Review

Non-Specific Adsorption Reduction Methods in Biosensing

1
Department of Electrical and Computer Engineering, School of Engineering, Baylor University, Waco, TX 76798, USA
2
Department of Mechanical Engineering, School of Engineering, Baylor University, Waco, TX 76798, USA
3
Department of Electrical and Computer Engineering, School of Engineering, Baylor University, Waco, TX 76798, USA
*
Author to whom correspondence should be addressed.
Submission received: 1 May 2019 / Revised: 24 May 2019 / Accepted: 27 May 2019 / Published: 31 May 2019
(This article belongs to the Section Biosensors)

Abstract

:
Non-specific adsorption (NSA) is a persistent problem that negatively affects biosensors, decreasing sensitivity, specificity, and reproducibility. Passive and active removal methods exist to remedy this issue, by coating the surface or generating surface forces to shear away weakly adhered biomolecules, respectively. However, many surface coatings are not compatible or effective for sensing, and thus active removal methods have been developed to combat this phenomenon. This review aims to provide an overview of methods of NSA reduction in biosensing, focusing on the shift from passive methods to active methods in the past decade. Attention is focused on protein NSA, due to their common use in biosensing for biomarker diagnostics. To our knowledge, this is the first review to comprehensively discuss active NSA removal methods. Lastly, the challenges and future perspectives of NSA reduction in biosensing are discussed.

1. Introduction

Creating sensitive, selective, and stable biosensors for the timely identification of disease biomarkers is greatly important [1]. In recent years, copious research has been focused on the development of new methods for early disease detection [1]. Numerous in-depth reviews on biosensors exist for readers of interest [2,3,4,5]. The goal for a biosensor is to achieve the maximum sensitivity, specificity and selectivity for a particular application [6]. In order to accomplish this, both the sensor element and the neighboring surfaces have to be unreactive to non-specific adsorption [6]. This is imperative for limiting the attachment of target molecules to the sensitive areas of the sensor by subsequent specific functionalization [6]. Especially for small-scale sensors, it is crucial to accomplish controlled patterning of functional molecules. This is because the size of the molecules used for passivation and capture, as well as the analytes of interest, have similar dimensions to the sensor element or at least comparable to the size of the sensitive area [6].
Surface-based sensing is a common method of biosensing, exemplified by immunosensors (e.g., ELISA and SPR), microfluidic biosensors, and electrochemical biosensors. A model biosensor contains functional areas coinciding with the sensing area and inactive areas everywhere else to promote minimum sample consumption [6]. However, biomolecular areas of surface-based sensors often come into contact with complex mixtures of proteins and other molecules during their use [7]. Most surfaces are particularly prone to non-specific and irreversible adsorption of proteins, known as non-specific adsorption (NSA) [7,8,9,10]. NSA happens when a molecule adsorbs to a sensor’s surface, resulting in high background signals that are indiscernible from the specific binding [9]. This phenomenon occurs because of physisorption and can decrease the sensor’s performance [1,9].
It is important to note that for some sensors, NSA can be distinguished from the analyte signal. This is the case for biodetection with attenuated total internal reflection Fourier transform infrared (ATR-FTIR) or ellipsometry. ATR-FTIR spectroscopy have been used for biotin detection, where the specific binding was able to be differentiated from the non-specific protein binding [11]. Ellipsometry, which measures the change of polarization upon reflection or transmission, has been used to study in vitro polyclonal antibody adsorption [12]. This technique can be used to identify the specifically adsorbed proteins from a complex solution. However, this method is not successful for monoclonal antibodies, and requires an optically reflective, flat surface. Infrared (IR) ellipsometry has been demonstrated as a label-free imaging method to map different materials on a biochip [13]. Compared to conventional chemiluminescence imaging, this method was found to be able to map larger sample areas with greater resolution (300 μm × 300 μm), attained by using radiation from an IR synchrotron beamline. While imaging techniques are useful for discriminating between specific and non-specific adsorption, their implantation is not applicable to all sensors. A majority of surface-based sensors suffer from the effects of NSA, particularly microfluidic biosensors.
Microfluidics is the field of fluid handling technology involving very small volumes of liquids in the range of 10−6 to 10−18 L [4]. Microfluidics have numerous applications including biomedicine, cell analysis, drug screening, cell biology, and biosensing [4]. In recent years, momentous work has been done to reduce the size of conventional biosensors using microfluidics for lab-on-a-chip practices [8]. Numerous benefits exist for using microfluidics such as the low consumption of costly reagents, minimal handling of hazardous materials, short reaction time, multiple sample detection, increased portability and design versatility [8]. Decreasing the sensor dimensions also enhances signal-to-noise ratio by increasing the signal density and reducing background signals [14]. Literature on the design and applications of microfluidic biosensors is vast [5,15,16,17,18]. Microfluidic biosensors often have immobilized bioreceptors, such as antibodies [19], enzymes [20], DNA [21], and linker molecules such as SAMs (self-assembled monolayers) to improve surface immobilization. However, the linker molecules are prone to NSA, resulting in decreasing sensitivity and false responses [8].
The interaction between the bioreceptor and analyte can be categorized into affinity or catalytic [4]. Affinity biosensors are based on the binding affinity between the analyte and bioreceptor (e.g., antibody-antigen). In catalytic biosensors a chemical reaction produces a product which can be correlated with the concentration of target analyte [4]. An example of a catalytic sensor is a glucose biosensor which uses the enzyme glucose oxidase that catalyzes the oxidation of glucose to hydrogen peroxide. This review focuses more on the NSA of affinity biosensors due to their prevalence in diagnostic biomarker protein detection.
The balance between simultaneous sensing and NSA removal is a key obstacle in biosensing [22]. Numerous methods exist to reduce non-specific interactions on sensing surfaces. Blocking methods aim to reduce NSA by coating the surface, while removal methods generate surface shear forces to overpower the adhesive forces of the non-specifically adsorbed molecules [9]. Physical removal methods use a transducer, and are typically categorized as electromechanical or acoustic devices. Another less commonly used technique is hydrodynamic removal, which relies solely on the fluid flow to generate the shear forces.

2. Non-Specific Adsorption

Adsorption is the adhesion of atoms, ions or molecules from a gas, liquid or dissolved solid to a surface [23]. There are two types of adsorption: physical and chemical. Physical adsorption, or physisorption, results from intermolecular forces including hydrophobic forces, ionic interactions, van der Waals forces, and hydrogen bonding. Chemical adsorption, or chemisorption, results from chemical interactions such as covalent binding. In NSA (Figure 1), molecules are adsorbed to a surface through the weaker adsorption, physisorption [9].
Non-specific adsorption (NSA), also known as non-specific binding or biofouling is a persistent challenge in biosensing [5,8,25,26,27]. While NSA is an issue in other biological fields including implantable biomedical devices [28,29,30], and marine equipment [31,32], this report will focus on NSA for biosensing applications. Most biomolecular surfaces, whether comprised of antibodies, enzymes or proteins, experience the common issue of hindrance from non-specific species [25]. For biosensors, the molecules typically adsorb to the sensing surface from a liquid medium.
Immunosensors, a common type of biosensor, use antibodies or antigens as the bioreceptor. For antibodies, there are two types of non-specific adsorption: immunological and methodological. Immunological has more to do with the affinity between the antibody and antigens, and cannot be improved without using different proteins. Methodological non-specificity can occur due to a variety of reasons, but is usually a combination of protein-protein interactions, surface protein denaturation or mis-orientation, substrate stickiness, non-specific electrostatic binding to charged surface, and adsorption of molecules in free spaces [25]. This phenomenon can result in four different types of NSA: (1) molecules adsorbed on vacant spaces (2) molecules adsorbed on non-immunological sites (3) molecules adsorbed on immunological sites, allowing access to antigens and (4) molecules adsorbed on immunological sites. Non-specific adsorption leads to elevated background signals that cannot be discriminated from specific binding [9]. These false-positive signals affect the dynamic range [25], limit of detection [2,9], reproducibility, selectivity, and sensitivity [9]. The reduction of NSA is crucial in the development of biosensors, especially for point-of-care clinical diagnostics [5].
Limiting NSA has been investigated for decades [33]. Recent reviews focus on NSA reduction for certain sensor types such as the suppression of protein NSA in the development of electrochemical immunosensors [26], antifouling materials for SPR sensors [34], nanoscale sensors [6], and whispering gallery mode optical biosensors [35]. With the increased trend in micro/nano-scale biosensors, a shift from passive methods to active removal methods is noticed in literature, which is the focus of this review.

3. Methods of NSA Reduction

Non-specific adsorption can be reduced by a number of approaches with varying complexity [7]. NSA reduction methods can be generally divided into two categories: passive methods and active methods (Figure 2). The passive methods aim to prevent undesired adsorption by coating the surface, and can broadly be sub-categorized as physical and chemical. These methods have existed for several decades and are vast in the literature. In contrast, active methods dynamically remove adsorption post-functionalization and are a more recent technique. Active removal methods can be further categorized as transducer-based or fluid-based. The transducers used are typically electromechanical or acoustic. Fluid-based methods utilize the pressure-driven flow of microfluidics to shear away biomolecules. This review will briefly describe the passive methods, which have been thoroughly described in the literature already [8,25,26,31,36], while focusing on the active methods which have yet to be thoroughly investigated.

3.1. Passive Methods

Blocking methods can be categorized into two broad categories: physical (i.e., protein blockers) and chemical (i.e., linker molecules). A recent review has provided an in-depth review of physical and chemical surface modification methods [6]. Therefore, here we will only briefly summarize the main mechanisms in this review and note key examples. Passive methods aim to create a thin hydrophilic and non-charged boundary layer to thwart protein adsorption [9]. The goal of anti-fouling coatings is to minimize the intermolecular forces and interactions between the adsorbing molecules and the host substrate so the molecules can be easily detached and released under low shear stresses, like washing [37]. Materials for non-fouling coatings are usually neutral or weakly negative and well hydrated [36].

3.1.1. Physical

Physical surface modification does not typically change the chemical composition of the surface. The most popular and easiest method to prevent NSA is to use blocker proteins that adsorb to surfaces. For example, serum albumins (e.g., BSA) [38,39], casein [39,40], and other milk proteins [36] are commonly used as blocking agents for enzyme-linked immunosorbent assay (ELISA), Western blotting, and other enzyme based assays [39,41,42]. The underlying assumption of these methods is the idea that a controlled patterning of the surface-active protein will inhibit the immobilization of other molecules to the substrate [7]. Commonly used bovine serum albumin (BSA), protein derived from cows, is easily accessible and inexpensive [38]. However, the usage of BSA often leads to a non-uniform layer on the surface, and studies have reported proteins can still adsorb to a BSA layer [7]. Additionally, while protein blockers can be applied to a variety of surface materials, they can have high lot-to-lot variability, cross-reactivity [43] and alter the original surface properties [44].
For catalytic biosensors, a permselective membrane, with preferential permeation of certain ionic species through ion-exchange membranes, can be used to reduce NSA [14,45,46]. In an amperometric glucose sensor, the ion permselectivity avoids interfaces of the detection of hydrogen peroxide by ascorbate, urate and acetaminophen [14,47]. Additionally, the use of a permselective membrane can reduce the influence of buffer concentration on the sensor signal [48]. Another method of dealing with protein NSA utilizes the Vroman effect [14,49,50]. This is when a weak affinity protein bound to the surface is displaced by a strong affinity protein [17]. The reverse operation does not occur. This phenomenon is due to the fact that protein adsorption capability is largely dependent on the molecular weight [51]. Therefore, in general, a high molecular weight protein adsorbs more strongly than a low molecular weight protein. The displacement of these proteins can be measured by SPR [52].

3.1.2. Chemical

Chemical methods of anti-fouling include using a poly(ethylene glycol) (PEG)/oligo(ethylene glycol) (OEG) based coating [9,37,53,54], functionalized self-assembled monolayers (SAMS) [8,9,55,56,57,58], or zwitterionic polymers [1,9,36,59,60,61]. PEG is a non-ionic, water soluble polymer commonly used to form protein-resistant layers [54]. PEG and OEG have been widely used antifouling materials due to their surface hydration [1], and weakly basic ether linkages [37]. However, they are vulnerable under oxidative conditions, which limits their efficacy for long-term applications [1,54]. Additionally, the auto-oxidation of the terminal hydroxyl group in PEG into aldehydes can deteriorate the specific proteins used for sensing [62]. Some types of PEG are only useful for negatively-charged surfaces, and are most successful with metal oxide surfaces [7]. These compounds are better suited for single proteins, but most often fail in complex mixtures of proteins (e.g., blood plasma [63,64]), or under in vivo conditions [65]. Furthermore, the density and chain length of PEG need to be optimized for minimizing non-specific interactions [53]. The ability of PEG to prevent NSA depends on other parameters, such as temperature [64], and can be unstable [66]. Researchers have used nanostructured surface coatings to reduce protein adsorption as an alternative to polymer modification [67].
Self-assembled monolayers (SAMs) offer an oligoethylene glycol functional group at the solid-water interface [7]. Alkanethiol SAMs are one of the most popularly used bioreceptor linker molecules, due to the two major advantages: plentiful functional groups to immobilize bioreceptors and a dense monolayer that allows little NSA [8]. Sensing surfaces with alkanethiol SAMs suffer from NSA due to three main reasons: polycrystalline gold grain structure, imperfectly formed monolayers, and non-immobilized bioreceptors [8]. Additionally, factors such as surface roughness, annealing, and the use of short or long chain SAMs can affect the results [8]. Furthermore, oligoethylene glycol-terminated SAMs are only compatible with gold and silver surfaces [7], which limits their use considering the variety of substrates used today in biosensing [7,62].
Zwitterion polymers or polyzwitterions are polymers made up of zwitterions, molecules which have both a positive and negative charge [61]. Zwitterion polymers have the same number of anionic and cationic groups, so the overall charge is zero under normal conditions. These ionic groups repel unwanted adhesion with a strong hydration layer through ionic solvation [1,37] and are functional over a large pH window [61]. Zwitterionic polymer brushes may be grafted to or grafted from surface. However, most zwitterionic reagents that can be presented onto substrate surfaces for NSA reduction are usually restricted to betaine and phosphorylcholine related chemicals, whose synthetic processes are slow and difficult [1]. A promising subclass of zwitterionic polymers is polyampholyte polymers, composed of mixtures of charged monomer subunits, whose mechanical properties of the polymers can be tailored. Surface grafted polyampholyte polymer brushes have been demonstrated as an anti-fouling coating for sensing applications in complex media [68].
Challenges with chemical methods include laborious functionalization processes, long-term chemical stability, raising the sensor background signal [9], and the prospect to damage active surface [9,62]. Additionally, some chemical methods have limitations in transducer interfaces as previously described [7,9]. The prevention of NSA with passivation methods often involves the use of harsh chemicals that are normally not appropriate for many biological uses [7].
Manipulating surface chemistry is only part of controlling protein adsorption. The surface topography is key in determining the structure of bound proteins [69] and can be used to control specific protein adsorption [70]. Alternative methods to control NSA is through the patterning or imprinting the surface with a unique surface topography or synthetic polymer. Shi et al. reported a method of imprinting surfaces with nanocavities to specifically recognize template proteins [70]. The authors suggest non-template proteins will be less likely to attach to the imprinted sites because they lack the ability to interlock in the cavities and form a strong bond. The imprinted surface topography was evaluated using tapping-mode atomic force microscopy (AFM). They evaluated three imprinted blood proteins (BSA, IgG, and fibrinogen) compared to a flat control surface, and found a higher number of proteins were adsorbed on the imprints. After using a detergent to clean the surface, the researchers found the protein imprint retained more of the proteins than the control surface, implying a greater affinity at the imprinted cavities. Specific binding sites can be limited to the nanocavities, while elsewhere is blocked with BSA passivation. Roach et al. investigated the effect of surface topography on the binding of proteins [69]. They found differences in how certain protein geometries were affected by surface structure. For example, a globular protein like albumin, is more compatible with a surface with high curvature. However, a rod-like protein such as fibrinogen, is distorted by a curved surface, which promotes structure loss. These effects are important when considering how to optimize specific adsorption and prevent NSA [69].
A synthetic recognition system can also be implemented as an alternative to the typical biological receptor used in sensing [71], such as molecularly imprinted polymers (MIPs). This synthetic recognition system is inexpensive, easy to implement, and allows target species to rebind to the imprinted sites. However, challenges exist the uncontrolled random polymerization produces uneven imprinted sites. Additionally, the use of MIPs is limited to chemical detection or bioassays due to the inability to generate signal output at the surface of the transducer. Molecularly imprinted nanomaterials, such as nanowires, nanotubes, nanofibers have been investigated for chemosensors as a solution to the challenges faced by traditionally imprinted materials. While improvement in achieving sensitive detection has been achieved with these methods, there are still limitations to this method including the need to enhance the specific binding, reduce NSA, and allow for multiplexing and easy integration [71]. Due to these limitations, active methods have emerged as an alternative to combat NSA.

3.2. Active Methods

Methods of active NSA removal have emerged in the past decade as a promising approach to reduce NSA in biosensors. These methods produce surface shear forces, which are stronger than the adhesive force of the weakly bound NSA molecules. Meanwhile, the specific proteins are not removed because the affinity of specific proteins toward a ligand is several orders of magnitude larger than that of non-specific proteins [22,72]. Transducer-based approaches can be used for numerous types of proteins, making them time-saving, easy operating, and efficient methods for removing NSA [9]. Transducer devices to remove NSA can be categorized as electromechanical or acoustic (Figure 2) [9]. A summary table of active NSA removal methods is provided in Table 1.

3.2.1. Transducer-Based

The shearing forces in electromechanical methods are generated from the electrical or mechanical transducers. Some examples of electromechanical methods include ac-EHD (alternating current electrohydrodynamics) induced nanoshearing [10,73,74,75,76], hypersonic resonance [9], and resonant cantilever vibration [44].
A sensor has been developed that uses alternating current electrohydynamics (ac-EHD) induced surface shear forces, or nanoshearing, to improve the analyte capture of the sensor and remove weakly (non-specifically) bound molecules from the electrode surface [10]. The working principle is that the ac-EHD induced flow generates shearing forces within nanometer distances of the electrode surface to produce lateral fluid flow (Figure 3a). This not only mixes the fluid around the sensor surface to enhance capture efficiency through increased sensor-target collisions, but the ac-electric field can enable selection of specifically bound proteins over non-specifically bound proteins (Figure 3b). This produced 100 fg·mL−1 naked eye detection of multiple protein targets spiked in human serum [10]. The results were compared to that of a pressure-driven flow system with the same flow rate. The authors found the absorbance measurements were 1000 times better for the ac-EHD system than the hydrodynamic. Additionally, they noted limitations of pressure-driven flow including the stationary boundary layer of fluid, which does not allow for surface shear force manipulation, and the external syringe pump needed that is not easily integrated into portable devices. This system is multiplexible via several individual channels for distinct functionalization of different bioreceptors. This device has been applied for sensing and biofouling of exosomes [75] as well as being incorporated with a surface-enhanced Raman scattering (SERS)-based immunoassay [74]. A complete look at ac-EHD use in microsystems has been recently published for the interested reader [84].
Since the shearing forces typically come from electrical or mechanical vibration of the devices, the NSA removal is limited to the proteins adsorbed on the transducer surface [9]. Pan et al. proposed a solution to this problem by using a microfabricated 2.5 GHz hypersonic resonator to generate microvortexes in the fluid, which results in the attenuation of acoustic energy into the liquid (Figure 4) [9]. The resulting drag and lift forces along the liquid-solid interfaces were used to control the NSA removal. The experimental results supported the mathematical and simulated models. The proteins were modeled as spherical proteins with a diameter of 100 nm, which correlates to a 500-mW power to remove them. However, if the proteins are 1–10 nm, the required power is very high, ~5 W. The mass-sensitive hypersonic resonator displayed a shift in the resonance frequency as each step was performed. However, frequency change cannot differentiate between specific or nonspecific binding. Therefore, fluorescent microscopy was used to verify the NSA removal with a specific immunoglobulin G (IgG) and nonspecific IgG. The quantitative testing showed an 83.8% fluorescence intensity decrease for the nonspecific binding sites after hypersonic treatment and only 6.1% with specific binding sites. Additionally, the vertical orientation of the vortexes allows a contactless setup, which enables the resonator to be combined with other surface based biosensors. The detachable actuator enables reuse. The authors determined the drag force was the primary force in removing NSA.
Johnson and Mutharasan demonstrated that transverse modes in electrochemical piezoelectric-excited millimeter-sized cantilevers (ePEMCs) caused the release of adsorbed bovine and human serum albumin proteins from sensor surfaces [44]. They first determined ePEMC transverse mode vibration does cause streaming and tested different excitations voltages and particle concentrations. Then, they tested the release of BSA and human serum albumin (HSA), and repeated the experiments with thiolated single-stranded DNA strands to Au sites, which has a 4 times higher binding energy than BSA. Videos of the experiments were recorded to calculate the streaming velocities and visualize the ePEMC vibration-induced flow. The results were determined through electrochemical and mass change sensing techniques, and further verified with wet chemical-based spectroscopic analysis. The study found the release of the NSA proteins was caused by a combination of surface strain energy, body forces, and acoustic streaming associated with hydrodynamic effects. Furthermore, this work could be applied to removing specific binding for bioseparation and diagnostic applications. Before this work, few studies had investigated the non-specific interactions of the non-target species [85]. In recent years, the use of micro cantilever-based sensing systems has increased due to their anticipated high sensitivity [85,86,87,88,89,90,91].
Acoustic wave sensors use acoustic energy to disturb the bonds between the sensing layer and analyte, coercing only the molecules with the higher affinity (i.e., specific proteins) to stay attached. The fluid motion induced from high-intensity sound waves is called acoustic streaming [62,92,93]. The produced sound fields cause tangential fluid motion along the fluid boundaries. The fluid motion applies a steady viscous stress on the boundary layer, causing liquid circulation near the boundaries. These stresses are strong enough to remove loosely bound molecules on the surface of the device. Acoustic streaming enables higher accuracy measurements and reusability of the devices [62]. Examples of acoustic transducer devices include surface acoustic wave (SAW) [62,77,78], orthogonal surface acoustic wave [21], longitudinal acoustic wave (LAW) [80], and bulk acoustic wave [81,82]. Detailed information on acoustic sensing has been recently published [94,95,96,97,98,99].
Combined electromechanical systems are under investigation too. The nanomolecule desorption mechanism was demonstrated using a combination of electric field and mechanical vibration to remove BSA and IgG from metal coated lead zirconate titanate (PZT) surfaces (Figure 5) [65]. Surface acoustic waves propagating on the surface of a piezoelectric device can be used to induce acoustic streaming within the fluid [62]. Previous studies demonstrated the ability of Rayleigh wave modes to release NSA proteins from the surfaces of SAW devices [78]. The acoustic waves generated from the piezoelectric coupling successfully induced fluid motion (acoustic streaming) and physically forced weakly bound non-specific protein species from their binding sites, improving the biosensor response [78]. The work was furthered by using fluid-solid interaction models to explore parameters for elevating induced acoustic-streaming velocities and forces to eradicate biofouling, while minimizing the effect of streaming-induced removal forces on the antibody sensing layer in immuno-SAW sensors [62]. Singh et al. reported multi-directional transducer-based SAW devices on a single piezoelectric platform produces the benefit of carefully using aspects which are particular to acoustic waves spreading along a given crystallographic orientation [22].
Hsu et al. used a piezoelectric transducer to produce varied amplitude and swept-frequency longitudinal acoustic waves remove non-specifically bound biomolecules (Figure 6) [81]. They used a sample chip composed of a diced glass slide chip (GSC) modified with gold nanoparticles (GNPs) and immobilized biotin as a probe. The absorbance spectra of the localized surface plasmon resonance (SPR) band were used for quantitative analysis of the removal efficiency. They demonstrated that even strongly bound streptavidin molecules were able to be removed from the biotin-functionalized GNP-GSC surface [81]. The study also aimed to investigate the regeneration of specific binding sites. The device was found to be reusable at least five times for regeneration, and the NSA removal ability was tested with four different adsorbing molecules: BSA, streptavidin, anti-biotin, and anti-fibrinogen. A range of voltages was tested to optimize removal efficiency for of the non-specific biomolecules. This non-contact system can remove non-specific or specifically bound molecules from the surface.

3.2.2. Fluid-Based

Fluid-based methods of NSA removal utilize the pressure-driven flow already integral in microfluidic lab-on-chip systems to remove NSA on the sensing surface along the bottom channel wall (Figure 7). Other applications of fluid manipulation in channels includes the use of hydrodynamic forces to manipulate membrane-bound proteins in a cell membrane [100] and for polystyrene sphere removal on quartz surfaces [101,102]. Another example is using a confining sheath fluid within a microfluidic channel to prevent NSA [103]. However, this method is limited in scope to adsorption-free T-sensor measurements to be made within the core of the flow channel [103] and is not applicable for surface-based sensing.
Hydrodynamic shearing is a transducer-free method of non-specific binding removal, not commonly found in literature. Li et al. applied this technique to effectively remove non-specifically bound carbon nanotubes (CNT) on a multiplexible biosensor after electrophoresis alignment using a 1X PBS solution shearing at 500 µL/min (Figure 8) [84]. The CNTs bridge across the two parallel electrodes to bind with targets. Due to the large hydrodynamic drag imposed by the cross-flow of the cylindrical CNTs with high aspect ratio, a critical hydrodynamic shear rate removed the non-target linkers of the aligned CNTs. Detection limits of 100 aM and 10 fM in pure samples for two ELISA tests: biotin/streptavidin and HER2 (Human epidermal growth factor receptor 2)/HER2 antibody were achieved. For both models, the dynamic range was tuned up to five orders of magnitude by increasing the CNT numbers, with high sensitivity and specificity. The researchers used biotin-streptavidin as the antigen-antibody pair to optimize the flow rate [84]. The key step in this process is the DC electrophoresis and AC dielectrophoresis to align the CNTs before shearing. This requires gold electrodes fabricated into the device and additional time (20 min) for the CNTs assembly. This system is proposed as an improved sensitivity ELISA assay, with a cost estimate of $20 USD per test.

4. Discussion

In summarizing the various methods of NSA reduction in sensing, the main disadvantage of the passive methods is their inconsistency and tedious processes required. The main disadvantage of the transducer methods is that they require extra and equipment. The main disadvantage of the fluid-based methods is the precise fluid manipulation. Ultimately, there is not a single ideal method of NSA reduction for all sensors. However, we suggest the optimal method to be chosen based on several factors: (1) Integration, (2) Multiplexibility, and (3) Tunability. First, the method of NSA reduction should be easily incorporated with the existing sensor structure. Note that many of the active removal methods are designed for microfluidic devices [10] while the passive methods could be applied to most surface-based sensors. Second, the method ideally can be applied to different sensing areas, since most sensors have or strive to have multiplexibility. Third, the method should be able to be adjusted for different experimental conditions, i.e., analyte/receptor combinations or device size. Other factors like ease of operation, and cost will play a role too.
When comparing biosensor performance, it is important to recognize the sensitivity depends on the analyte, recognition molecule, and surface immobilization method used [85]. The non-specific adsorption is highly dependent on all of those things, and repeatable results are of course necessary for successful demonstration of an NSA reduction method. Also, the authors point out one key limitation to measuring the efficiency of NSA removal methods is how researchers evaluate their devices’ anti-fouling ability. The most common method for evaluating adsorption is to compare the residual material left on the device after a typical use [105]. While this is direct and simple, it ignores transient adsorption. Hawkins et al. present a solution to this issue, developing a microfluidic device for transient monitoring of adsorption [105].

5. Conclusions and Future Trends

Non-specific adsorption is a critical problem in biosensing. While various methods exist to reconcile the problem, none of them are sufficient to create an ideal biosensor. In this review, we discussed the various passive and active methods of NSA reduction for biosensors. There is a trend present in the past decade of transitioning from surface chemistry NSA prevention to actively removing using transducer-based or fluid-based methods. A comprehensive table summarizing the reported methods of active NSA removal in biosensing and criteria for choosing a method of NSA reduction was given. Despite the extensive published literature, several important areas remain relatively unexplored. The first is investigating the use of various bioreceptor–analyte combinations in conjunction with transducer-based active NSA methods. The second is the ability of the NSA reduction method to not only limit NSA, but also improve specific binding, which a few studies claim. The last is combining active and passive NSA reduction methods in hopes of achieving further improved sensitivity. The research on methods to combat non-specific adsorption will likely continue to grow with the ever-expanding field of biosensor development and the increase in miniaturization technologies. With the shift from passive methods to active removal methods, the ability to use the more complex methods of NSA removal for a multitude of sensor types should be explored further.

Author Contributions

Writing—original draft preparation, J.Y.L.; Writing—review and editing, Y.L. and S.K.

Funding

This work was supported in part by NSF Grant Nos. CBET-1706472 and ECCS-1707049.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Wang, G.; Han, R.; Su, X.; Li, Y.; Xu, G.; Luo, X. Zwitterionic peptide anchored to conducting polymer PEDOT for the development of antifouling and ultrasensitive electrochemical DNA sensor. Biosens. Bioelectron. 2017, 92, 396–401. [Google Scholar] [CrossRef] [PubMed]
  2. Ronkainen, N.J.; Halsall, H.B.; Heineman, W.R. Electrochemical biosensors. Chem. Soc. Rev. 2010, 39, 1763. [Google Scholar] [CrossRef] [PubMed]
  3. Alhadrami, H.A. Biosensors: Classifications, medical applications, and future prospective. Biotechnol. Appl. Biochem. 2018, 65, 497–508. [Google Scholar] [CrossRef] [PubMed]
  4. Chen, S.; Shamsi, M.H. Biosensors-on-chip: A topical review. J. Micromech. Microeng. 2017, 27, 083001. [Google Scholar] [CrossRef]
  5. Choi, S.; Goryll, M.; Sin, L.Y.M.; Wong, P.K.; Chae, J. Microfluidic-based biosensors toward point-of-care detection of nucleic acids and proteins. Microfluid. Nanofluid. 2011, 10, 231–247. [Google Scholar] [CrossRef]
  6. Reimhult, E.; Höök, F. Design of Surface Modifications for Nanoscale Sensor Applications. Sensors 2015, 15, 1635–1675. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  7. Ratner, B.D.; Hoffman, A.S.; Schoen, F.J.; Lemons, J.E. Biomaterials Science: An Introduction to Materials in Medicine; Academic Press: Cambridge, MA, USA, 2012. [Google Scholar]
  8. Choi, S.; Chae, J. Methods of reducing non-specific adsorption in microfluidic biosensors. J. Micromech. Microeng. 2010, 20, 075015. [Google Scholar] [CrossRef]
  9. Pan, S.; Zhang, H.; Liu, W.; Wang, Y.; Pang, W.; Duan, X. Biofouling Removal and Protein Detection Using a Hypersonic Resonator. ACS Sens. 2017, 2, 1175–1183. [Google Scholar] [CrossRef]
  10. Vaidyanathan, R.; van Leeuwen, L.M.; Rauf, S.; Shiddiky, M.J.A.; Trau, M. A Multiplexed Device Based on Tunable Nanoshearing for Specific Detection of Multiple Protein Biomarkers in Serum. Sci. Rep. 2015, 5, 9756. [Google Scholar] [CrossRef] [Green Version]
  11. Schartner, J.; Güldenhaupt, J.; Mei, B.; Rögner, M.; Muhler, M.; Gerwert, K.; Kötting, C. Universal Method for Protein Immobilization on Chemically Functionalized Germanium Investigated by ATR-FTIR Difference Spectroscopy. J. Am. Chem. Soc. 2013, 135, 4079–4087. [Google Scholar] [CrossRef]
  12. Tengvall, P.; Lundström, I.; Liedberg, B. Protein adsorption studies on model organic surfaces: An ellipsometric and infrared spectroscopic approach. Biomaterials 1998, 19, 407–422. [Google Scholar] [CrossRef]
  13. Hinrichs, K.; Gensch, M.; Esser, N.; Schade, U.; Rappich, J.; Kröning, S.; Portwich, M.; Volkmer, R. Analysis of biosensors by chemically specific optical techniques. Chemiluminescence-imaging and infrared spectroscopic mapping ellipsometry. Anal. Bioanal. Chem. 2007, 387, 1823–1829. [Google Scholar] [CrossRef]
  14. Soleymani, L.; Li, F. Mechanistic Challenges and Advantages of Biosensor Miniaturization into the Nanoscale. ACS Sens. 2017, 2, 458–467. [Google Scholar] [CrossRef]
  15. Rackus, D.G.; Shamsi, M.H.; Wheeler, A.R. Electrochemistry, biosensors and microfluidics: A convergence of fields. Chem. Soc. Rev. 2015, 44, 5320–5340. [Google Scholar] [CrossRef]
  16. Luka, G.; Ahmadi, A.; Najjaran, H.; Alocilja, E.; DeRosa, M.; Wolthers, K.; Malki, A.; Aziz, H.; Althani, A.; Hoorfar, M. Microfluidics Integrated Biosensors: A Leading Technology towards Lab-on-a-Chip and Sensing Applications. Sensors 2015, 15, 30011–30031. [Google Scholar] [CrossRef] [Green Version]
  17. Prakash, S.; Pinti, M.; Bhushan, B. Theory, fabrication and applications of microfluidic and nanofluidic biosensors. Philos. Trans. Math. Phys. Eng. Sci. 2012, 370, 2269–2303. [Google Scholar] [CrossRef]
  18. Rivet, C.; Lee, H.; Hirsch, A.; Hamilton, S.; Lu, H. Microfluidics for medical diagnostics and biosensors. Chem. Eng. Sci. 2011, 66, 1490–1507. [Google Scholar] [CrossRef]
  19. Felix, F.S.; Angnes, L. Electrochemical immunosensors—A powerful tool for analytical applications. Biosens. Bioelectron. 2018, 102, 470–478. [Google Scholar] [CrossRef]
  20. Min, J.; Nothing, M.; Coble, B.; Zheng, H.; Park, J.; Im, H.; Weber, G.F.; Castro, C.M.; Swirski, F.K.; Weissleder, R.; et al. Integrated Biosensor for Rapid and Point-of-Care Sepsis Diagnosis. ACS Nano 2018, 12, 3378–3384. [Google Scholar] [CrossRef]
  21. Ramanavičius, A.; Ramanavičienė, A.; Malinauskas, A. Electrochemical sensors based on conducting polymer—Polypyrrole. Electrochim. Acta 2006, 51, 6025–6037. [Google Scholar] [CrossRef]
  22. Singh, R.; Sankaranarayanan, S.K.R.S.; Bhethanabotla, V.R. Orthogonal surface acoustic wave device based on langasite for simultaneous biosensing and biofouling removal. Appl. Phys. Lett. 2009, 94, 263503. [Google Scholar] [CrossRef]
  23. U.S. EPA. Glossary. BTSC. Available online: https://brownfieldstsc.org/glossary.cfm?q=1 (accessed on 15 April 2019).
  24. Zhang, Y.; Lyons, V.; Pappas, D. Fundamentals of affinity cell separations. Electrophoresis 2018, 39, 732–741. [Google Scholar] [CrossRef]
  25. Ahluwalia, A.; Giusto, G.; de Rossi, D. Non-specific adsorption on antibody surfaces for immunosensing. Mater. Sci. Eng. C 1995, 3, 267–271. [Google Scholar] [CrossRef]
  26. Contreras-Naranjo, J.E.; Aguilar, O. Suppressing Non-Specific Binding of Proteins onto Electrode Surfaces in the Development of Electrochemical Immunosensors. Biosensors 2019, 9, 15. [Google Scholar] [CrossRef]
  27. Ogi, H.; Fukunishi, Y.; Nagai, H.; Okamoto, K.; Hirao, M.; Nishiyama, M. Nonspecific-adsorption behavior of polyethylenglycol and bovine serum albumin studied by 55-MHz wireless–electrodeless quartz crystal microbalance. Biosens. Bioelectron. 2009, 24, 3148–3152. [Google Scholar] [CrossRef]
  28. Khalili, A.; Ahmad, M.R. A Review of Cell Adhesion Studies for Biomedical and Biological Applications. Int. J. Mol. Sci. 2015, 16, 18149–18184. [Google Scholar] [CrossRef] [Green Version]
  29. Ettelt, V.; Ekat, K.; Kämmerer, P.W.; Kreikemeyer, B.; Epple, M.; Veith, M. Streptavidin-coated surfaces suppress bacterial colonization by inhibiting non-specific protein adsorption. J. Biomed. Mater. Res. A 2018, 106, 758–768. [Google Scholar] [CrossRef]
  30. Kenna, J.G.; Major, G.N.; Williams, R.S. Methods for reducing non-specific antibody binding in enzyme-linked immunosorbent assays. J. Immunol. Methods 1985, 85, 409–419. [Google Scholar] [CrossRef]
  31. Salta, M.; Wharton, J.A.; Stoodley, P.; Dennington, S.P.; Goodes, L.R.; Werwinski, S.; Mart, U.; Wood, R.J.; Stokes, K.R. Designing biomimetic antifouling surfaces. Philos. Trans. R. Soc. Math. Phys. Eng. Sci. 2010, 368, 4729–4754. [Google Scholar] [CrossRef] [Green Version]
  32. Manov, D.V.; Chang, G.C.; Dickey, T.D. Methods for Reducing Biofouling of Moored Optical Sensors. J. Atmos. Ocean. Technol. 2004, 21, 958–968. [Google Scholar] [CrossRef] [Green Version]
  33. Brandtzaeg, F. Conjugates of Immunoglobulin G with Different Fluorochromes. Scand. J. Immunol. 1973, 2, 333–348. [Google Scholar] [CrossRef]
  34. Liu, B. Design and mechanisms of antifouling materials for surface plasmon resonance sensors. Acta Biomater. 2016, 40, 100–118. [Google Scholar] [CrossRef]
  35. Soteropulos, C.E.; Zurick, K.M.; Bernards, M.T.; Hunt, H.K. Tailoring the Protein Adsorption Properties of Whispering Gallery Mode Optical Biosensors. Langmuir 2012, 28, 15743–15750. [Google Scholar] [CrossRef]
  36. Schlenoff, J.B. Zwitteration: Coating Surfaces with Zwitterionic Functionality to Reduce Nonspecific Adsorption. Langmuir 2014, 30, 9625–9636. [Google Scholar] [CrossRef]
  37. Zhang, H.; Chiao, M. Anti-fouling Coatings of Poly(dimethylsiloxane) Devices for Biological and Biomedical Applications. J. Med. Biol. Eng. 2015, 35, 143–155. [Google Scholar] [CrossRef] [Green Version]
  38. Sheikh, S.; Blaszykowski, C.; Thompson, M. Sacrificial BSA to block non-specific adsorption on organosilane adlayers in ultra-high frequency acoustic wave sensing. Surf. Interface Anal. 2013, 45, 1781–1784. [Google Scholar] [CrossRef]
  39. Riquelme, M.V.; Zhao, H.; Srinivasaraghavan, V.; Pruden, A.; Vikesland, P.; Agah, M. Optimizing blocking of nonspecific bacterial attachment to impedimetric biosensors. Sens. Bio-Sens. Res. 2016, 8, 47–54. [Google Scholar] [CrossRef] [Green Version]
  40. Tacha, D.E.; McKinney, L. Casein Reduces Nonspecific Background Staining in Immunolabeling Techniques. J. Histotechnol. 1992, 15, 127–132. [Google Scholar] [CrossRef]
  41. Steinitz, M. Quantitation of the Blocking Effect of Tween 20 and Bovine Serum Albumin in ELISA Microwells. Anal. Biochem. 2000, 282, 232–238. [Google Scholar] [CrossRef]
  42. Muñoz-Berbel, X.; Godino, N.; Laczka, O.; Baldrich, E.; Muñoz, F.X.; del Campo, F.J. Impedance-Based Biosensors for Pathogen Detection. In Principles of Bacterial Detection: Biosensors, Recognition Receptors and Microsystems; Zourob, M., Elwary, S., Turner, A., Eds.; Springer: New York, NY, USA, 2008; pp. 341–376. [Google Scholar]
  43. Practical Tips of ELISA. Creative Diagnostics. Available online: https://www.creative-diagnostics.com/Practical-Tips-of-ELISA.htm (accessed on 25 April 2019).
  44. Mutharasan, B.N.J.R. Biosensing using dynamic-mode cantilever sensors: A review. Biosens. Bioelectron. 2012, 32, 1–18. [Google Scholar] [CrossRef]
  45. Palmisano, F.; Malitesta, C.; Centonze, D.; Zambonin, P.G. Correlation between Permselectivity and Chemical Structure of Overoxidized Polypyrrole Membranes Used in Electroproduced Enzyme Biosensors. Anal. Chem. 1995, 67, 2207–2211. [Google Scholar] [CrossRef]
  46. Lin, Y.; Lu, F.; Tu, Y.; Ren, Z. Glucose Biosensors Based on Carbon Nanotube Nanoelectrode Ensembles. Nano Lett. 2004, 4, 191–195. [Google Scholar] [CrossRef]
  47. Chen, X.; Matsumoto, N.; Hu, Y.; Wilson, G.S. Electrochemically Mediated Electrodeposition/Electropolymerization to Yield a Glucose Microbiosensor with Improved Characteristics. Anal. Chem. 2002, 74, 368–372. [Google Scholar] [CrossRef]
  48. Soldatkin, P. Glucose sensitive conductometric biosensor with additional Nafion membrane: Reduction of influence of buffer capacity on the sensor response and extension of its dynamic range. Anal. Chim. Acta 1994, 288, 197–203. [Google Scholar] [CrossRef]
  49. Jung, S.-Y. The Vroman Effect:  A Molecular Level Description of Fibrinogen Displacement. J. Am. Chem. Soc. 2003, 125, 12782–12786. [Google Scholar] [CrossRef]
  50. Vroman, L.; Adams, A.L. Findings with the recording ellipsometer suggesting rapid exchange of specific plasma proteins at liquid/solid interfaces. Surf. Sci. 1969, 16, 438–446. [Google Scholar] [CrossRef]
  51. Biosensing Instrument. Application Note: 110—Studying Protein Adsorption Properties with SPR | Biosensing Instrument. 2019. Available online: http://biosensingusa.com/application-notes/application-note-110-studying-protein-adsorption-properties-spr/ (accessed on 15 April 2019).
  52. Choi, S.; Chae, J. Surface plasmon resonance biosensor based on Vroman effect: Towards cancer biomarker detection. In Proceedings of the IEEE 15th International Mixed-Signals, Sensors, and Systems Test Workshop, Scottsdale, AZ, USA, 10–12 June 2009. [Google Scholar] [CrossRef]
  53. Charles, P.; Stubbs, V.R.; Soto, C.M.; Martin, B.D.; White, B.J.; Taitt, C.R. Reduction of Non-Specific Protein Adsorption Using Poly(ethylene) Glycol (PEG) Modified Polyacrylate Hydrogels in Immunoassays for Staphylococcal Enterotoxin B Detection. Sensors 2009, 9, 645–655. [Google Scholar] [CrossRef]
  54. Cerruti, M.; Fissolo, S.; Carraro, C.; Ricciardi, C.; Majumdar, A.; Maboudian, R. Poly(ethylene glycol) Monolayer Formation and Stability on Gold and Silicon Nitride Substrates. Langmuir 2008, 24, 10646–10653. [Google Scholar] [CrossRef]
  55. Wink, T.; van Zuilen, S.J.; Bult, A.; van Bennekom, W.P. Self-assembled Monolayers for Biosensors. Analyst 1997, 122, 43R–50R. [Google Scholar] [CrossRef]
  56. Dahint, R.; Seigel, R.R.; Harder, P.; Grunze, M.; Josse, F. Detection of non-specific protein adsorption at artificial surfaces by the use of acoustic plate mode sensors. Sens. Actuators B Chem. 1996, 36, 497–505. [Google Scholar] [CrossRef]
  57. Zuo, G. Dual-SAM functionalization on integrated cantilevers for specific trace-explosive sensing and non-specific adsorption suppression. Nanotechnology 2007, 18, 255501. [Google Scholar] [CrossRef]
  58. Bolduc, R.; Masson, J.-F. Monolayers of 3-Mercaptopropyl-amino Acid to Reduce the Nonspecific Adsorption of Serum Proteins on the Surface of Biosensors. Langmuir 2008, 24, 12085–12091. [Google Scholar] [CrossRef] [PubMed]
  59. Ladd, J.; Zhang, Z.; Chen, S.; Hower, J.C.; Jiang, S. Zwitterionic Polymers Exhibiting High Resistance to Nonspecific Protein Adsorption from Human Serum and Plasma. Biomacromolecules 2008, 9, 1357–1361. [Google Scholar] [CrossRef]
  60. van Andel, E. Systematic Comparison of Zwitterionic and Non-Zwitterionic Antifouling Polymer Brushes on a Bead-Based Platform. Langmuir 2019, 35, 1181–1191. [Google Scholar] [CrossRef]
  61. Laschewsky, A. Structures and Synthesis of Zwitterionic Polymers. Polymers 2014, 6, 1544–1601. [Google Scholar] [CrossRef]
  62. Sankaranarayanan, S.K.R.S.; Cular, S.; Bhethanabotla, V.R.; Joseph, B. Flow induced by acoustic streaming on surface-acoustic-wave devices and its application in biofouling removal: A computational study and comparisons to experiment. Phys. Rev. E 2008, 77, 066308. [Google Scholar] [CrossRef]
  63. Gref, R.; Lück, M.; Quellec, P.; Marchand, M.; Dellacherie, E.; Harnisch, S.; Blunk, T.; Müller, R.H. Stealth’ corona-core nanoparticles surface modified by polyethylene glycol (PEG): Influences of the corona (PEG chain length and surface density) and of the core composition on phagocytic uptake and plasma protein adsorption. Colloids Surf. B Biointerfaces 2000, 18, 301–313. [Google Scholar] [CrossRef]
  64. Prime, K.L.; Whitesides, G.M. Adsorption of proteins onto surfaces containing end-attached oligo(ethylene oxide): A model system using self-assembled monolayers. J. Am. Chem. Soc. 1993, 115, 10714–10721. [Google Scholar] [CrossRef]
  65. Yeh, P.-Y.J.; Kizhakkedathu, J.N.; Madden, J.D.; Chiao, M. Electric field and vibration-assisted nanomolecule desorption and anti-biofouling for biosensor applications. Colloids Surf. B Biointerfaces 2007, 59, 67–73. [Google Scholar] [CrossRef]
  66. Dykstra, P.H.; Roy, V.; Byrd, C.; Bentley, W.E.; Ghodssi, R. Microfluidic Electrochemical Sensor Array for Characterizing Protein Interactions with Various Functionalized Surfaces. Anal. Chem. 2011, 83, 5920–5927. [Google Scholar] [CrossRef]
  67. Ainslie, K.M.; Sharma, G.; Dyer, M.A.; Grimes, C.A.; Pishko, M.V. Attenuation of Protein Adsorption on Static and Oscillating Magnetostrictive Nanowires. Nano Lett. 2005, 5, 1852–1856. [Google Scholar] [CrossRef]
  68. Bernards, M.; He, Y. Polyampholyte polymers as a versatile zwitterionic biomaterial platform. J. Biomater. Sci. Polym. Ed. 2014, 25, 1479–1488. [Google Scholar] [CrossRef]
  69. Roach, L.S.; Song, H.; Ismagilov, R.F. Controlling Nonspecific Protein Adsorption in a Plug-Based Microfluidic System by Controlling Interfacial Chemistry Using Fluorous-Phase Surfactants. Anal. Chem. 2005, 77, 785–796. [Google Scholar] [CrossRef]
  70. Shi, H.; Tsai, W.-B.; Garrison, M.D.; Ferrari, S.; Ratner, B.D. Template-imprinted nanostructured surfaces for protein recognition. Nature 1999, 398, 593. [Google Scholar] [CrossRef]
  71. Guan, G.; Liu, B.; Wang, Z.; Zhang, Z. Imprinting of Molecular Recognition Sites on Nanostructures and Its Applications in Chemosensors. Sensors 2008, 8, 8291–8320. [Google Scholar] [CrossRef] [Green Version]
  72. Lee, C.-K.; Wang, Y.-M.; Huang, L.-S.; Lin, S. Atomic force microscopy: Determination of unbinding force, off rate and energy barrier for protein-ligand interaction. Micron 2007, 38, 446–461. [Google Scholar] [CrossRef]
  73. Shiddiky, M.J.A.; Vaidyanathan, R.; Rauf, S.; Tay, Z.; Trau, M. Molecular Nanoshearing: An Innovative Approach to Shear off Molecules with AC-Induced Nanoscopic Fluid Flow. Sci. Rep. 2014, 4, 3716. [Google Scholar] [CrossRef] [Green Version]
  74. Wang, Y.; Vaidyanathan, R.; Shiddiky, M.J.A.; Trau, M. Enabling Rapid and Specific Surface-Enhanced Raman Scattering Immunoassay Using Nanoscaled Surface Shear Forces. ACS Nano 2015, 9, 6354–6362. [Google Scholar] [CrossRef]
  75. Vaidyanathan, R.; Naghibosadat, M.; Rauf, S.; Korbie, D.; Carrascosa, L.G.; Shiddiky, M.J.; Trau, M. Detecting Exosomes Specifically: A Multiplexed Device Based on Alternating Current Electrohydrodynamic Induced Nanoshearing. Anal. Chem. 2014, 86, 11125–11132. [Google Scholar] [CrossRef]
  76. Vaidyanathan, R.; Shiddiky, M.J.A.; Rauf, S.; Dray, E.; Tay, Z.; Trau, M. Tunable ‘Nano-Shearing’: A Physical Mechanism to Displace Nonspecific Cell Adhesion during Rare Cell Detection. Anal. Chem. 2014, 86, 2042–2049. [Google Scholar] [CrossRef]
  77. Johnson, B.N.; Mutharasan, R. Reduction of nonspecific protein adsorption on cantilever biosensors caused by transverse resonant mode vibration. Analyst 2014, 139, 1112–1120. [Google Scholar] [CrossRef]
  78. Sankaranarayanan, S.K.R.S.; Singh, R.; Bhethanabotla, V.R. Acoustic streaming induced elimination of nonspecifically bound proteins from a surface acoustic wave biosensor: Mechanism prediction using fluid-structure interaction models. J. Appl. Phys. 2010, 108, 104507. [Google Scholar] [CrossRef]
  79. Cular, S.; Branch, D.W.; Bhethanabotla, V.R.; Meyer, G.D.; Craighead, H.G. Removal of Nonspecifically Bound Proteins on Microarrays Using Surface Acoustic Waves. IEEE Sens. J. 2008, 8, 314–320. [Google Scholar] [CrossRef]
  80. Richardson, M.B.; Bhethanabotla, V.R. Orthogonal SAW Transducers on Substrate for Simultaneous Biofouling Removal and Biosensing. U.S. Patent 10,031,140, 24 July 2018. [Google Scholar]
  81. Hsu, W.-T.; Feng, G.-H.; Cho, C.-L.; Chau, L.-K. Nonspecific binding removal and specific binding regeneration using longitudinal acoustic waves. RSC Adv. 2013, 3, 16159–16166. [Google Scholar] [CrossRef]
  82. Duan, X.; Pan, S.; Pang, W. Development of high-frequency bulk acoustic wave (BAW) resonators as biosensors and bioactuators. In Proceedings of the 2017 IEEE International Electron Devices Meeting (IEDM), San Francisco, CA, USA, 2–6 December 2017. [Google Scholar]
  83. Zhang, M.; Huang, J.; Cui, W.; Pang, W.; Zhang, H.; Zhang, D.; Duan, X. Kinetic studies of microfabricated biosensors using local adsorption strategy. Biosens. Bioelectron. 2015, 74, 8–15. [Google Scholar] [CrossRef]
  84. Li, D.; Wang, C.; Sun, G.; Senapati, S.; Chang, H.-C. A shear-enhanced CNT-assembly nanosensor platform for ultra-sensitive and selective protein detection. Biosens. Bioelectron. 2017, 97, 143–149. [Google Scholar] [CrossRef]
  85. Vaidyanathan, R.; Dey, S.; Carrascosa, L.G.; Shiddiky, M.J.A.; Trau, M. Alternating current electrohydrodynamics in microsystems: Pushing biomolecules and cells around on surfaces. Biomicrofluidics 2015, 9, 061501. [Google Scholar] [CrossRef] [Green Version]
  86. Dufour, I. Effect of hydrodynamic force on microcantilever vibrations: Applications to liquid-phase chemical sensing. Sens. Actuators B Chem. 2014, 192, 664–672. [Google Scholar] [CrossRef] [Green Version]
  87. Goeders, M.; Colton, J.S.; Bottomley, L.A. Microcantilevers:  Sensing Chemical Interactions via Mechanical Motion. Chem. Rev. 2008, 108, 522–542. [Google Scholar] [CrossRef]
  88. Zhu, Q. Microcantilever Sensors in Biological and Chemical Detections. Sens. Transducers 2011, 125, 1–21. [Google Scholar]
  89. Boisen, A.; Dohn, S.; Keller, S.S.; Schmid, S.; Tenje, M. Cantilever-like micromechanical sensors. Rep. Prog. Phys. 2011, 74, 036101. [Google Scholar] [CrossRef]
  90. Li, X.; Lee, D.-W. Integrated microcantilevers for high-resolution sensing and probing. Meas. Sci. Technol. 2011, 23, 022001. [Google Scholar] [CrossRef]
  91. Haring, P.; Cesewski, E.; Johnson, B.N. Piezoelectric Cantilever Biosensors for Label-free, Real-time Detection of DNA and RNA. In Biosensors and Biodetection: Methods and Protocols, Volume 2: Electrochemical, Bioelectronic, Piezoelectric, Cellular and Molecular Biosensors; Prickril, B., Rasooly, A., Eds.; Springer: New York, NY, USA, 2017; pp. 247–262. [Google Scholar]
  92. Brereton, G.J.; Qi, Q. Mechanisms of removal of micron-sized particles by high-frequency ultrasonic waves. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 1995, 42, 619–629. [Google Scholar] [CrossRef]
  93. Lei, J.; Glynne-Jones, P.; Hill, M. Acoustic streaming in the transducer plane in ultrasonic particle manipulation devices. Lab Chip 2013, 13, 2133–2143. [Google Scholar] [CrossRef] [Green Version]
  94. RLec, M.; Lewin, P.A. Acoustic wave biosensors. In Proceedings of the 20th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Hong Kong, China, 1 November 1998; Volume 6, pp. 2779–2784. [Google Scholar]
  95. Länge, B.; Rapp, E.; Rapp, M. Surface acoustic wave biosensors: A review. Anal. Bioanal. Chem. 2008, 391, 1509–1519. [Google Scholar] [CrossRef]
  96. Fogel, R.; Limson, J.; Seshia, A.A. Acoustic biosensors. Essays Biochem. 2016, 60, 101–110. [Google Scholar] [CrossRef] [Green Version]
  97. Yeo, Y.; Friend, J.R. Surface Acoustic Wave Microfluidics. Annu. Rev. Fluid Mech. 2014, 46, 379–406. [Google Scholar] [CrossRef] [Green Version]
  98. Friend, J.; Yeo, L.Y. Microscale acoustofluidics: Microfluidics driven via acoustics and ultrasonics. Rev. Mod. Phys. 2011, 83, 647–704. [Google Scholar] [CrossRef] [Green Version]
  99. Bruus, H.; Dual, J.; Hawkes, J.; Hill, M.; Laurell, T.; Nilsson, J.; Radel, S.; Sadhal, S.; Wiklund, M. Forthcoming Lab on a Chip tutorial series on acoustofluidics: Acoustofluidics—Exploiting ultrasonic standing wave forces and acoustic streaming in microfluidic systems for cell and particle manipulation. Lab Chip 2011, 11, 3579–3580. [Google Scholar] [CrossRef]
  100. Jönsson, P.; Gunnarsson, A.; Höök, F. Accumulation and Separation of Membrane-Bound Proteins Using Hydrodynamic Forces. Anal. Chem. 2011, 83, 604–611. [Google Scholar] [CrossRef]
  101. Burdick, G.M.; Berman, N.S.; Beaudoin, S.P. Hydrodynamic particle removal from surfaces. Thin Solid Films 2005, 488, 116–123. [Google Scholar] [CrossRef]
  102. Burdick, G.M.; Berman, N.S.; Beaudoin, S.P. Describing Hydrodynamic Particle Removal from Surfaces Using the Particle Reynolds Number. J. Nanopart. Res. 2001, 3, 453–465. [Google Scholar] [CrossRef]
  103. Munson, S.; Hasenbank, M.S.; Fu, E.; Yager, P. Suppression of non-specific adsorption using sheath flow. Lab Chip 2004, 4, 438–445. [Google Scholar] [CrossRef] [PubMed]
  104. Sivagnanam, V.; Yang, H.; Gijs, M.A.M. Study of Spatio-Temporal Immunofluorescence on Bead Patterns in a Microfluidic Channel. AIP Conf. Proc. 2010, 1311, 198–203. [Google Scholar] [CrossRef]
  105. Hawkins, K.R.; Steedman, M.R.; Baldwin, R.R.; Fu, E.; Ghosal, S.; Yager, P. A method for characterizing adsorption of flowing solutes to microfluidic device surfaces. Lab Chip 2007, 7, 281–285. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Target molecules (blue) in a sample containing background molecules (red) flow through a microchannel. The surface is covered with bioreceptors. Affinity capture is when the receptors on the target molecule are recognized by the capture ligand. Non-specific binding occurs when a background molecule binds to the ligand or the channel surface without forming ligand-receptor bonds. Adapted with permission from ref [24]. Copyright 2018 John Wiley and Sons.
Figure 1. Target molecules (blue) in a sample containing background molecules (red) flow through a microchannel. The surface is covered with bioreceptors. Affinity capture is when the receptors on the target molecule are recognized by the capture ligand. Non-specific binding occurs when a background molecule binds to the ligand or the channel surface without forming ligand-receptor bonds. Adapted with permission from ref [24]. Copyright 2018 John Wiley and Sons.
Sensors 19 02488 g001
Figure 2. Chart outlining the main mechanisms of NSA reduction (passive and active) and sub-categories of each.
Figure 2. Chart outlining the main mechanisms of NSA reduction (passive and active) and sub-categories of each.
Sensors 19 02488 g002
Figure 3. Ac-EHD induced fluid flow (a) Mechanism of how the electric field induces charges in the double layer of each electrode experience. Adapted with permission from [70]. Copyright 2014 Springer Nature. (b) Schematic of ac-EHD induced nanoshearing. The target proteins (blue) are bound to the bioreceptors while the non-specifically bound molecules (green and red) are removed. Adapted with permission from [10]. Copyright 2015 Springer Nature.
Figure 3. Ac-EHD induced fluid flow (a) Mechanism of how the electric field induces charges in the double layer of each electrode experience. Adapted with permission from [70]. Copyright 2014 Springer Nature. (b) Schematic of ac-EHD induced nanoshearing. The target proteins (blue) are bound to the bioreceptors while the non-specifically bound molecules (green and red) are removed. Adapted with permission from [10]. Copyright 2015 Springer Nature.
Sensors 19 02488 g003
Figure 4. Hypersonic resonator for biofouling removal and protein detection: (a) Illustration of microvortexes in liquid triggered by the hypersonic resonator; (b) Simulation of the fluid motion triggered by the hypersonic resonator. Adapted with permission from ref [9]. Copyright 2017 the American Chemical Society.
Figure 4. Hypersonic resonator for biofouling removal and protein detection: (a) Illustration of microvortexes in liquid triggered by the hypersonic resonator; (b) Simulation of the fluid motion triggered by the hypersonic resonator. Adapted with permission from ref [9]. Copyright 2017 the American Chemical Society.
Sensors 19 02488 g004
Figure 5. Schematic diagram of the piezoelectric membrane as an implantable sensor coating. The membrane is composed of two driving electrodes with a piezoelectric material in between. Adsorbed proteins can be desorbed by an electric field and carried away by acoustic streaming generated by the vibration. Reprinted from [65] with permission from Elsevier.
Figure 5. Schematic diagram of the piezoelectric membrane as an implantable sensor coating. The membrane is composed of two driving electrodes with a piezoelectric material in between. Adsorbed proteins can be desorbed by an electric field and carried away by acoustic streaming generated by the vibration. Reprinted from [65] with permission from Elsevier.
Sensors 19 02488 g005
Figure 6. Schematic representation and working principle of the experimental setup for generating LAWs and transmitting to a GNP-GSC through buffer solution. Biomolecular binding at the surface of the functionalized GNPs removed by the acoustic wave effect. Reproduced from [81] with permission from The Royal Society of Chemistry.
Figure 6. Schematic representation and working principle of the experimental setup for generating LAWs and transmitting to a GNP-GSC through buffer solution. Biomolecular binding at the surface of the functionalized GNPs removed by the acoustic wave effect. Reproduced from [81] with permission from The Royal Society of Chemistry.
Sensors 19 02488 g006
Figure 7. Diagram of how pressure-driven flow creates a parabolic velocity profile, which can behaves on immobilized bioreceptors and bound analytes in microchannel. Reproduced from [104] with permission of AIP Publishing.
Figure 7. Diagram of how pressure-driven flow creates a parabolic velocity profile, which can behaves on immobilized bioreceptors and bound analytes in microchannel. Reproduced from [104] with permission of AIP Publishing.
Sensors 19 02488 g007
Figure 8. Schematic of CNTs switch sensor for protein detection. Reproduced from [84] with permission of Elsevier.
Figure 8. Schematic of CNTs switch sensor for protein detection. Reproduced from [84] with permission of Elsevier.
Sensors 19 02488 g008
Table 1. Summary of Active NSA Removal Methods.
Table 1. Summary of Active NSA Removal Methods.
CategoryTypeAnalyteReference
Transducer-based
ElectromechanicalHypersonic ResonanceCy3 labeled human IgG antigen[9]
ac-EHDPSA [10];
IgG [10];
HER2 [10,73,74];
HER2 & PSA exosomes [75];
MCF7 & T-47D cells [76]
[10,73,74,75,76]
Resonant Cantilever
Vibration
BSA[44,77]
Acoustic WaveSurfaceNone (simulated) [62,78]
IgG antigen [79]
[62,78,79]
Orthogonal SurfaceNone (simulated) [22][22,80]
LongitudinalBiotinylated GNP-GSC[81]
PiezoelectricAvidin[38]
BulkProteins [82]
None (simulated) [83]
[82,83]
Fluid-based
HydrodynamicHER2[84]

Share and Cite

MDPI and ACS Style

Lichtenberg, J.Y.; Ling, Y.; Kim, S. Non-Specific Adsorption Reduction Methods in Biosensing. Sensors 2019, 19, 2488. https://0-doi-org.brum.beds.ac.uk/10.3390/s19112488

AMA Style

Lichtenberg JY, Ling Y, Kim S. Non-Specific Adsorption Reduction Methods in Biosensing. Sensors. 2019; 19(11):2488. https://0-doi-org.brum.beds.ac.uk/10.3390/s19112488

Chicago/Turabian Style

Lichtenberg, Jessanne Y., Yue Ling, and Seunghyun Kim. 2019. "Non-Specific Adsorption Reduction Methods in Biosensing" Sensors 19, no. 11: 2488. https://0-doi-org.brum.beds.ac.uk/10.3390/s19112488

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop