Quartz-Enhanced Photoacoustic Sensor Based on a Multi-Laser Source for In-Sequence Detection of NO2, SO2, and NH3

1
PolySense Lab-Dipartimento Interateneo di Fisica, Politecnico and University of Bari, Via Amendola 173, I-70100 Bari, Italy
2
Polysense Innovations Srl 2, Via Amendola 173, I-70100 Bari, Italy
3
Walter Schottky Institute, Technical University of Munich, Am Coulombwall 4, D-85748 Garching, Germany
*
Authors to whom correspondence should be addressed.
Submission received: 23 September 2023 / Revised: 27 October 2023 / Accepted: 4 November 2023 / Published: 6 November 2023
(This article belongs to the Special Issue Photonics for Advanced Spectroscopy and Sensing)

Abstract

:
In this work, we report on the implementation of a multi-quantum cascade laser (QCL) module as an innovative light source for quartz-enhanced photoacoustic spectroscopy (QEPAS) sensing. The source is composed of three different QCLs coupled with a dichroitic beam combiner module that provides an overlapping collimated beam output for all three QCLs. The 3λ-QCL QEPAS sensor was tested for detection of NO2, SO2, and NH3 in sequence in a laboratory environment. Sensitivities of 19.99 mV/ppm, 19.39 mV/ppm, and 73.99 mV/ppm were reached for NO2, SO2, and NH3 gas detection, respectively, with ultimate detection limits of 9 ppb, 9.3 ppb, and 2.4 ppb for these three gases, respectively, at an integration time of 100 ms. The detection limits were well below the values of typical natural abundance of NO2, SO2, and NH3 in air.

1. Introduction

Air pollution refers to the modification of the characteristics of the atmosphere induced by contaminants [1]. Sources of air pollution are, for example, any kind of combustion, vehicles, and industrial processes that send the waste of production into the atmosphere [2]. Air pollutants can be classified as either primary or secondary air pollutants [3]. Primary air pollutants include constituents emitted directly into the atmosphere from the source. Secondary air pollutants are generated by chemical reactions involving primary pollutants and other atmospheric constituents [4]. Every country establishes regulations for maximum acceptable concentrations as well as desiderated ones. Pollutants of major public health concern include nitrogen dioxide, sulfur dioxide, and ammonia.
Ammonia (NH3) is a pungent-smelling and toxic gas naturally present in the environment, with a typical concentration of a few parts per billion (ppb). Its excessive release into the atmosphere through human activities has resulted in it being considered a significant air pollutant. Its concentrations should be controlled below 20 parts per million (ppm), otherwise it can lead to an irritating effect on the eyes and on the respiratory system, which becomes critical at a concentration of 110 ppm [5]. Above 1500 ppm, NH3 is very harmful or even lethal to humans [6].
Nitrogen dioxide (NO2) is a reddish-brown and pungent-smelling gas, primarily produced by the combustion of fossil fuels, including combustion in vehicle engines and industrial processes. It reacts with water to produce nitric acid and nitric oxide as well as with other pollutants in the atmosphere to form fine particulate matter (PM2.5). Different limit values of NO2 have applied to outdoor air. For a one hour exposure period, the European Union has specified a value of 100 ppb as the maximum concentration limit [7]. Inhalation of NO₂ can cause dizziness, headaches, and irritation of the respiratory system. Prolonged exposure to elevated levels of NO2 can exacerbate respiratory conditions such as asthma and shortness of breath.
Sulfur dioxide (SO2) is a colorless, pungent-smelling, and irritant gas mainly produced in fossil fuel combustion and industrial processes. SO2 is responsible for acid rain and can contribute to the formation of PM2.5. SO2 occurs in concentrations typically of up to several tens of ppb and in extreme cases up to several hundreds of ppb [8]. The exposure to SO2 can have serious health implications, causing acute severe effects in the bronchi at concentrations higher than 4 ppm, and prolonged exposure can compromise lung function [9].
Real-time pollution monitoring with high sensitivity and selectivity as well as public alerts is vital to minimize the exposure of the population, particularly the vulnerable categories, to air pollution [10,11]. Miniature, low-cost electrochemical sensors have found widespread use in industrial settings but generally are still not stable or sensitive enough for monitoring ambient pollutants reliably. Photonics sensors based on laser absorption spectroscopy are of growing interest due to the development of powerful infrared lasers, such as quantum cascade lasers (QCLs) tunable over wide wavelength ranges, which permit the unambiguous detection of numerous substances at low concentrations [12,13]. Several optical-based sensors have been developed to detect air pollutants, with the final aim of on-field, real-time detection. Trace concentrations of NH3, NO2, and SO2 were measured by employing detection techniques based on direct detection of absorbed light, such as tunable diode laser absorption spectroscopy [14,15,16] and cavity ringdown spectroscopy [17,18,19]. Among the optical techniques for gas sensing, Quartz-Enhanced Photoacoustic Spectroscopy (QEPAS) exploits the photoacoustic effect occurring in a gas sample when a modulated, resonant light is absorbed by the target analytes. A weakly damped propagating acoustic (pressure) wave with wavelengths in the centimeter range is generated in the proximity of the exciting light beam [20]. In QEPAS, these sound waves are detected by a spectrophone, composed of a quartz-tuning fork (QTF) transducer and a pair of millimeter-size resonator tubes, aligned on both sides of the QTF. The laser beam is focused between the QTF prongs while passing through both tubes. In this way, the sound wave is generated between QTF’s prongs and confined in the resonator tubes. The generated standing wave vibrational pattern puts prongs in vibration and a sinusoidal electrical signal is generated because of the piezoelectric effect occurring in the quartz crystal [21]. Avoiding the use of a photodetector, QEPAS is a wavelength-independent technique. Indeed, the same QTF can operate with laser sources emitting in the spectral range from UV to THz. This classifies QEPAS as an ideal technique for multi-gas detection when it is combined with a multi-laser source [22,23,24,25].
In this work, a multi-QCL QEPAS sensor for multi-gas detection was developed and tested in a laboratory environment. A custom-designed three-wavelength laser module was employed as a light source for the multi-color QEPAS sensor. This module consists of three different QCLs whose output beams are combined into a single collimated output beam by means of dichroitic beam combiners mounted inside the module. In this configuration, the developed QEPAS sensor was calibrated and tested for sequential detection of gaseous samples containing trace concentrations of NO2, SO2, and NH3.

2. Assembly of the Multi-QCL Source

The three-wavelength laser module integrates three distinct QCL chips and is designed to emit a single collimated beam output by using dichroitic beam combiners. The three Distributed Feedback QCLs from Thorlabs GmbH (Newton, MA, USA) assembled in the module are the QD7385HHLH-B device designed for operation at 7.38 μm, the QD9062HHLH-C operating at 9.06 μm, and the QD6250HHLH-A emitting at 6.25 μm. The three QCLs were selected to target absorption features of SO2, NH3, and NO2, respectively.
A schematic illustration of the 3λ-QCL module prototype is depicted in Figure 1a.
F1 is a low-pass filter (Spectrogon SP-7500, Täby, Sweden) with a transmission spectrum that allows the transmission of the 6.25 μm beam and the reflection of the 9.06 μm beam. On the other hand, F2 is a band-pass filter (Spectrogon BBP-7000-8750c, Täby, Sweden) used for the transmission of the 7.38 μm beam and efficient reflection the other two beams. For this reason, F2 is placed in front of the 7.38 μm QCL. F1, F2, and the mirror M1 are attached to kinematic mounts (IM10.C2 from Siskiyou Corporation, Redwood, CA, USA) with two adjusters with an angular range of ±4°. Moreover, the F1 and M1 kinematic mounts are fixed on a 1-axis translation stage (50.25 dt from Siskiyou Corporation) with a 6 mm adjustment range to fine control the distance between the filter and the mirror with the QCLs. Thus, fine tilting of M1, F1, and F2 can steer both the 9.06 μm and the 6.25 μm beam to make them collinear with the 7.38 μm beam. In this way, the three QCL beams can come out of the output module window collinearly. The three QCL sources are mounted on three custom-made heatsinks for extra heating dissipation when the lasers are turned on. The module is mounted over a back plate with four holes for easy fitting with an optical table. The lateral sides can be easily removed to ensure quick access to the internal components for fine adjustments of the optical alignment. A 3D-sketch of the module is represented in Figure 1b.
The alignment of the three QCL beams plays a crucial role in their use as a light source in a QEPAS sensor. It is mandatory to reach high collinearity of the three QCL beams in order to use a single aspheric lens to simultaneously focus all of them in an 800 µm-wide gap between the prongs of the QTF without touching the QTF prongs. The exact alignment and focus allows for the noise contribution coming from light touching the QTF prong to be minimized and for increased accuracy and precision of measurements.
A pyrocamera (Pyrocam III, Ophir Spiricon, Darmstadt, Germany) with high-resolution pixel dimensions of 0.08 × 0.08 mm was used for the beam alignment and for acquiring and analyzing far field beam profiles at different distances from the 3λ-QCL module. The collinearity of the three beams was firstly evaluated by analyzing their spatial overlap at 25 cm from the 3λ-QCL module, where the focusing lens of the QEPAS sensor was supposed to be located. The beam profiles were acquired in pairs, namely, the 7.38 μm beam together with the 9.06 μm beam and the 7.38 μm beam with the 6.25 μm beam, as shown in Figure 2a,b.
The results show a very good overlap of the profiles at a 25 cm distance from the 3λ-QCL module. Then, a 1′′-diameter ZnSe Plano-Convex lens with a focal length of 50 mm and an AR coating in the range of 7–12 μm (model LA7656-E3 from Thorlabs; the same lens was mounted in the QEPAS sensor) was placed at a 25 cm distance from the 3λ-QCL module, with the pyrocamera moved into the focal plane of the lens. To investigate the characteristics of the laser beams in the focal plane, beam spots were acquired in pairs, as well as for the measurements reported in Figure 2a,b. The results are shown in Figure 3a,b. The beam waists were measured by extracting the intensity values along the x- and y-directions of the 2D-intensitiy distribution. For both directions, the extracted datapoints were fitted with a Gaussian function. For each gaussian fit, the beam waist was evaluated as the radial distance at which the light intensity dropped to 1/e2 of its maximum central value. The extracted beam waist values were 350 μm, 430 μm, and 310 μm for the 7.38 μm, 9.06 μm, and 6.25 μm beam spots, respectively. For ease of viewing, each spot is sketched in Figure 3c as a circumference, with the mean value of beam waists along the x- and y-directions as the diameter and the coordinates of the peak value those of the center. The data in Figure 3c clearly show that the distance between the three peak values of the three beams was less than 50 µm, comparable to the pyrocamera resolution. The excellent degree of overlap as well as the similarity of the beam spot sizes themselves allowed for easy alignment of the three beams through the spectrophone of the QEPAS sensor, as discussed in the next session.

3. Architecture of the 3λ-QEPAS Sensor

The 3λ-QEPAS sensor architecture is depicted in Figure 4.
The 3λ-QCL module was used as a light source exciting the analytes within the acoustic detection module (ADM01, provided by Thorlabs GmbH, Newton, USA). It is composed of the QEPAS spectrophone enclosed in stainless-steel housing with an inlet and outlet connector for gas flow. The QEPAS spectrophone consists of a T-shaped QTF and a pair of resonator tubes aligned in an on-beam configuration. Each tube has a length of 12.4 mm and an internal diameter of 1.59 mm. The laser beam was focused into the ADM and fixed on a five-axis stage for alignment purposes using a 50 mm focal-length ZnSe lens with a 3–12 µm AR coating. QEPAS measurements were performed using the wavelength modulation and dual-frequency detection method (2f-WM): A sinusoidal dither matching half of the QTF resonance frequency of the employed spectrophone was applied to the QCL current driver (ITC4002QCL, Benchtop Laser Driver and Temperature Controller, Thorlabs) and the transduced QTF signal was demodulated by the lock-in amplifier (MFI 500 kHz Lock-in Amplifier, Zurich Instruments, Zurich, Switzerland) at the QTF resonance frequency [26].
The lock-in integration time was set to 100 ms. The demodulated signal was thus digitized and stored on a personal computer by means of a data acquisition board, with the card-sampling time set to three times the lock-in integration time. The pressure of the gas mixtures flowing inside the ADM was regulated using a pressure controller, a valves system, and a vacuum pump, whereas the flow rate was set by the gas mixer (MCQ Instruments, Gas Blender 103, Rome Italy), with a setpoint accuracy of 1% for each channel. All measurements were performed by fixing the mass flow rate to 90 sccm and the pressure to 400 Torr, resulting from preliminary measurements of the optimal operating pressure for the sequence detection of the three gas species, namely, NH3, SO2, and NO2, of 9.06 μm, 7.38 μm, and 6.25 μm QCL, respectively. At 400 Torr, the spectrophone had a resonance frequency of 12,439.4 Hz with a quality factor of 14,650.

4. Results

4.1. Single-Analyte Calibration

The 3λ-QCL QEPAS sensor was calibrated to detect the selected absorption species, i.e., NH3, NO2, and SO2. First, the sensor was operated to target each absorber independently. The optimal absorption lines were selected to provide both the highest intensity and the lowest interference with spectral lines of other absorbers.
For each of three analytes, the absorption cross-section was reconstructed within the spectral dynamic range of the related QCL by using the HITRAN database to simulate a mixture of 10 ppm of the analyte in N2 [27]. To ensure that the other two analytes, H2O, and other standard air components did not interfere with the detection of the selected analyte, mixtures consisting of 10 ppm of the other two analytes in N2 and a mixture of 1% of H2O in standard air were also simulated within the same spectral range of the QCL. The results of the HITRAN simulations are reported in Figure 5a–c for each target analyte.
For each analyte, the 3λ-QCL QEPAS sensor calibration was performed in controlled humidity conditions by setting the water vapor concentration to 1.5 ‰. The humidity level and the temperature of the gas mixture within the gas line were continuously monitored by a hygrometer, i.e., iST humidity module HYT 271(Innovative Sensor technology, Ebnat-Kappel, Switzerland). The analyte concentration inside the humidified line was varied by diluting a certified concentration of the analyte in N2 with pure N2. When a mixture with a fixed analyte concentration is injected into the gas line for sensor calibration, polar molecules with a permanent dipole moment have a strong tendency to stick to surfaces with adsorption and/or desorption processes, leading to a time-dependent gas-phase concentration within the gas line. For example, an extensive study of the effect of the stickiness of NH3 molecules on the estimation of their concentration is reported in Ref. [23]. The study demonstrated that, starting from a fixed concentration injected into the gas line, a dynamic equilibrium between the gas flux and the adsorption processes could be established after a transient, leading to a stable concentration flowing into the ADM. A similar situation could be established in the sensor-cleaning process. For the three gas species, the equilibrium condition was reached in less than 20 min, mainly due to the high flow rate employed (90 sccm). For each gas species, before passing to the next concentration, the gas line was purged with pure nitrogen for about 1 h. For each analyte concentration value, a spectral scan around the peak was performed with a 100 ms lock-in integration time. The measured 2f-QEPAS spectral scans are shown in Figure 6a–c.
For each spectral scan, peak values of the strongest feature (occurring at 392 mA, 298 mA, and 257 mA for NH3, NO2, and SO2, respectively) were extracted and plotted as a function of the concentration for each analyte to obtain the calibration curves reported in Figure 7a–c with the best linear fits (red solid lines) of the experimental data. A strong correlation between the data and the best linear fit was derived for all the three calibration curves, with a coefficient of determination of R2 > 0.999. Sensitivities (slopes of the linear fits) of 19.99 ± 0.30 mV/ppm, 19.39 ± 0.19 mV/ppm, and 73.99 ± 0.49 mV/ppm were estimated for NO2, SO2, and NH3 detection, respectively, with a measured 1-σ noise level of 0.18 mV. The best linear fit of the NO2 QEPAS sensor calibration (Figure 7b) indicated a negative intercept (−15.13 ± 1.85 mV) far from the zero value. This is ascribed to the chemical reaction of the NO2 molecules with the H2O molecules within the gas line, which resulted in an effective decrease in the actual concentration of NO2 molecules reaching the ADM [28].
This phenomenon showed up as a shift of the calibration curve along the x-axis, causing the negative intercept. An accurate estimation of the NO2 concentration considering this behavior required a concentration correction of 760 ppb. The minimum detection limit (MDL) was calculated as the concentration corresponding to a signal-to-noise ratio equal to 1. Therefore, MDLs of 9 ppb, 9.3 ppb, and 2.4 ppb was estimated for NO2, SO2, and NH3 detection, respectively, with a lock-in integration time of 100 ms.
To investigate the stability of the developed sensor, an Allan–Werle deviation analysis of the QEPAS signal was performed [29]. Each laser source was operated at a fixed current far from the absorption features, with a 2 h-long acquisition of QEPAS signals in pure N2 performed at a 0.1 s lock-in integration time. The measurements were taken under the same experimental condition set for the sensor’s calibration. The results are shown in Figure 8.
When each QCL was turned on, the noise level decreased as the integration time increased up to 100 s, following the expected trend of ~ 1 / t , which indicates that the main contribution to the sensor’s noise was due to the QTF thermal noise. For integration times higher than 100 s, the noise started to deteriorate. This behavior is ascribed to mechanical instabilities of the sensor, i.e., laser instability, mechanical vibrations, etc., which affect the performance of the sensor during long integration times. Indeed, the turning point at 100 ms was not observed when the three QCLs were simultaneously off, strengthening the hypothesis that the turning point is due to sensor instabilities induced by operation of the laser. At an average time of 10 s, MDLs of 0.4 ppb, 1.5 ppb, and 1.4 ppb were achieved for NH3, NO2, and SO2, respectively.

4.2. Multi-Gas Detection

The 3λ-QCL multi-gas detection employing the 3λ-QCL QEPAS sensor was performed by flowing three different gas mixtures in the ADM:
  • Mix #1: 5 ppm NH3, 5 ppm NO2, 2.1 ppm SO2 in N2;
  • Mix #2: 10 ppm NH3, 2.5 ppm NO2, 2.1 ppm SO2 in N2;
  • Mix #3: 5 ppm NH3, 2.5 ppm NO2, 4.3 ppm SO2 in N2.
For each gas mixture, the spectral scan of each gas species was acquired with a 100 ms lock-in integration time under the same experimental conditions described in the previous sections. The measured 2f-QEPAS spectral scans are shown in Figure 9a–i.
The spectral scans acquired in the multi-gas mixtures retraced those acquired in the single-gas mixtures (see Figure 6a–c). No relevant variations in the spectra due to the presence of the other components in the samples were observed. The calibration curves extracted in Figure 7a–c were employed to estimate the target gas concentrations in the three mixtures. The peak values for each species were extracted, and the QEPAS signals were converted into gas concentrations using the sensitivity values, for each analyte. The achieved results are reported in Table 1 together with the expected concentrations for each analyte.
The errors associated with the estimated concentrations were calculated by propagating the uncertainties associated with the retrieved sensitivities and with the collected signals. The latter were evaluated as 1-σ standard deviation of long-term acquisition (~30 min) performed by flushing the multi-gas mixtures in the QEPAS sensor (relative fluctuations below 1% of the corresponding signal). The errors associated with the expected concentrations were calculated by considering the uncertainties of certified concentrations for the gas cylinders with relative uncertainties of 4% at 3-σ and the uncertainties provided by the gas mixer. The collected results clearly indicate that the retrieved concentrations were in excellent agreement with the expected ones. For each gas sample, the relative deviation of the predicted concentrations was calculated, resulting in an average relative deviation of 96.6%, 98.6%, and 99.9% for Mix #1, #2, and #3, respectively.

5. Conclusions

In this work, a multi-QCL QEPAS sensor employing a three-wavelength laser module was successfully implemented and experimentally validated in a controlled laboratory environment. The module incorporates three Distributed Feedback QCLs supplied by Thorlabs GmbH, each operating at distinct wavelengths of 7.38 μm, 9.06 μm, and 6.25 μm. The 3λ-QCL module was employed as a light source in a QEPAS sensor for the detection of NH3, NO2, and SO2 in sequence. Sensitivities of 73.99 mV/ppm, 19.99 mV/ppm, and 19.39 mV/ppm were measured for NH3, NO2, and SO2, respectively, with a measured 1-σ noise level of 0.18 mV. Therefore, MDLs of 2.4 ppb, 9 ppb, and 9.3 ppb were estimated, respectively, which were well below their typical natural abundance in air, even after demodulating the collected signal at an integration time as low as 0.1 s.
Multi-gas detection employing the 3λ-QCL QEPAS sensor was also performed by targeting three different gas mixtures, retracing the same experimental conditions employed for the single-analyte mixtures. The presence of other components in the gas samples did not introduce any substantial variations in the spectra. The developed 3λ-QCL QEPAS sensor demonstrated its capability to accurately determine the composition of the gas samples, with an average accuracy of >96.6%.
The compact multi-QCL assembly, as well as the compactness of the QEPAS sensors, promotes the development of a multi-species QEPAS sensor for accurate, precise, and reliable real-time outdoor air pollutant monitoring.

Author Contributions

Conceptualization, P.P., V.S. and M.A.B.; methodology, P.P. and N.A.; multi-QCL unit design and construction: E.D.T., D.B., V.T. and M.A.B.; validation, N.A.; formal analysis, N.A. and P.P.; data curation, N.A. and P.P.; writing—original draft preparation, P.P.; writing—review and editing, P.P., V.S. and M.A.B.; funding acquisition, P.P. All authors have read and agreed to the published version of the manuscript.

Funding

This work is supported by funding from the European Union’s Horizon 2020 research and innovation program under grant agreement No. 101016956 PASSEPARTOUT in the context of the Photonics Public Private Partnership. The authors from Dipartimento Interateneo di Fisica di Bari acknowledge financial support from PNRR MUR project PE0000023-NQSTI and THORLABS GmbH within the PolySenSe joint research laboratory.

Informed Consent Statement

Not applicable.

Data Availability Statement

Row data will be available on demand.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Tiwary, A.; Williams, I. Air Pollution: Measurement, Modelling and Mitigation, 4th ed.; CRC Press: Boca Raton, FL, USA, 2019; ISBN 9781138503663. [Google Scholar]
  2. Saxena, P.; Naik, V. Air Pollution: Sources, Impacts and Controls; CAB International: Wallingford, UK, 2019. [Google Scholar]
  3. Brusseau, M.L.; Pepper, I.L.; Gerba, C.P. Environmental and Pollution Science; Academic Press: Cambridge, MA, USA, 2019; ISBN 9780128147207. [Google Scholar]
  4. Saxena, P.; Sonwani, S. Criteria Air Pollutants and Their Impact on Environmental Health; Springer: Singapore, 2019; ISBN 9789811399923. [Google Scholar]
  5. NRC (National Research Council); Committee on Acute Exposure Guideline Levels. Ammonia Acute Exposure Guideline Levels. In Acute Exposure Guideline Levels for Selected Airborne Chemicals: Volume 6; National Academies Press: Washington, DC, USA, 2008. [Google Scholar]
  6. Barsan, M.E. NIOSH Pocket Guide to Chemical Hazards; U.S. Department of Health & Human Services: Washington, DC, USA, 2007. [Google Scholar]
  7. World Health Organization. WHO Global Air Quality Guidelines: Particulate Matter (PM2.5 and PM10), Ozone, Nitrogen Dioxide, Sulfur Dioxide and Carbon Monoxide: Executive Summary; World Health Organization: Geneva, Switzerland, 2021; ISBN 9789240034433. [Google Scholar]
  8. Kaufmann, R.K.; Davidsdottir, B.; Garnham, S.; Pauly, P. The determinants of atmospheric SO2 concentrations: Reconsidering the environmental Kuznets curve. Ecol. Econ. 1998, 25, 209–220. [Google Scholar] [CrossRef]
  9. Orellano, P.; Reynoso, J.; Quaranta, N. Short-term exposure to sulphur dioxide (SO) and all-cause and respiratory mortality: A systematic review and meta-analysis. Environ. Int. 2021, 150, 106434. [Google Scholar] [CrossRef] [PubMed]
  10. Bales, E.; Nikzad, N.; Quick, N.; Ziftci, C.; Patrick, K.; Griswold, W.G. Personal pollution monitoring: Mobile real-time air quality in daily life. Pers. Ubiquitous Comput. 2019, 23, 309–328. [Google Scholar] [CrossRef]
  11. Adams, M.D.; Kanaroglou, P.S. Mapping real-time air pollution health risk for environmental management: Combining mobile and stationary air pollution monitoring with neural network models. J. Environ. Manag. 2016, 168, 133–141. [Google Scholar] [CrossRef]
  12. Li, J.; Yu, Z.; Du, Z.; Ji, Y.; Liu, C. Standoff chemical detection using laser absorption spectroscopy: A review. Remote Sens. 2020, 12, 2771. [Google Scholar] [CrossRef]
  13. Lin, S.; Chang, J.; Sun, J.; Xu, P. Improvement of the Detection Sensitivity for Tunable Diode Laser Absorption Spectroscopy: A Review. Front. Phys. 2022, 10, 853966. [Google Scholar] [CrossRef]
  14. Miller, D.J.; Sun, K.; Tao, L.; Khan, M.A.; Zondlo, M.A. Open-path, quantum cascade-laser-based sensor for high-resolution atmospheric ammonia measurements. Atmos. Meas. Tech. 2014, 7, 81–93. [Google Scholar] [CrossRef]
  15. Yu, Y.; Sanchez, N.P.; Yi, F.; Zheng, C.; Ye, W.; Wu, H.; Griffin, R.J.; Tittel, F.K. Dual quantum cascade laser-based sensor for simultaneous NO and NO2 detection using a wavelength modulation-division multiplexing technique. Appl. Phys. B 2017, 123, 164. [Google Scholar] [CrossRef]
  16. Li, J.; Ding, Y.; Li, Z.; Peng, Z. Simultaneous measurements of SO2 and SO3 in the heterogeneous conversions of SO2 using QCL absorption spectroscopy. Appl. Phys. B 2022, 128, 61. [Google Scholar] [CrossRef]
  17. Oh, M.-K.; Lee, J. Real-time ultrasensitive detection of ammonia gas using a compact CRDS spectrometer. Appl. Opt. 2023, 62, 1357–1363. [Google Scholar]
  18. Medina, D.S.; Liu, Y.; Wang, L.; Zhang, J. Detection of sulfur dioxide by cavity ring-down spectroscopy. Environ. Sci. Technol. 2011, 45, 1926–1931. [Google Scholar] [CrossRef] [PubMed]
  19. Li, Z.; Hu, R.; Xie, P.; Chen, H.; Liu, X.; Liang, S.; Wang, D.; Wang, F.; Wang, Y.; Lin, C.; et al. Simultaneous measurement of NO and NO2 by a dual-channel cavity ring-down spectroscopy technique. Atmos. Meas. Tech. 2019, 12, 3223–3236. [Google Scholar] [CrossRef]
  20. Kosterev, A.A.; Bakhirkin, Y.A.; Curl, R.F.; Tittel, F.K. Quartz-enhanced photoacoustic spectroscopy. Opt. Lett. 2002, 27, 1902. [Google Scholar] [CrossRef] [PubMed]
  21. Dello Russo, S.; Giglio, M.; Sampaolo, A.; Patimisco, P.; Menduni, G.; Wu, H.; Dong, L.; Passaro, V.M.N.; Spagnolo, V. Acoustic coupling between resonator tubes in quartz-enhanced photoacoustic spectrophones employing a large prong spacing tuning fork. Sensors 2019, 19, 4109. [Google Scholar] [CrossRef] [PubMed]
  22. Sampaolo, A.; Patimisco, P.; Giglio, M.; Zifarelli, A.; Wu, H.; Dong, L.; Spagnolo, V. Quartz-enhanced photoacoustic spectroscopy for multi-gas detection: A review. Anal. Chim. Acta 2022, 1202, 338894. [Google Scholar] [CrossRef] [PubMed]
  23. Menduni, G.; Zifarelli, A.; Kniazeva, E.; Dello Russo, S.; Ranieri, A.C.; Ranieri, E.; Patimisco, P.; Sampaolo, A.; Giglio, M.; Manassero, F.; et al. Measurement of methane, nitrous oxide and ammonia in atmosphere with a compact quartz-enhanced photoacoustic sensor. Sens. Actuators B Chem. 2023, 375, 132953. [Google Scholar] [CrossRef]
  24. Sampaolo, A.; Menduni, G.; Patimisco, P.; Giglio, M.; Passaro, V.M.N.; Dong, L.; Wu, H.; Tittel, F.K.; Spagnolo, V. Quartz-enhanced photoacoustic spectroscopy for hydrocarbon trace gas detection and petroleum exploration. Fuel 2020, 277, 118118. [Google Scholar] [CrossRef]
  25. Giglio, M.; Zifarelli, A.; Sampaolo, A.; Menduni, G.; Elefante, A.; Blanchard, R.; Pfluegl, C.; Witinski, M.F.; Vakhshoori, D.; Wu, H.; et al. Broadband detection of methane and nitrous oxide using a distributed-feedback quantum cascade laser array and quartz-enhanced photoacoustic sensing. Photoacoustics 2020, 17, 100159. [Google Scholar] [CrossRef]
  26. Patimisco, P.; Sampaolo, A.; Bidaux, Y.; Bismuto, A.; Scott, M.; Jiang, J.; Muller, A.; Faist, J.; Tittel, F.K.; Spagnolo, V. Purely wavelength- and amplitude-modulated quartz-enhanced photoacoustic spectroscopy. Opt. Express 2016, 24, 25943. [Google Scholar] [CrossRef]
  27. Gordon, I.E.; Rothman, L.S.; Hargreaves, R.J.; Hashemi, R.; Karlovets, E.V.; Skinner, F.M.; Conway, E.K.; Hill, C.; Kochanov, R.V.; Tan, Y.; et al. The HITRAN2020 molecular spectroscopic database. J. Quant. Spectrosc. Radiat. Transf. 2022, 277, 107949. [Google Scholar] [CrossRef]
  28. Sakamaki, F.; Hatakeyama, S.; Akimoto, H. Formation of nitrous acid and nitric oxide in the heterogeneous dark reaction of nitrogen dioxide and water vapor in a smog chamber. Int. J. Chem. Kinet. 1983, 15, 1013–1029. [Google Scholar] [CrossRef]
  29. Giglio, M.; Patimisco, P.; Sampaolo, A.; Scamarcio, G.; Tittel, F.K.; Spagnolo, V. Allan Deviation Plot as a Tool for Quartz-Enhanced Photoacoustic Sensors Noise Analysis. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2016, 63, 555–560. [Google Scholar] [CrossRef] [PubMed]
Figure 1. (a) Schematic of the internal structure of the 3λ-QCL module. F2 is a band-pass filter; F1 is a low-pass filter; M1 is a mirror. (b) Top view of the Solidworks 3D model of the 3λ-QCL module without the top and lateral sides.
Figure 1. (a) Schematic of the internal structure of the 3λ-QCL module. F2 is a band-pass filter; F1 is a low-pass filter; M1 is a mirror. (b) Top view of the Solidworks 3D model of the 3λ-QCL module without the top and lateral sides.
Sensors 23 09005 g001
Figure 2. (a) Combined 7.38 μm and 9.06 μm beam profiles overlapped at 25 cm from the 3λ-QCL module. (b) Combined 7.38 μm and 6.25 μm beam profiles overlapped at 25 cm from the 3λ-QCL module.
Figure 2. (a) Combined 7.38 μm and 9.06 μm beam profiles overlapped at 25 cm from the 3λ-QCL module. (b) Combined 7.38 μm and 6.25 μm beam profiles overlapped at 25 cm from the 3λ-QCL module.
Sensors 23 09005 g002
Figure 3. (a) Combined 7.38 μm and 9.06 μm beam profiles overlapped at the focal plane of the lens. (b) Combined 7.38 μm and 6.25 μm beam profiles overlapped at the focal plane of the lens. (c) Representation of the three beam spots as circumferences. The 7.41 µm, 6.25 µm, and 9.06 µm beam waists are depicted as green, blue, and red solid circumferences, respectively. The radii of the circumferences are equal to the mean value of widths of the beam waists along the x- and y-directions. The coordinates of the center of the circumferences are those of the peak values of the three beam spots.
Figure 3. (a) Combined 7.38 μm and 9.06 μm beam profiles overlapped at the focal plane of the lens. (b) Combined 7.38 μm and 6.25 μm beam profiles overlapped at the focal plane of the lens. (c) Representation of the three beam spots as circumferences. The 7.41 µm, 6.25 µm, and 9.06 µm beam waists are depicted as green, blue, and red solid circumferences, respectively. The radii of the circumferences are equal to the mean value of widths of the beam waists along the x- and y-directions. The coordinates of the center of the circumferences are those of the peak values of the three beam spots.
Sensors 23 09005 g003
Figure 4. Schematic of the 3λ-QCL-based QEPAS sensor for NH3, SO2, and NO2 detection. L—lens; ADM—acoustic detection module; DAQ—data acquisition board; PC—personal computer; PM—power meter.
Figure 4. Schematic of the 3λ-QCL-based QEPAS sensor for NH3, SO2, and NO2 detection. L—lens; ADM—acoustic detection module; DAQ—data acquisition board; PC—personal computer; PM—power meter.
Sensors 23 09005 g004
Figure 5. (a) HITRAN simulation of an absorption cross-section of a mixture of 10 ppm of NH3 in N2, a mixture of 10 ppm of SO2 in N2, a mixture of 10 ppm of NO2 in N2, and a mixture of 1% of water vapor in standard air within the emission spectral range of the 9.06 μm QCL. (b) Simulation of an absorption cross-section of a mixture of 10 ppm of NO2 in N2, a mixture of 10 ppm of SO2 in N2, a mixture of 10 ppm of NH3 in N2, and a mixture of 1% of water vapor in N2 within the emission spectral range of the 6.25 μm QCL. (c) Simulation of an absorption cross-section of a mixture of 10 ppm of SO2 in N2, a mixture of 10 ppm of NH3 in N2, a mixture of 10 ppm of NO2 in N2, and a mixture of 1% of water vapor in N2 within the emission spectral range of the 7.38 μm QCL.
Figure 5. (a) HITRAN simulation of an absorption cross-section of a mixture of 10 ppm of NH3 in N2, a mixture of 10 ppm of SO2 in N2, a mixture of 10 ppm of NO2 in N2, and a mixture of 1% of water vapor in standard air within the emission spectral range of the 9.06 μm QCL. (b) Simulation of an absorption cross-section of a mixture of 10 ppm of NO2 in N2, a mixture of 10 ppm of SO2 in N2, a mixture of 10 ppm of NH3 in N2, and a mixture of 1% of water vapor in N2 within the emission spectral range of the 6.25 μm QCL. (c) Simulation of an absorption cross-section of a mixture of 10 ppm of SO2 in N2, a mixture of 10 ppm of NH3 in N2, a mixture of 10 ppm of NO2 in N2, and a mixture of 1% of water vapor in N2 within the emission spectral range of the 7.38 μm QCL.
Sensors 23 09005 g005
Figure 6. (a) QEPAS spectral scans measured for different concentrations of NH3 in N2 and pure N2 using the 9.06 μm QCL. (b) QEPAS spectral scans measured for different concentrations of NO2 in N2 and pure N2 obtained when the 6.25 μm QCL is turned on. (c) QEPAS spectral scans measured for different concentrations of SO2 in N2 and pure N2 using the 7.38 μm QCL The peak at 275 mA observed for pure N2 is due to residual H2O in the gas line.
Figure 6. (a) QEPAS spectral scans measured for different concentrations of NH3 in N2 and pure N2 using the 9.06 μm QCL. (b) QEPAS spectral scans measured for different concentrations of NO2 in N2 and pure N2 obtained when the 6.25 μm QCL is turned on. (c) QEPAS spectral scans measured for different concentrations of SO2 in N2 and pure N2 using the 7.38 μm QCL The peak at 275 mA observed for pure N2 is due to residual H2O in the gas line.
Sensors 23 09005 g006
Figure 7. (a) QEPAS signal as a function of the NH3 concentration (black squares) with the corresponding best linear fit (red line). (b) QEPAS signal as a function of the NO2 concentration (black squares) with the corresponding best linear fit (red line). (c) QEPAS signal as a function of the SO2 concentration (black squares) with the corresponding best linear fit (red line).
Figure 7. (a) QEPAS signal as a function of the NH3 concentration (black squares) with the corresponding best linear fit (red line). (b) QEPAS signal as a function of the NO2 concentration (black squares) with the corresponding best linear fit (red line). (c) QEPAS signal as a function of the SO2 concentration (black squares) with the corresponding best linear fit (red line).
Sensors 23 09005 g007
Figure 8. Allan deviation of the QEPAS signal as a function of the lock-in integration time.
Figure 8. Allan deviation of the QEPAS signal as a function of the lock-in integration time.
Sensors 23 09005 g008
Figure 9. QEPAS spectral scan of NH3 (a), NO2 (b), and SO2 (c) in Mix #1 NO2; NH3 (d), NO2 (e), and SO2 (f) in Mix #2; NH3 (g), NO2 (h), and SO2 (i) in Mix #3. The peak at 275 mA observed for pure N2 is due to residual H2O in the gas line.
Figure 9. QEPAS spectral scan of NH3 (a), NO2 (b), and SO2 (c) in Mix #1 NO2; NH3 (d), NO2 (e), and SO2 (f) in Mix #2; NH3 (g), NO2 (h), and SO2 (i) in Mix #3. The peak at 275 mA observed for pure N2 is due to residual H2O in the gas line.
Sensors 23 09005 g009
Table 1. Expected and estimated analyte concentrations in the three analyzed mixtures.
Table 1. Expected and estimated analyte concentrations in the three analyzed mixtures.
MixTarget GasExpected Concentration (ppm)Estimated Concentration (ppm)
#1NH35.0 ± 0.14.90 ± 0.03
#1NO25.0 ± 0.15.06 ± 0.08
#1SO22.1 ± 0.11.97 ± 0.03
#2NH310.0 ± 0.29.96 ± 0.07
#2NO22.5 ± 0.12.62 ± 0.05
#2SO22.1 ± 0.12.01 ± 0.02
#3NH35.0 ± 0.15.06 ± 0.04
#3NO22.5 ± 0.12.60 ± 0.05
#3SO24.3 ± 0.14.06 ± 0.05
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Patimisco, P.; Ardito, N.; De Toma, E.; Burghart, D.; Tigaev, V.; Belkin, M.A.; Spagnolo, V. Quartz-Enhanced Photoacoustic Sensor Based on a Multi-Laser Source for In-Sequence Detection of NO2, SO2, and NH3. Sensors 2023, 23, 9005. https://0-doi-org.brum.beds.ac.uk/10.3390/s23219005

AMA Style

Patimisco P, Ardito N, De Toma E, Burghart D, Tigaev V, Belkin MA, Spagnolo V. Quartz-Enhanced Photoacoustic Sensor Based on a Multi-Laser Source for In-Sequence Detection of NO2, SO2, and NH3. Sensors. 2023; 23(21):9005. https://0-doi-org.brum.beds.ac.uk/10.3390/s23219005

Chicago/Turabian Style

Patimisco, Pietro, Nicoletta Ardito, Edoardo De Toma, Dominik Burghart, Vladislav Tigaev, Mikhail A. Belkin, and Vincenzo Spagnolo. 2023. "Quartz-Enhanced Photoacoustic Sensor Based on a Multi-Laser Source for In-Sequence Detection of NO2, SO2, and NH3" Sensors 23, no. 21: 9005. https://0-doi-org.brum.beds.ac.uk/10.3390/s23219005