Next Article in Journal
Estimation of Infiltration Volumes and Rates in Seasonally Water-Filled Topographic Depressions Based on Remote-Sensing Time Series
Previous Article in Journal
Software Fault Localization through Aggregation-Based Neural Ranking for Static and Dynamic Features Selection
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Compact Exposimeter Device for the Characterization and Recording of Electromagnetic Fields from 78 MHz to 6 GHz with Several Narrow Bands (300 kHz)

by
Marco Xavier Rivera González
1,*,
Nazario Félix González
1,
Isabel López
1,
Juan Sebastián Ochoa Zambrano
2,
Andrés Miranda Martínez
1 and
Ceferino Maestú Unturbe
1,3
1
Center for Biomedical Technology (CTB), Escuela Tecnica Superior de Ingenieros de Telecomunicaciones (ETSIT), Universidad Politécnica de Madrid (UPM), 28040 Madrid, Spain
2
School of Computer Systems Engineering (ETSISI), Universidad Politécnica de Madrid (UPM), 28040 Madrid, Spain
3
Networking Research Center of Bioengineering, Biomaterials and Nanomedicine (CIBER-BBN), Universidad Politécnica de Madrid (UPM), 28040 Madrid, Spain
*
Author to whom correspondence should be addressed.
Submission received: 12 October 2021 / Revised: 31 October 2021 / Accepted: 4 November 2021 / Published: 7 November 2021
(This article belongs to the Section Environmental Sensing)

Abstract

:
A novel compact device with spectrum analyzer characteristics has been designed, which allows the measuring of the maximum power received in multiple narrow frequency bands of 300 kHz, recording the entire spectrum from 78 MHz to 6 GHz; the device is capable of measuring the entire communications spectrum and detecting multiple sources of electromagnetic fields using the same communications band. The proposed device permits the evaluation of the cross-talk effect that, in conventional exposimeters, generates a mistake estimation of electromagnetic fields. The device was calibrated in an anechoic chamber for far-fields and was validated against a portable spectrum analyzer in a residential area. A strong correlation between the two devices with a confidence higher than 95% was obtained; indicating that the device could be considered as an important tool for electromagnetic field studies.

1. Introduction

With the advancement of technology, the population is increasingly exposed to radiation by non-ionizing electromagnetic fields in the radio frequency range (RF-EMF), which is from 100 kHz to 300 GHz; these frequencies correspond to wireless technologies, such as Wifi, mobile phone, television, radio, etc.
According to global statistics reported by the International Telecommunication Union (ITU), carried out from 2001 to 2018, there is a constant growth in the use of wireless technologies [1], which has generated concern in society about the possible health effects associated with RF-EMF exposures [2]. The regulatory organizations limit the amount of radiated energy following the International Commission on Non-Ionizing Radiation Protection (ICNIRP) guidelines that only consider the so-called thermal effects using the criterion of the maximum rate of radiation and Specific Absorption Rate (SAR) as the only evaluation parameter, which consists of the measurement of the maximum RF-EMF power absorbed by living tissue [3]. Nevertheless, the effects that could be considered non-thermal are not recognized within the scientific evidence by ICNIRP [3,4,5,6,7,8,9]. The recommendations of the ICNIRP establish the maximum levels of radiation at 450 µW/cm2 [3,9,10,11]. However, the BioInitiative working group, together with other researchers [8,10,12,13,14,15,16], suggest that adverse health effects are observed at low levels of exposure 0.1 µW/cm2. Studies suggest that RF-EMF exposures with powers below the recommendations of the ICNIRP have effects related to changes in brain activity [17], affecting cognitive and motor performance [12,13], infertility problems in the male reproductive system [18,19], DNA damage [20,21], association to different brain tumors and intensity of RF-EMF, and having a greater effect in children and teenagers than in adults [4,6,12,22,23,24,25]. These studies suggest that exposure to RF-EMF is an important factor to consider as a “possible carcinogen” classified in group 2B by the International Agency for Research on Cancer (IARC) [26]. On the other hand, using the same evaluation criteria, they recommend that RF-EMF exposures should belong to group 2A in the IARC as “probable carcinogenic” [8,15].
The lack of knowledge of the levels of RF-EMF radiation to which the population is exposed raises greater concern due to the possible adverse effects on health resulting from continuous exposure to these electromagnetic fields [1,2,8,12,27]. At the moment, the solution for this problem is the use of systems, such as spectrum analyzers and exposimeters, that allow the measurement and characterization of RF-EMF [28]. According to the criteria of the International Telecommunication Union (ITU), the measurements must be carried out continuously, allowing the power intensity to be measured and discriminate between each frequency corresponding to the RF-EMF emission sources [1]; therefore, measurements made with a spectrum analyzer are the most appropriate for the characterization of RF-EMF. However, it is an expensive method and also requires experts for the correct use and interpretation of the equipment’s data; for this reason, the use of exposimeters is considered as the best option [16]. Different exposimeters perform measurements in broadband [29], which does not allow correct discrimination between the various frequencies and, therefore, the proper identification of the RF-EMF emission sources [30]. There is a clear need to provide to the population a small and low-cost exposimeter that allows monitoring the peak-to-peak maximum electromagnetic radiation levels of the entire radioelectric spectrum in several narrow frequency bands to differentiate between multiple sources in the same band of telecommunications.

2. Materials and Methods

The exposimeters used for epidemiological studies of telecommunication bands, such as radio, television, mobile telephony, Wifi, are governed by the regulations and standards for electromagnetic fields, such as the ICNIRP guidelines or IEEE standard C95.1 [3,31], developed by the international committee on Electromagnetic safety [32]. The use of exposimeters arises from spot measurements performed with spectrum analyzers at telecommunication base stations in order to corroborate that the transmitted power is according to the standards. [33]
The most common narrowband exposimeters used in studies are: EME Spy-120, EME Spy-121, EME Spy-140, EME Spy-120 (SATIMO, Courtaboeuf, France), ESM 140 (Maschek Electronik, Bad Wörishofen, Germany), and ExpoM-RF (Fields at Work GmbH, Zürich, Switzerland). All these devices have different characteristics in the number of frequency bands, sampling interval, detector type, size, and weight [29,30]. Table 1 lists the main characteristics of these devices.
Spectrum analyzers allow, with great precision, the exposure to electromagnetic fields both in frequency and power to be determined, allowing the identification of each of the sources of electromagnetic field emission, is a great advantage over the aforementioned exposimeters; however, these have some disadvantages, such as expensive equipment, low precision in dynamic environments, and need expert personnel for their correct operation. On the other hand, exposimeters are cheaper, allowing their use without the need of an expert, are faster in dynamic places, and, due to their size and easy use, allow their implementation in various epidemiological studies [16]. The sensitivity of exposimeters could be improved by performing separate measurements in various fractions of the RF spectrum, where it detects sources that contribute significantly to exposures, considering between 12 and 20 radio frequency bands as the most significant in the entire radio spectrum. However, this is not supported by a systematic investigation of the entire radio frequency spectrum, and it is possible that more bands should be measured or that some of them can be ignored [33].
The major limitation of exposimeters is that these do not allow the measurement of the entire spectrum and are not suitable for differentiating between multiple electromagnetic field sources because its resolution bandwidth is determined by the full desired frequency band to be measured; using a single measurement for the entire selected band avoids adequately showing the user the different sources that occupy the same band [34]. The use of cell phones close to the exposimeter could increase the measured value in the band, and a clear increase, or an overestimation, of the measurement, is observed. [35] Another drawback is cross-talk, a signal that is recorded in a nearby band, providing a false measurement; it usually happens in mobile telephony between Uplink and Downlink bands [30]. The comparison features between conventional exposimeters and spectrum analyzers is shown in Table 2.
The main disadvantage of exposimeters is the inability to measure the entire radioelectric spectrum in several narrow frequency bands in such a way that allows properly differentiating between multiple radiations sources. Therefore, an exposimeter system with similar characteristics to a spectrum analyzer is proposed in this paper without losing the advantages of the traditional exposimeters.

2.1. Proposed System

The design and manufacture of the device comprise three phases that are explained below:
  • The first phase covers the design and development of hardware for the detection of electromagnetic (EM) signals in the frequency range between 78 MHz and 6 GHz with bandwidths of 300 kHz.
  • The second phase consists of the calibration of the device in a controlled environment.
  • The third phase is the development of an algorithm for data processing and an interface for visualization of the power density in different frequency bands.

2.1.1. Data Acquisition System

The proposed system consists of an array of fractal antennas (FRACTUSANTENNAS, Barcelona, Spain) to cover the frequency range between 78 MHz and 6 GHz. These antennas are suitable for this design because they can cover large bandwidths and have small dimensions compared to other types of commercial antennas. This makes it possible to optimize space on the printed circuit board (PCB). Table 3 shows the models and bandwidth of the fractal antennas used for the proposed exposimeter. The plots of radiation pattern of each antenna are shown in Appendix B.
Figure 1 shows the general block diagram of the system operation, where the fractal antennas from Table 3 are multiplexed by an Analog Devices HMC321ALP4E RF mux (Analog Devices, DE, USA), which has eight inputs and one output that enables 3:8 binary TTL decoding control; the switching time (TS) is 125 nsec; and the RF mux switch to the next antenna when the device has measured all its bandwidth. The signals from the antenna are amplified by an Analog Devices ADL5542 low noise amplifier (LNA), which has a maximum gain of 20 dB. The Analog Devices ADF4355-3 fractional-N phase-locked Loop (PLL) works as an RF signal generator, covering the range from 54 to 6600 MHz; the total lock time of the PLL to generate a signal (TPLL) is 45 µs. The signal generated by the PLL and the signal received from the antennas are multiplied in the mixer in order to lower the frequency of the antenna signal to a frequency that matches the 315 MHz center frequency and 300 kHz bandwidth of the SAW bandpass filter; the Analog devices ADL 5802 mixer and the AFS315E-T filter from Abracon LLC were chosen. The filtered signal is measured by an Analog devices AD8309 logarithmic RF detector that provides a voltage output signal proportional to the power of the input signal in dBm; the output settling time of the logarithmic detector (TLD) is 220 ns. The output voltage of the logarithmic detector is sampled using an Analog-to-Digital Converter (ADC) and recorded in the microcontroller; the ADC rate (TADC) is 1 Msamples/s (1 µs for each sample). The time to measure one sample (T′) is described as follows:
T = T P L L + T L D + T A D C ,
T = 45   µ s + 0.22   µ s + 1   µ s = 46.22   µ s ,
The proposed system measures the received power in the spectrum from 78 MHz to 6 GHz with 300 kHz of resolution bandwidth, measuring approximately 19,500 narrow frequency bands. Therefore, the total time to measure all the bands is described as follow:
T T = n S T + n A T S ,
T T = 19 , 500 × 46.22   µ s + 5 0.125   µ s = 901.915   µ s ,
where TT = 901.915 µs is the total time to measure all the bands; nS is the number of samples (19,500); nA is the number of antennas; finally, TS is the switching time of the RF mux.
Finally, the measured signals are stored with the GPS geolocation data in a non-volatile SD memory that can store up to 2 Gbytes of data. The system has a USB communication port to communicate with the computer and download the set of records made. In Figure 2, the flow chart for the measurement process is shown.
The PLL works as a tuner of the signals recorded by the antennas; therefore, consider that the PLL generates a signal S1, and the antenna array receives a signal S2. These can be described as follows:
S 1 = A 1 cos ( 2 π f 1 t + θ 1 ) ,
S 2 = k A 2 ( 2 π f 2 t + θ 2 ) ,
where k is the amplification value due to the low noise amplifier, A1 and A2 are the amplitudes corresponding to each of the signals, and likewise, their frequencies are f1, f2 and phases are θ1 and θ2, respectively. These are multiplied by the mixer, and the following equation is obtained with the signal mixer theory [36].
S 1 S 2 = k A 1 A 2 2 [ cos ( 2 π ( f 1 + f 2 ) t + θ 1 + θ 2 ) + cos ( 2 π ( f 1 f 2 ) t + θ 1 θ 2 ) ] ,
The signal generated by the PLL signal synthesizer is 315 MHz different from the received signal; this allows it to match with the center frequency of the bandpass filter; therefore, f1f2 is equal to 315 MHz, and Equation (7) would be as follows:
S 1 S 2 = k A 1 A 2 2 [ cos ( 2 π ( f 1 + f 2 ) t + θ 1 + θ 2 ) + cos ( 2 π ( 315 ( 10 6 ) ) t + θ 1 θ 2 ) ] ,
Then, through the bandpass filter, the signals with a frequency different than 315 MHz are filtered. Giving the next equation:
S 1 S 2 = k A 1 A 2 2 [ cos ( 2 π ( 315 ( 10 6 ) ) t + θ 1 θ 2 ) ] ,
Then, by the signal theory [37], the power of the PS1S2 signal is expressed with the following equation:
P S 1 S 2 = lim T 1 T ( k A 1 A 2 2 ) 2 T T cos ( 2 π ( 315 ( 10 6 ) ) t + θ 1 θ 2 ) 2 t ,
P S 1 S 2 ( W ) = lim T 1 T ( k A 1 A 2 2 ) 2 T = ( k A 1 A 2 2 ) 2 ,
According to Equation (11), the power of the signal depends only on its amplitude; therefore, the logarithmic detector detects the power of the PS1S2 signal expressed in dBm. This value is registered and stored in non-volatile memory using communication with the microprocessor, which also has the function of controlling the RF mux, the PLL, and the GPS to obtain the geolocation of each measurement performed.
As shown in Figure 3, the proposed system consists of two PCBs fitting one on top of the other, incorporating all the components mentioned above. The first PCB (Figure 3a) integrates the radio frequency components; the second PCB (Figure 3b) includes the microcontroller, GPS, SD memory, the components for battery charging, and voltage regulators. The antennas share space between the two PCBs and are connected using coaxial cables. The PCBs were designed with enough free space in the PCB in order to avoid electromagnetic interference; this is specified in the datasheets from FRACTUS ANTENNAS. The system uses 2-cell lithium-ion (Li-ion) batteries of 3500 mAh in series, with low noise linear regulators 3.3 and 5.0 V for power supply. The average current consumption is estimated at 450 mA, in continuous operation mode, allowing the system to run for several hours without the need to charge the batteries.
The proposed system has a sampling rate of 20,000 samples/s, taking less than one second to measure the entire frequency spectrum. The memory provides up to one week of non-volatile memory space. The system has a dynamic power measurement range of 90 dB with RF input power from −70 to +20 dBm and a 0.04 dBm resolution. Table 4 shows the specifications of the proposed system.

2.1.2. System Calibration

The calibration of the device was performed in the anechoic chamber of the Universidad Politécnica de Madrid. The measurements were taken at a distance of 2 m between the transmitting antenna and the proposed exposimeter, evaluating only the far-field from 698 MHz to 6 GHz. The PNA network analyzer (Agilent Technologies E8362B, Santa Clara, CA, USA) was used to generate the signals, and 10 dBm of power was configured. Figure 4 shows a block diagram of the exposimeter calibration system.
As shown in the block diagram in Figure 4, a computer controls the PNA to generate the signals to be measured and communicates with the exposimeter to perform the corresponding measurement, to obtain: first, the data of return losses delivered by the PNA and, second, the measurements of power received by the exposimeter. According to the antenna theory, have the following equations [38]:
R L = 20 l o g 10 | Γ | ,
where RL is the return loss expressed in dB generated by the PNA, and Γ is the reflection coefficient, which indicates the percentage of the reflected signal. Solving Equation (12), the reflection coefficient is obtained:
Γ = ( 10 ( R L 20 ) ) ,
To find the percentage of transmitted power, the following equation was used:
T = ( 1 10 ( R L 20 ) )
In Equation (14), T is the percentage of transmitted power and allows the output power of the transmitting antenna to be calculated. In order to obtain the power received by the antennas, the free space attenuation equation was used [39]:
L b f = 32.4 + 20 l o g 10 ( f ) + 20 l o g 10 ( d ) ,
where Lbf is in dB, f is the frequency in MHz, and d is the distance in meters. Therefore, the power received by the antennas Pr as a function of the transmitted power PT, is expressed as follows:
P r = P T × T L b f ,
The signals obtained by the PNA are processed using Equation (16) to obtain the received power value and relate it with the measurements obtained by the exposimeter to create the calibration curves.

2.1.3. Software Development: Data Processing and Visualization System

The stored data by the exposimeter is transmitted to the PC using LabVIEW software (National Instruments Company, Austin, TX, USA) and, the data is processed using Matlab software (MathWorks Company, Natick, MA, USA), where each power measurement is corrected through the calibration wave. To obtain the data in power density (W/m2), the following equations according to the theory of radiation parameters were used [40].
S = P r A ,
A = λ 2 4 π G ,
P r ( W ) = 10 ( P r ( d B m ) 10 ) 1000 ,
where S is the power density, Pr is the received power in dBm, λ is the wavelength expressed in meters, and G is the maximum orientation of the antenna that is considered to be 1.5 so that the effective area A is a function of the frequency. Therefore, Equation (17) can be written as:
S = 10 ( P r ( d B m ) 10 ) λ 2 4 π G = ( 4 π f 2 c 2 ) 10 ( P r ( d B m ) 10 ) ,
Equation (20), for power density, is expressed in mW/m2 and is a function of the frequency, which is applied to each measure in order to obtain the final result.

2.2. Case Study for Validation of the Proposed Design

A case study was carried out to validate the proposed exposimeter against a portable spectrum analyzer. The regulations established by ICNIRP [3] were not followed, the measurements were performed in 10-min periods, and the maximum received power levels were considered (not the average); therefore, the data analysis considers only the peak values or the maximum power recorded at each measurement point. The measurements were performed to validate only the far-field, not on the body, and all measurements were performed outdoors. The validation study was deployed in a residential area in Madrid, Spain; seven different measurement points have been proposed in this urban area. According to the Ministry of Industry, Energy, and Tourism data [41], Figure 5 shows the location of the telephone stations (red) that are near the measurement points (blue). There are nine telephone base stations in this area that operate on the frequencies indicated in Table 5.
The FSH8 portable spectrum analyzer with a TSEMF-B2 omnidirectional antenna was used (Munich, Germany, ROHDE&SCHWARDZ), which allows measurements in bandwidths from 700 MHz to 6 GHz, registering maximum power levels in 10-min intervals. Figure 6 shows the spectrum analyzer and TSEMF-B2 omnidirectional antenna.
The professional equipment (spectrum analyzer) and the proposed exposimeter perform the measurements at each point at the same time. The measures of both devices were carried out at 1.5 m above the ground using a tripod, and personal equipment (mobile phone) was turned off in order to avoid stimulating the signals from the base stations. The proposed exposimeter measured from 78 MHz to 6 GHz; the data from 700 MHz to 6 GHz were analyzed.
Student’s t-test was used to evaluate independent variables to verify if there are significant differences between the measurements of the two devices, where the number of samples corresponds to the maximum peak power recorded at each frequency band of the measurement points from 1 to 7.
Then, Pearson’s correlation factor is calculated to find the relationship between the FSH8 spectrum analyzer and the proposed exposimeter, evaluating the linear dependence measure between two variables from the mean and standard deviation of each variable [42], as follows in Equation (21).
ρ ( A , B ) = 1 N 1 i = 1 N ( A i µ A σ A ) ( B i µ B σ B )
where A and B correspond to the variables to be measured, µA and σA correspond to the mean and standard deviation of A, µB and σB correspond to the mean and standard deviation of B, and finally, N corresponds to the number of observations or samples.

3. Results

The measurements at each point in Figure 5 were performed with the proposed exposimeter and FSH8 spectrum analyzer at the same time. The plots obtained for each measurement point are shown in Appendix A. Table 6 shows the values of the highest power peaks for each working frequency of the telephone base stations (Table 5) and the frequencies corresponding to 2G and 5G Wifi, where SA and E correspond to the FSH8 Spectrum Analyzer and the proposed Exposimeter, respectively. Finally, the last row indicates the correlation factor between the two pieces of equipment for each measurement point.
Table 7, shows the frequencies corresponding to the maximum peaks of the spectrum analyzer and the exposimeter from Table 6.
The data obtained by the spectrum analyzer are compared with the data registered by the proposed exposimeter using Student’s t-test. The Kolmogorov Smirnov test has been used to verify the normality of the measurements. The number of samples (N = 49) corresponds to the maximum power recorded at each frequency band of the measurement points from 1 to 7 in Table 6. The resulting data is transformed into logarithmic units, obtaining the results presented in Table 8. In Table 9, the Levene test for equality of variances and the Student’s t-test results are obtained.
Figure 7, shows the error graph concerning to the means values analyzed for the spectrum analyzer and the proposed exposimeter, observing that the error graphs are within the same range, and there is no significant differences between both devices. In Figure 8, the correlation graph between the FSH8 spectrum analyzer equipment and the proposed exposimeter at each frequency band of all measurement points from 1 to 7 in Table 6 is shown.

4. Discussion

In this study, the regulations established by ICNIRP [3] were not followed, the measurements of the maximum received power levels were considered (not the average). The non-thermal effects are not considered in the establishment of the exposure limits, and the BioInitiative working group, together with other researchers [8,10,12,13,14,15,16], suggest that adverse health effects are observed at low levels of exposure and suggest that using the SAR criterion alone is not the most appropriate for this purpose. Conventional exposimeters are not suitable for differentiating between multiple electromagnetic field sources because their resolution bandwidth is determined by the full desired frequency band to be measured, avoiding the detection of the sources responsible for the greatest contribution of electromagnetic fields [34]. Biological effects due to prolonged radiofrequency radiation exposure are considered to increase when energy peaks are detected.
The proposed system follows the ICNIRP requirement to measure the entire radioelectric spectrum in one second or less [3]. The sweeping time is 901.29 ms, and the switching between each fractal antenna does not cause any delay or uncertainty that could affect the measurements. However, the sweeping time could generate the differences in the power measurements between the spectrum analyzer and the proposed exposimeter, as is observed in Table 6. On the other hand, restricting and performing measurements only in the communication bands used by the base stations (Table 5) will significantly increase the sampling rate, obtaining a better estimation of the received power. However, information from the rest of the radioelectric spectrum could be lost. Conventional exposimeters measure up to 20 frequency bands [33]; therefore, these are very sensitive issues about sweeping time.
Table 7 shows the frequencies that correspond to the maximum power values measured in Table 6. Showing that a high correlation exists at each measurement point. However, there are differences in frequency that are attributed to the different resolution bandwidths of the spectrum analyzer and the proposed exposimeter. The designed system measures all the radioelectric spectrum in multiple narrow bands (300 kHz), allowing the identification of different sources of electromagnetic fields, even within the same communications band, for example, if there are two slots occupied in the same frequency band, the proposed exposimeter will be able to detect both sources, and it will also provide the information of the rest of the spectrum in order to detect the cross-talk effect, which in conventional exposimeters could generate a wrong estimation of the measured power [30,34,35].
Concerning the ICNIRP recommendations, all of the measurements in Table 6 exceed the established limit of 450 µW/cm2; however, several of the measurements exceed 0.1 µW/cm2, which according to the suggestions of the BioInitiative working group [8], could be related to adverse health effects.
In Table 8 of the Smirnov Kolmogorov test with a 95% confidence level, a significant factor of 0.2 is obtained; therefore, the measurements of both devices have a normal distribution. In Table 9, equal variances in the Levene test and a significant factor of 0.948 for Student’s t-test were obtained, indicating there is no significant difference between the measurements between the spectrum analyzer and the exposimeter.
In Figure 8, the Pearson’s correlation factor of 0.9682 is obtained, indicating that there is a strong correlation between the data obtained by the FSH8 spectrum analyzer and the proposed exposimeter, corroborating the results obtained with Student’s t-test. Therefore, the proposed exposimeter could be used as a tool to measure electromagnetic fields with reliability comparable to a spectrum analyzer.

5. Conclusions

A compact exposimeter system has been proposed, which has similar characteristics to a spectrum analyzer, without losing the advantages of conventional exposimeters, and can be a useful tool for electromagnetic field studies analyzing a far-field. The proposed exposimeter measures the received power in the spectrum between 78 MHz and 6 GHz with a resolution bandwidth of 300 KHz, measuring approximately 19,500 narrow bands; the proposed exposimeter has passed the testing phase compared with spectrum analyzers with a significant factor of 0.948 for Student’s t-test and a correlation factor of 0.9682 between the measurements of both devices.
The proposed exposimeter requires one second to measure the entire radioelectric spectrum; therefore, it has less sensitivity to variations in electromagnetic fields. This could be improved by restricting the measurements to the communication bands of interest; however, information from the rest of the radio spectrum is lost. It is important to define the relationship between the communication bands to be measured and the sensitivity to electromagnetic field variations with respect to time.
In future work, it is expected to use several of the proposed devices in electromagnetic field surveys to cover residential areas and provide information as an electromagnetic field sensor network to be used in epidemiological studies in the deployed area.

Author Contributions

Conceptualization, C.M.U., N.F.G., and M.X.R.G.; methodology, C.M.U., N.F.G., and M.X.R.G.; software, J.S.O.Z. and M.X.R.G.; validation, I.L. and M.X.R.G.; formal analysis, N.F.G., A.M.M., J.S.O.Z., and M.X.R.G.; data curation, I.L. and M.X.R.G.; writing—original draft preparation N.F.G.; J.S.O.Z., and M.X.R.G.; writing—review and editing, N.F.G., J.S.O.Z., and M.X.R.G.; supervision, N.F.G. and C.M.U.; project administration N.F.G. and C.M.U.; funding acquisition, C.M.U. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The data presented in this study are available on request from the corresponding author.

Conflicts of Interest

The authors declare no conflict of interest.

Appendix A

Plots obtained by the FSH8 spectrum analyzer and the proposed exposimeter at each of the measurement points in Figure 5. (Red, data obtained by the proposed exposimeter and blue, data obtained by the spectrum analyzer.)
Figure A1. Power density measurement at the first measuring point.
Figure A1. Power density measurement at the first measuring point.
Sensors 21 07395 g0a1
Figure A2. Power density measurement at the second measuring point.
Figure A2. Power density measurement at the second measuring point.
Sensors 21 07395 g0a2
Figure A3. Power density measurement at the third measuring point.
Figure A3. Power density measurement at the third measuring point.
Sensors 21 07395 g0a3
Figure A4. Power density measurement at the fourth measuring point.
Figure A4. Power density measurement at the fourth measuring point.
Sensors 21 07395 g0a4
Figure A5. Power density measurement at the fifth measuring point.
Figure A5. Power density measurement at the fifth measuring point.
Sensors 21 07395 g0a5
Figure A6. Power density measurement at the sixth measuring point.
Figure A6. Power density measurement at the sixth measuring point.
Sensors 21 07395 g0a6
Figure A7. Power density measurement at the seventh measuring point.
Figure A7. Power density measurement at the seventh measuring point.
Sensors 21 07395 g0a7

Appendix B

Plots of radiation patterns of each fractal antenna used for the proposed exposimeter.
Figure A8. Radiation pattern of the FR01-B3-W-0-055 antenna at 98 MHz (FRACTUS ANTENNAS, Barcelona, Spain).
Figure A8. Radiation pattern of the FR01-B3-W-0-055 antenna at 98 MHz (FRACTUS ANTENNAS, Barcelona, Spain).
Sensors 21 07395 g0a8
Figure A9. Radiation pattern of the FR01-B3-V-0-054 antenna at 200 MHz (FRACTUS ANTENNAS, Barcelona, Spain).
Figure A9. Radiation pattern of the FR01-B3-V-0-054 antenna at 200 MHz (FRACTUS ANTENNAS, Barcelona, Spain).
Sensors 21 07395 g0a9
Figure A10. Radiation pattern of the FR01-B1-S-0-047antenna at 890 MHz (FRACTUS ANTENNAS, Barcelona, Spain).
Figure A10. Radiation pattern of the FR01-B1-S-0-047antenna at 890 MHz (FRACTUS ANTENNAS, Barcelona, Spain).
Sensors 21 07395 g0a10
Figure A11. Radiation pattern of the FR01-S4-250 antenna at 1.71, 2.2, and 2.69 GHz (FRACTUS ANTENNAS, Barcelona, Spain).
Figure A11. Radiation pattern of the FR01-S4-250 antenna at 1.71, 2.2, and 2.69 GHz (FRACTUS ANTENNAS, Barcelona, Spain).
Sensors 21 07395 g0a11
Figure A12. Radiation pattern of the FR05-S1-N0-1-004 antenna at 2.45 and 5.4 GHz (FRACTUS ANTENNAS, Barcelona, Spain).
Figure A12. Radiation pattern of the FR05-S1-N0-1-004 antenna at 2.45 and 5.4 GHz (FRACTUS ANTENNAS, Barcelona, Spain).
Sensors 21 07395 g0a12

References

  1. ITU. Monitoring of Electromagnetic Field Levels; Recomendation ITU-T K.83. International Telecommunication Union (ITU) Std., 2011; Volume 83. Available online: https://www.itu.int/rec/T-REC-K.83-201103-S (accessed on 21 January 2021).
  2. Eurobarometer TNS Opinion & Social. (2010, June). Eurobarometer 73.3, Electromagnetic Fields. 2010. Available online: http://ec.europa.eu/public_opinion/archives/ebs/ebs_347_en.pdf (accessed on 21 May 2020).
  3. International Commission on Non-Ionizing Radiation Protection. ICNIRP Guidelines for limiting exposure to time-varying electric, magnetic, and electromagnetic fields (up to 300 GHz). Health Phys. 1998, 74, 494–521. [Google Scholar]
  4. Belpomme, D.; Hardell, L.; Belyaev, I.; Burgio, E.; Carpenter, D.O. Thermal and non-thermal health effects of low intensity non-ionizing radiation: An international perspective. Environ. Pollut. 2018, 242, 643–658. [Google Scholar] [CrossRef]
  5. Marko, M. Dosimetry in Bioelectromagnetics, 1st ed.; Marko, M., Ed.; Taylor & Francis Group, LLC CRC: New York, NY, USA, 2017; ISBN 9781315154572. [Google Scholar]
  6. Hardell, L. World health organization, radiofrequency radiation and health—A hard nut to crack (Review). Int. J. Oncol. 2017, 51, 405–413. [Google Scholar] [CrossRef] [Green Version]
  7. Blackman, C. Cell phone radiation: Evidence from ELF and RF studies supporting more inclusive risk identification and assessment. Pathophysiology 2009, 16, 205–216. [Google Scholar] [CrossRef] [PubMed]
  8. Biolnitiative Working Group; Cindy, S.; Carpenter, D.O. Biolnitiative Report: A Rationale for Biologically-Based Exposure Standards for Low-Intensity Electromagnetic Radiation. 2012. Available online: www.bioinitiative.org (accessed on 14 July 2021).
  9. European Commission. Non-binding Guide to Good Practice for Implementing Directive 2013/35/EU Electromagnetic Fields; European Commission: Luxemburg, 2014; ISBN 978-92-79-45869-9. [Google Scholar]
  10. Johansson, O. Disturbance of the immune system by electromagnetic fields-A potentially underlying cause for cellular damage and tissue repair reduction which could lead to disease and impairment. Pathophysiology 2009, 16, 157–177. [Google Scholar] [CrossRef] [PubMed]
  11. SCENIHR (Scientific Committee on Emerging and Newly Identified Health Risks). Opinion on: Potential Health Effects of Exposure to Electromagnetic Fields (EMF); European Commission DG SANTE: Luxemburg, 2015; ISBN 978-92-79-30134-6. [Google Scholar]
  12. Hardell, L.; Sage, C. Biological effects from electromagnetic field exposure and public exposure standards. Biomed. Pharmacother. 2008, 62, 104–109. [Google Scholar] [CrossRef]
  13. Vecchio, F.; Buffo, P.; Sergio, S.; Iacoviello, D.; Rossini, P.M.; Babiloni, C. Mobile phone emission modulates event-related desynchronization of alpha rhythms and cognitive-motor performance in healthy humans. Clin. Neurophysiol. 2012, 123, 121–128. [Google Scholar] [CrossRef] [PubMed]
  14. Havas, M.; Marrongelle, J.; Pollner, B.; Kelley, E.; Rees, C.R.G.; Tully, L. Provocation study using heart rate variability shows microwave radiation from 2.4 GHz cordless phone affects autonomic nervous system. Eur. J. Oncol. Library 2010, 5, 273–300. [Google Scholar]
  15. Morgan, L.L.; Miller, A.B.; Sasco, A.; Davis, D.L. Mobile phone radiation causes brain tumors and should be classified as a probable human carcinogen (2A) (Review). Int. J. Oncol. 2015, 46, 1865–1871. [Google Scholar] [CrossRef] [Green Version]
  16. Celaya-Echarri, M.; Azpilicueta, L.; Ramos, V.; Lopez-Iturri, P.; Falcone, F. Empirical and Modeling Approach for Environmental Indoor RF-EMF Assessment in Complex High-Node Density Scenarios: Public Shopping Malls Case Study. IEEE Access 2021, 9, 46755–46775. [Google Scholar] [CrossRef]
  17. Schmid, M.R.; Loughran, S.P.; Regel, S.J.; Murbach, M.; Bratic Grunauer, A.; Rusterholz, T.; Bersagliere, A.; Kuster, N.; Achermann, P. Sleep EEG alterations: Effects of different pulse-modulated radio frequency electromagnetic fields. J. Sleep Res. 2012, 21, 50–58. [Google Scholar] [CrossRef]
  18. Sepehrimanesh, M.; Davis, D.L. Proteomic impacts of electromagnetic fields on the male reproductive system. Comp. Clin. Path. 2017, 26, 309–313. [Google Scholar] [CrossRef]
  19. Jelodar, G.; Talebzadeh, M.R.; Lari, M.A. Effect of short-term exposure to radio frequency emitted by base transceiver station (BTS) antenna on epididymal sperms. Comp. Clin. Path. 2012, 21, 1285–1290. [Google Scholar] [CrossRef]
  20. Phillips, J.L.; Singh, N.P.; Lai, H. Electromagnetic fields and DNA damage. Pathophysiology 2009, 16, 79–88. [Google Scholar] [CrossRef]
  21. Havas, M. When theory and observation collide: Can non-ionizing radiation cause cancer? Environ. Pollut. 2017, 221, 501–505. [Google Scholar] [CrossRef] [PubMed]
  22. Hardell, L.; Carlberg, M.; Hansson Mild, K. Use of mobile phones and cordless phones is associated with increased risk for glioma and acoustic neuroma. Pathophysiology 2013, 20, 85–110. [Google Scholar] [CrossRef]
  23. Hardell, L.; Carlberg, M.; Söderqvist, F.; Mild, K.H. Case-control study of the association between malignant brain tumours diagnosed between 2007 and 2009 and mobile and cordless phone use. Int. J. Oncol. 2013, 43, 1833–1845. [Google Scholar] [CrossRef] [Green Version]
  24. Hardell, L.; Carlberg, M.; Söderqvist, F.; Mild, K.H. Pooled analysis of case-control studies on acoustic neuroma diagnosed 1997-2003 and 2007-2009 and use of mobile and cordless phones. Int. J. Oncol. 2013, 43, 1036–1044. [Google Scholar] [CrossRef] [Green Version]
  25. Hardell, L.; Carlberg, M.; Hansson Mild, K.; Eriksson, M. Case-control study on the use of mobile and cordless phones and the risk for malignant melanoma in the head and neck region. Pathophysiology 2011, 18, 325–333. [Google Scholar] [CrossRef] [PubMed]
  26. International Agency for Research on Cancer (IARC). Non-Ionizing Radiation, Part 2: Radiofrequency Electromagnetic Fields; International Agency for Research on Cancer: Lyon, France, 2012; ISBN 978-92-832-1325-3. [Google Scholar]
  27. Ruediger, H.W. Genotoxic effects of radiofrequency electromagnetic fields. Pathophysiology 2009, 16, 89–102. [Google Scholar] [CrossRef]
  28. López, I.; Félix, N.; Rivera, M.; Alonso, A.; Maestú, C. What is the radiation before 5G? A correlation study between measurements in situ and in real time and epidemiological indicators in Vallecas, Madrid. Environ. Res. 2021, 194, 110734. [Google Scholar] [CrossRef] [PubMed]
  29. Bhatt, C.R.; Redmayne, M.; Abramson, M.J.; Benke, G. Instruments to assess and measure personal and environmental radiofrequency-electromagnetic field exposures. Australas. Phys. Eng. Sci. Med. 2016, 39, 29–42. [Google Scholar] [CrossRef] [PubMed]
  30. Bolte, J.F.B. Lessons learnt on biases and uncertainties in personal exposure measurement surveys of radiofrequency electromagnetic fields with exposimeters. Environ. Int. 2016, 94, 724–735. [Google Scholar] [CrossRef] [PubMed]
  31. International Commission on Non – Ionizing Radiation Protection (ICNIRP). Guidelines for Limiting Exposure to Electromagnetic Fields (100 kHz to 300 GHz). Health Phys. 2020, 118, 483–524. [Google Scholar] [CrossRef]
  32. IEEE. IEEE Standard for Safety Levels with Respect to Human Exposure to Radio Frequency Electromagnetic Fields, 3 kHz to 300 GHz; IEEE International Committee on Electromagnetic Safety (SCC39): New York, NY, USA, 2006; ISBN 0-7381-4835-0-SS95389. [Google Scholar]
  33. Mann, S. Assessing personal exposures to environmental radiofrequency electromagnetic fields. Comptes Rendus Phys. 2010, 11, 541–555. [Google Scholar] [CrossRef]
  34. Röösli, M.; Frei, P.; Bolte, J.; Neubauer, G.; Cardis, E.; Feychting, M.; Gajsek, P.; Heinrich, S.; Joseph, W.; Mann, S.; et al. Conduct of a personal radiofrequency electromagnetic field measurement study: Proposed study protocol. Environ. Heal. A Glob. Access Sci. Source 2010, 9, 1–14. [Google Scholar] [CrossRef] [Green Version]
  35. Frei, P.; Mohler, E.; Bürgi, A.; Fröhlich, J.; Neubauer, G.; Braun-Fahrländer, C.; Röösli, M. Classification of personal exposure to radio frequency electromagnetic fields (RF-EMF) for epidemiological research: Evaluation of different exposure assessment methods. Environ. Int. 2010, 36, 714–720. [Google Scholar] [CrossRef] [Green Version]
  36. Bretchko, P.; Ludwig, R. RF Circuit Design—Theory and Applications; Prentice-Hall, Inc.: Hoboken, NJ, USA, 2007; pp. 201–253. ISBN 978-0-13-147137-5. [Google Scholar]
  37. Oppenheim, A.V.; Willsky, A.S.; Nawab, S.H. Signals and Systems, 2nd ed.; Prentice-Hall, Inc.: Hoboken, NJ, USA, 1998; ISBN 0-13-814757-4. [Google Scholar]
  38. Kishk, A.A. Fundamentals of Antennas; McGraw Hill: New York, NY, USA, 2009; pp. 1–30. [Google Scholar]
  39. ITU. Calculation of Free-Space Attenuation; Recommendation ITU-R P.525-4. International Telecommunication Union (ITU) Std., 2019; Volume 4. Available online: http://www.itu.int/publ/R-REC/en/ (accessed on 25 July 2021).
  40. Aznar, Á.C.; Robert, J.R.; Casals, J.M.R.; Roca, L.J.; Boris, S.B.; Bataller, M.F. Antenas; Univ. Politèc. de Catalunya: Barcelona, Spain, 1998; ISBN 84-8301-625-7. [Google Scholar]
  41. Ministerio de Industria Energia y Turismo Información pública de telefonía móvil. Available online: https://geoportal.minetur.gob.es/VCTEL/vcne.do (accessed on 29 March 2021).
  42. Benesty, J.; Chen, J.; Huang, Y.; Cohen, I. Pearson Correlation Coefficient. In Noise Reduction in Speech Processing; Springer: Berlin/Heidelberg, Germany, 2009; ISBN 978-3-642-00296-0. [Google Scholar]
Figure 1. Block diagram of the data acquisition system.
Figure 1. Block diagram of the data acquisition system.
Sensors 21 07395 g001
Figure 2. Flow chart for the measurement process.
Figure 2. Flow chart for the measurement process.
Sensors 21 07395 g002
Figure 3. Top view of the proposed system: (a) RF PCB; (b) control PCB.
Figure 3. Top view of the proposed system: (a) RF PCB; (b) control PCB.
Sensors 21 07395 g003
Figure 4. Block diagram of the calibration system.
Figure 4. Block diagram of the calibration system.
Sensors 21 07395 g004
Figure 5. Measurement points in the residential area: In blue, the measurement points, and in red, the location of the base stations.
Figure 5. Measurement points in the residential area: In blue, the measurement points, and in red, the location of the base stations.
Sensors 21 07395 g005
Figure 6. FSH8 portable spectrum analyzer and TSEMF-B2 omnidirectional antenna (Munich, Germany, ROHDE&SCHWARDZ).
Figure 6. FSH8 portable spectrum analyzer and TSEMF-B2 omnidirectional antenna (Munich, Germany, ROHDE&SCHWARDZ).
Sensors 21 07395 g006
Figure 7. Comparative error bar graph between the FSH8 spectrum analyzer and the proposed exposimeter.
Figure 7. Comparative error bar graph between the FSH8 spectrum analyzer and the proposed exposimeter.
Sensors 21 07395 g007
Figure 8. Correlation between the commercial equipment FSH8 and the exposimeter.
Figure 8. Correlation between the commercial equipment FSH8 and the exposimeter.
Sensors 21 07395 g008
Table 1. Characteristics of different types of exposimeters.
Table 1. Characteristics of different types of exposimeters.
Frequency Band (MHz)ESM-140Expo M-RFEME Spy-120EME Spy-121EME Spy-140EME Spy-200
FM (87–107)
TV3 (174–223)
TETRA I (380–400)
TETRA II (410–430)
TETRA III (450–470)
TV4&5 (470–770)
LTE 800 DL (791–821)
LTE 800 UL (832–862)
GSM UMTS 900 UL (880–925)
GSM UMTS 900 DL (925–960)
GSM 1800 UL (1710–1785)
GSM 1800 DL (1805–1880)
DECT (1880–1900)
UMTS 2100 UL (1920–1980)
UMTS 2100 DL (2110–2170)
Wifi 2G (2400–2484)
LTE 2600 UL (2500–2570)
LTE 2600 DL (2620–2690)
WiMAX (3300–3900)
Wifi 5G (5150–5850)
DetectorLogRMSLogLogRMSRMS
Sampling Interval (s)0.5–103–60004–2554–2554–2554–255
Range (V/m)0.01–700.003–50.05–100.05–100.005–100.005–5
Lower detection limit (V/m) 0.0030.050.050.010.005
✗: Does not cover the band ✓: Cover the band.
Table 2. Features between exposimeters and spectrum analyzers.
Table 2. Features between exposimeters and spectrum analyzers.
FeatureSpectrum AnalyzerExposimeters
Full spectrum measurement.YesNo (20 bands only)
Detection of multiple sources in the same communications band.HighNo (estimation)
Cross-talk detectionHighNo (estimation)
SizeBigCompact
CostHighLow
PortabilityLowHigh
User-friendlinessDifficult in dynamic environmentsSuitable for dynamic environments
Table 3. Fractal antennas used and their respective bandwidths.
Table 3. Fractal antennas used and their respective bandwidths.
FRACTUSANTENAS ModelsBandwidth (MHz)
FR01-B3-W-0-05578–108
FR01-B3-V-0-054180–220
FR01-B1-S-0-047470–1675
FR01-S4-250698–2690
FR05-S1-N0-1-0042400–5875
Table 4. Specifications of the proposed system.
Table 4. Specifications of the proposed system.
SpecificationValue
Operating Range78 MHz–6 GHz
Sampling Rate20k samples/s
Dynamic Range90 dB
Minimum Input Power−70 dBm
Maximum Input Power20 dBm
Sweeping time902 ms
Time memory space1 week
Current Consumption450 mA
Battery autonomyhours
Table 5. Frequency bands of the telephone base stations near the measurement points.
Table 5. Frequency bands of the telephone base stations near the measurement points.
Frequency BandFrequency (MHz)
GSM 800791–862
GSM 900880–960
GSM 18001710–1879
GSM 19001900–1980
GSM 21002110–2170
Table 6. Highest power peaks at each measurement point and correlation factor.
Table 6. Highest power peaks at each measurement point and correlation factor.
Frequency BandMP 1 1 (µW/cm2)MP 1 2 (µW/cm2)MP 1 3 (µW/cm2)MP 1 4 (µW/cm2)MP 1 5 (µW/cm2)MP 1 6 (µW/cm2)MP 1 7 (µW/cm2)
GSM 800SA 20.0051360.0053210.0002150.0012530.000037 *0.0020380.1437
E 30.0056610.0029270.0002080.0012010.000021 *0.0013640.1431
GSM 900SA 20.0000170.0000150.0001350.0007810.0011480.0038050.08302
E 30.000002 *0.000002 *0.0001170.0012210.0017090.0029870.0729
GSM 1800SA 20.000006 *0.021610.0045230.000013 *0.000003 *0.0014770.4644
E 30.000008 *0.012530.002680.00001 *0.000008 *0.0018520.4025
GSM 1900SA 20.016910.0047120.000022 *0.0002830.0001230.000013 *0.000015 *
E 30.013090.002630.000055 *0.0002530.0000830.000059 *0.000048 *
GSM 2100SA 20.0001810.000023 *0.0001980.0002140.000005 *0.0002130,187
E 30.000013 *0.000019 *0.000063 *0.000078 *0.000012 *0.000085 *0.1252
Wifi 2GSA 20.0019930.0017880.0001640.0003020.0002960.0011920.000076 *
E 30.0018910.0072440.0004550.0001120.0002810.0030390.1043
Wifi 5GSA 20.000072 *0.0020440.0001850.00085 *0.0010240.0081160.000089 *
E 30.000056 *0.0022460.000046 *0.00058 *0.0007860.0068910.000055 *
Correlation factor0.99160.91950.92620.92900.89710.89700.9680
* Noise. 1 MP: Measuring Point. 2 SA: Spectrum Analyzer. 3 E: proposed Exposimeter.
Table 7. Frequencies of the power peaks of each measurement point and its correlation factor.
Table 7. Frequencies of the power peaks of each measurement point and its correlation factor.
Frequency BandMP 1 1 (MHz)MP 1 2 (MHz)MP 1 3 (MHz)MP 1 4 (MHz)MP 1 5 (MHz)MP 1 6 (MHz)MP 1 7 (MHz)
GSM 800SA 2784776809809824 *809809
E 3745736810800738 *800809
GSM 900SA 2935 *927 *927852893935927
E 3949 *957 *928851864951928
GSM 1800SA 21841 *171817681777 *1815 *18361853
E 31865 *172717891725 *1731 *18471851
GSM 1900SA 2188618951931 *186119031903 *1908 *
E 3188218881975 *189218511947 *1969 *
GSM 2100SA 22135 *2144 *212221222158 *21302147
E 32169 *2169 *2138 *21012169 *21412141
Wifi 2GSA 22416244124162438247524672441
E 32469244824612441246124362481
Wifi 5GSA 25735 *522651925293 *522652775345 *
E 35775 *52155623 *5585 *521852705658 *
Correlation factor0.99990.99990.99840.99860.99860.99870.9987
* Noise. 1 MP: Measuring Point. 2 SA: Spectrum Analyzer. 3 E: proposed Exposimeter.
Table 8. Kolmogorov Smirnov normality test.
Table 8. Kolmogorov Smirnov normality test.
InstrumentNFactor Significance
Spectrum analyzer490.200
Exposimeter490.200
Table 9. Levene test for equality of variances and Student’s t-test.
Table 9. Levene test for equality of variances and Student’s t-test.
Equal Variances
Levene TestSignificance Factor0.525
Student’s t-testSignificance Factor0.948
Mean Difference0.01671
Standard error difference0.2569
95% CI 1 Lower−0.4933
95% CI 1 Higher0.5267
1 95% CI: 95% Confidence Interval.
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Rivera González, M.X.; Félix González, N.; López, I.; Ochoa Zambrano, J.S.; Miranda Martínez, A.; Maestú Unturbe, C. Compact Exposimeter Device for the Characterization and Recording of Electromagnetic Fields from 78 MHz to 6 GHz with Several Narrow Bands (300 kHz). Sensors 2021, 21, 7395. https://0-doi-org.brum.beds.ac.uk/10.3390/s21217395

AMA Style

Rivera González MX, Félix González N, López I, Ochoa Zambrano JS, Miranda Martínez A, Maestú Unturbe C. Compact Exposimeter Device for the Characterization and Recording of Electromagnetic Fields from 78 MHz to 6 GHz with Several Narrow Bands (300 kHz). Sensors. 2021; 21(21):7395. https://0-doi-org.brum.beds.ac.uk/10.3390/s21217395

Chicago/Turabian Style

Rivera González, Marco Xavier, Nazario Félix González, Isabel López, Juan Sebastián Ochoa Zambrano, Andrés Miranda Martínez, and Ceferino Maestú Unturbe. 2021. "Compact Exposimeter Device for the Characterization and Recording of Electromagnetic Fields from 78 MHz to 6 GHz with Several Narrow Bands (300 kHz)" Sensors 21, no. 21: 7395. https://0-doi-org.brum.beds.ac.uk/10.3390/s21217395

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop