Next Article in Journal
FLT PET Radiomics for Response Prediction to Chemoradiation Therapy in Head and Neck Squamous Cell Cancer
Previous Article in Journal
[18F] FDG Positron Emission Tomography (PET) Tumor and Penumbra Imaging Features Predict Recurrence in Non–Small Cell Lung Cancer
 
 
Tomography is published by MDPI from Volume 7 Issue 1 (2021). Previous articles were published by another publisher in Open Access under a CC-BY (or CC-BY-NC-ND) licence, and they are hosted by MDPI on mdpi.com as a courtesy and upon agreement with Grapho, LLC.
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Bias in PET Images of Solid Phantoms Due to CT-Based Attenuation Correction

by
Darrin W. Byrd
1,
John J. Sunderland
2,
Tzu-Cheng Lee
1 and
Paul E. Kinahan
1,*
1
Department of Radiology, University of Washington, Box 357987, Seattle, WA 98195, USA
2
Department of Radiology, University of Iowa, Iowa City, IA, USA
*
Author to whom correspondence should be addressed.
Submission received: 7 December 2018 / Revised: 11 January 2019 / Accepted: 4 February 2019 / Published: 1 March 2019

Abstract

The use of computed tomography (CT) images to correct for photon attenuation in positron emission tomography (PET) produces unbiased patient images, but it is not optimal for synthetic materials. For test objects made from epoxy, image bias and artifacts have been observed in well-calibrated PET/CT scanners. An epoxy used in commercially available sources was infused with long-lived 68Ge/68Ga nuclide and measured on several PET/CT scanners as well as on older PET scanners that measured attenuation with 511-keV photons. Bias in attenuation maps and PET images of phantoms was measured as imaging parameters and methods varied. Changes were made to the PET reconstruction to show the influence of CT-based attenuation correction. Additional attenuation measurements were made with a new epoxy intended for use in radiology and radiation treatment whose photonic properties mimic water. PET images of solid phantoms were biased by between 3% and 24% across variations in CT X-ray energy and scanner manufacturer. Modification of the reconstruction software reduced bias, but object-dependent changes were required to generate accurate attenuation maps. The water-mimicking epoxy formulation showed behavior similar to water in limited testing. For some solid phantoms, transformation of CT data to attenuation maps is a major source of PET image bias. The transformation can be modified to accommodate synthetic materials, but our data suggest that the problem may also be addressed by using epoxy formulations that are more compatible with PET/CT imaging.
Keywords: PET; calibration; phantoms PET; calibration; phantoms

Share and Cite

MDPI and ACS Style

Byrd, D.W.; Sunderland, J.J.; Lee, T.-C.; Kinahan, P.E. Bias in PET Images of Solid Phantoms Due to CT-Based Attenuation Correction. Tomography 2019, 5, 154-160. https://0-doi-org.brum.beds.ac.uk/10.18383/j.tom.2018.00043

AMA Style

Byrd DW, Sunderland JJ, Lee T-C, Kinahan PE. Bias in PET Images of Solid Phantoms Due to CT-Based Attenuation Correction. Tomography. 2019; 5(1):154-160. https://0-doi-org.brum.beds.ac.uk/10.18383/j.tom.2018.00043

Chicago/Turabian Style

Byrd, Darrin W., John J. Sunderland, Tzu-Cheng Lee, and Paul E. Kinahan. 2019. "Bias in PET Images of Solid Phantoms Due to CT-Based Attenuation Correction" Tomography 5, no. 1: 154-160. https://0-doi-org.brum.beds.ac.uk/10.18383/j.tom.2018.00043

Article Metrics

Back to TopTop