Communication

A Sensor for Characterisation of Liquid Materials with High Permittivity and High Dielectric Loss

1
School of Physics and Electronic Information, Anhui Normal University, Wuhu 241002, China
2
School of Electronic and Information, Anhui University, Hefei 230601, China
3
The State Key Laboratory of Complex Electromagnetic Environment Effects on Electronic and Information System, Luoyang 471004, China
4
The National Physical Laboratory, Teddington TW11 0LW, UK
*
Author to whom correspondence should be addressed.
Submission received: 18 January 2022 / Revised: 21 February 2022 / Accepted: 22 February 2022 / Published: 24 February 2022
(This article belongs to the Special Issue RF Sensors: Design, Optimization and Applications)

Abstract

:
This paper reports on a sensor based on multi-element complementary split-ring resonator for the measurement of liquid materials. The resonator consists of three split rings for improved measurement sensitivity. A hole is fabricated at the centre of the rings to accommodate a hollow glass tube, through which the liquid sample can be injected. Electromagnetic simulations demonstrate that both the resonant frequency and quality factor of the sensor vary considerably with the dielectric constant and loss tangent of the liquid sample. The volume ratio between the liquid sample and glass tube is 0.36, yielding great sensitivity in the measured results for high loss liquids. Compared to the design based on rectangular split rings, the proposed ring structure offers 37% larger frequency shifts and 9.1% greater resonant dips. The relationship between dielectric constant, loss tangent, measured quality factor and resonant frequency is derived. Experimental verification is conducted using ethanol solution with different concentrations. The measurement accuracy is calculated to be within 2.8%, and this validates the proposed approach.

1. Introduction

Complex permittivity is one fundamental electromagnetic property of dielectric materials [1]. Accurate dielectric measurements play an important role in applications such as manufacturing [2], processing [3], antenna design [4] and aerospace technology [5], etc. Usually, the permittivity needs to be known as a priori in, for example, the design of microwave planar circuits. Compared to solid materials, the permittivity of liquid materials is prone to impacts from factors such as temperature, moisture, contaminants in the measurement sample holder, air pressure, and so on [6,7]. In addition, the measurement of a liquid sample is less convenient due to its fluid nature. Many liquid samples usually consist of polar molecules, leading to high dielectric constant and loss of properties. In this regard, in situ measurement of dielectric constant and loss tangent for liquid materials is in active demand.
Complex permittivity may be expressed as ε r = ε r j ε r = ε r ( 1 j tan δ ) , where ε r is the real part and often referred to as dielectric constant, ε r is the imaginary part and is called the loss factor, while tan δ is the loss tangent characterising the dielectric loss. The precise measurement of complex permittivity requires the determination of both the dielectric constant and loss tangent with sufficiently high accuracy.
There exist various types of measurement methods for liquid samples, including capacitance method, transmission line method, free-space method, and resonant cavity method. A cross capacitance method was developed for the accurate measurement of liquids of low dielectric constants in Ref. [8]. In general, capacitance methods are suitable for low frequency measurements. The transmission line method requires that the material be placed in a part of the enclosed transmission line during measurements [9,10]. It also requires precisely manufactured samples to fit the transmission line [11]. In addition, the injected liquid sample in an enclosed transmission line always faces the challenge of sealing the sample without leakage. The free space method uses free space as the transmission medium and is more suitable for dielectric measurement in the millimeter wave range [12]. The resonant cavity method is sensitive to low-loss and low-permittivity samples with high accuracy. The traditional resonant method usually creates a higher mode for high permittivity samples [13]. Moreover, high-loss samples usually degrade the quality factor, resulting in a considerable reduction of measurement accuracy [14]. In summary, new dielectric measurement techniques shall be developed for high-permittivity and high-loss liquid samples with sufficiently high sensitivity.
In recent years, microwave sensors using the resonance method have become increasingly popular. Structures such as the split-ring resonator (SRR), complementary spiral resonator (CSR), and complementary split ring resonator (CSRR) are the most commonly used ones in the design of microwave liquid sensors [15,16,17,18,19,20,21,22,23,24,25,26,27,28,29,30]. Examples of SRR structure can be found in Refs. [15,16,17,18,19,20]. The sensor in Ref. [15] must maintain the fluidity of the liquid during measurement, which may cause a lot of loss of liquid samples. Sensors reported by Kiani et al. [16] can measure the dielectric constant of liquids, but not the loss tangent. This work was later further developed by the same group to realize dual-frequency operation [18]. Juan et al. [17] developed a concentration sensor using the Q factor and the maximum S21 of the resonance as sensing magnitudes. Juan et al. [19] modified this design using certified biocompatible material. The structure was fabricated by 3D printing, which provides stronger interaction between the electromagnetic fields and the sample. The microfluid sensor reported in Ref. [20] can measure both the real and imaginary parts by using both magnitude and phase of S21. However, the accuracy is not good enough, which may be attributed to the fact that that phase is more prone to measurement errors. Interestingly, Su et al. [21] made a systematic study on defect-ground-plane based resonators. This general method provides guidelines for the design of microstrip line resonators. Refs. [22,23] reported on sensors based on CSR structures. The sensor reported by Su et al. [22] uses flexible materials but can only measure low-loss materials, while, the sensitivity of the sensor in Ref. [23] is too low and much noise can be observed during in vivo measurement. The CSRR structure can be found in Refs. [24,25,26,27,28,29,30]. The substrate integrated waveguide structure reported by Cai et al. [24] can only measure the real part of the permittivity. The sensor developed by Chen et al. [25] measured permittivity through the influence of liquid on the operating frequency of an antenna. However, the measurement was greatly influenced by its environment and the deviation is as high as 11.0%. Good sensitivity can be obtained using the sensors in Refs. [26,27,28], but the measurement range of the loss tangent is narrow, and they cannot measure corrosive liquids for direct contact between the sensors and samples. In Ref. [29], liquid samples pass through the central hole of the CSRR sensor, which maximizes the sample’s exposure to the electric field. However, it is still difficult to distinguish small changes in permittivity. Another CSRR-based liquid sensor working at 200–330 MHz is reported for dielectric measurement of different density ethanol [30]. Again, it is not able to measure corrosive liquids.
For the resonance-based method, the resonant structure acts as an LC resonator. The general description of this type of sensor can be found in Ref. [21] and a good review on this topic can be found by Juan et al. [31]. When materials are placed on top of the resonator, interaction between fields and samples will modify the resonant frequency f r and the corresponding quality factor Q. The permittivity can be retrieved from the variation in f r and Q. However, the difficulty in the measurement of high-permittivity high-loss measurement lies in its dilemma with high sensitivity. Being inspired by two pioneered designs in Ref. [29] and Ref. [32], we recognize that two approaches may be beneficial to handle this dilemma. One is to increase the field intensity in the resonating area under the condition of unit power input. The other is to use partial filled sample. By combining these methods, the sensitivity and measurement range can be simultaneously improved.
To enable measurement of high-loss polar liquids, a multi-element CSRR structure is proposed. In this design, three circular split rings are used, which will prove that ahigher regional electric field can be created. A glass tube passes through the central hole of the coaxial rings, see Figure 1. The glass tube is made of low loss fused quartz, functioning as a liquids holder and partial filling the resonator. These methods enable high sensitivity measurement for high-permittivity and high-loss liquids.
In the following sections, the performance of the circular CSRR structure and the rectangular CSRR structure will be compared in detail in Section 2. It is concluded that the circular CSRR structure has higher sensitivity than the rectangular CSRR structure. In Section 3 and Section 4, a theoretical model is established based on the simulation results, and extensive simulation analysis is carried out. Measurements and results are presented in Section 5. Section 6 summarizes this work.

2. Resonant Properties of CSRR Structures

This section provides some physical insights into CSRR. A comparison between circular and rectangular CSRR is also presented.
The CSRR structure in Ref. [29] is shown in Figure 2a, where the resonator uses three rectangular split rings. In the middle of the CSRR, a via hole is drilled to cater a glass tube. The main parameters of this structure are given in Table 1. The initial values of these parameters are chosen so that the outer ring is half wavelength at the resonant frequency. The final values are obtained through parameter sweep in HFSS software. Such a structure can be modelled as a hybrid LC resonator [33,34]. The corresponding values of each lumped element can be retrieved from full wave simulations, and the extracted values are given in Table 2. The response of the LC resonant circuit is evaluated using both circuit method and full wave simulation. It has to be mentioned that the central frequency is largely determined by the length of the split ring. Changing the length will modify the central frequency. To make a comparison to the design in Ref. [29], we chose the diameter of the circular ring to be the same as the side length of the rectangular ring.
According to the equivalent circuit diagram, the resonant frequency of the structure can be deduced as [35]
f r = 1 2 π L r ( C c + C r )
The equivalent circuit model gives good prediction to the resonant frequencies, compared to the full wave simulated results by commercial software HFSS, as shown in Figure 3. The resonant dips of the circular CSRR are several dB larger than that of the rectangular CSRR. The field distribution around the resonator is plotted in Figure 4. It can be seen from the scale bar that the peak field of the circular CSRR is 7 dB higher than that of the rectangular CSRR, indicating that better interaction between fields and sample can take place for circular CSRR. This may explain why circular CSRR has deeper resonance.

3. Sensitivity Analysis

The sensor structure and some dimensions are shown in Figure 1. The substrate is 20 mm wide and 28 mm long. Port 1 and Port 2 are connected to coaxial lines. The quartz glass tube has a length of 75 mm, an outer diameter of 1.5 mm, and an inner diameter of 0.9 mm. The volume ratio between liquid sample and glass tube is 0.36. Simulations of liquid samples of different concentrations are undertaken. During the measurement, the samples are injected into the glass tube using a syringe. Each glass tube is used for only one concentration to avoid contamination.

3.1. Resonant Frequency

To compare the sensitivities of the two structures, the resonant frequencies ( f r ) and frequency shifts ( Δ f r = f r f 0 , with f 0 defined as the resonant frequency at unloaded case) versus dielectric constant ( ε r ) are plotted in Figure 5. Firstly, the unloaded sensor with an empty glass tube is simulated. Then the glass tube filled with samples of different ε r in the range of 1–90 is simulated. The resonant frequencies and the frequency shift relative to the unloaded case are recorded.
As seen in Figure 5a, the resonant frequencies for the circular and rectangular CSRR structure are 2960 MHz and 2440 MHz, respectively, which is consistent with the results of the equivalent circuit in Figure 2. The frequency shift when ε r = 90 of the circular CSRR structure is 110 MHz, and it is 80 MHz for the rectangular CSRR structure. The frequency shift is increased by 37%. According to the definition by Juan et al. [31], the relative sensitivity (RS) in this work can be written as
R S = Δ f r f 0 ( ε r 1 ) × 100
where f 0 is the resonant frequency at unloaded case, Δ f r is the frequency shift, and ε r is the dielectric constant. Sensitivity (S) may be defined by S = ( Δ f r / f 0 ) × 100 , but the relative sensitivity seems better since RS involves the quantity to be measured. The calculated RS of circular structure is 0.041, and it is 0.036 for the rectangular structure, as shown in Figure 5b. It is seen that the sensitivity is increased by 13.9%. Therefore, it can be deduced that the sensitivity of the circular CSRR structure is better than that of the rectangular one. All other parameters being equivalent, higher RS means higher sensitivity. Actually, the increase in sensitivity may be attributed to the increase in electric intensity, as shown in Figure 4. The circular sensor concentrates more power in the central hole, where the liquid sample is placed, so that the field interaction with sample can be observed. Therefore, for measuring liquid samples with ε r in the range of 1–90, the circular CSRR structure is a better choice, particularly in the high-permittivity range.

3.2. Quality Factor

In many cases, the permittivity is expressed using dielectric constant and imaginary part. The imaginary part of the complex permittivity of the material under test (MUT) can be expressed by loss tangent, which can be written as [19]:
Q MUT = 1 tan δ = ε r ε r
where, Q MUT can be calculated by unloaded quality factor Q U and the corresponding S 21 [36]
Q MUT = Q U [ 1 10 S 21 ( dB ) 20 ]
Therefore, the loss tangent can be calculated by using the quality factor Q and the transmission coefficient S 21 . It is deduced that the larger variation in S 21 ( Δ S 21 ), the higher the measurement sensitivity.
In Figure 6, the quality factors Q and Δ S 21 of the circular CSRR structure and the rectangular CSRR structure to different loss tangent under ε r = 60 are plotted. It can be seen that the quality factors of the two structures are almost the same. However, the Δ S 21 of the circular structure is 1 dB higher than the Δ S 21 of the rectangular structure, increased by 9.7%. Considering Equation (4), it can be seen that the circular CSRR structure has higher measurement sensitivity when measuring the loss tangent.
It has to be mentioned that the comparison in this section is only for the specific case between Ref. [29] and this work, where the diameter of the circular sensor equals the side length of the square sensor. More general sensitivity needs further investigation in terms of resonance, the geometrical structure of the sensor, and even the volume ratio of the sample exposed to the sensor.

4. Simulation Analysis

Before extracting the permittivity from measured data, a systematic analysis has to be conducted so as to build a mapping model. This section describes simulation work using full wave simulation to conduct feasibility analysis and establish a mathematical model, so that the complex permittivity of the liquid sample under test can be expressed by the measured S 21 (resonant frequency and quality factor). Since the proposed CSRR sensor is designed for high-permittivity and high-loss liquid samples, ε r is set to 5–90 during the simulation, with a step size of 5, and the loss tangent is set to 0–1 with a step size of 0.1.
To ensure that the transmission coefficient S 21 is a single value function of ε r and tan δ , two groups of values ε r = ( 20 , 70 ) and tan δ = ( 0 , 0.3 , 0.6 , 0.9 ) are chosen as examples. The simulated results are plotted in Figure 7. It can be observed that: (1) higher permittivity creates lower resonant frequency, and (2) higher loss tangent reduces the resonant dips; (3) each pair of ( ε r , tan δ ) only corresponds to a single curve of S 21 . These facts verify that it is feasible to extract the complex permittivity from S 21 .

4.1. Building the Frequency Function for Lossless Samples

Since there are too many curves, only a portion of the simulated results are presented in Figure 8. For the unloaded case, the resonant frequency is 2.974 GHz. For other values of ε r , the resonant frequency shifts to lower side. This reminds us that the real part ε r of the permittivity can be retrieved by observing the frequency shift Δ f r relative to the unloaded condition for the lossless sample. From the simulated results, the frequency shift Δ f r versus ε r is plotted in Figure 9, under the condition of the lossless case ( tan δ = 0 ). The fitted quartic equation is
Δ f r = f 0 f r = 2.00691 × 10 4 + 0.00342 ε r 6.27316 × 10 5 ε r 2 + 6.83277 × 10 7 ε r 3 2.91958 × 10 9 ε r 4
Here, the coefficient of determination R 2 is 0.9998, showing a sufficiently good fit for data retrieval.

4.2. Building the Retrieval Function for Lossy Samples

In contrast, the permittivity of lossy samples involves real part and imaginary part (or loss tangent). To retrieve their permittivity, one needs to take into account both the quality factor and the corresponding resonance frequency. It can be seen from Figure 7 and Figure 8 that the resonant frequency is almost determined by the real part ε r . In light of this, only Q MUT 1 versus tan δ is shown in Figure 10. Clearly, higher loss tangent causes smaller Q MUT . The linearity for all cases is pretty good. Such an observation will make the retrieval results sufficiently accurate.
To reach a two-variable function, a three-dimensional plot is shown in Figure 11. The fitted quartic equation is:
ε r = ε r F ( Q MUT 1 , ε r ) G ( Q MUT 1 , ε r )
where
{ F ( Q MUT 1 , ε r ) = 2.75 × 10 4 3.79 × 10 2 ε r + 7.50 × 10 6 Q MUT 1 5.09 Q MUT 2 + 5.50 × 10 4 ε r Q MUT 1 G ( Q MUT 1 , ε r ) = 1 + 1.84 × 10 2 ε r + 2.189 × 10 5 Q MUT 1 + 0.26 ( ε r ) 2 3.00 Q MUT 2 1.98 × 10 4 ε r Q MUT 1
The RationalTaylor model [37] was selected to fit all points, so that the nonlinear surface covers all points in the coordinates as much as possible. In this case, the coefficient of determination R2 is 0.99943. This is also good enough for data retrieval. By using Equations (6) and (7), the complex permittivity of the sample can be readily restored.

5. Measurements and Results

Figure 12a,b shows photos of the sensor, which was fabricated using planar PCB technology. The sensor was made from RO3035 substrate with a thickness of 0.75 mm and a dielectric constant of 3.5 and loss tangent of 0.0015. The middle glass tube is made of fused quartz having a dielectric constant of 3.78. The tube has a diameter of 1.5 mm, a length of 75 mm, and an inner diameter of 0.3 mm. The bottom of the glass tube is sealed, and the upper end is open, which can be used to inject liquid samples into the tube. A pair of 50 Ω SMA connectors are soldered onto both ends of the sensor.
Measurement was conducted using a vector network analyzer (VNA) Ceyear AV3672D. The VNA was calibrated using a standard short-load-open-thru (SLOT) method with a 3.5 mm calibration kit. The measurement frequency range was from 2.75 GHz to 3 GHz. After calibration, the S 21 of unloaded situation was first measured. Then, the liquid samples were injected into the glass tube through syringes. In this study, the liquid sample was diluted ethanol solution with a mass concentration of 20%, 50%, 70% and 95%. These samples were injected into different glass tubes to avoid contamination. Placing the quartz glass tube filled with the measured sample to the CSRR center through the hole of the sensor and recording the S 21 for each sample. The measured data are plotted in Figure 13. From the measured data, the resonant frequency and the quality factor can be deduced. Therefore, the dielectric constant and loss tangent can be calculated by using the fitting Equations (5)–(7).
The measured data are plotted to Figure 14 to better show the tendency of S 21 with the concentration. With the increase in ethanol concentration, the resonant frequency shifts upwards, and the loss shows a decreasing tendency, which is consistent with simulations.
The dielectric constant change due to the concentration of the liquid sample can be explained using mixture theory. The mixture of water and methanol shows concentration dependent property. Water has larger permittivity and methanol has smaller permittivity. Therefore, with the increase in methanol concentration, the effective permittivity will decrease. Considering that the resonant frequency is inversely proportional to the square of the permittivity, it is seen that the resonant frequency moves towards a lower frequency side with the increase in concentration.
The values are substituted into Equations (5)–(7) and the one that obtained the complex permittivity of the sample, as shown in Table 3. The measured results have a good agreement with values reported in Ref. [38]. The calculated accuracy in comparison to the cited values is as good as 2.8%, in contrast to the values in Ref. [38]. It is noted that the worst case is the imaginary part for pure water, as seen in Table 3. This figure (2.8%) is used to represent the accuracy.
Table 4 shows the performance comparison between the proposed sensor of this work and other similar microwave dielectric sensors reported in the literature. For Refs. [24,25], the main drawback is that only the real part can be measured. The accuracy of the sensor in Ref. [25] is 11%, which is even worse than a non-resonance method. For Refs. [26,27,28], samples are in direct contact with the sensor, making them unable to measure corrosive samples. It is observed that the resonance dip becomes worse when the loss tangent increases. In comparison, this design uses a partial-filled glass tube, which is helpful to increase the resonance dip for high loss media. Compared to Ref. [38], this work shows higher relative sensitivity and provides 2.8% measurement accuracy.
From the comparison, it is seen that the proposed design is suitable for measurement of high-permittivity and high loss liquid samples. However, it has to be mentioned that for low-loss liquid samples, the accuracy is yet to be verified. In addition, since the sample is contained in a glass tube, samples are not in direct contact with sensors, which will decrease the sensitivity for low-loss samples. However, for high-loss samples, this is a good method. With this connection, there are many aspects for further study, such as the development of a sensor for the measurement of both high-loss and low-loss samples. Other topics may include the design of a sensor for very small volume samples.

6. Conclusions

A multi-element split-ring resonator has been designed, fabricated and verified using diluted ethanol samples for the dielectric measurement of liquids. This method used three circular ring resonators to enhance the local field so that more field interaction with samples can take place. A low-loss fused quartz was used to hold liquid so that high-loss liquid only occupies 36% volume, which enables the measurement of high-loss samples. By combining these two measures, it was demonstrated that water and diluted samples can be measured with an accuracy of 2.8%. In addition, an analytical function has been developed, facilitating the direct recovery of permittivity from measured S21. This method is also useful for the measurement of biological and corrosion samples. Further development may include systematic study on influential factors on sensitivity, and the development of sensors for the measurement of samples covering a wider range of loss and dielectric constants.

Author Contributions

Conceptualization, X.L. and X.S.; methodology, C.W.; validation, S.Y.; investigation, Z.H.; data curation, C.W.; writing—original draft preparation, C.W.; writing—review and editing, X.L. and X.S.; visualization, C.W.; software, X.Y.; funding acquisition, X.L., X.Y. and Z.H. All authors have read and agreed to the published version of the manuscript.

Funding

This work is funded in part by the University Synergy Innovation Program of Anhui Province, GXXT-2021-027, the National Natural Science Foundation of China under the contract number of 61871003 and the open project of the state key laboratory of complex electromagnetic environment effects on electronics and information system under grant number CEMEE2022Z0201B. The work of X. Shang was supported by the National Measurement System Programme of the UK government’s Department for Business, Energy & Industrial Strategy (BEIS).

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Not applicable.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Paul, R.B.; Engin, A.E.; Aguirre, J. Dielectric and Underfill Characterization Using Cavity Resonators for Millimeter-Wave Applications. IEEE Lett. Electromagn. Compat. Pract. Appl. 2020, 2, 76–80. [Google Scholar] [CrossRef]
  2. Bor, J.; Lafond, O.; Merlet, H.; Le Bars, P.; Himdi, M. Technological Process to Control the Foam Dielectric Constant Application to Microwave Components and Antennas. IEEE Trans. Compon. Packag. Manuf. Technol. 2014, 4, 938–942. [Google Scholar] [CrossRef] [Green Version]
  3. Okada, K.; Kurimoto, K.; Suzuki, M. Anomalous TDDB Statistics of Gate Dielectrics Caused by Charging-Induced Dynamic Stress Relaxation Under Constant–Voltage Stress. IEEE Trans. Electron Devices 2016, 63, 2268–2274. [Google Scholar] [CrossRef]
  4. Monkevich, J.M.; Le Sage, G.P. Design and Fabrication of a Custom-Dielectric Fresnel Multi-Zone Plate Lens Antenna Using Additive Manufacturing Techniques. IEEE Access 2019, 7, 61452–61460. [Google Scholar] [CrossRef]
  5. Zhengyi, Z.; Wei, F.; Yunfei, C.; Ning, G. Adaptive Scheduling for Millimeter Wave Multi-Beam Satellite Communication Systems. J. Commun. Inf. Netw. 2016, 1, 42–50. [Google Scholar] [CrossRef]
  6. Abedian, B.; Baker, K.N. Temperature Effects on the Electrical Conductivity of Dielectric Liquids. IEEE Trans. Dielectr. Electr. Insul. 2008, 15, 888–892. [Google Scholar] [CrossRef]
  7. Bobowski, J.S.; Clements, A.P. Permittivity and Conductivity Measured Using a Novel Toroidal Split-Ring Resonator. IEEE Trans. Microw. Theory Tech. 2017, 65, 2132–2138. [Google Scholar] [CrossRef] [Green Version]
  8. Rahman, O.; Islam, T.; Ahmad, A.; Parveen, S.; Khera, N.; Khan, S.A. Cross Capacitance Sensor for Insulation Oil Testing. IEEE Sens. J. 2021, 21, 20980–20989. [Google Scholar] [CrossRef]
  9. Piekarz, I.; Sorocki, J.; Wincza, K.; Gruszczynski, S. Liquids Permittivity Measurement Using Two-Wire Transmission Line Sensor. IEEE Sens. J. 2018, 18, 7458–7466. [Google Scholar] [CrossRef]
  10. Meyne née Haase, N.; Fuge, G.; Trieu, H.K.; Zeng, A.-P.; Jacob, A.F. Miniaturized Transmission-Line Sensor for Broadband Dielectric Characterization of Biological Liquids and Cell Suspensions. IEEE Trans. Microw. Theory Tech. 2015, 63, 3026–3033. [Google Scholar] [CrossRef]
  11. Ye, D.; Islam, M.S.; Yu, G.; Wang, P. A Single-Line Single-Channel Method with Closed-Form Formulas for the Characterization of Dielectric Liquids. IEEE Trans. Microw. Theory Tech. 2019, 67, 2443–2450. [Google Scholar] [CrossRef]
  12. Sasaki, K.; Segawa, H.; Mizuno, M.; Wake, K.; Watanabe, S.; Hashimoto, O. Development of the Complex Permittivity Measurement System for High-Loss Biological Samples Using the Free Space Method in Quasi-Millimeter and Millimeter Wave Bands. Phys. Med. Biol. 2013, 58, 1625–1633. [Google Scholar] [CrossRef]
  13. Alvarez, J.O.; Peñaranda-Foix, F.L.; Catalá-Civera, J.M.; Gutiérrez-Cano, J.D. Permittivity Spectrum of Low-Loss Liquid and Powder Geomaterials Using Multipoint Reentrant Cavities. IEEE Trans. Geosci. Remote Sens. 2020, 58, 3097–3112. [Google Scholar] [CrossRef]
  14. Massoni, E.; Siciliano, G.; Bozzi, M.; Perregrini, L. Enhanced Cavity Sensor in SIW Technology for Material Characterization. IEEE Microw. Wirel. Compon. Lett. 2018, 28, 948–950. [Google Scholar] [CrossRef]
  15. Withayachumnankul, W.; Jaruwongrungsee, K.; Tuantranont, A.; Fumeaux, C.; Abbott, D. Metamaterial-Based Microfluidic Sensor for Dielectric Characterization. Sens. Actuators A Phys. 2013, 189, 233–237. [Google Scholar] [CrossRef] [Green Version]
  16. Kiani, S.; Rezaei, P.; Navaei, M. Dual-Sensing and Dual-Frequency Microwave SRR Sensor for Liquid Samples Permittivity Detection. Measurement 2020, 160, 107805. [Google Scholar] [CrossRef]
  17. Juan, C.G.; Bronchalo, E.; Potelon, B.; Quendo, C.; Ávila-Navarro, E.; Sabater-Navarro, J.M. Concentration Measurement of Microliter-Volume Water–Glucose Solutions Using Q Factor of Microwave Sensors. IEEE Trans. Instrum. Meas. 2019, 68, 2621–2634. [Google Scholar] [CrossRef]
  18. Kiani, S.; Rezaei, P.; Fakhr, M. Dual-Frequency Microwave Resonant Sensor to Detect Noninvasive Glucose-Level Changes Through the Fingertip. IEEE Trans. Instrum. Meas. 2021, 70, 6004608. [Google Scholar] [CrossRef]
  19. Juan, C.G.; Potelon, B.; Quendo, C.; García-Martínez, H.; Ávila-Navarro, E.; Bronchalo, E.; Sabater-Navarro, J.M. Study of Qᵤ-Based Resonant Microwave Sensors and Design of 3-D-Printed Devices Dedicated to Glucose Monitoring. IEEE Trans. Instrum. Meas. 2021, 70, 8005716. [Google Scholar] [CrossRef]
  20. Vélez, P.; Su, L.; Grenier, K.; Mata-Contreras, J.; Dubuc, D.; Martín, F. Microwave Microfluidic Sensor Based on a Microstrip Splitter/Combiner Configuration and Split Ring Resonators (SRRs) for Dielectric Characterization of Liquids. IEEE Sens. J. 2017, 17, 6589–6598. [Google Scholar] [CrossRef] [Green Version]
  21. Su, L.; Muñoz-Enano, J.; Vélez, P.; Martel, J.; Medina, F.; Martí, F. On the Modeling of Microstrip Lines Loaded With Dumbbell Defect-Ground-Structure (DB-DGS) and Folded DB-DGS Resonators. IEEE Access 2021, 9, 150878–150888. [Google Scholar] [CrossRef]
  22. Su, L.; Huang, X.; Guo, W.; Wu, H. A Flexible Microwave Sensor Based on Complementary Spiral Resonator for Material Dielectric Characterization. IEEE Sens. J. 2020, 20, 1893–1903. [Google Scholar] [CrossRef]
  23. Kandwal, A.; Nie, Z.; Igbe, T.; Li, J.; Liu, Y.; Liu, L.W.; Hao, Y. Surface Plasmonic Feature Microwave Sensor with Highly Confined Fields for Aqueous-Glucose and Blood-Glucose Measurements. IEEE Trans. Instrum. Meas. 2021, 70, 8000309. [Google Scholar] [CrossRef]
  24. Cai, J.; Zhou, Y.J.; Yang, X.M. A Metamaterials-Loaded Quarter Mode SIW Microfluidic Sensor for Microliter Liquid Characterization. J. Electromagn. Waves Appl. 2019, 33, 261–271. [Google Scholar] [CrossRef]
  25. Chen, Y.; Hua, C.; Shen, Z. Circularly Polarized UHF RFID Tag Antenna for Wireless Sensing of Complex Permittivity of Liquids. IEEE Sens. J. 2021, 21, 26746–26754. [Google Scholar] [CrossRef]
  26. Jiang, Q.; Yu, Y.; Zhao, Y.; Zhang, Y.; Liu, L.; Li, Z. Ultra-Compact Effective Localized Surface Plasmonic Sensor for Permittivity Measurement of Aqueous Ethanol Solution with High Sensitivity. IEEE Trans. Instrum. Meas. 2021, 70, 6008709. [Google Scholar] [CrossRef]
  27. Fan, L.-C.; Zhao, W.-S.; Wang, D.-W.; Liu, Q.; Chen, S.; Wang, G. An Ultrahigh Sensitivity Microwave Sensor for Microfluidic Applications. IEEE Microw. Wirel. Compon. Lett. 2020, 30, 1201–1204. [Google Scholar] [CrossRef]
  28. Buragohain, A.; Mostako, A.T.T.; Das, G.S. Low-Cost CSRR Based Sensor for Determination of Dielectric Constant of Liquid Samples. IEEE Sens. J. 2021, 21, 27450–27457. [Google Scholar] [CrossRef]
  29. Chuma, E.L.; Iano, Y.; Fontgalland, G.; Bravo Roger, L.L. Microwave Sensor for Liquid Dielectric Characterization Based on Metamaterial Complementary Split Ring Resonator. IEEE Sens. J. 2018, 18, 9978–9983. [Google Scholar] [CrossRef]
  30. Lee, C.-S.; Bai, B.; Song, Q.-R.; Wang, Z.-Q.; Li, G.-F. Open Complementary Split-Ring Resonator Sensor for Dropping-Based Liquid Dielectric Characterization. IEEE Sens. J. 2019, 19, 11880–11890. [Google Scholar] [CrossRef]
  31. Juan, C.G.; Potelon, B.; Quendo, C.; Bronchalo, E. Microwave Planar Resonant Solutions for Glucose Concentration Sensing: A Systematic Review. Appl. Sci. 2021, 11, 7018. [Google Scholar] [CrossRef]
  32. Ansari, M.A.H.; Jha, A.K.; Akhtar, M.J. Design and Application of the CSRR-Based Planar Sensor for Noninvasive Measurement of Complex Permittivity. IEEE Sens. J. 2015, 15, 7181–7189. [Google Scholar] [CrossRef]
  33. Liu, J.-C.; Lin, H.-C.; Zeng, B.-H.; Yeh, K.-D.; Chang, D.-C. An Improved Equivalent Circuit Model for CSRR-Based Bandpass Filter Design With Even and Odd Modes. IEEE Microw. Wirel. Compon. Lett. 2010, 20, 193–195. [Google Scholar] [CrossRef]
  34. Gil, I.; Bonache, J.; Gil, M.; García-García, J.; Martín, F.; Marqués, R. Accurate Circuit Analysis of Resonant-Type Left Handed Transmission Lines with Inter-Resonator Coupling. J. Appl. Phys. 2006, 100, 074908. [Google Scholar] [CrossRef] [Green Version]
  35. Ebrahimi, A.; Withayachumnankul, W.; Al-Sarawi, S.F.; Abbott, D. Dual-Mode Behavior of the Complementary Electric-LC Resonators Loaded on Transmission Line: Analysis and Applications. J. Appl. Phys. 2014, 116, 083705. [Google Scholar] [CrossRef] [Green Version]
  36. Bray, J.R.; Roy, L. Measuring the Unloaded, Loaded, and External Quality Factors of One- and Two-Port Resonators Using Scattering-Parameter Magnitudes at Fractional Power Levels. IEE Proc. Microw. Antennas Propag. 2004, 151, 345–350. [Google Scholar] [CrossRef]
  37. Maione, G. A Rational Discrete Approximation to the Operator s0.5. IEEE Signal Processing Lett. 2006, 13, 141–144. [Google Scholar] [CrossRef]
  38. Bao, J.; Swicord, M.L.; Davis, C.C. Microwave Dielectric Characterization of Binary Mixtures of Water, Methanol, and Ethanol. J. Chem. Phys. 1996, 104, 4441–4450. [Google Scholar] [CrossRef]
Figure 1. Circular CSRR (a) a 3D illustration; (b) Top view; (c) Rear view.
Figure 1. Circular CSRR (a) a 3D illustration; (b) Top view; (c) Rear view.
Sensors 22 01764 g001
Figure 2. CSRR based on (a) rectangular structures; and (b) circular structures.
Figure 2. CSRR based on (a) rectangular structures; and (b) circular structures.
Sensors 22 01764 g002
Figure 3. The simulated transmission results of (a) Rectangular rings; and (b) Circular rings. The insets are their corresponding equivalent circuit models.
Figure 3. The simulated transmission results of (a) Rectangular rings; and (b) Circular rings. The insets are their corresponding equivalent circuit models.
Sensors 22 01764 g003
Figure 4. The simulated field distribution of (a) Rectangular rings; and (b) Circular rings.
Figure 4. The simulated field distribution of (a) Rectangular rings; and (b) Circular rings.
Sensors 22 01764 g004
Figure 5. Comparison of sensitivity of the two resonators. (a) frequency and frequency shift, (b) sensitivity versus change in dielectric constant.
Figure 5. Comparison of sensitivity of the two resonators. (a) frequency and frequency shift, (b) sensitivity versus change in dielectric constant.
Sensors 22 01764 g005
Figure 6. Variation of quality factor versus loss tangent for ε r = 60 .
Figure 6. Variation of quality factor versus loss tangent for ε r = 60 .
Sensors 22 01764 g006
Figure 7. Variation of S 21 (dB) magnitude of the sensor with loss tangent value ranging from 0 to 0.9 for ε r = 20 and ε r = 70 .
Figure 7. Variation of S 21 (dB) magnitude of the sensor with loss tangent value ranging from 0 to 0.9 for ε r = 20 and ε r = 70 .
Sensors 22 01764 g007
Figure 8. (dB) magnitude of sensor for ε r range from 5–90.
Figure 8. (dB) magnitude of sensor for ε r range from 5–90.
Sensors 22 01764 g008
Figure 9. The relationship between ε r and Δ f r .
Figure 9. The relationship between ε r and Δ f r .
Sensors 22 01764 g009
Figure 10. The relationship between Q MUT 1 and tan δ under different ε r .
Figure 10. The relationship between Q MUT 1 and tan δ under different ε r .
Sensors 22 01764 g010
Figure 11. Nonlinear surface fitting for loss tangent and quality factor.
Figure 11. Nonlinear surface fitting for loss tangent and quality factor.
Sensors 22 01764 g011
Figure 12. Experimental verification. (a) Top view of the CSRR; (b) The sensor is fixed on a PTFE plate; (c) Experimental set up.
Figure 12. Experimental verification. (a) Top view of the CSRR; (b) The sensor is fixed on a PTFE plate; (c) Experimental set up.
Sensors 22 01764 g012
Figure 13. Comparison between measurement and simulation of S21 (a) unloaded; (b) 0% ethanol (water); (c) 20% ethanol; (d) 50% ethanol; (e) 70% ethanol; (f) 95% ethanol.
Figure 13. Comparison between measurement and simulation of S21 (a) unloaded; (b) 0% ethanol (water); (c) 20% ethanol; (d) 50% ethanol; (e) 70% ethanol; (f) 95% ethanol.
Sensors 22 01764 g013
Figure 14. S 21 of different concentration of ethanol.
Figure 14. S 21 of different concentration of ethanol.
Sensors 22 01764 g014
Table 1. The Main Parameters for two types of CSRRs shown in Figure 1.
Table 1. The Main Parameters for two types of CSRRs shown in Figure 1.
Parametersabcdefg
Values (mm)0.450.30.40.30.37.30.5
Table 2. The Lumped Parameters for the Equivalent Circuits.
Table 2. The Lumped Parameters for the Equivalent Circuits.
ParametersRectangular CSRRCircular CSRR
L5.512 nH5.048 nH
Cc1.563 pF1.210 pF
Cr2.231 pF2.187 pF
Lr1.141 nH0.838 nH
Table 3. Comparison between measured results of this work and data in the literature [38].
Table 3. Comparison between measured results of this work and data in the literature [38].
SampleRef. [38]This WorkAccuracy
ε r ε r ε r ε r ε r ε r
Water79.010.980.011.21.3%2.8%
20% ethanol64.017.363.817.50.3%1.2%
50% ethanol43.018.043.117.90.2%0.6%
70% ethanol28.517.128.617.20.7%1.7%
95% ethanol11.09.010.99.10.9%1.1%
Table 4. Comparison of various liquid microwave microfluidic sensors.
Table 4. Comparison of various liquid microwave microfluidic sensors.
ReferfenceFrequency (GHz)RSAccuracyComplex PermittivityCorrosive Sample
[24]2.620–3.8160.348/noyes
[25]0.86–0.910.06911.0%nono
[26]3.1800.488/yesno
[27]2.1890.98/yesno
[28]1.200–2.3350.623–0.8794%yesno
[29]2.30–2.350.024/yesyes
This work2.850–2.9600.0412.8%yesyes
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Wang, C.; Liu, X.; Huang, Z.; Yu, S.; Yang, X.; Shang, X. A Sensor for Characterisation of Liquid Materials with High Permittivity and High Dielectric Loss. Sensors 2022, 22, 1764. https://0-doi-org.brum.beds.ac.uk/10.3390/s22051764

AMA Style

Wang C, Liu X, Huang Z, Yu S, Yang X, Shang X. A Sensor for Characterisation of Liquid Materials with High Permittivity and High Dielectric Loss. Sensors. 2022; 22(5):1764. https://0-doi-org.brum.beds.ac.uk/10.3390/s22051764

Chicago/Turabian Style

Wang, Chen, Xiaoming Liu, Zhixiang Huang, Shuo Yu, Xiaofan Yang, and Xiaobang Shang. 2022. "A Sensor for Characterisation of Liquid Materials with High Permittivity and High Dielectric Loss" Sensors 22, no. 5: 1764. https://0-doi-org.brum.beds.ac.uk/10.3390/s22051764