Next Article in Journal
Enhancing Compression Level for More Efficient Compressed Sensing and Other Lessons from NMR Spectroscopy
Next Article in Special Issue
Low-Humidity Sensing Properties of Multi-Layered Graphene Grown by Chemical Vapor Deposition
Previous Article in Journal
Rheotaxially Grown and Vacuum Oxidized SnOx Nanolayers for NO2 Sensing Characteristics at ppb Level and Room Temperature
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Determination of Trace Lead and Cadmium in Decorative Material Using Disposable Screen-Printed Electrode Electrically Modified with Reduced Graphene Oxide/L-Cysteine/Bi-Film

1
Research Institute of Forestry New Technology and Research Institute of Wood Industry, Chinese Academy of Forestry, Beijing 100091, China
2
State Key Laboratory of Animal Nutrition, Institute of Animal Science, Chinese Academy of Agricultural Sciences, Beijing 100193, China
3
Geographic Information Center of Yulin City, Shannxi 719000, China
*
Author to whom correspondence should be addressed.
Submission received: 31 January 2020 / Revised: 23 February 2020 / Accepted: 26 February 2020 / Published: 28 February 2020
(This article belongs to the Special Issue Graphene Based Chemical Sensors)

Abstract

:
Cadmium (Cd) and lead (Pb) in decorative materials threaten human health. To determine the content of Cd(II) and Pb(II), a disposable screen-printed electrode (DSPE) electrically modified with reduced graphene oxide (rGO) and L-cysteine (LC) was fabricated, which was further electroplated with bismuth film (Bi/LC-rGO/DSPE) in situ. The electrochemical properties of this electrode were studied using cyclic voltammetry, electrochemical impedance spectroscopy, linear sweep voltammetry and differential pulse voltammetry. The results indicated that the Bi/LC-rGO/DSPE had excellent sensitivity, selectivity and stability with low cost and easy production. After optimizing the detection parameters, the linear range of the Bi/LC-rGO/DSPE was from 1.0 to 30.0 μg/L for Cd(II) and Pb(II), and the detection limits were 0.10 μg/L for Cd(II) and 0.08 μg/L for Pb(II). Finally, the Bi/LC-rGO/DSPE was applied to determine the concentrations of Cd(II) and Pb(II) in different decorative materials where the recoveries were in the range from 95.86% to 106.64%.

1. Introduction

Wood, as a renewable and recycled green material, is widely used in daily life. Due to its pore structure with hydrophilic groups [1], wood can absorb moisture, causing dry shrinkage, moisture expansion, warping, deformation, cracking or other defects [2]. To improve the practicability, surface decorative materials [3]—especially paint—are applied to coat the wood surface, which mainly prevents damage from moisture or scratches. However, some decorative materials contain soluble heavy metals [4]. These heavy metals can easily enter the human body through the skin or mouth [5] and threaten human health. In decorative materials, lead (Pb) and cadmium (Cd) are the two common pollutants, leading to some severe diseases [6,7], such as hypertension, cardiovascular disease, mental retardation and dementia.
To date, there is still no decorative material standard for heavy metals in China. Many developed countries, including the European Union, United States, Canada and Japan, have formulated relevant standards. According to the European Union Toy Safety Directive, the maximum contents of Pb(II) and Cd(II) are 90 mg/kg and 75 mg/kg, respectively, which is suitable to evaluate China’s decoration material industry.
There are many methods to analyze the concentrations of Pb(II) and Cd(II). Traditional methods based on spectroscopic methodology, including atomic absorption spectrometry (AAS), ultraviolet–visible spectrophotometry (UV-Vis), atomic fluorescence spectrometry (AFS), inductively coupled plasma mass spectrometry (ICP-MS) and X-ray fluorescence spectrometry (XRF) have high sensitivity and selectivity [8]. However, these methods need expensive devices, a professional operator and complex sample pretreatment, so they can be applied in laboratories and professional testing institutions [9]. Therefore, it is highly urgent to develop a low-cost method to determine the concentrations of Pb(II) and Cd(II) that can conveniently analyze decorative materials in the home environment. Electrochemical stripping analysis (ESA) [10,11] is recognized as a powerful tool for the simultaneous determination of metal ions. In comparison to spectroscopic instruments, the relevant device is easy-to-use, portable, compact and inexpensive [12,13].
A disposable screen-printed electrode (DSPE) is a promising electrochemical electrode in view of its low cost, mass production and low background current. It can overcome some shortcomings of glassy carbon electrodes and carbon paste electrodes [14], including memory effects and tedious cleaning processes. In addition, the inherent disadvantages of DSPEs containing low sensitivity and reproducibility limit their practical application, but their surface is prone to modify different sensing materials to achieve excellent detection performance [15,16]. Therefore, DSPEs are suitable for Pb(II) and Cd(II) detection in the home environment.
Graphene [17] is a two-dimensional carbon material with atoms arranged in a honeycomb lattice. Owing to its excellent physical, chemical, optical and electrical properties, it can be employed to decorate a DSPE’s surface to improve the sensitivity. However, graphene hardly dissolves in water to form a homogeneous solution, making it difficult to modify the DSPE’s surface directly. Graphene oxide (GO) is a vital derivative of graphene that can dissolve well in water because of the various oxygen-containing functional groups on the base and edge of graphene [18,19]. Since its discovery, graphene oxide has had unceasing applications in the field of sensors [20,21]. L-cysteine (LC) is a naturally occurring amino acid with good biocompatibility and water solubility. One LC molecule contains three functional groups (-SH, -NH2, -COOH) [22] that are easy to coordinate with the heavy metal ions. LC can be functionalized on the surface of electrochemical electrodes, which will increase the amount of Cd(II) and Pb(II) on the electrode surface [23,24,25,26]. Bismuth (Bi) is a low-toxicity element with low-temperature melting. It can form binary or multicomponent fusible alloys with numerous heavy metals, including lead, cadmium, thallium, antimony, indium or gallium, which can facilitate the nucleation process during the deposition of heavy metals [27].
Here, the DSPE was modified with reduced graphene oxide (rGO), LC and Bi film to improve the detection performance for Cd(II) and Pb(II). These three materials provided the DSPE with a large specific surface area, conductivity and high mechanical stability, which were investigated through different electrochemical methods. Finally, a sample pre-treatment method was established to extract the Cd(II) and Pb(II) from the decorative materials.

2. Experimental

2.1. Reagents

GO was obtained from Nanjing Xianfeng Nanomaterial Technology Co., Ltd. Standard solutions of Bi(III), Cd(II) and Pb(II) (1000 mg/L) were provided by the National Standard Substances Center (China), which were diluted into the required concentrations. LC was offered by Shanghai Civil Chemical Technology Co., Ltd. A disposable screen-printed electrode was bought by Suzhou mayor’s triangle system, Interdisciplinary Research Institute Co., Ltd. Potassium ferricyanide and potassium ferrocyanide were offered by Beijing Chemical Industry Group Co., Ltd (Beijing, China). The supporting electrolyte obtained from Xiamen Haibiao Technology Co., Ltd. (Xiamen, China) was an acetate buffer solution (0.1 mol/L). Millipore-Q (18.2 MΩ cm) water was used for all experiments. The rest of the chemicals and reagents were of analytical grade without any further purification.

2.2. Apparatus

The electrochemical measurements were collected through a CHI 760 electrochemical workstation (CH Instruments, Austin, TX, USA). The conventional three-electrode system (a saturated Ag/AgCl reference electrode, a Pt wire auxiliary electrode and the Bi/LC-rGO/DSPE working electrode) was immersed into an electrochemical cell containing 10 mL of analyte liquor. During the deposition step, the analyte liquor was stirred using a magnetic stirrer (Shanghai Lichen Technology Co., Ltd, Shanghai, China). A PHS-3C digital pH meter was purchased from Shanghai Instrument Scientific Instrument Co., Ltd, Shanghai, China. All the experiments were performed at room temperature.

2.3. Electrode Preparation

The electrochemically reduced graphene oxide (rGO) and L-cysteine (LC) were modified on the DSPE’s surface in Figure 1 using the following protocol [9,23]. Firstly, the DSPE was rinsed with Millipore-Q water and dried in an N2 atmosphere. Then, 1 mg of GO was put into 5.0 mL of Millipore-Q water to form a homogeneous solution through ultrasonic treatment. The GO/DSPE was obtained by adding 3 μL of the dispersed solution on the DSPE’s surface directly. The GO/DSPE was heated at 50 ℃ for 60 min, and then immersed in a phosphate buffer solution (pH 7.0) to reduce the GO using cyclic voltammetry for 5 cycles with a scan rate of 20 mV/s. In addition, the rGO/DSPE was put into the 2.5 mol/L LC solution and scanned using cyclic voltammetry in the potential range of -0.4~+2.4 V with a scan rate of 40 mV/s. The modified electrodes (LC/DSPE and rGO/DSPE) were fabricated using parts of these steps. Bi can be corroded easily by oxygen in air or hydrolyzed in an alkaline and neutral environment, so bismuth film decorated on the surface of the LC-rGO/DSPE was usually carried out in an acidic environment. Thus, Bi film was deposited along with heavy metal detection.

2.4. Sample Extract Preparation

The heavy metals were extracted from the decorative materials of the floor surface using the following protocol. Firstly, 1.0 g of decorative materials were scraped from the floor surface, and then ground to a powder. After that, the powder was added into 100 mL of a hydrochloric acid solution for 60 min. When the powder was deposited on the bottom of the bottle, the upper layer was transferred into another tube and filtered through a pore size of 0.22 μm. Prior to the measurement, the pH of the filter solution was adjusted to 4.5 using a 0.1 mol/L acetate buffer solution and 0.1 mol/L sodium hydroxide solution.

2.5. Measurement Procedure

The three electrodes were immersed in the acetic acid buffer or the sample extracts, and differential pulse voltammetry was applied to simultaneously measure Cd(II) and Pb(II). The LC-rGO/DSPE was enriched with the heavy metal ions at −1.2 V for 240 s by stirring and then equilibrated for 10 s. The parameters for the differential pulse voltammetry were as follows: amplitude, 0.004 V; amplitude, 0.05 V; pulse period, 0.5 s; sample width, 0.0157 s.

3. Results and Discussion

3.1. The Characteristics

The electrochemical properties of the DSPE before and after it was decorated with reduced graphene oxide and L-cysteine were investigated by cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS). The CV curves are shown in Figure 2 with a scan rate of 50 mV/s. The redox current of the bare DSPE was about 100 μA, and the potential between the oxidation peak and reduction peak reached 220 mV. On the LC/DSPE, the current redox response decreased to 72 μA with a peak-to-peak separation of 340 mV, ascribing that L-cysteine was a non-conductive material that prevented the transmission of electrons [28]. Moreover, the current redox responses of the rGO/DSPE were significantly higher than those of the DSPE and LC/DSPE. The reason was that the rGO provided a large specific surface area and electrical conductivity [29], which made the electrode surface have high electronic conductivity and accelerated the transfer rate of electroactive substances at the interface between the electrode and the solution. When L-cysteine was immobilized on the rGO/DSPE, the electrochemical performance of the LC-rGO/DSPE was decreased.
The equivalent circuit model of the LC-rGO/DSPE was employed to analyze the impedance spectroscopy in Figure 3. The EIS curve consists of two parts: a linear region at the lower frequencies and a semicircle region at higher frequencies, which represent the diffusion process and the electron transfer resistance (Rct), respectively. The Rct value of the commercial DSPE was about 1.25 kΩ. After L-cysteine was immobilized on the DSPE, the impedance between the LC/DSPE and electrolyte was increased, and the Rct value increased to 2.1 kΩ, which was ascribed to the poor chemical conductivity [30]. Due to the reduced graphene oxide, the Rct value of the rGO/DSPE was about 310 Ω. This small semicircle appeared on the rGO/DSPE, which indicates that the dynamic performance of the electronic transmission was poor. For the LC-rGO/DSPE, the Rct value increased to 700 Ω, which made the electroanalytical probe unable to reach the electrode surface to participate in the reaction. The results were in agreement with the conclusion obtained from the CV.
The hydrogen evolution has a profound effect on the electrochemical response. Hence, this property at the electrode surface was studied. As shown in Figure 4, the background current of the LC-rGO/DSPE had almost no change in the range from −1.0 V to −0.4 V and increases rapidly while the potential was lower than −1.0 V, indicating that hydrogen evolution occurred. It can be concluded that the electrochemical window of the LC-rGO/DSPE in the acetic acid buffer solution was about −1.0 V. After a bismuth film was deposited on the LC-rGO/DSPE surface, the hydrogen evolution potential of the Bi/LC-rGO/DSPE was shifted to the negative direction and reached about −1.3 V. This was mainly due to the formation of a bismuth film on the LC-rGO/DSPE, which expanded the electrochemical window [31]. Therefore, the Bi/LC-rGO/DSPE was much more suitable for the detection of Cd(II) and Pb(II) using the differential pulse anodic stripping voltammetric method.
Figure 5 shows the differential pulse voltammetry of 30.0 μg/L of Cd(II) and Pb(II) of the Bi/DSPE, Bi/LC/DSPE, Bi/rGO/DSPE and Bi/LC-rGO/DSPE. Two small peaks on the bare DSPE were located at −0.8 V and −0.6 V, which presented the stripping responses of Cd(II) and Pb(II), respectively. This can be attributed to the fact that bismuth can form an alloy with cadmium and lead that more readily reduces to Cd(II) and Pb(II) [32]. On the Bi/LC/DSPE, the stripping responses increased, which is mainly due to the -SH group that can bind strongly to heavy metal ions. The stripping responses of the Bi/rGO/DSPE were much higher than the Bi/DSPE owing to the high conductivity, large specific surface area and high adsorption capacity of the rGO. Furthermore, the stripping response of the Bi/LC-rGO/DSPE was better than that of the Bi/rGO/DSPE. This improved performance depended on the synergistic effects between the rGO and LC.

3.2. Analytical Performance

To investigate the analytical performance of the Bi/LC-rGO/DSPE, differential pulse voltammetry was used to determine different concentrations of Cd(II) and Pb(II) under the optimized parameters discussed in Supplementary Materials Figures S1–S4 (deposition time: 240 s; deposition potential: −1.2 V; supporting electrolyte: 0.1 mol/L pH 4.5 acetate buffer solution; Bi(III) concentration: 2 mg/L). Figure 6 shows the stripping voltammetric curves for a series of different concentrations of Cd(II) and Pb(II) under the optimized parameters. It was clear that the stripping responses of Cd(II) and Pb(II) had a positive correlation with the concentrations of Cd(II) and Pb(II), but the striping peaks were shifted to the lower voltage with increasing concentrations. For Cd(II) or Pb(II), the voltage of the initial stripping reaction was the same, and the reaction time changed with the concentration. The electrochemical reaction for the high concentration was longer than for the low concentration. With the scanning voltage of the differential pulse voltammetry (DPV), the stripping peaks shifted to the scanning direction of the DPV. Comparing by calculation, the stripping responses of Cd(II) and Pb(II) had a positive correlation with the concentrations of Cd(II) and Pb(II), which exhibited a linear relationship in the range from 1.0 to 30.0 μg/L. The linear regression equations and correlation coefficients were:
Cd ( II ) :   y = 0.6511   x 0.615   ( R 2 = 9978 )
Pb ( II ) :   y = 0.5020   x 0.3619   ( R 2 = 9951 )
Based on the calculation of the triple signal-to-noise ratio (S/N = 3), the limits of detection of the Bi/LC-rGO/DSPE were 0.1 μg/L and 0.08 μg/L for Cd(II) and Pb(II), respectively. The proposed electrode can meet the requirements for Cd(II) and Pb(II) detection in the decorated materials.
Table 1 shows the properties of the Bi/LC-rGO/DSPE for Cd(II) and Pb(II) detection in comparison with the electrodes published in previous papers. The linear ranges and limit of detection are much lower than the modified glassy carbon electrodes but better than a majority of the modified screen-printed electrodes. The sensitivity is superior to most electrodes. Even though the Bi/LC-rGO/DSPE did not have excellent properties, it can meet the requirements for Cd(II) and Pb(II) detection in decorative material.

3.3. Interference Effects

The decorated materials might coexist with different metal ions (Na(I), K(I), Hg(II), Cu(II), Zn(II), Mg(II), Fe(III) and Al(III)), which were extracted at the sample pre-treatment stage. These metal ions probably interfered with the stripping responses of Cd(II) and Pb(II) on the Bi/LC-rGO/DSPE. The interference effects were studied through the Bi/LC-rGO/DSPE and explored against a standard solution containing 25 µg/L of Cd(II) and Pb(II) by adding 10-fold coexisting ions. In Figure 7, the highest relative currents were 2.07% and 3.61% for Cd(II) and Pb(II) when the Bi/LC-rGO/DSPE was determined as the standard solution adding 10-fold Hg(II). Therefore, different ions had no obvious interference, indicating the Bi/LC-rGO/DSPE has good selectivity.

3.4. Application for the Sample Analysis

To evaluate the application, three samples of decorative materials were collected, which were pre-treated following the protocol in Section 2.4. The Cd(II) and Pb(II) concentrations in the extracted solutions were determined by the Bi/LC-rGO/DSPE and AAS. After each extracted solution was measured, the solution was added to 5.0 μg/L of Cd(II) and Pb(II), and then measured again. The results are shown in Table 2, indicating that the recoveries for Cd(II) and Pb(II) were in the range from 95.86% to 106.64%, and the relative standard deviations were less than 10%. This demonstrates that these two methods had no significant difference for Cd(II) and Pb(II) detection.

4. Conclusions

The DSPE was electrically modified with rGO, LC and Bi film for Cd(II) and Pb(II) detection with good stability. After the study, we found that the prepared electrode combined the advantages of the DSPE and the three materials, exhibiting excellent analytical performances, such as high sensitivity, stability (Supplementary Materials Figure S5) and anti-interference. The optimized parameters were pH 4.5, 2 mg/L of Bi(III) concentration, 240 s of deposition time and −1.2 V deposition potential. Under these conditions, the Bi/LC-rGO/DSPE had a wide linear response range from 1.0 to 30.0 μg/L for Cd(II) and Pb(II) with lower detection limits of 0.1 and 0.08 μg/L. The pre-treatment of the decorative sample was established, and the extraction was measured by the Bi/LC-rGO/DSPE. The recoveries for the Cd(II) and Pb(II) determination were in the range from 95.86% to 106.64%. Even though the established method possessed high precision and simple operation, the sample pre-treatment and detection times still need to be decreased.

Supplementary Materials

The following are available online at https://0-www-mdpi-com.brum.beds.ac.uk/1424-8220/20/5/1322/s1: Figure S1: The stripping responses of Cd(II) and Pb(II) affected by the different pH of the acetic acid buffer solution, Figure S2: The stripping responses of Cd(II) and Pb(II) affected by the different Bi(III) concentrations, Figure S3: The stripping responses of Cd(II) and Pb(II) affected by the deposition times, Figure S4: The stripping responses of Cd(II) and Pb(II) affected by the deposition potentials, Figure S5: The stability of the Bi/LC-rGO/DSPE measured with 20 μg/L of Cd(II) and Pb(II) every two days.

Author Contributions

Data curation, Y.W., L.W. and H.W.; formal analysis, L.W.; funding acquisition, X.H. and B.X.; investigation, Y.W.; writing (original draft), H.W.; writing (review and editing), X.H., B.X. and H.W. All authors have read and agreed to the published version of the manuscript.

Funding

This work was supported by the Science and Technology Major Project of Guangxi (GUIKE AA17204087-13), Key Realm R&D Program of Guangdong Province (No. 2019B020215004) and Chinese National Natural Science Foundation (No. 31671578).

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Baltrėnas, P.; Baltrėnaitė, E.; Spudulis, E. Biochar from pine and birch morphology and pore structure change by treatment in biofilter. Water Air Soil Pollut. 2015, 226, 1–14. [Google Scholar] [CrossRef]
  2. Lautzenhiser, L.L.; LeBlanc, S.S.; LeBlanc, M.; Huang, L. Modular Magnetically Receptive Wood and Engineered Wood Surface Units and Magnetic Box System for Covering Floors, Walls, and Other Surfaces. U.S. Patent 20,190,063,075, 28 February 2019. [Google Scholar]
  3. Wang, H.; Wang, J.; Liu, G.; Zhang, Z.; Hou, X. Electrochemical sensing of Pb (II) and Cd (II) in decorative material of wood panel using nano-cellulose paper-based electrode modified using Graphene/Multi-walled Carbon Nanotubes/Bismuth film. Int. J. Electrochem. Sci 2019, 14, 11253–11266. [Google Scholar] [CrossRef]
  4. Rasheed, T.; Bilal, M.; Nabeel, F.; Iqbal, H.M.; Li, C.; Zhou, Y. Fluorescent sensor based models for the detection of environmentally-related toxic heavy metals. Sci. Total Environ. 2018, 615, 476–485. [Google Scholar] [CrossRef] [PubMed]
  5. Wu, X.; Cobbina, S.J.; Mao, G.; Xu, H.; Zhang, Z.; Yang, L. A review of toxicity and mechanisms of individual and mixtures of heavy metals in the environment. Environ. Sci. Pollut. Res. Int. 2016, 23, 8244–8259. [Google Scholar] [CrossRef] [PubMed]
  6. Zhong, W.S.; Ren, T.; Zhao, L.J. Determination of Pb (Lead), Cd (Cadmium), Cr (Chromium), Cu (Copper), and Ni (Nickel) in Chinese tea with high-resolution continuum source graphite furnace atomic absorption DSPEctrometry. J. Food Drug Anal. 2016, 24, 46–55. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  7. Oliveira, T.; Batista, P.; Leal, M.; Campagnaro, B.; Nogueira, B.; Vassallo, D.; Meyrelles, S.; Padilha, A.S. Chronic Cadmium Exposure Accelerates the Development of Atherosclerosis and Induces Vascular Dysfunction in the Aorta of ApoE−/− Mice. Biol. Trace Elem. Res. 2019, 187, 163–171. [Google Scholar] [CrossRef]
  8. Wang, H.; Yin, Y.; Gang, L. Single-gap Microelectrode Functionalized with Single-walled Carbon Nanotubes and Pbzyme for the Determination of Pb2+. Electroanalysis 2019, 31, 1–9. [Google Scholar] [CrossRef]
  9. Hui, W.; Yin, Y.; Zhao, G.; Bienvenido, F.; Flores-Parrad, I.M.; Wang, Z.; Liu, G. Graphene oxide/multi-walled carbon nanotubes/gold nanoparticle hybridfunctionalized disposable screen-printed carbon electrode to determine Cd (II) and Pb (II) in soil. Int. J. Agric. Biol. Eng. 2019, 12, 194–200. [Google Scholar]
  10. Hernandez-Vargas, G.; Sosa-Hernández, J.E.; Saldarriaga-Hernandez, S.; Villalba-Rodríguez, A.M.; Parra-Saldivar, R.; Iqbal, H. Electrochemical biosensors: A solution to pollution detection with reference to environmental contaminants. Biosensors 2018, 8, 29. [Google Scholar] [CrossRef] [Green Version]
  11. Honeychurch, K.C.; Piano, M. Electrochemical (bio) Sensors for Environmental and Food Analyses. Available online: https://0-www-mdpi-com.brum.beds.ac.uk/2079-6374/8/3/57 (accessed on 28 February 2020).
  12. Lazanas, A.C.; Tsirka, K.; Paipetis, A.S.; Prodromidis, M.I. 2D bismuthene/graphene modified electrodes for the ultra-sensitive stripping voltammetric determination of lead and cadmium. Electrochim. Acta 2020, 336, 135726. [Google Scholar] [CrossRef]
  13. Rasheed, T.; Nabeel, F.; Adeel, M.; Bilal, M.; Iqbal, H.M. “Turn-on” fluorescent sensor-based probing of toxic Hg (II) and Cu (II) with potential intracellular monitoring. Biocatal. Agric. Biotechnol. 2019, 17, 696–701. [Google Scholar] [CrossRef]
  14. Wang, H.; Wang, Z.; Zhang, Z.; Ma, X.; Liu, G.; Zhang, M. Graphene and DCHP Modified Glassy Carbon Electrode for Sensitive Determination of Rutin in Tea, American Society of Agricultural and Biological Engineers. Available online: https://elibrary.asabe.org/abstract.asp?aid=44707 (accessed on 28 February 2020).
  15. Wang, H.; Duan, Y.; Zhao, G.; Wang, Z.; Liu, G. Sensitive Determination of Rutin in Pharmaceuticals Using an Ionic Liquid and MWNT Modified Screen Printed Electrode. Int. J. Electrochem. Sci. 2015, 10, 8759–8769. [Google Scholar]
  16. Economou, A. Screen-printed electrodes modified with “green” metals for electrochemical stripping analysis of toxic elements. Sensors 2018, 18, 1032. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  17. Li, X.; Yu, J.; Wageh, S.; Al-Ghamdi, A.A.; Xie, J. Graphene in photocatalysis: A review. Small 2016, 12, 6640–6696. [Google Scholar] [CrossRef]
  18. Guex, L.G.; Sacchi, B.; Peuvot, K.F.; Andersson, R.L.; Pourrahimi, A.M.; Ström, V.; Farris, S.; Olsson, R.T. Experimental review: Chemical reduction of graphene oxide (GO) to reduced graphene oxide (rGO) by aqueous chemistry. Nanoscale 2017, 9, 9562–9571. [Google Scholar] [CrossRef] [Green Version]
  19. Thakur, S.; Karak, N. Alternative methods and nature-based reagents for the reduction of graphene oxide: A review. Carbon 2015, 94, 224–242. [Google Scholar] [CrossRef]
  20. Wang, J.; Chen, B. Adsorption and coadsorption of organic pollutants and a heavy metal by graphene oxide and reduced graphene materials. Chem. Eng. J. 2015, 281, 379–388. [Google Scholar] [CrossRef]
  21. Jiao, C.; Xiong, J.; Tao, J.; Xu, S.; Zhang, D.; Lin, H.; Chen, Y. Sodium alginate/graphene oxide aerogel with enhanced strength–toughness and its heavy metal adsorption study. Int. J. Biol. Macromol. 2016, 83, 133–141. [Google Scholar] [CrossRef]
  22. Bin, P.; Huang, R.; Zhou, X. Oxidation resistance of the sulfur amino acids: methionine and cysteine. BioMed Res. Int. 2017, 2017, 9584932. [Google Scholar] [CrossRef]
  23. Zhou, W.; Li, C.; Sun, C.; Yang, X. Simultaneously determination of trace Cd2+ and Pb2+ based on l-cysteine/graphene modified glassy carbon electrode. Food Chem. 2016, 192, 351–357. [Google Scholar] [CrossRef]
  24. Zhou, S.F.; Wang, J.J.; Gan, L.; Han, X.J.; Fan, H.L.; Mei, L.Y.; Huang, J.; Liu, Y.Q. Individual and simultaneous electrochemical detection toward heavy metal ions based on L-cysteine modified mesoporous MnFe2O4 nanocrystal clusters. J. Alloys Compd. 2017, 721, 492–500. [Google Scholar] [CrossRef]
  25. Abeysinghe, A.; Sirimuthu, N. Rapid and Highly Selective Detection of Heavy Metal Ions in Drinking Water Using Nanosensor Based Colorimetric Assay. Available online: http://journals.sjp.ac.lk/index.php/fesympo/article/view/4395 (accessed on 28 February 2020).
  26. Morton, J.; Havens, N.; Mugweru, A.; Wanekaya, A.K. Detection of trace heavy metal ions using carbon nanotube-modified electrodes. Electroanal. Int. J. Devoted Fundam. Pract. ADSPEcts Electroanal. 1597, 21, 1597–1603. [Google Scholar] [CrossRef]
  27. Wang, J. Stripping analysis at bismuth electrodes: A review. Electroanal. Int. J. Devoted Fundam. Pract. ADSPEcts Electroanal. 2005, 17, 1341–1346. [Google Scholar] [CrossRef]
  28. Zhang, Q.; Hu, Y.; Wu, D.; Ma, S.; Wang, J.; Rao, J.; Xu, L.; Xu, H.; Shao, H.; Guo, Z. Protein-mimicking nanowire-inspired electro-catalytic biosensor for probing acetylcholinesterase activity and its inhibitors. Talanta 2018, 183, 258–267. [Google Scholar] [CrossRef]
  29. Lee, K.; Shin, S.; Degen, T.; Lee, W.; Yoon, Y.S. In situ analysis of SnO2/Fe2O3/RGO to unravel the structural collapse mechanism and enhanced electrical conductivity for lithium-ion batteries. Nano Energy 2017, 32, 397–407. [Google Scholar] [CrossRef]
  30. Ricciardulli, A.G.; Yang, S.; Wetzelaer, G.J.A.; Feng, X.; Blom, P.W. Hybrid silver nanowire and graphene-based solution-processed transparent electrode for organic optoelectronics. Adv. Funct. Mater. 2018, 28, 1–6. [Google Scholar] [CrossRef]
  31. Wang, Z.; Wang, H.; Zhang, Z.; Liu, G. Electrochemical determination of lead and cadmium in rice by a disposable bismuth/electrochemically reduced graphene/ionic liquid composite modified screen-printed electrode. Sens. Actuators B 2014, 199, 7–14. [Google Scholar] [CrossRef]
  32. Wang, H.; Zhao, G.; Zhang, Z.; Yi, Y.; Wang, Z.; Liu, G. A portable electrochemical workstation using disposable screen-printed carbon electrode decorated with multiwall carbon nanotube-ionic liquid and bismuth film for Cd (II) and Pb (II) determination. Int. J. Electrochem. Sci. 2017, 12, 4702–4713. [Google Scholar] [CrossRef]
  33. Zinoubi, K.; Majdoub, H.; Barhoumi, H.; Boufi, S.; Jaffrezic-Renault, N. Determination of trace heavy metal ions by anodic stripping voltammetry using nanofibrillated cellulose modified electrode. J. Electroanal. Chem. 2017, 799, 70–77. [Google Scholar] [CrossRef]
  34. Philips, M.F.; Gopalan, A.I.; Lee, K.P. Development of a novel cyano group containing electrochemically deposited polymer film for ultrasensitive simultaneous detection of trace level cadmium and lead. J. Hazard. Mater. 2012, 237, 46–54. [Google Scholar] [CrossRef]
  35. Aragay, G.; Pons, J.; Merkoçi, A. Enhanced electrochemical detection of heavy metals at heated graphite nanoparticle-based screen-printed electrodes. J. Mater. Chem. 2011, 21, 4326–4331. [Google Scholar] [CrossRef]
  36. Sosa, V.; Barceló, C.; Serrano, N.; Ariño, C.; Díaz-Cruz, J.M.; Esteban, M. Antimony film screen-printed carbon electrode for stripping analysis of Cd (II), Pb (II), and Cu (II) in natural samples. Anal. Chim. Acta 2015, 855, 34–40. [Google Scholar] [CrossRef] [PubMed]
  37. Riman, D.; Jirovsky, D.; Hrbac, J.; Prodromidis, M.I. Green and facile electrode modification by spark discharge: Bismuth oxide-screen printed electrodes for the screening of ultra-trace Cd (II) and Pb (II). Electrochem. Commun. 2015, 50, 20–23. [Google Scholar] [CrossRef]
  38. Ruecha, N.; Rodthongkum, N.; Cate, D.M.; Volckens, J.; Chailapakul, O.; Henry, C.S. Sensitive electrochemical sensor using a graphene–polyaniline nanocomposite for simultaneous detection of Zn (II), Cd (II), and Pb (II). Electrochim. Acta 2015, 874, 40–48. [Google Scholar] [CrossRef] [PubMed]
  39. Zhu, L.; Xu, L.; Huang, B.; Jia, N.; Tan, L.; Yao, S. Simultaneous determination of Cd (II) and Pb (II) using square wave anodic stripping voltammetry at a gold nanoparticle-graphene-cysteine composite modified bismuth film electrode. Electrochim. Acta 2014, 115, 471–477. [Google Scholar] [CrossRef]
  40. Chen, L.; Su, Z.; He, X.; Liu, Y.; Qin, C.; Zhou, Y.; Li, Z.; Wang, L.; Xie, Q.; Yao, S. Square wave anodic stripping voltammetric determination of Cd and Pb ions at a Bi/Nafion/thiolated polyaniline/glassy carbon electrode. Electrochem. Commun. 2012, 15, 34–37. [Google Scholar] [CrossRef]
Figure 1. The process of a disposable screen-printed electrode (DSPE) modified to reduce graphene oxide and L-cysteine.
Figure 1. The process of a disposable screen-printed electrode (DSPE) modified to reduce graphene oxide and L-cysteine.
Sensors 20 01322 g001
Figure 2. Cyclic voltammograms of the DSPE, L-cysteine (LC)/DSPE, reduced graphene oxide (rGO)/DSPE and LC-rGO/DSPE in 5 mmol/L [Fe(CN)6]3−/4− containing 0.1 mol/L of KCl with a scan rate of 50 mV/s.
Figure 2. Cyclic voltammograms of the DSPE, L-cysteine (LC)/DSPE, reduced graphene oxide (rGO)/DSPE and LC-rGO/DSPE in 5 mmol/L [Fe(CN)6]3−/4− containing 0.1 mol/L of KCl with a scan rate of 50 mV/s.
Sensors 20 01322 g002
Figure 3. Electrochemical impedance spectroscopy of the DSPE, LC/DSPE, rGO/DSPE and LC-rGO/DSPE in 5 mmol/L [Fe(CN)6]3-/4− and 0.1 mol/L KCl with frequencies from 1 to 105 Hz.
Figure 3. Electrochemical impedance spectroscopy of the DSPE, LC/DSPE, rGO/DSPE and LC-rGO/DSPE in 5 mmol/L [Fe(CN)6]3-/4− and 0.1 mol/L KCl with frequencies from 1 to 105 Hz.
Sensors 20 01322 g003
Figure 4. Linear sweep voltammetry of (a) the LC-rGO/DSPE and (b) Bi/LC-rGO/DSPE in 0.1mol/L acetic acid buffer solution with a scan rate of 50 mV/s (supporting electrolyte: 0.1 mol/L pH 4.5 acetate buffer solution; Bi(III) concentration: 2 mg/L).
Figure 4. Linear sweep voltammetry of (a) the LC-rGO/DSPE and (b) Bi/LC-rGO/DSPE in 0.1mol/L acetic acid buffer solution with a scan rate of 50 mV/s (supporting electrolyte: 0.1 mol/L pH 4.5 acetate buffer solution; Bi(III) concentration: 2 mg/L).
Sensors 20 01322 g004
Figure 5. Differential pulse voltammetry of 30.0 μg/L of Cd(II) and Pb(II) by different electrodes: the Bi/DSPE, Bi/LC/DSPE, Bi/rGO/DSPE and Bi/LC-rGO/DSPE (deposition time: 240 s; deposition potential: −1.2 V; supporting electrolyte: 0.1 mol/L pH 4.5 acetate buffer solution; Bi(III) concentration: 2 mg/L).
Figure 5. Differential pulse voltammetry of 30.0 μg/L of Cd(II) and Pb(II) by different electrodes: the Bi/DSPE, Bi/LC/DSPE, Bi/rGO/DSPE and Bi/LC-rGO/DSPE (deposition time: 240 s; deposition potential: −1.2 V; supporting electrolyte: 0.1 mol/L pH 4.5 acetate buffer solution; Bi(III) concentration: 2 mg/L).
Sensors 20 01322 g005
Figure 6. Differential pulse voltammetry of Cd(II) and Pb(II) on the Bi/LC-rGO/DSPE with various concentrations from a to g: 1, 5, 10, 15, 20, 25 and 30 μg/L. The inset shows the calibration curves for the determination of Cd(II) and Pb(II).
Figure 6. Differential pulse voltammetry of Cd(II) and Pb(II) on the Bi/LC-rGO/DSPE with various concentrations from a to g: 1, 5, 10, 15, 20, 25 and 30 μg/L. The inset shows the calibration curves for the determination of Cd(II) and Pb(II).
Sensors 20 01322 g006
Figure 7. The selectivity of the Bi/LC-rGO/DSPE measured with 25 μg/L of Cd(II) and Pb(II) by adding 10-fold coexisting ions.
Figure 7. The selectivity of the Bi/LC-rGO/DSPE measured with 25 μg/L of Cd(II) and Pb(II) by adding 10-fold coexisting ions.
Sensors 20 01322 g007
Table 1. The performances of different electrodes for Cd(II) and Pb(II) detection.
Table 1. The performances of different electrodes for Cd(II) and Pb(II) detection.
ElectrodeLinear Range (μg/L)Limit of Detection (μg/L)Sensitivity (μA/(μg/L))Reference
Cd(II)Pb(II)Cd(II)Pb(II)Cd(II)Pb(II)
NFC/GCE0.64–6400.2–20000.320.110.640.13[33]
Bi/Polyaniline/GCE1.26–9070.26–58.730.260.171.133.25[34]
Bi/MWNTs/GO/GCE-5–400.310.21-0.06[35]
Sb/SPCE11.5–72.416.8–62.63.45.00.660.23[36]
BiO/SPCE0.5–120.5–120.20.20.830.67[37]
G/PANI/SPCE1–3001–3000.10.10.030.02[38]
Bi/Au-GN-Cys/GCE0.5–400.5–400.10.051.520.58[39]
Bi/Nafion/PANI-MES/GCE0.1–200.1–300.040.050.40.2[40]
Bi/LC-rGO/DSPE1–301–300.10.080.650.50This work
NFC: Nanofibrillated Cellulose; Bi/MWNTs/GO: Bi film/Multiwalled carbon nanotubes/Graphene oxide; G/PANI: graphene–polyaniline nanocomposite; Sb: Antimony film; Au-GN-Cys: Gold nanoparticles/Graphene/L-Cysteine; Bi/Nafion PANI-MES: Bi film/Nafion/2-mercaptoethanesulfonate thiolated polyaniline.
Table 2. Results for the determination of Cd(II) and Pb(II) in the extracted solutions.
Table 2. Results for the determination of Cd(II) and Pb(II) in the extracted solutions.
SampleAdd (μg/L)Bi/LC-rGO/DSPE (μg/L) aAAS (μg/L) aRecovery (%)
Cd(II)Pb(II)Cd(II)Pb(II)Cd(II)Pb(II)
1-1.57 ± 0.12 b3.24 ± 0.27 b1.633.3896.3295.86
56.45 ± 0.528.54 ± 0.41--97.23101.91
2-2.34 ± 0.474.67 ± 0.432.234.78104.9397.70
57.46 ± 0.659.36 ± 0.69--103.1895.71
3-5.65 ± 0.314.98 ± 0.315.594.67101.07106.64
510.28 ± 0.539.81 ± 0.53--97.07101.45
a Atomic absorption spectrometry. b Value is the mean of three measurements ± standard deviation.

Share and Cite

MDPI and ACS Style

Hou, X.; Xiong, B.; Wang, Y.; Wang, L.; Wang, H. Determination of Trace Lead and Cadmium in Decorative Material Using Disposable Screen-Printed Electrode Electrically Modified with Reduced Graphene Oxide/L-Cysteine/Bi-Film. Sensors 2020, 20, 1322. https://0-doi-org.brum.beds.ac.uk/10.3390/s20051322

AMA Style

Hou X, Xiong B, Wang Y, Wang L, Wang H. Determination of Trace Lead and Cadmium in Decorative Material Using Disposable Screen-Printed Electrode Electrically Modified with Reduced Graphene Oxide/L-Cysteine/Bi-Film. Sensors. 2020; 20(5):1322. https://0-doi-org.brum.beds.ac.uk/10.3390/s20051322

Chicago/Turabian Style

Hou, Xiaopeng, Benhai Xiong, Yue Wang, Li Wang, and Hui Wang. 2020. "Determination of Trace Lead and Cadmium in Decorative Material Using Disposable Screen-Printed Electrode Electrically Modified with Reduced Graphene Oxide/L-Cysteine/Bi-Film" Sensors 20, no. 5: 1322. https://0-doi-org.brum.beds.ac.uk/10.3390/s20051322

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop