Article

A Three-Parameter Weibull Distribution Method to Determine the Fracture Property of PMMA Bone Cement

1
School of Civil Engineering and Communication, North China University of Water Resources and Electric Power, Zhengzhou 450045, China
2
State Key Laboratory of Hydraulics and Mountain River Engineering, Sichuan University, Chengdu 610065, China
*
Author to whom correspondence should be addressed.
Submission received: 9 July 2022 / Revised: 22 August 2022 / Accepted: 24 August 2022 / Published: 30 August 2022

Abstract

:
Poly (methyl methacrylate) (PMMA) bone cement is an excellent biological material for anchoring joint replacements. Tensile strength f t and fracture toughness K IC have a considerable impact on its application and service life. Considering the variability of PMMA bone cement, a three-parameter Weibull distribution method is suggested in the current study to evaluate its tensile strength and fracture toughness distribution. The coefficients of variation for tensile strength and fracture toughness were the minimum when the characteristic crack of PMMA bone cement was α ch = 8 d av . Using the simple equation α ch = 8 d av and fictitious crack length Δ α fic = 1.0 d av , the mean value μ (= 43.23 MPa ), minimum value f t min (= 26.29 MPa ), standard deviation σ (= 6.42 MPa ) of tensile strength, and these values of fracture toughness ( μ = 1.77 MPa m 1 / 2 , K IC min = 1.02 MPa m 1 / 2 , σ = 0.2644 MPa m 1 / 2 ) were determined simultaneously through experimental data from a wedge splitting test. Based on the statistical analysis, the prediction line between peak load P max and equivalent area A e 1 A e 2 was obtained with 95% reliability. Nearly all experimental data are located within the scope of a 95% confidence interval. Furthermore, relationships were established between tensile strength, fracture toughness, and peak load P max . Consequently, it was revealed that peak load might be used to easily obtain PMMA bone cement fracture characteristics. Finally, the critical geometric dimension value of the PMMA bone cement sample with a linear elastic fracture was estimated.

1. Introduction

Poly (methyl methacrylate) (PMMA) bone cement is a typical quasi-brittle material commonly used in total joint replacements (TJRs) [1,2,3,4,5,6,7]. It can be anchored to continuous bone for applications such as fixation of the implant, delivery of antibiotics, filling of bone defects, etc. [8,9]. Fracture toughness K IC and tensile strength f t of PMMA bone cement are two essential indicators for determining stability and service life performance. The relevant information and mechanical characteristics of PMMA bone cement in the existing literature are presented in Table 1. As shown in Table 1, the test results of K IC and f t of PMMA bone cement have significant discreteness. Therefore, it is urgent and important to define a reasonable distribution function for statistical analysis to help obtain the exact parameter properties of PMMA bone cement.
The Weibull distribution function, as a skewed function, is commonly used to investigate the discreteness of the parameters of quasi-brittle materials. For example, Amaral investigated the flexural strength of granite through using the Weibull distribution function. Based on statistical results, engineers can specify specific types of brittle building materials to improve the safety of construction projects [22]. Blasi demonstrated, using Weibull statistics, that the size effect has a significant impact on the flexural strength of marble [23]. Li utilized the Weibull function to calculate the fracture toughness and tensile strength of concrete and predicted peak loads with 95% reliability [24]. Xu investigated the size effect of concrete by using the Weibull function. The better the uniformity of concrete is, the more obvious the size effect will be [25]. Lei reformulated and validated the three-parameter Weibull statistical fracture theory for uniaxial flexure of prismatic beams by analyzing the strength and experimental data of different ceramics [26]. Hu used the Weibull distribution to investigate the microcrack toughness of Yttria-stabilized zirconia (3Y-TZP) [27]. Gorjan indicated that the Weibull distribution provides the best accuracy for strength scattering with high alumina ceramics, outperforming the normal, log-normal, and Gamma distributions [28]. The Weibull function has also been used to investigate the fatigue properties of concrete [29,30]. According to Bala’s research, the fatigue life of a composite asphalt mixture follows the Weibull distribution [31]. Jin investigated the compressive strength variability of tungsten particle (Wp)-reinforced Zr-based bulk metallic glass composites using the Weibull distribution [32]. Furthermore, the Weibull function was employed to analyze polymer properties. In this context, Carmona investigated the failure probability of natural Luffa cylindrica fibers using the Weibull distribution function [33], whereas Sia utilized the Weibull distribution to quantify variations in the tensile strength of pineapple leaf fibers [34]. Wang investigated the effect of the length and diameter of bamboo fibers on tensile strength by using a modified Weibull model. It was shown that the accuracy of the Weibull model in terms of strength and predicted size correlation of bamboo fibers was satisfactory [35]. Wang described the statistical distribution of the critical energy release rate (Gc) of the transverse layer (the transverse layer) by using a two-parameter Weibull function. Combined with the numerical model, the mechanical behavior of the laminated composite formed by unidirectional fiber reinforced laminae can be calculated [36]. Equivalent fracture toughness of EPOLAM 2025 CT epoxy resin was also analyzed using the Weibull distribution [37].
Previous studies have clearly revealed that the parameter characteristics of quasi-brittle materials and high polymers can be statistically analyzed with the Weibull distribution function. However, PMMA bone cement is a high polymer quasi-brittle material with uncertain distribution characteristics regarding its fracture toughness and tensile strength parameters. In this paper, based on the fracture test results for wedge-splitting specimens (WS) [38], K IC and f t of PMMA bone cement were investigated using the three-parameter Weibull distribution function; the mean and variance of its fracture toughness and tensile strength parameters were obtained. According to the findings, the minimum coefficient of variation ( CV ) was obtained when the relative characteristic crack was α ch = 8 d av . Using the constant value of β (= 1.0), the quasi-brittle fracture of PMMA bone cement was predicted. Furthermore, the peak loads P max with a specified 95% reliability were predicted by using three-parameter Weibull distribution analysis.

2. The Theoretical Background of PMMA Distribution Characteristics

2.1. Intrinsic Causes of PMMA Fracture Discreteness

A previous experimental study on the subject revealed that, since the same batch of PMMA bone cement specimens with the same size and loading procedure were used on the same equipment, the recorded peak load P max maintained its discreteness [38]. As illustrated in Figure 1, the fictitious crack growth in PMMA bone cement specimens of the same size was discrete due to the random distribution of material particles, which caused the unpredictability of P max . This meant that separation was the key feature of PMMA bone cement. The variation dispersion coefficient β is introduced in the current paper to investigate this discreteness, and Equation (1) was utilized to establish the relationship between the fictitious crack growth Δ a fic and the average particle size d av at the peak load P max [39]. As shown in Figure 2, the three-parameter Weibull distribution function was used to characterize the distribution characteristics of the variation dispersion coefficient β to assess the statistical characteristics of the fracture strength of PMMA bone cement. Further, Δ a fic and P max related to β match the three-parameter Weibull distribution.
Δ α fic = β d av
The Weibull distribution function is commonly used to characterize the statistical distribution of parameters of materials such as rocks, concrete, ceramics, and polymers. This section provides a brief description of the three-parameter Weibull distribution function. The parameters x ( x 1 , x 2 , x 3 ... x i ... x N ) were adjusted to fit the three-parameter Weibull distribution function. The probability density of parameter x is calculated by Equation (2) as follows:
P f ( x x i ) = F i = N i N
where P f ( x x i ) is equal to or less than the cumulative frequency of x , N represents the total number of parameters, and N i is equal to or less than the value of x i .
The probability density function of the three-parameter Weibull distribution can be expressed as follows:
P f ( x ) = f ( x ) = K ω x x min ω K 1 exp x x min ω K x x min
The corresponding probability distribution function is shown in Equation (4)
P f ( x x i ) = F ( x x i ) = 1.0 exp x x min ω K x x min
where ω is the size parameter of the Weibull distribution, K is the shape parameter of the Weibull distribution, and x is the position parameter (it is the minimum value obtained by fitting the parameter x , no parameter is smaller than this point).
The mean value μ and standard deviation σ σ of the three-parameter Weibull distribution function is obtained by Equations (5) and (6):
μ = E ( x ) = x min + ω Γ 1 + 1 K
σ 2 = ω 2 Γ 1 + 1 K Γ 2 1 + 1 K
where Γ is the gamma function.

2.2. A Brief Description of the Boundary Effect Model (BEM)

Wedge-splitting specimens (WS) [11,13], three-point bending specimens [10], and chevron-notched short-rod specimens [14,15,16] are commonly used in laboratory testing to examine the fracture parameters of PMMA bone cement. The tensile strength of PMMA bone cement was obtained through four-point bending specimens [19] and dumb-bell specimens [18,21]. A large number of experiments are required to obtain the necessary parameters, which increases costs and the time required for the experiments. It is proposed in this section that the BEM model be adapted to PMMA bone cement specimens; thus, the two parameters of fracture toughness and tensile strength can be calculated simultaneously only based on WS test results. In this paper, the distribution form of the nominal stress at the crack tip was assumed to be rectangular [39], as shown in Figure 3. Furthermore, the BEM considering the variation dispersion coefficient β , the average aggregate size d av , and the peak load P max is given in Equation (7) [39,40,41,42].
1 σ n 2 P max , Δ α fic = 1 σ n 2 P max , β d av = 1 f t 2 + 4 a e K IC 2
where σ n represents nominal stress.
For the WS specimens, the equivalent crack a e can be determined by the following Equation [39,40,41,42]:
a e = 2 ( 1 α ) 2 2 + α × Y α 1.12 2 × a 0
where α is the ratio of a 0 to W, α = a 0 / W , a 0 is the initial crack length, and Y α is the geometric factor. The expression of Y α is as follows [39,40,41,42]:
Y α = 2 + α × 0.866 + 4.64 α 13.32 α 2 + 14.72 α 3 5.6 α 4 4 π α 1 α 3 / 2
According to the stress distribution form in Figure 4, the expression for nominal stress σ n can be derived by balancing the force and moment of force according to the following Equation [39,40,41,42]:
σ n P max , β d av = P max 3 W 2 + W 1 6 B W 1 2 6 + W 1 6 β d av + W a 0 2 β d av
where B is the width of the specimen and W is the height of the specimen.
W 1 = W a 0 β d av
W 2 = W + a 0 + β d av
In the BEM, the ratio of characteristic crack length α ch to average aggregate size d av is a constant C, obtained as follows [39]:
C = α ch / d av = 0.25 K IC / f t 2 / d av
Combining Equations (7), (13), and (14), (15) provides the following:
f t = σ n P max , β d av 1 + a e C d av
K IC = 2 σ n P max , β d av a e + C d av
Substituting Equation (10) into Equations (14)–(17) obtains:
f t = P max 3 W 2 + W 1 6 B W 1 2 6 + W 1 6 β d av + W a 0 2 β d av × 1 + a e C d av
K IC = P max 3 W 2 + W 1 3 B W 1 2 6 + W 1 6 β d av + W a 0 2 β d av × a e + C d av
Individual fracture toughness and tensile strength can be calculated according to C, β , d av , P max , and the geometric parameters of the specimens, as seen from Equations (16) and (17). The dispersion coefficient β , fictitious crack growth Δ a fic , and peak load P max all provide the three-parameter Weibull function distributions. Therefore, tensile strength f t and fracture toughness K IC also provide the three-parameter Weibull function distributions. Additionally, the mean value μ and standard deviation σ of the statistical distribution for PMMA bone cement strength fracture parameters were obtained according to Equations (5) and (6).

3. Statistical Analysis of Test Data

3.1. Raw Data of the Experiment

A specific quantity of the experimental sample data set is needed to test the validity of the statistical analysis results. In this paper, experimental data from PMMA bone cement wedge separation samples by Merta were selected for analysis as a data set [38]. The geometric information of the specimens and the test results of the load peak value P max are presented in Table 2. The width B of the test specimens was 6 mm. The α (the ratio of a 0 to W) variation range of the specimen height W (= 15 mm) was 0.13~0.53. The α (the ratio of a 0 to W) variation range of the specimen height W (= 22 mm) was 0.09~0.64. The α (the ratio of a 0 to W) variation range of the specimen height W (= 29 mm) was 0.07~0.69. The α (the ratio of a 0 to W) variation range of the specimen height W (= 36 mm) was 0.05~0.81. For each combination of a 0 and W, 5 to 9 specimens were produced, with a total of 160 specimens, and all specimens with cracks larger than 2 mm were removed, thus leaving 96 specimens. The numbering rule for specimens in this paper is E + “height” + “-initial crack length” for the convenience of expression. As an example, E15-2 represents a specimen with a height of 15 mm and a 2 mm initial crack length. In the analysis, the constant C was taken as 0.5, 1, 1.5, 2, 2.5, 3, 4, 5, 6, 8, 9, and 10, respectively; β adopted the constant method to take a value (= 1, 2, 3, 4). According to reports in the relevant literature, the aggregate size d of PMMA bone cement ranges from 5 to 100 μ m 40, with average aggregate size d av equal to 52.5 μ m . Fracture toughness K IC and tensile strength f t were calculated according to Equations (16) and (17). The statistical distribution results of fracture toughness K IC and tensile strength f t were then obtained using the three-parameter Weibull distribution function.

3.2. Raw Data of the Experiment

Based on the findings of the peak load P max test on 96 PMMA bone cement WS specimens and using Equations (16) and (17), 96 values of tensile strength f t and fracture toughness K IC were calculated for each combination of C, β , and d av values, respectively. Further analyses were performed using the three-parameter Weibull function, and Equations (5) and (6) were used to obtain the statistical mean value μ , standard deviation σ , and coefficient of variation CV (= σ / μ ) for fracture strength parameters of PMMA bone cement. The three-parameter Weibull distributions of tensile strength f t and fracture toughness K IC for C = 8.0 at d av = 52.5   μ m are shown in Figure 4 and Figure 5, which are constrained by article space. It can be clearly seen that the distribution curve is in good agreement with the experimental results.
The statistical findings for the experimental data obtained from Merta using the three-parameter Weibull distribution are listed in Table 3 [38]. The coefficient of variation CV curves of tensile strength f t and fracture toughness K IC , with the respective C value, are shown in Figure 6. As can be observed, the coefficient of variation CV of f t reaches a lesser or minimal value when C = 8.0, regardless of the value of β . As shown in Table 3, the value β has little impact on the statistical characteristics of fracture strength when the value of C is constant. For the convenience of calculation, the constant C for the PMMA bone cement specimen was 8.0, and the variation dispersion coefficient β was 1.0. The calculated average tensile strength f t of PMMA bone cement was 43.23 MPa . It is located in the range of 33.0 MPa ~ 51.4 MPa for tensile strength f t which has been reported in related literature [17,18,19,20,21]. The average value of fracture toughness K IC was 1.77 MPa m 1 / 2 , which likewise falls within the range of 1.0 MPa m 1 / 2 ~ 2.7 MPa m 1 / 2 as reported by related literature [10,11,12,13,14,15,16,17]. In addition, the minimum value f t min of tensile strength was 26.29 MPa . The minimum value K IC min of fracture toughness was 1.02 MPa m 1 / 2 . Minimum values of tensile strength and fracture toughness provided a basis for evaluating the safety of PMMA bone cement.

4. Three-Parameter Weibull Distribution Prediction of PMMA Bone Cement Fracture

The constant C was taken as 8.0 and the variation dispersion coefficient β as 1.0 based on the analysis results in Section 3. The statistical distribution results of tensile strength f t and fracture toughness K IC were obtained from the three-parameter Weibull distribution function. Combined with Equation (7), the complete fracture prediction curve of PMMA bone cement can be predicted.
The lower limit of the material parameter can be established based on the minimum value of the parameter, according to the three-parameter Weibull distribution principle. Namely, according to the minimum value of tensile strength f t and fracture toughness K IC , a minimum safety line is determined as the minimum safety control index. According to the complete fracture prediction line of PMMA bone cement, the tensile strength ( f t ) control zone ( a e / a ch 0.1 ), the quasi-brittle fracture control zone ( 0.1 a e / a ch 10 ), and the fracture toughness ( K IC ) control zone ( a e / a ch 10 ) can be obtained.
Figure 7 depicts the predicted results of the 95% confidence interval ( μ ± 2 σ ) of the PMMA bone cement WS specimen based on the mean value and standard deviation. By analyzing the minimum value, the lower limit of the material parameter was determined. As shown in Figure 7, the PMMA bone cement WS data points in the laboratory experiments are all in the quasi-brittle fracture control zone and above the lower limit fracture failure curve fitted by the minimum value. Almost all experimental data are within the 95% confidence interval.
By transforming Equations (16) and (17), a simplified prediction model for the tensile strength f t and fracture toughness K IC of PMMA bone cement WS specimens can be obtained. Tensile strength f t = P max / A e 1 and fracture toughness K IC = P max / A e 2 , with respect to the equivalent area A e 1 A e 2 and the peak load P max , can be obtained as shown in Equations (18) and (19). Conversely, the peak load P max of the test piece can also be calculated from the equivalent area A e 1 A e 2 and f t ( K IC ). Mean value μ and standard deviation σ of fracture strength parameters were obtained according to three-parameter Weibull distribution statistics. As shown in Figure 8, the predicted results of the peak load P max for the PMMA bone cement WS specimen with a 95% confidence interval ( μ ± 2 σ ) were obtained at the lower limit of the prediction of peak load P max . As can be seen in Figure 8, the vast majority of data are within the 95% confidence range, and all experimental data were above the minimum fit’s lower limit prediction curve.
P max = f t 6 B W 1 2 6 + W 1 6 β d av + W a 0 2 β d av 3 W 2 + W 1 × 1 + a e C d av = f t A e 1
P max = K IC 6 B W 1 2 6 + W 1 6 β d av + W a 0 2 β d av 1.12 × π 3 W 2 + W 1 × a e + C d av = K IC A e 2

5. Discussion

The situation of β = 1.0, 2.0, 3.0, and 4.0 is discussed in this section according to Table 3. When C = 8 , the mean value, minimum value, standard deviation σ , coefficient of variation CV of tensile strength f t , and these values of fracture toughness K IC presented slight variations with increases in β . For example, when β was increased from 1 to 4, the mean values of f t only varied from 42.43 MPa to 43.23 MPa , and values of fracture toughness K IC only varied from 1.74 MPa m 1 / 2 to 1.77 MPa m 1 / 2 . Limited by space, only the destruction full curve and prediction curves for β = 4 were plotted, as shown in Figure 9 and Figure 10. Nearly all experimental data are also located within the scope of a 95% confidence interval. However, as shown in Figure 4, when β = 1 the correlation coefficients of f t and K IC both have the maximum value of 0.9867 and 0.9191, respectively. With the increase in β , the correlation coefficients of f t and K IC present a downtrend. Thus, it is reasonable to take β = 1 for the convenience of calculation and application. In addition, more test data are required to demonstrate the correctness of three-parameter Weibull distribution of material parameters in the future.

6. Conclusions

A three-parameter Weibull distribution approach was used to assess the fracture property of PMMA bone cement based on the experimental results of wedge-splitting specimens of PMMA bone cement. During the process, the characteristic crack of PMMA bone cement ( α ch = C d av = 0.5 d av , 1.0 d av , 1.5 d av , 2.0 d av , 2.5 d av , 3.0 d av , 4.0 d av , 5.0 d av , 6.0 d av , 7.0 d av , 8.0 d av , 9.0 d av , 10.0 d av ) at peak loads, and at a constant value of β (= 1.0, 2.0, 3.0, 4.0), was investigated.
The results obtained from the study are as follows:
  • Due to the random distribution of PMMA bone cement particles, a three-parameter Weibull distribution method was employed to analyze the discrete characteristic of the fracture property. For different values of β , when the characteristic crack was α ch = 8 d av , tensile strength f t and fracture toughness K IC with the smallest coefficient of variation ( CV ) were obtained. The minimum CV values for f t and K IC were 0.1485 and 0.1492, respectively.
  • The mean value μ (= 43.23 MPa ), minimum value f t min (= 26.29 MPa ), standard deviation σ (= 6.42 MPa ) of tensile strength f t , and these values of fracture toughness K IC ( μ = 1.77 MPa m 1 / 2 , K IC min = 1.02 MPa m 1 / 2 , σ = 0.2644 MPa m 1 / 2 ) were determined simultaneously from the three-parameter Weibull distribution method by using the characteristic crack α ch = 8 d av and the fictitious crack growth Δ α fic = 1.0 d av . Furthermore, the lower safety control index of PMMA bone cement was obtained based on the statistical properties of the minimum value of f t and K IC .
  • The whole prediction breaking curve with 95% reliability for PMMA bone cement was obtained. Additionally, based on the simplified prediction model, the prediction line between peak load P max and equivalent area A e 1 A e 2 was obtained with 95% reliability. Nearly all experimental data are located within the scope of a 95% confidence interval. All experimental data were gained above the lower limit for safe prediction value.

Author Contributions

Conceptualization, J.G. and L.L.; methodology, J.G.; software, L.L.; validation, L.L., H.C. and J.G.; formal analysis, H.C.; investigation, L.N.; resources, S.H.; data curation, L.L.; writing—original draft preparation, H.C.; writing—review and editing, H.C. and J.G.; visualization, L.L.; supervision, L.N.; project administration, H.L.; funding acquisition, S.H. All authors have read and agreed to the published version of the manuscript.

Funding

Program for Science & Technology Innovation Talents in Universities of Henan Province (20HASTIT013), Sichuan Univ, State Key Lab Hydraul & Mt River Engn (SKHL2007), The National Natural Science Foundation of China (Grant No. 52179132). Innovation Ability Enhancement project of the 13th postgraduate of North China University of Water Resources and Electric Power (YK-2021-20). Key scientific research projects of Henan Province colleges and universities (15A560025). The Plan for Tackling Key Problems in Science and Technology of Henan Province in 2021 (212102310057), High Level Talents Scientific Research Start Project of North China University of Water Resources and Electric Power (202110016).

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The data presented in this study are available on request from the corresponding author.

Conflicts of Interest

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

References

  1. Lewis, G. Properties of acrylic bone cement: State of the art review. J. Biomed. Mater. Res. A 1997, 38, 155–182. [Google Scholar]
  2. Yi, S.; Rim, D.C.; Park, S.W.; Murovic, J.A.; Lim, J.; Park, J. Biomechanical Comparisons of Pull Out Strengths After Pedicle Screw Augmentation with Hydroxyapatite, Calcium Phosphate, or Polymethylmethacrylate in the Cadaveric Spine. World Neurosurg. 2015, 83, 976–981. [Google Scholar] [CrossRef] [PubMed]
  3. He, Z.; Zhai, Q.; Hu, M.; Cao, C.; Wang, J.; Yang, H.; Li, B. Bone cements for percutaneous vertebroplasty and balloon kyphoplasty: Current status and future developments. J. Orthop. Transl. 2015, 3, 1–11. [Google Scholar]
  4. Ali, U.; Juhanni, K.; Abd, B.; Buang, N.A. A Review of the Properties and Applications of Poly (Methyl Methacrylate) (PMMA). Polym. Rev. 2015, 55, 678–705. [Google Scholar] [CrossRef]
  5. Roozbahani, M.; Kharaziha, M.; Emadi, R. Fabrication and characterization of laponite-calcium phosphate based cement for filling bone defects. Mater. Today Proc. 2018, 5, 15754–15760. [Google Scholar] [CrossRef]
  6. Roozbahani, M.; Alehosseini, M.; Kharaziha, M. Nano-calcium phosphate bone cement based on Si-stabilized α-tricalcium phosphate with improved mechanical properties. Mat. Sci. Eng. C-Mater. 2017, 81, 532–541. [Google Scholar]
  7. Dunne, N.J.; Orr, J.F. Curing characteristics of acrylic bone cement. J. Mater. Sci. Mater. Med. 2002, 13, 17–22. [Google Scholar] [CrossRef]
  8. Azuara, G.; García-García, J.; Ibarra, B.; Parra-Ruiz, F.J.; Asúnsolo, A.; Ortega, M.A.; Vázquez-Lasa, B.; Buján, J.; San Román, J.; de la Torre, B. Experimental study of the application of a new bone cement loaded with broad spectrum antibiotics for the treatment of bone infection. Rev. Esp. Cir. Ortop. Traumatol. 2019, 63, 95–103. [Google Scholar]
  9. Thielen, T.; Maas, S.; Waldmann, D.; Anagnostakos, K.; Kelm, J. Development of a reinforced PMMA-based hip spacer adapted to patients’ needs—ScienceDirect. Med. Eng. Phys. 2009, 31, 930–936. [Google Scholar] [CrossRef]
  10. Kim, S.; Baril, C.; Rudraraju, S.; Ploeg, H.L. Influence of Porosity on Fracture Toughness and Fracture Behavior of Antibiotic-Loaded PMMA Bone Cement. J. Biomech. Eng. 2022, 144, 011006–011014. [Google Scholar]
  11. Biggs, P.; Jones II, L.; Lewis, G. An autonomically-healed PMMA bone cement: Influence of the crystal size of Grubbs’ catalyst on fracture toughness and polymerisation rate. Int. J. Nano Biomater. 2009, 2, 494–504. [Google Scholar] [CrossRef]
  12. Bokam, P.; Germaneau, A.; Breque, C.; Rigoard, P.; Vendeuvre, T.; Valle, V. Fracture behavior of cancellous bone and cancellous bone-PMMA bone cement interface: An experimental study using an integrated methodology (wedge splitting test and Heaviside-based digital image correlation). J. Mech. Behav. Biomed. Mater. 2021, 122, 104663. [Google Scholar] [PubMed]
  13. Lewis, G.; Mladsi, S. Correlation between impact strength and fracture toughness of PMMA-based bone cements. Biomaterials 2000, 21, 775–781. [Google Scholar] [CrossRef]
  14. Lewis, G.; Nyman, J.; Hai, H.T. The apparent fracture toughness of acrylic bone cement: Effect of three variables. Biomaterials 1998, 19, 961–967. [Google Scholar] [CrossRef]
  15. Lewis, G. Effect of lithotriptor treatment on the fracture toughness of acrylic bone cement. Biomaterials 1992, 13, 225–229. [Google Scholar] [CrossRef]
  16. Wang, C.T.; Pilliar, R.M. Fracture toughness of acrylic bone cements. J. Mater. Sci. 1989, 24, 3725–3738. [Google Scholar] [CrossRef]
  17. Kindt-Larsen, T.; Smith, D.B.; Jensen, J.S. Innovations in acrylic bone cement and application equipment. J. Appl. Biomater. 1995, 6, 75–83. [Google Scholar] [CrossRef]
  18. Harper, E.J.; Braden, M.; Bonfield, W.; Wahlig, E.D. Influence of sterilization upon a range of properties of experimental bone cements. J. Mater. Sci. Mater. Med. 1997, 8, 849–853. [Google Scholar]
  19. Kjellson, F.; Abdulghani, S.; Tanner, K.E.; Mccarthy, I.D.; Lidgren, L. Effect of iodixanol particle size on the mechanical properties of a PMMA based bone cement. J. Mater. Sci. Mater. Med. 2007, 18, 1043–1051. [Google Scholar]
  20. Harper, E.J.; Braden, M.; Bonfield, W. Mechanical properties of hydroxyapatite reinforced poly(ethylmethacrylate) bone cement after immersion in a physiological solution: Influence of a silane coupling agent. J. Mater. Sci. Mater. Med. 2000, 11, 491–497. [Google Scholar] [CrossRef]
  21. Kjellson, F.; Wang, J.S.; Almén, T.; Mattsson, A.; Lidgren, L. Tensile properties of a bone cement containing non-ionic contrast media. J. Mater. Sci. Mater. Med. 2001, 12, 889–894. [Google Scholar] [CrossRef] [PubMed]
  22. Amaral, P.M.; Fernandes, J.C.; Rosa, L.G. Weibull statistical analysis of granite bending strength. Rock Mech. Rock Engng. 2008, 41, 917–928. [Google Scholar] [CrossRef]
  23. Blasi, P.; Frisa Morandini, A.; Mancini, R.; Marini, P. An experimental investigation on marble testing for bending strength. RocMaquina 2000, 38, 20–25. [Google Scholar]
  24. Li, L.; Guan, J.; Yuan, P.; Yin, Y.; Li, Y. A Weibull distribution-based method for the analysis of concrete fracture. Eng. Fract. Mech. 2021, 256, 107964. [Google Scholar] [CrossRef]
  25. Xu, J.; He, X. Size effect on the strength of a concrete member. Eng. Fract. Mech. 1990, 35, 687–695. [Google Scholar] [CrossRef]
  26. Lei, W.S.; Zhang, P.; Yu, Z.; Qian, G. Statistics of ceramic strength: Use ordinary Weibull distribution function or Weibull statistical fracture theory? Ceram. Int. 2020, 46, 20751–20768. [Google Scholar] [CrossRef]
  27. Hu, X.Z. Determination of pore distribution n Yttria-stabilized zirconia from the Weibull strength distribution. Mater. Forum 1992, 16, 313. [Google Scholar]
  28. Gorjan, L.; Ambrožič, M. Bend strength of alumina ceramics: A comparison of Weibull statistics with other statistics based on very large experimental data set. J. Eur. Ceram. Soc. 2012, 30, 1221–1227. [Google Scholar] [CrossRef]
  29. Goel, S.; Singh, S.P.; Singh, P. Fatigue Analysis of Plain and Fiber-Reinforced Self-Consolidating Concrete. ACI Mater. J. 2012, 109, 573–582. [Google Scholar]
  30. Li, H.; Zhang, M.; Ou, J. Flexural fatigue performance of concrete containing nano-particles for pavement. Int. J. Fatigue 2007, 29, 1292–1301. [Google Scholar] [CrossRef]
  31. Bala, N.; Napiah, M. Fatigue life and rutting performance modelling of nanosilica/polymer composite modified asphalt mixtures using Weibull distribution. Int. J. Pavement Eng. 2018, 21, 497–506. [Google Scholar]
  32. Jin, J.; Ma, Y.; Wen, H.; Gong, P.; Wang, X. Mechanical properties and Weibull reliability analysis of tungsten-particle reinforced Zr-based bulk metallic glass composites. Mater. Lett. 2022, 306, 130879. [Google Scholar] [CrossRef]
  33. Carmona, A.R.; Colorado, H. Luffa fibers as promising reinforcement for polymer composites: Mechanical characterization of NaOH treated and untreated dumbbell test-pieces with Weibull statistics. J. Compos. Mater. 2020, 55, 1667–1681. [Google Scholar] [CrossRef]
  34. Sia, C.; Wong, J.S.Y.; Thangavelu, S.K.; Chong, K.H. Weibull Strength Analysis of Pineapple Leaf Fiber. Mater. Sci. Forum 2021, 1030, 45–52. [Google Scholar] [CrossRef]
  35. Wang, F.; Shao, J.X. Modified Weibull Distribution for Analyzing the Tensile Strength of Bamboo Fibers. Polymers 2014, 6, 3005–3018. [Google Scholar] [CrossRef]
  36. Wang, F.; Ding, J.; Chen, Z.Q. Statistical Analysis of the Progressive Failure Behavior for Fiber-Reinforced Polymer Composites under Tensile Loading. Polymers 2014, 6, 145–159. [Google Scholar]
  37. Álvarez-Vázquez, A.; Muñiz-Calvente, M.; Fernández Fernández, P.; Fernández-Canteli, A.; Lamela-Rey, M.J.; Pintado, J.M. Probabilistic Assessment of Fracture Toughness of Epoxy Resin EPOLAM 2025 Including the Notch Radii Effect. Polymers 2021, 13, 1857. [Google Scholar] [CrossRef]
  38. Merta, I.; Berger, L.; Heidfogel, G.; Kuhn, K.D.; Lewis, G.; Tschegg, E.K. Size and boundary effects on notch tensile strength and fracture properties of PMMA bone cement. Polym. Test. 2017, 59, 441–448. [Google Scholar]
  39. Guan, J.; Yuan, P.; Hu, X.; Qing, L.; Yao, X. Statistical analysis of concrete fracture using normal distribution pertinent to maximum aggregate size. Theor. Appl. Fract. Mech. 2019, 101, 236–253. [Google Scholar] [CrossRef]
  40. Mousa, W.F.; Kobayashi, M.; Shinzato, S.; Kamimura, M. Biological and mechanical properties of PMMA-based bioactive bone cements. Biomaterials 2000, 21, 2137–2146. [Google Scholar] [CrossRef]
  41. Guan, J.; Hu, X.; Li, Q. In-depth analysis of notched 3-p-b concrete fracture. Eng. Fract. Mech. 2016, 165, 57–71. [Google Scholar] [CrossRef]
  42. Hu, X.; Guan, J.; Wang, Y.; Keating, A.; Yang, S. Comparison of Boundary and Size Effect Models Based on New Developments. Eng. Fract. Mech. 2017, 175, 146–167. [Google Scholar] [CrossRef]
Figure 1. Discreteness of fictitious crack growth at peak loads.
Figure 1. Discreteness of fictitious crack growth at peak loads.
Polymers 14 03589 g001
Figure 2. Distribution characteristics of the fracture strength parameters of PMMA bone cement.
Figure 2. Distribution characteristics of the fracture strength parameters of PMMA bone cement.
Polymers 14 03589 g002
Figure 3. BEM analysis for PMMA bone cement.
Figure 3. BEM analysis for PMMA bone cement.
Polymers 14 03589 g003
Figure 4. Three-parameter Weibull distribution of fracture toughness K IC at d av = 52.5   μ m .
Figure 4. Three-parameter Weibull distribution of fracture toughness K IC at d av = 52.5   μ m .
Polymers 14 03589 g004
Figure 5. Three-parameter Weibull distribution of tensile strength f t at d av = 52.5   μ m .
Figure 5. Three-parameter Weibull distribution of tensile strength f t at d av = 52.5   μ m .
Polymers 14 03589 g005aPolymers 14 03589 g005b
Figure 6. Variation curve of coefficient of variation CV for fracture toughness and tensile strength with C value at d av = 52.5   μ m .
Figure 6. Variation curve of coefficient of variation CV for fracture toughness and tensile strength with C value at d av = 52.5   μ m .
Polymers 14 03589 g006
Figure 7. Destruction prediction full curve for d av = 52.5   μ m , β = 1 , and C = 8 .
Figure 7. Destruction prediction full curve for d av = 52.5   μ m , β = 1 , and C = 8 .
Polymers 14 03589 g007
Figure 8. (a) Prediction curve of f t for d av = 52.5   μ m , β = 1 , and C = 8 . (b) Prediction curve of K IC for d av = 52.5   μ m , β = 1 , and C = 8 .
Figure 8. (a) Prediction curve of f t for d av = 52.5   μ m , β = 1 , and C = 8 . (b) Prediction curve of K IC for d av = 52.5   μ m , β = 1 , and C = 8 .
Polymers 14 03589 g008
Figure 9. Destruction prediction full curve for d av = 52.5   μ m , β = 4 , and C = 8 .
Figure 9. Destruction prediction full curve for d av = 52.5   μ m , β = 4 , and C = 8 .
Polymers 14 03589 g009
Figure 10. (a) Prediction curve of f t for d av = 52.5   μ m , β = 4 , and C = 8 . (b) Prediction curve of K IC for d av = 52.5   μ m , β = 4 , and C = 8 .
Figure 10. (a) Prediction curve of f t for d av = 52.5   μ m , β = 4 , and C = 8 . (b) Prediction curve of K IC for d av = 52.5   μ m , β = 4 , and C = 8 .
Polymers 14 03589 g010aPolymers 14 03589 g010b
Table 1. PMMA bone cement research table.
Table 1. PMMA bone cement research table.
No.Main IngredientType of Test SpecimenTensile Strength
( MPa )
Fracture Toughness
( MPa m 1 / 2 )
SourceRemark
1Palacos®R3-p-b * 2.70 ± 0.22Kim [10]
2Simplex®PWS * 2.15 ± 0.11Biggs [11]
3Kyphon Xpede Bone CementWS * 1.11 ± 0.03Pba [12]
4Palacos®RWS * 1.85 ± 0.12Lewis [13]
5Palacos®R
CMW3
Osteopal
CNSR * 1.75 ± 0.07
1.92 ± 0.10
Lewis [14]
6Palacos®RCNSR * 1.81 ± 0.14Lewis [15]
7Simplex®PCNSR * 1.30 ± 0.26
1.00 ± 0.05
1.01 ± 0.04
Wang [16]① UHM specimens
② CHM specimens
③ CFG specimens
8Palacos®R
CMW-1
Simplex P
Zimmer D
CNSR *33 ± 2
40 ± 8
1.59 ± 0.07
1.73 ± 0.17
Kindt-
Larsen [17]
④ Open bowl mixing
⑤ Vacuum mixing
9Palacos®RDB *51.4 ± 3.47 Harper [18]
10IDX
Palacos®R
4-p-b *43.4 ± 1.6 Jellson [19]
11commercially available PMMADB *44.7 ± 4.3 Harper [20]
12Palacos®R
IDX
IHX
DB *42 Kjellson [21]
Note *: 3-p-b: three-point bending specimens; WS: wedge-splitting specimens; CNSR: chevron notched short-rod specimens; DB: dumb-bell specimens; 4-p-b: four-point bending specimens; UHM: uncontrolled hand-mixed; CHM: controlled hand-mixed; CFG: centrifuged.
Table 2. Experimental results of WS samples.
Table 2. Experimental results of WS samples.
Height
(W) (mm)
Initial Crack Length
( a 0 )   ( mm )
Peak Load ( P max )   ( N ) Height
(W) (mm)
Initial Crack Length
( a 0 )   ( mm )
Peak   Load   ( P max )   ( N ) Height
(W) (mm)
Initial Crack Length
( a 0 )   ( mm )
Peak   Load   ( P max )
(N)
152352.47295521.233611318.81
325.47518.00298.50
328.45453.23290.32
372.29447.03291.82
5211.808422.5714285.38
207.34361.23282.65
173.90356.99254.17
171.68354.99231.63
889.1911314.4817216.53
91.43292.38218.52
71.36283.60194.00
69.13272.27194.49
222499.7814223.6520171.69
457.27219.08160.56
463.97215.19133.80
450.63201.33113.25
5361.5117140.9223107.06
314.95129.47102.60
299.04118.8592.95
287.90122.8483.53
8245.13362096.192682.54
245.7383.06
232.1075.2071.40
230.3172.18
11216.922693.652942.91
158.99691.56
152.10580.1035.23
152.80528.34
1484.775515.45\\\
85.47443.64
78.76418.11
74.65398.80
292670.708381.21\\\
637.41376.26
635.20358.93
621.61332.43
Table 3. PMMA three-parameter Weibull distribution calculation results at d av = 52.5   μ m .
Table 3. PMMA three-parameter Weibull distribution calculation results at d av = 52.5   μ m .
C = α ch / d av β = Δ a fic / d av 1234
0.5 f t 155.61154.64153.69152.76
f t min 85.4083.8882.3080.77
σ 23.8223.7623.7023.65
CV 0.15310.15360.15420.1548
K IC 1.591.581.571.57
K IC min 0.860.790.780.77
σ 0.24610.24540.24470.2441
CV 0.15430.15490.15540.1559
1.0 f t 110.90110.21109.53108.87
f t min 61.5960.5659.4358.39
σ 16.8916.8416.8016.76
CV 0.15230.15280.15340.1539
K IC 1.611.601.591.58
K IC min 0.820.810.800.78
σ 0.24680.24600.24530.2446
CV 0.15360.15400.15450.1550
1.5 f t 91.2590.6890.1289.57
f t min 51.3250.4149.5148.65
σ 13.8313.7913.7513.71
CV 0.15160.15210.15260.1531
K IC 1.621.611.601.59
K IC min 0.840.820.810.80
σ 0.24760.24670.24600.2452
CV 0.15290.15330.15380.1542
2.0 f t 79.6279.1378.6478.16
f t min 45.2644.5043.7142.95
σ 12.0211.9811.9411.91
CV 0.15100.15140.15180.1523
K IC 1.631.621.611.60
K IC min 0.850.840.830.81
σ 0.24850.24760.24670.2459
CV 0.15230.15270.15310.1537
2.5 f t 71.7571.3070.8670.43
f t min 41.1740.4839.8139.16
σ 10.7910.7510.7210.68
CV 0.15040.15080.15120.1516
K IC 1.641.631.621.61
K IC min 0.870.850.840.83
σ 0.24940.24850.24760.2467
CV 0.15170.15210.15250.1532
3.0 f t 65.9865.5765.1664.77
f t min 38.1837.5837.0036.39
σ 9.909.859.829.78
CV 0.15000.15030.15060.1510
K IC 1.661.651.641.63
K IC min 0.880.870.860.84
σ 0.25050.24950.24850.2476
CV 0.15120.15160.15190.1519
4.0 f t 57.9757.6057.2556.90
f t min 34.0633.5633.0732.53
σ 8.658.618.578.54
CV 0.14920.14950.14970.1500
K IC 1.681.671.661.65
K IC min 0.910.900.880.87
σ 0.25280.25160.25050.2495
CV 0.15050.15070.15090.1512
5.0 f t 52.5752.2451.9251.60
f t min 31.2830.8530.4129.97
σ 7.827.787.747.70
CV 0.14880.14890.14910.1493
K IC 1.701.691.681.67
K IC min 0.940.930.910.90
σ 0.25540.25400.25280.2517
CV 0.14990.15000.15030.1505
6.0 f t 48.6548.3448.0447.74
f t min 29.2328.8728.4828.10
σ 7.227.187.147.10
CV 0.14850.14850.14860.1487
K IC 1.731.721.711.69
K IC min 0.970.960.940.93
σ 0.25820.25670.25530.2540
CV 0.14950.14960.14970.1499
7.0 f t 45.6345.3445.0644.78
f t min 27.6227.3027.0026.64
σ 6.776.736.686.64
CV 0.14840.14830.14830.1484
K IC 1.751.741.731.72
K IC min 1.000.980.970.95
σ 0.26120.25960.25810.2566
CV 0.14930.14930.14940.1494
8.0 f t 43.2342.9642.6942.43
f t min 26.2926.0525.7525.47
σ 6.426.376.336.29
CV 0.14850.14830.14820.1482
K IC 1.771.761.751.74
K IC min 1.021.010.990.98
σ 0.26440.26260.26100.2594
CV 0.14920.14910.14910.1492
9.0 f t 41.2741.0140.7540.50
f t min 25.1824.9524.7124.47
σ 6.146.096.046.00
CV 0.14870.14850.14830.1481
K IC 1.791.781.771.76
K IC min 1.031.031.021.00
σ 0.26830.26590.26400.2623
CV 0.14950.14910.14900.1490
10.0 f t 39.6339.3839.1338.89
f t min 24.2124.0223.8323.61
σ 5.915.865.815.76
CV 0.14910.14870.14840.1482
K IC 1.821.801.791.78
K IC min 1.051.041.041.03
σ 0.27230.26980.26740.2654
CV 0.14990.14950.14910.1489
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Li, L.; Cao, H.; Guan, J.; He, S.; Niu, L.; Liu, H. A Three-Parameter Weibull Distribution Method to Determine the Fracture Property of PMMA Bone Cement. Polymers 2022, 14, 3589. https://0-doi-org.brum.beds.ac.uk/10.3390/polym14173589

AMA Style

Li L, Cao H, Guan J, He S, Niu L, Liu H. A Three-Parameter Weibull Distribution Method to Determine the Fracture Property of PMMA Bone Cement. Polymers. 2022; 14(17):3589. https://0-doi-org.brum.beds.ac.uk/10.3390/polym14173589

Chicago/Turabian Style

Li, Lielie, Hekai Cao, Junfeng Guan, Shuanghua He, Lihua Niu, and Huaizhong Liu. 2022. "A Three-Parameter Weibull Distribution Method to Determine the Fracture Property of PMMA Bone Cement" Polymers 14, no. 17: 3589. https://0-doi-org.brum.beds.ac.uk/10.3390/polym14173589

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop