned conflicting evidence regarding PEEK cages and the literature comparing the advantages and disadvantages of PEEK, metals, and biological supplements [53,54,55]. Surface-coated interbody cages increase bone-to-implant contact ratio and bioactivity [53]. Interbody models may be covered with a thin layer of various metals including hydroxyapatite, titanium, gold, titanium dioxide, diamond-like carbon, and tert-butoxides [56]. Hydroxyapatite is the most commonly bioactive material used, with several studies demonstrating increased osteoconductivity of hydroxyapatite-coated PEEK cages, in addition to potentially having osteoinductive properties [56].
Surface-coated interbody cages are thought to induce bony overgrowth and arthrodesis starting at the bone–cage interface due to the rough nanometer surface [55]. This surface promotes bony fusion and induces calcium phosphate deposit [55,56,57]. Titanium and gold coating, for example, has been shown to promote osteoblast adhesion on the PEEK interbody graft [58]. Despite these promising results, surface-coated interbody cages have been subject to scrutiny because their modulus of elasticity can range from 10 GPa to 100 GPa (compared to 1.0–2.4 GPa in cortical bone), depending on the density of the coat [53]. Titanium-coated interbody cages have been shown to increase shear strength between implant and bone, reducing the risk of pseudarthrosis, though there is risk of delamination [53]. Kienle et al. performed a biomechanical study to simulate the impaction process in titanium-coated PEEK cages [59]. In contrast to surface-etched implants, the titanium-coated PEEK implants were susceptible to impaction-related wear debris, half of which was of a size range that allows phagocytosis, thus promulgating a systemic inflammatory reaction and possibly hindering arthrodesis [59].

4. Surface Modifications

4.1. Additive and Subtractive Manufacturing

The two primary types of manufacturing spine implants are additive and subtractive [60]. Additive manufacturing, often termed 3-dimensional (3D) printing, involves the application using computer software or 3D material coating on the implant [61]. Subtractive manufacturing involves generation of surface features through removal of material, producing submicron surface textures; examples of which include acid etching and grit blasting of titanium surfaces, which have been shown to increase osteoblastic differentiation and improve osseointegration and bone formation [62].

4.2. Microroughness and Nanostructures

Nanoscale surface modifications represent a developing subfield of fusion science whereby host cells are able to interact with implants on a molecular level via cellular membrane receptors to trigger osteoblastic-lineage [60]. This process occurs on nanoscale (10−9 m) surfaces, as microscale (10−6 m) surface texture does not interact with cellular membranes [60].
Implant surface roughness can be manipulated to influence particular protein families that stimulate certain types of cells to attach. This has been shown to occur with metallic surfaces, PEEK, and hyaluronic acid [52]. In addition, surface roughness can influence initial implant fixation by increasing friction and limiting micromotion [52]. Most commercially available implants contain a surface modification to increase roughness, as this has been demonstrated to have beneficial results with in vitro and in vivo analyses [12,26]. Moreover, surfaces with complex microtopography appear to be even more osteogenic than surfaces with only one type of roughness. Surface modification techniques to increase microroughness include acid etching, sand blasting, heat treatments, and anodic oxidation [7].
Olivares-Navarrete et al. previously evaluated the effects of roughened surfaces on implant types [38]. One study demonstrated that osteoblasts exhibit a more differentiated phenotype when grown on machined or grit-blasted titanium aluminum vanadium alloys than when grown on smoother titanium surfaces [38]. This was supported by Sykaras et al., whom found that the highest levels of osteoclast inhibitors (transforming growth factor beta and osteoprotegerin), angiogenesis factors (fibroblast growth factor 2, vascular endothelial growth factor A, and angiopoietin-1), and bone morphogenetic proteins on roughened titanium surfaces compared to smooth titanium and PEEK [12]. In addition, multiple studies have shown that nanoengineered implants increase stimulation of local growth factors, including bone morphogenetic proteins, vascular endothelial growth, and transforming growth factor beta [63,64].

4.3. Bioabsorbable Interbody Cages

The notion of a bioabsorbable interbody fusion is appealing because these implants serve to provide structure before being resorbed over time and replaced by host bone. These structures are intended to recreate the extracellular matrix of bone [64]. However, skepticism regarding the value of bioabsorbable implants and their limited use in clinical practice stem from poor osteoconductivity and low primary stability with development of cracks and foreign body reactions [65]. To improve osteoconductivity, nanosized β-tricalcium phosphate (β-TCP) has been incorporated into polylactide (PLA) cages [66]. Cao et al. developed a bioabsorbable cervical fusion cage from PLA and β-TCP that was shown to have greater biomechanical stability in a sheep model, as compared with tricortical iliac crest grafts and PEEK cages, allowing for resorption over time [66]. A subsequent in vivo study by the same group compared the use of a novel polylactide/nano-sized β-tricalcium phosphate bioabsorbable self-retaining cervical fusion cage (BCFC) to autologous bone graft and PEEK cages [66]. The authors found that at 12 weeks post-operatively the BCFC group yielded a significantly lower range of motion in axial rotation than both the autologous bone graft and PEEK cage group. Histologic evaluation revealed an increased intervertebral bone volume/total volume ratio and better interbody fusion in the BCFC group than in the other groups [66]. However, this was proven in an animal model with no clinical data to date [66].

4.4. Hydroxyapatite Coating

It is well established that bone demonstrates a strong affinity to implants composed of sintered hydroxyapatite (HA) [67]. HA interbody spaces have been used, but the mechanical properties of HA on its own are not well suited to this application because resistance to fatigue failure is very low [67]. However, it is possible to utilize additive techniques to attach HA to titanium. This HA coating technology has also demonstrated increased bone apposition, increased resistance to pull-out forces, and increased extraction torque for HA-coated stainless steel pedicle screws, though this has not been shown for interbody spacers [68].

5. 3D Printing

3D printing is a form of additive manufacturing in which multiple 2D layers are formed atop one another to create a 3-dimensional product (Figure 4). 3D printing technology in spine surgery can be used to create models of pathology for pre-operative planning, individualized pedicle screw guides, and most commonly customized implants [69,70]. There are many techniques for 3D printing with the most common being extrusion printing, in which a solid starting material is extruded as either a liquid or semi-liquid and then rapidly cooled. However, orthopedic implants are often printed using powder bed fusion, in which a thin layer of powdered material is deposited on a platform and an electron beam is used to fuse the material to form the implant design [70]. While a predominant amount of the literature on 3D implants utilize titanium, there have been promising reports of PEEK printed implants [71,72].
3D printed implants can be separated into two groups: patient specific (PS) and off-the-shelf (OTS). OTS implants are similar to standard implants but utilize 3D printing technology to impart a customizable porosity and stiffness [73]. Alternatively, PS implants are designed using a patient’s pre-operative CT or MRI scans and allow for superior endplate matching, as well as customized amounts of lordosis and height [70]. OTS implants are desirable as they allow for mass manufacturing and thus decreased costs. Current studies investigating OTS 3D implants have been promising. Mokawem et al. utilized OTS 3D printed TLIF and LLIF titanium interbodies impregnated with silicate-substituted calcium phosphate and found solid fusion on CT at one year in 92 of 93 patients [74]. McGilvray et al. found increased fusion mass for OTS 3D printed titanium alloy cages at 16 weeks compared to PEEK and plasma sprayed PEEK in an ovine model [73].
Several case reports of successful implementation of PS implants in the cervical, thoracic, and lumbar spine for neoplasia, degenerative disease, infection, congenital anomaly, and trauma have been published [72,73,75]. One of the greatest theorized benefits of customized 3D printed implants is increased endplate-cage contact resulting in decreased force point-loading. Mobbs et al created a customized ALIF cage and used finite element analysis to compare pressure loading across the endplate [76]. They found that the custom implant resulted in a more even distribution of force along the endplate compared to a generic ALIF implant [76]. In corpectomy surgery, custom printed cages can be created in concert with planned pedicle screw trajectories to allow for interposition of the posterior construct with the corpectomy cage. 3D printing technology also allows for the creation of multilevel cages made to match a corrected sagittal profile for patients undergoing large corpectomies.
Custom 3D printed implants have also been successfully used in atlantoaxial fusion. Phan et al successfully used a 3D printed posterior implant with integrated screw holes with a predetermined trajectory for a C1–C2 fusion [77]. While there have been no studies in spine surgery comparing outcomes between OTS and PS implants, there is current literature in knee arthroplasty suggesting PS implants have been associated with less intraoperative blood loss and greater patient reported outcome measures when compared to OTS implants [77]. Further research is warranted to determine if a similar effect is seen in 3D printed spine implants.
While preliminary reports of 3D printed spinal implants have been promising, this technology is not without its limitations. 3D printed implants, especially PS implants, are more expensive than traditional implants, take longer to produce, and require the technical skills to design and print the final product. As the use of this technology increases, costs are likely to go down, however, it is uncertain if customized implants will gain traction for more common cases with standard patient anatomy.

6. Clinical Implications and Future Perspective

As 3D printing continues to generate more sophisticated interbody shapes and additive potential, the clinical applicability of 3D cages will continue to broaden. This is expected to have significant clinical applicability, as augmenting implants on microscales and nanoscales to improve fusion rates and decrease subsidence are likely to produce better patient outcomes and satisfaction. Positive in vivo results are a promising first step to bringing surface modifications to clinical practice, but long-term clinical studies are needed to ascertain the full clinical implications of these difference surface features on implant performance. Further prospective clinical studies with longer-term outcomes are needed comparing OTS to custom 3D implants.

7. Clinical Implications

The purpose of this review was to evaluate the key biological processes that occur around implants, discuss the role that surface structure plays on osseointegration, and discuss current literature on custom 3D printed cages and their impact on fusion rates. This review was deemed necessary due to rapidly evolving methods of augmenting implants and clinical needs to decrease subsidence and generate quicker fusion rates. Cages afford excellent load bearing of the vertebral column and height restoration of the intervertebral space [4]. While they reduce stress on adjacent vertebral bodies, cages do not inherently provide the stimulus for bone remodeling necessary for fusion [4]. Successful spinal fusion utilizing interbody implants is dependent on several characteristics, including surface roughness, material properties, and adequate endplate preparation. Osseointegration relies on a biologic response of osteoblasts, fibroblasts and integrins to allow for new bone to form within the interbody. This review demonstrates PEEK implants offer the benefit of better visualization of fusion on x-ray with less subsidence than titanium implants. The downside of using PEEK has been the lack of roughness for osseointegration. In the future, 3D printed PEEK implants may be able to correct this downside with further clinical studies pending.

Author Contributions

Conceptualization, C.K., C.L.A., Writing—C.K., I.W., H.S.M., S.M.E. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

No Institional Review Board Required.

Data Availability Statement

Not applicable.

Conflicts of Interest

The authors declare no conflict of interests.

References

  1. Yelin, E.; Weinstein, S.; King, T. The burden of musculoskeletal diseases in the United States. Semin. Arthritis Rheum. 2016, 46, 259–260. [Google Scholar] [CrossRef] [PubMed]
  2. Hanley, E.N. The indications for lumbar spinal fusion with and without instrumentation. Spine 1995, 20, 143S–153S. [Google Scholar] [CrossRef] [PubMed]
  3. Hacker, R.J.; Cauthen, J.C.; Gilbert, T.J.; Griffith, S.L. A Prospective Randomized Multicenter Clinical Evaluation of an Anterior Cervical Fusion Cage. Spine 2000, 25, 2646–2655. [Google Scholar] [CrossRef] [PubMed]
  4. Zdeblick, T.A.; Phillips, F.M. Interbody Cage Devices. Spine 2003, 28, S2–S7. [Google Scholar] [CrossRef] [PubMed]
  5. Agrawal, C.M.; Attawia, M.; Borden, M.D.; Boyan, B.D.; Bruder, S.P.; Bucholz, R.W. Bone Graft Substitutes; ASTM International/AAOS: Rosemont, IL, USA, 2003. [Google Scholar]
  6. Ray, C.D. Threaded fusion cages for lumbar interbody fusions: An economic comparison with 360 degrees fusions. Spine 1997, 22, 681–685. [Google Scholar] [CrossRef] [PubMed]
  7. Gittens, R.A.; Olivares-Navarrete, R.; Schwartz, Z.; Boyan, B.D. Implant osseointegration and the role of microroughness and nanostructures: Lessons for spine implants. Acta Biomater. 2014, 10, 3363–3371. [Google Scholar] [CrossRef] [Green Version]
  8. Li, Y.; Brånemark, R. Osseointegrated prostheses for rehabilitation following amputation: The pioneering Swedish model. Unfallchirurg 2017, 120, 285–292. [Google Scholar] [CrossRef] [Green Version]
  9. Brånemark, P.I.; Hansson, B.O.; Adell, R.; Breine, U.; Lindström, J.; Hallén, O.; Ohman, A. Osseointegrated implants in the treatment of the edentulous jaw. Experience from a 10-year period. Scand. J. Plast. Reconstr. Surgery Suppl. 1977, 16, 1–132. [Google Scholar]
  10. Bobbio, A. The first endosseous alloplastic implant in the history of man. Bull. Hist. Dent. 1972, 20, 1–6. [Google Scholar]
  11. Lang, B.R.; Chiappa, A. Mandibular implants: A new method of attachment. J. Prosthet. Dent. 1969, 22, 261–267. [Google Scholar] [CrossRef]
  12. Sykaras, N.; Iacopino, A.M.; Marker, V.A.; Triplett, R.G.; Woody, R.D. Implant materials, designs, and surface topographies: Their effect on osseointegration. A literature review. Int. J. Oral Maxillofac. Implant. 2000, 15, 675–690. [Google Scholar]
  13. Cook, H.P. Immediate reconstruction of the mandible by metallic implant following resection for neoplasm. Ann. R. Coll. Surg. Engl. 1968, 42, 233–259. [Google Scholar] [PubMed]
  14. Thomas, K.A.; Kay, J.F.; Cook, S.D.; Jarcho, M. The Effect of Surface Macrotexture and Hydroxylapatite Coating on the Mechanical Strengths and Histologic Profiles of Titanium Implant Materials. J. Biomed Mater. Res. 1987, 21, 1395–1414. [Google Scholar] [CrossRef] [PubMed]
  15. Hallab, N.J.; Cunningham, B.W.; Jacobs, J.J. Spinal Implant Debris-Induced Osteolysis. Spine 2003, 28, S125–S138. [Google Scholar] [CrossRef] [Green Version]
  16. Kurtz, S.M.; Devine, J.N. PEEK biomaterials in trauma, orthopedic, and spinal implants. Biomaterials 2007, 28, 4845–4869. [Google Scholar] [CrossRef] [Green Version]
  17. Santos, E.R.G.; Goss, D.G.; Morcom, R.K.; Fraser, R.D. Radiologic Assessment of Interbody Fusion Using Carbon Fiber Cages. Spine 2003, 28, 997–1001. [Google Scholar] [CrossRef]
  18. Anjarwalla, N.K.; Morcom, R.K.; Fraser, R.D. Supplementary stabilization with anterior lumbar intervertebral fusion—A radi-ologic review. Spine 2006, 31, 1281–1287. [Google Scholar] [CrossRef]
  19. Cunningham, B.W.; Orbegoso, C.M.; E Dmitriev, A.; Hallab, N.J.; Sefter, J.C.; Asdourian, P.; McAfee, P.C. The effect of spinal instrumentation particulate wear debris: An in vivo rabbit model and applied clinical study of retrieved instrumentation cases. Spine J. 2003, 3, 19–32. [Google Scholar] [CrossRef]
  20. Lohmann, C.H.; Schwartz, Z.; Koster, G.; Jahn, U.; Buchhorn, G.H.; MacDougall, M.J. Phagocytosis of wear debris by osteoblasts affects differentiation and local factor production in a manner dependent on particle composition. Biomaterials 2000, 21, 551–561. [Google Scholar] [CrossRef]
  21. Pioletti, D.P.; Takei, H.; Kwon, S.Y.; Wood, D.; Sung, K.L.P. The cytotoxic effect of titanium particles phagocytosed by osteoblasts. J. Biomed. Mater. Res. 1999, 46, 399–407. [Google Scholar] [CrossRef]
  22. Athanasou, N.A.; Quinn, J.; Bulstrode, C.J. Resorption of bone by inflammatory cells derived from the joint capsule of hip arthroplasties. J. Bone Joint. Surg. Br. 1992, 74, 57–62. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  23. Maniatopoulos, C.; Pilliar, R.M.; Smith, D.C. Threaded versus porous-surfaced designs for implant stabilization in bone-endodontic implant model. J. Biomed. Mater. Res. 1986, 20, 1309–1333. [Google Scholar] [CrossRef] [PubMed]
  24. Piattelli, A.; Scarano, A.; Favero, L.; Iezzi, G.; Petrone, G.; Favero, G.A. Clinical and Histologic Aspects of Dental Implants Removed Due to Mobility. J. Periodontol. 2003, 74, 385–390. [Google Scholar] [CrossRef] [PubMed]
  25. Sul, Y.-T.; Johansson, C.; Wennerberg, A.; Cho, L.-R.; Chang, B.-S.; Albrektsson, T. Optimum surface properties of oxidized implants for reinforcement of osseointegration: Surface chemistry, oxide thickness, porosity, roughness, and crystal structure. Int. J. Oral Maxillofac. Implant. 2005, 20, 349–359. [Google Scholar]
  26. Schwarz, F.; Wieland, M.; Schwartz, Z.; Zhao, G.; Rupp, F.; Geis-Gerstorfer, J.; Schedle, A.; Broggini, N.; Bornstein, M.M.; Buser, D.; et al. Potential of chemically modified hydrophilic surface characteristics to support tissue integration of titanium dental implants. J. Biomed. Mater. Res. Part Appl. Biomater. 2008, 88, 544–557. [Google Scholar] [CrossRef]
  27. Puleo, D.A.; Nanci, A. Understanding and controlling the bone-implant interface. Biomaterials 1999, 20, 2311–2321. [Google Scholar] [CrossRef]
  28. Kokubo, T.; Kim, H.-M.; Kawashita, M.; Nakamura, T. REVIEW Bioactive metals: Preparation and properties. J. Mater. Sci. Mater. Electron. 2004, 15, 99–107. [Google Scholar] [CrossRef]
  29. Brånemark, P. I, Adell, R.; Breine, U.; Hansson, B.O.; Lindström, J.; Ohlsson, A. Intra-osseous anchorage of dental prostheses. I. Experimental studies. Scand. J. Plast Reconstr. Surg. 1969, 3, 81–100. [Google Scholar]
  30. Andrade, J.D.; Hlady, V. Protein adsorption and materials biocompatibility: A tutorial review and suggested hypotheses. In Biopolymers/Non-Exclusion HPLC; Springer: Berlin/Heidelberg, Germany, 1986; pp. 1–63. [Google Scholar] [CrossRef]
  31. Wilson, C.J.; Clegg, R.E.; Leavesley, D.I.; Pearcy, M.J. Mediation of Biomaterial–Cell Interactions by Adsorbed Proteins: A Review. Tissue Eng. 2005, 11, 1–18. [Google Scholar] [CrossRef]
  32. Jansson, E.; Tengvall, P. In vitro preparation and ellipsometric characterization of thin blood plasma clot films on silicon. Biomaterials 2001, 22, 1803–1808. [Google Scholar] [CrossRef]
  33. Keselowsky, B.G.; Collard, D.M.; Garcia, A.J. Surface chemistry modulates fibronectin conformation and directs integrin bind-ing and specificity to control cell adhesion. J. Biomed. Mater. Res. 2003, 66A, 247–259. [Google Scholar] [CrossRef]
  34. Marx, R.E. Platelet-rich plasma: Evidence to support its use. J. Oral Maxillofac. Surg. 2004, 62, 489–496. [Google Scholar] [CrossRef] [PubMed]
  35. Babensee, J.E.; Anderson, J.M.; McIntire, L.V.; Mikos, A.G. Host response to tissue engineered devices. Adv. Drug Deliv. Rev. 1998, 33, 111–139. [Google Scholar] [CrossRef]
  36. Schindeler, A.; McDonald, M.; Bokko, P.; Little, D.G. Bone remodeling during fracture repair: The cellular picture. Semin. Cell Dev. Biol. 2008, 19, 459–466. [Google Scholar] [CrossRef] [PubMed]
  37. Bruder, S.P.; Fink, D.J.; Caplan, A. Mesenchymal stem cells in bone development, bone repair, and skeletal regenaration therapy. J. Cell. Biochem. 1994, 56, 283–294. [Google Scholar] [CrossRef] [PubMed]
  38. Olivares-Navarrete, R.; Gittens, R.A.; Schneider, J.M.; Hyzy, S.L.; Haithcock, D.A.; Ullrich, P.F.; Schwartz, Z.; Boyan, B.D. Osteoblasts exhibit a more differentiated phenotype and increased bone morphogenetic protein production on titanium alloy substrates than on poly-ether-ether-ketone. Spine J. 2012, 12, 265–272. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  39. Davies, J.E. Mechanisms of endosseous integration. Int. J. Prosthodont. 1998, 11, 391–401. [Google Scholar] [PubMed]
  40. Saruwatari, L.; Aita, H.; Butz, F.; Nakamura, H.K.; Ouyang, J.; Yang, Y. Osteoblasts generate harder, stiffer, and more de-lamination-resistant mineralized tissue on titanium than on polystyrene, associated with distinct tissue micro- and ultra-structure. J. Bone Miner. Res. 2005, 20, 2002–2016. [Google Scholar] [CrossRef]
  41. Owen, T.A.; Aronow, M.; Shalhoub, V.; Barone, L.M.; Wilming, L.; Tassinari, M.S. Progressive development of the rat osteo-blast phenotype invitro-Reciprocal relationships in expression of genes associated with osteoblast proliferation and dif-ferentiation during formation of the bone extracellular-matrix. J. Cell Physiol. 1990, 143, 420–430. [Google Scholar] [CrossRef]
  42. Mulari, M.T.K.; Qu, Q.; Harkonen, P.L.; Vaananen, H.K. Osteoblast-like cells complete osteoclastic bone resorption and form new mineraliz, ed bone matrix in vitro. Calcif Tissue Int. 2004, 75, 253–261. [Google Scholar] [CrossRef]
  43. Textor, M.; Sittig, C.; Frauchiger, V.; Tosatti, S.; Brunette, D.M. Properties and Biological Significance of Natural Oxide Films on Titanium and Its Alloys. In Titanium in Medicine; Springer: Berlin/Heidelberg, Germany, 2001; pp. 171–230. [Google Scholar] [CrossRef]
  44. Ramakrishna, S.; Mayer, J.; Wintermantel, E.; Leong, K.W. Biomedical applications of polymer-composite materials: A review. Comp. Sci. Technol. 2001, 61, 1189–1224. [Google Scholar] [CrossRef]
  45. Rao, P.J.; Pelletier, M.H.; Walsh, W.R.; Mobbs, R. Spine Interbody Implants: Material Selection and Modification, Functionalization and Bioactivation of Surfaces to Improve Osseointegration. Orthop. Surg. 2014, 6, 81–89. [Google Scholar] [CrossRef] [PubMed]
  46. Suh, P.B.; Puttlitz, C.; Lewis, C.; Bal, B.S.; McGilvray, K. The Effect of Cervical Interbody Cage Morphology, Material Composition, and Substrate Density on Cage Subsidence. J. Am. Acad. Orthop. Surg. 2017, 25, 160–168. [Google Scholar] [CrossRef] [PubMed]
  47. Lied, B.; Roenning, P.A.; Sundseth, J.; Helseth, E. Anterior cervical discectomy with fusion in patients with cervical disc degeneration: A prospective outcome study of 258 patients (181 fused with autologous bone graft and 77 fused with a PEEK cage). BMC Surg. 2010, 10, 10. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  48. König, S.A.; Spetzger, U. Distractable titanium cages versus PEEK cages versus iliac crest bone grafts for the replacement of cervical vertebrae. Minim. Invas. Ther. Allied Technol. 2013, 23, 102–105. [Google Scholar] [CrossRef] [PubMed]
  49. Kao, T.-H.; Wu, C.-H.; Chou, Y.-C.; Chen, H.-T.; Chen, W.-H.; Tsou, H.-K. Risk factors for subsidence in anterior cervical fusion with stand-alone polyetheretherketone (PEEK) cages: A review of 82 cases and 182 levels. Arch. Orthop. Trauma. Surg. 2014, 134, 1343–1351. [Google Scholar] [CrossRef] [Green Version]
  50. Pelletier, M.H.; Cordaro, N.; Punjabi, V.M.; Waites, M.; Lau, A.; Walsh, W.R. PEEK Versus Ti Interbody Fusion Devices: Resultant Fusion, Bone Apposition, Initial and 26-Week Biomechanics. Clin. Spine Surg. 2016, 29, E208–E214. [Google Scholar] [CrossRef] [Green Version]
  51. Olivares-Navarrete, R.; Hyzy, S.L.; Gittens, R.A. Rough titanium alloys regulate osteoblast production of angiogenic factors. Spine J. 2013, 13, 1563–1570. [Google Scholar] [CrossRef] [Green Version]
  52. Sagomonyants, K.B.; Jarman-Smith, M.L.; Devine, J.N.; Aronow, M.S.; Gronowicz, G.A. The in vitro response of human osteoblasts to polyetheretherketone (PEEK) substrates compared to commercially pure titanium. Biomaterials 2008, 29, 1563–1572. [Google Scholar] [CrossRef]
  53. Chen, Y.; Wang, X.; Lu, X.; Yang, L.; Yang, H.; Yuan, W.; Chen, D. Comparison of titanium and polyetheretherketone (PEEK) cages in the surgical treatment of multilevel cervical spondylotic myelopathy: A prospective, randomized, control study with over 7-year follow-up. Eur. Spine J. 2013, 22, 1539–1546. [Google Scholar] [CrossRef] [Green Version]
  54. Niu, C.C.; Liao, J.C.; Chen, W.J.; Chen, L.H. Outcomes of interbody fusion cages used in 1 and 2-levels anterior cervical discectomy and fusion: Titanium cages versus polyetheretherketone (PEEK) cages. J. Spinal Disord. Tech. 2010, 23, 310–316. [Google Scholar] [CrossRef] [PubMed]
  55. Basheer, A.; Macki, M.; La Marca, F. Options for interbody grafting. In De-generative Cervical Myelopathy and Radiculopathy; Kaiser, M., Haid, R., Shaffrey, C., Fehlings, M.G., Eds.; Springer: Berlin/Heidelberg, Germany, 2019; pp. 309–318. [Google Scholar]
  56. Lee, J.H.; Jang, H.L.; Lee, K.M.; Baek, H.R.; Jin, K.; Hong, K.S.; Noh, J.H.; Lee, H.K. In vitro and in vivo evaluation of the bioactivity of hydroxyapatite-coated polyetheretherketone biocomposites created by cold spray technology. Acta Biomater. 2013, 9, 6177–6187. [Google Scholar] [CrossRef] [PubMed]
  57. Devine, D.M.; Hahn, J.; Richards, R.G.; Gruner, H.; Wieling, R.; Pearce, S.G. Coating of carbon fiber-reinforced polyetheretherketone implants with titanium to improve bone apposition. J. Biomed Mater. Res. Appl Biomater. 2013, 101, 591–598. [Google Scholar] [CrossRef] [PubMed]
  58. Yao, C.; Storey, D.; Webster, T.J. Nanostructured metal coatings on polymers increase osteoblast attachment. Int. J. Nanomed. 2007, 2, 487–492. [Google Scholar]
  59. Kienle, A.; Graf, N.; Hans-Joachim, W. Does impaction of titanium-coated interbody fusion cages into the disc space cause wear debris or delamination? Spine J. 2016, 16, 235–242. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  60. Slosar, P.J. Spine Implant Surface Technology State of the Art. Spine 2018, 43, S10–S11. [Google Scholar] [CrossRef]
  61. Vyatskikh, A.; Delalande, S.; Kudo, A.; Zhang, X.; Portela, C.M.; Greer, J.R. Additive manufacturing of 3D nano-architected metals. Nat. Commun. 2018, 9, 1–8. [Google Scholar] [CrossRef] [Green Version]
  62. Katsuura, Y.; Wright-Chisem, J.; Wright-Chisem, A.; Virk, S.; McAnany, S. The Importance of Surface Technology in Spinal Fusion. HSS J. 2020, 16, 113–116. [Google Scholar] [CrossRef] [Green Version]
  63. Olivares-Navarrete, R.; Hyzy, S.L.; Slosar, P.J.; Schneider, J.M.; Schwartz, Z.; Boyan, B.D. Implant materials generate different peri-implant inflammatory factors: Poly-ether-ether-ketone promotes fibrosis and microtextured titanium promotes osteogenic factors. Spine 2015, 40, 399–404. [Google Scholar] [CrossRef]
  64. Polini, A.; Pisignano, D.; Parodi, M.; Quarto, R.; Scaglione, S. Osteoinduction of Human Mesenchymal Stem Cells by Bioactive Composite Scaffolds without Supplemental Osteogenic Growth Factors. PLoS ONE 2011, 6, e26211. [Google Scholar] [CrossRef]
  65. Kandziora, F.; Pflugmacher, R.; Scholz, M.; Eindorf, T.; Schnake, K.J.; Haas, N.P. Bioabsorbable Interbody Cages in a Sheep Cervical Spine Fusion Model. Spine 2004, 29, 1845–1855. [Google Scholar] [CrossRef] [PubMed]
  66. Cao, L.; Duan, P.G.; Li, X.-L.; Yuan, F.-L.; Zhao, M.-D.; Che, W.; Wang, H.-R.; Dong, J. -R.; Dong, J. Biomechanical stability of a bioabsorbable self-retaining polylactic acid/nano-sized β-tricalcium phosphate cervical spine interbody fusion device in single-level anterior cervical discectomy and fusion sheep models. Int. J. Nanomed. 2012, 7, 5875–5880. [Google Scholar] [CrossRef]
  67. Kim, H.M.; Himeno, T.; Kokubo, T.; Nakamura, T. Process and kinetics of bonelike apatite formation on sintered hydroxyapatite in a simulated body fluid. Biomaterials 2005, 26, 4366–4373. [Google Scholar] [CrossRef]
  68. Hasegawa, T.; Inufusa, A.; Imai, Y.; Mikawa, Y.; Lim, T.-H.; An, H.S. Hydroxyapatite-coating of pedicle screws improves resistance against pull-out force in the osteoporotic canine lumbar spine model: A pilot study. Spine J. 2005, 5, 239–243. [Google Scholar] [CrossRef]
  69. Wixted, C.M.; Peterson, J.R.; Kadakia, R.J.; Adams, S.B. Three-dimensional Printing in Orthopaedic Surgery: Current Applications and Future Developments. JAAOS Glob. Res. Rev. 2021, 5, e20.00230-11. [Google Scholar] [CrossRef] [PubMed]
  70. Amelot, A.; Colman, M.; Loret, J.-E. Vertebral body replacement using patient-specific three–dimensional-printed polymer implants in cervical spondylotic myelopathy: An encouraging preliminary report. Spine J. 2018, 18, 892–899. [Google Scholar] [CrossRef] [PubMed]
  71. Burnard, J.L.; Parr, W.C.H.; Choy, W.J.; Walsh, W.R.; Mobbs, R.J. 3D-printed spine surgery implants: A systematic review of the efficacy and clinical safety profile of patient-specific and off-the-shelf devices. Eur. Spine J. 2019, 29, 1248–1260. [Google Scholar] [CrossRef]
  72. McGilvray, K.C.; Easley, J.; Seim, H.B.; Regan, D.; Berven, S.H.; Hsu, W.K.; Mroz, T.E.; Puttlitz, C.M. Bony ingrowth potential of 3D-printed porous titanium alloy: A direct comparison of interbody cage materials in an in vivo ovine lumbar fusion model. Spine J. 2018, 18, 1250–1260. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  73. Mokawem, M.; Katzouraki, G.; Harman, C.L.; Lee, R. Lumbar interbody fusion rates with 3D-printed lamellar titanium cages using a silicate-substituted calcium phosphate bone graft. J. Clin. Neurosci. 2019, 68, 134–139. [Google Scholar] [CrossRef] [Green Version]
  74. Mobbs, R.J.; Coughlan, M.; Thompson, R.; Sutterlin, C.E.; Phan, K. The utility of 3D printing for surgical planning and patient-specific implant design for complex spinal pathologies: Case report. J. Neurosurg. Spine 2017, 26, 513–518. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  75. Mobbs, R.J.; Parr, W.C.H.; Choy, W.J.; McEvoy, A.; Walsh, W.R.; Phan, K. Anterior Lumbar Interbody Fusion Using a Personalized Approach: Is Custom the Future of Implants for Anterior Lumbar Interbody Fusion Surgery? World Neurosurg. 2019, 124, 452–458.e1. [Google Scholar] [CrossRef] [PubMed]
  76. Phan, K.; Sgro, A.; Maharaj, M.M.; D’Urso, P.; Mobbs, R.J. Application of a 3D custom printed patient specific spinal implant for C1/2 arthrodesis. J. Spine Surg. 2016, 2, 314. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  77. Culler, S.D.; Martin, G.M.; Swearingen, A. Comparison of adverse events rates and hospital cost between customized individually made implants and standard off-the-shelf implants for total knee arthroplasty. Arthroplast. Today 2017, 3, 257–263. [Google Scholar] [CrossRef] [PubMed] [Green Version]
Figure 1. Lateral X-ray of the lumbar spine. (A) Demonstrating L2–S1 anterior interbody fusion using a combination of PEEK and titanium cages that is supported posteriorly with pedicle screw fixation. One year post-op CT sagittal (B) and coronal slices (C) demonstrate bridging callus between the interbody spaces showing successful fusion with osseointegration.
Figure 1. Lateral X-ray of the lumbar spine. (A) Demonstrating L2–S1 anterior interbody fusion using a combination of PEEK and titanium cages that is supported posteriorly with pedicle screw fixation. One year post-op CT sagittal (B) and coronal slices (C) demonstrate bridging callus between the interbody spaces showing successful fusion with osseointegration.
Bioengineering 09 00108 g001
Figure 2. Cervical spine X-ray demonstrating a C5–C7 anterior discectomy and fusion using titanium interbodies with plate fixation.
Figure 2. Cervical spine X-ray demonstrating a C5–C7 anterior discectomy and fusion using titanium interbodies with plate fixation.
Bioengineering 09 00108 g002
Figure 3. Cervical spine X-ray demonstrating polyether ether ketone (PEEK) interbody after a C5–C6 anterior cervical discectomy and fusion.
Figure 3. Cervical spine X-ray demonstrating polyether ether ketone (PEEK) interbody after a C5–C6 anterior cervical discectomy and fusion.
Bioengineering 09 00108 g003
Figure 4. (A) 3D printed titanium cage (B) with interbody placement. Printed with permission from Medtronic PLC (Minneapolis, MN, USA).
Figure 4. (A) 3D printed titanium cage (B) with interbody placement. Printed with permission from Medtronic PLC (Minneapolis, MN, USA).
Bioengineering 09 00108 g004
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Kia, C.; Antonacci, C.L.; Wellington, I.; Makanji, H.S.; Esmende, S.M. Spinal Implant Osseointegration and the Role of 3D Printing: An Analysis and Review of the Literature. Bioengineering 2022, 9, 108. https://0-doi-org.brum.beds.ac.uk/10.3390/bioengineering9030108

AMA Style