Article

Towards an Optical Gas Standard for Traceable Calibration-Free and Direct NO2 Concentration Measurements

1
Physikalisch-Technische Bundesanstalt, Bundesallee 100, D-38116 Braunschweig, Germany
2
Technical University of Darmstadt, Department of Mechanical Engineering, Reactive Flows and Diagnostics, Otto-Berndt-Str. 3, 64287 Darmstadt, Germany
*
Authors to whom correspondence should be addressed.
Submission received: 30 April 2021 / Revised: 28 May 2021 / Accepted: 1 June 2021 / Published: 9 June 2021

Abstract

:
We report a direct tunable diode laser absorption spectroscopy (dTDLAS) instrument developed for NO2 concentration measurements without chemical pre-conversion, operated as an Optical Gas Standard (OGS). An OGS is a dTDLAS instrument that can deliver gas species amount fractions (concentrations), without any previous or routine calibration, which are directly traceable to the international system of units (SI). Here, we report NO2 amount fraction quantification in the range of 100–1000 µmol/mol to demonstrate the current capability of the instrument as an OGS for car exhaust gas application. Nitrogen dioxide amount fraction results delivered by the instrument are in good agreement with certified values of reference gas mixtures, validating the capability of the dTDLAS-OGS for calibration-free NO2 measurements. As opposed to the standard reference method (SRM) based on chemiluminescence detection (CLD) where NO2 is indirectly measured after conversion to NO, titration with O3 and the detection of the resulting fluorescence, a dTDLAS-OGS instrument has the benefit of directly measuring NO2 without distorting or delaying conversion processes. Therefore, it complements the SRM and can perform fast and traceable measurements, and side-by-side calibrations of other NO2 gas analyzers operating in the field. The relative standard uncertainty of the NO2 results reported in this paper is 5.1% (k = 1, which is dominated (98%) by the NO2 line strength), the repeatability of the results at 982.6 µmol/mol is 0.1%, the response time of the instrument is 0.5 s, and the detection limit is 825 nmol/mol at a time resolution of 86 s.

1. Introduction

Nitrogen dioxide (NO2) is a prominent air pollutant emitted from combustion processes, burning fuel, cars or power plants. Exposure to elevated levels of NO2 in the atmosphere causes serious respiratory diseases [1]. Accurate and reliable measurement methods for amount fraction (concentration) levels below 1 ppm are needed for atmospheric NO2 air quality control [2] and to monitor NO2 emitted directly from the exhausts of cars (1–2500 µmol/mol, note: µmol/mol is often called ppm) as required for the type approval emission limits quality control test (i.e., the EURO 5/6 emission standard states the range of NO2 emission levels, which need to be measured during type approval, to be in the 1–30 µmol/mol for cars with a functional exhaust catalyst, and 30–1000 µmol/mol for cars with a malfunctioning exhaust catalyst). The standard reference method (SRM) for measuring NO2 is based on chemiluminescence detection (CLD), as documented in the European standard EN 14211:2012 [3], where NO2 is only measured indirectly (i.e., NO2 in a sample is calculated from the difference of NOx and NO after the conversion (catalytic or photolytic) of NO2 to NO. The NO then reacts with the ozone (O3) to produce electronically excited NO*2 which decays to NO2 and emits light (luminescence) that is detected by, e.g., a photomultiplier tube. The NOx concentration is determined using the measured signals, and the NO2 concentration by subtracting NO (detected in an independent parallel channel) from NOx [4,5]). This indirect measurement of NO2 can lead to an overestimation of NO2, as other NOy gas species (HNO3, HNO2, N2O5) in the sample can also be converted into NO as well. It is worth noting that NO2 is the only air pollutant where the SRM is based on an indirect measurement. Due to the fact that NO2 is only indirectly measured, very reactive and that there is a lack of accurate static calibration reference materials CRM (gases), e.g., for car exhaust gas applications, in the 1–2500 ppm range with ≤1% uncertainties, there is the need to implement new accurate and reliable measurement methods (reference methods for ambient air and other applications) for direct NO2 measurements [2] and alternative calibration approaches (e.g., an instrument standard OGS) to complement CRMs [6].
Laser spectroscopic techniques provide an option to develop opportunities for accurate and reliable measurement methods for direct NO2 measurements [2]. Laser absorption spectroscopy has been demonstrated for direct measurements of the concentration of a variety of molecules in different applications such as atmospheric monitoring, combustion diagnostics and breath analysis [7,8,9,10,11,12,13,14,15,16,17,18,19,20,21], and therefore can serve as a good candidate to develop a reference method for direct NO2 measurements. There have been several studies to perform direct NO2 measurements and also addressing the shortcomings in the SRM-based measurements [2,22,23,24,25,26]. For atmospheric NO2 measurements, spectroscopic techniques such as tunable diode laser absorption spectroscopy (TDLAS), quantum cascade laser absorption spectroscopy (QCLAS), cavity ringdown spectroscopy (CRDS), photoacoustic spectroscopy (PAS) and cavity attenuated phase shift spectroscopy (CAPS) have been reported with sub ppb detection limits [22,24,27,28,29]. Differential optical absorption spectroscopy (DOAS), TDLAS, PAS and CRDS have been successfully demonstrated for the detection of NO2 emitted from automobile and aircraft engines, with detection limits in ppm down to the ppb range [23,30,31,32,33]. To date, for these spectroscopic NO2 measurements, information on the metrological traceability and uncertainty assessment following, e.g., the guide to the expression of uncertainty in measurement (GUM) [34] is rarely reported. Metrological traceability of results (i.e., traceability of the results to the SI) and uncertainty assessments according to the GUM [34] of the NO2 results derived using spectroscopic measurement methods are critical for quality control measurements [2] and type approval purposes [35].
Direct tunable diode laser absorption spectroscopy (dTDLAS) combined with the TILSAM (traceable infrared laser-spectrometric amount fraction measurement) method [36], has been demonstrated in a variety of applications addressing metrology data quality (SI-traceability and GUM compliant uncertainty assessments). dTDLAS is a variant of TDLAS that combines TDLAS with a special data evaluation procedure to derive concentration results that are directly traceable to the SI [7,37]. As reported on our previous work, a dTDLAS instrument is referred to as an optical gas standard (OGS) [7,8,9,10,37]. An OGS is similar to the ozone standard reference photometer (SRP) [38]. Employing dTDLAS, fast measurement frequencies (from sub-Hz to Hz) and response times (e.g., milliseconds to seconds) required in field measurements can be reached [39,40,41,42]. Focusing here on a new development for direct NO2 (reactive gas) measurements in car exhaust gases, our previous works in [7,8,9,10,37] focused on dTDLAS instrument developments for the measurement of H2O, CO and CO2 and H2O as well as reaching fast measurement frequencies [39,40,41,42].
In this paper, we report the details of a new dTDLAS spectrometer [43] specifically designed and developed for accurate direct NO2 measurements. The focus of this work was to report the development stages of the NO2 dTDLAS instrument towards an OGS (an instrument standard for direct NO2 measurements), the testing and validation steps performed, the potentials and the current limitations of the instrument. The capabilities for direct NO2 measurements (looking at all measured quantities desired for the results) are demonstrated via NO2 concentration measurements. A comparison of calibration-free dTDLAS NO2 concentration results (range: 100–1000 µmol/mol, within the car exhaust NO2 concentration range) versus certified reference gas mixtures is done as a validation step. We further show that the instrument is capable of direct NO2 measurements in car exhaust gas matrix. In order to compete with, and compliment CRMs and supplement the SRM for NO2 measurements, with other gas matrices such as the car exhaust gas matrix in focus, the traceability of the spectroscopic results to the SI is addressed and an uncertainty assessment following GUM principles is made for the NO2 results.

2. Experimental

Figure 1 depicts a schematic of the dTDLAS spectrometer for direct NO2 measurements [43]. The instrument employs an interband cascade DFB laser (Nanoplus) emitting at 3.4 µm, a 77 cm silconert© coated stainless-steel single pass gas cell and a photovoltaic mid-infrared detector (Vigo system, detector diameter: 1 mm2). The sampling lines as well as the gas cell are coated to minimize the potential adsorption effects of NO2 in the gas sample. For spectroscopic measurements, NO2 gas samples flow (typical flow rate: 1.5 L/min) through the gas cell controlled by a mass flow controller (MFC, Bronkhorst) at the gas input and a valve at the output (connected to a pump) of the sampling lines as shown in Figure 1. The laser wavelength is swept at a frequency of 139.8 Hz (similar to [7,8,9,10], and the collimated laser light is passed through the gas cell on to the detector. The signal from the detector, pre-amplified with a transimpedance amplifier (Femto DLPCA 200), is acquired using a data acquisition card (DAC, 18 bits, 625 KS/s NI PCI-6289) and is subsequently saved on a personal computer (PC) with a state-of-the-art operating system.
Figure 2a depicts typical signals (with NO2 absorption lines visible around 2919.6 cm−1) measured at 0.2 Hz (the measurement frequency for single scan measurements is 18 Hz, i.e., ~0.056 s time resolution which is longer than the acquisition time of approximately 0.00352 s (2200 points on the up ramp at 625KS/s) due to the extra time needed to save the data using our acquisition software and corresponds to a time resolution of about ~5.6 s, i.e., ~0.2 Hz measurement frequency for 100 averaged single scans) for a gas mixture with an NO2 amount fraction of 979 µmol/mol in the air. It could be noted here that the measurement frequency is different from the laser modulation frequency of 139 Hz. The absorbance data (Figure 2b derived using the data in Figure 2a, 100 averaged scans, 739 points in the fit window) are fitted by a set of 8 Voigt profiles [44] in order to derive the line areas (areas underneath the absorption lines), i.e., together with the NO2 line at 2919.59 cm−1, 7 other neighboring NO2 lines (see right y axis and the lines in red: line strength [45] plot in Figure 2b) are fitted in the spectral window selected for data analysis. The fitting was performed with the line positions, the line areas and the Lorentzian line widths kept free. The Doppler line widths were calculated at the measured gas temperature.
The bottom panel of Figure 2b shows the residuals from the multiline Voigt fit with 1σtotal = 2.7 × 10−4. Using the 1σtotal value and the peak absorbance value of 0.092, a signal-to-noise ratio of 341 was calculated for the data in Figure 2b. Due to the signal-to-noise ratio (SNR) around and below 100 and the complex multi-line structure of the NO2 line used, higher order line shape profiles like Galatry or Hartmann–Tran profiles [46,47,48] were not found to be suitable and advised to fit the data in this work. Using the line area of the NO2 line at 2919.59 cm−1, derived for the data in Figure 2b, the NO2 amount fraction (concentration) xNO2 can be derived using Equation (1) (which is derived from the Beer lambert law) [7]:
x NO 2 = k B T S T L p total A ( ν ˜ ) d ν ˜ = k B T S T L p total A line
where the quantity kB is the Boltzmann constant, ST is the line strength of the probed molecular transition at gas temperature T, L is the optical path length of the light beam transmitted through the absorbing medium, ptotal the total gas pressure, A ( ν ˜ ) the spectral absorbance and Aline the line area. The amount fraction (concentration) results xNO2 is directly traceable to the SI if all the quantities in the right-hand side of Eqn. 1 are traceable to the SI as well [7,8,9,37]. To this end, xNO2 results are derived without the calibration of the instruments with a reference gas mixture. As introduced, an instrument that can deliver amount fraction results that are directly traceable to the SI shall be referred to as an optical gas standard (OGS).
From the data in Figure 2b, a line area value of 0.0010999414 cm−1 is derived for the NO2 line 2919.59 cm−1, and a dTDLAS amount fraction value of (982.0 ± 49.1) µmol/mol calculated according to Eqn. 1 is combined with the TILSAM method [36]. This measured NO2 amount fraction is in good agreement with the certified value of (979 ± 20) µmol/mol reported for the probed gas mixture, demonstrating the capability of the instrument for an absolute (“calibration-free”) direct NO2 amount fraction measurements.
Table 1 holds an uncertainty budget for dTDLAS NO2 results. As shown in Table 1, the combined uncertainty of the NO2 results is 5.1% (k = 1). The quantity k is a coverage factor [34]. The uncertainty of the amount fraction results of 5.1% is dominated by the uncertainty of the measured line area (1%) and the literature line strength value (5% [45]) as indicated by the higher index values (% individual contribution) in Table 1. The uncertainties of the other significant input quantities (ptotal, T, kB, L) are all less than 0.2%.
In order to determine the repeatability of the measurements delivered by the instrument, Figure 3a shows repeated measurements of NO2 amount fractions performed with the 979 µmol/mol NO2 in synthetic air gas mixture. The mean value of the results in Figure 3a is 982.6 µmol/mol, the reproducibility (1σ, standard deviation) is 8.7 µmol/mol (0.9%) and the repeatability (standard deviation of the mean) is 0.5 µmol/mol (0. 1%). The excellent repeatability of 0.1% demonstrates the capability of a dTDLAS instrument for NO2 OGS applications. The inset in Figure 3a shows a histogram of the measurements, depicting a normal distribution of the results with a full width at half maximum (FWHM) of 9.2 µmol/mol (corresponding to an SNR of 107 for the amount fraction results at 979 µmol/mol). Figure 3b depicts a plot of the Allan deviation [49] of the NO2 amount fraction results (here, the measurements are done at 2 Hz, i.e., at 0.5 s with 10 averaged scans) as a function of time. From the Allan plot in Figure 3b, an optimal precision of 825 nmol/mol (detection limit) is evaluated (conservatively, as there is no minimum visible) for the instrument at a time resolution of 86 s.
Repeated measurements were performed at different amount fraction levels (switching from about 100 to 1000 µmol/mol as shown on Figure 4a) to study the behavior, response and linearity of the instrument. The gas mixture used for the measurements at the lowest amount fraction level of 100 µmol/mol was a primary gas mixture (SI traceable, prepared using gravimetry by a national metrology institute). At 300 µmol/mol we used a commercial gas mixture (secondary: SI-traceable, with a Deutscher Kalibrierdienst-DKD certificate [50]) and that at 1000 µmol/mol was a commercial gas mixture from the same manufacturer (internally assigned, with no SI-traceability information provided). The standard uncertainty reported for the NO2 amount fractions in these gas mixtures are all 1% (k = 1), corresponding to expanded uncertainties of 2% (k = 2).
As presented, the sampling lines and the gas cell were all coated by Silconert, resulting in a fast response time of 0.5 s for the NO2 measurements in Figure 4a. It could be noted here that this response time of 0.5 s does not also include the cylinder switching time and the time of manually opening valves to introduce the NO2 gas sample into the gas cell. The repeatability (calculated as the standard deviation of the mean) at the lowest NO2 results at the 100 µmol/mol level is 0.5 µmol/mol (0.5% relative). At the respective amount fraction levels, 1σ (standard deviation) values were evaluated, demonstrating the short-term reproducibility (<10 µmol/mol at 1σ) for the NO2 OGS results. Using the respective 1σ values, an SNR from 18 at 100 µmol/mol to 106 at 1000 µmol/mol levels, at a time resolution of 86 s, were evaluated for the measurements in Figure 4a.
Figure 4b shows a plot of the NO2 amount fraction results as a function of the certified values reported for the probed gas mixtures. A generalized linear regression (GLR) was applied to the data in Figure 4b, resulting in a slope of 0.984 ± 0.095 (k = 2) and an intercept value of (10.25 ± 16.10) µmol/mol, respectively. The slope value of 0.984 shows the good accuracy of the dTDLAS instrument, while the small deviation from the linear fit shows good linearity. The intercept value is consistent with zero, indicating an offset free measurement. The relative residuals from the linear model in the bottom panel of Figure 4b are all within the 1% range and thus well within the expanded uncertainty of the dTDLAS results of 5.1%, k = 1), which shows good agreement (compare well; validation step) with the certified values associated with the gas mixtures.

3. Discussion

The NO2 OGS spectrometer was tested in the concentration range from 100 to 1000 µmol/mol and was found to be in good agreement with the certified values of the reference gases, demonstrating the calibration-free OGS capability of the instrument. The detection limits of the dTDLAS instrument are 0.825 µmol/mol (at Δt = 86 s) and the relative uncertainty of the results delivered by the instrument is 5.1%, k = 1. This relative standard uncertainty was derived following GUM [34] principles. These results indicate that the instrument has the potential to measure NO2 (that can be in the 1 to 1000 µmol/mol: detection limit of 0.825 µmol/mol) emitted from the exhaust of cars as further supported by the simulation (HITRAN [45] line data used) in Figure 5 where the interference (due to absorption by other exhaust gas molecules) on NO2 absorption is insignificant (absorbance in the order of 10−5 for H2O, NO2, CO2, and CO: L = 77.4 cm, p = 100 hPa, T = 296 K). As presented, the relative repeatabilities of the NO2 results at 1000 µmol/mol and lowest amount fraction of 100 µmol/mol are 0.1% and 0.5%, respectively. Table 2 holds a summary of the capabilities of the NO2 OGS instrument presented in this work.
The relative uncertainties of the NO2 results of 5.1% (k = 1) are dominated (96% of the total uncertainty) by the uncertainty of the line strength value. Therefore, to further reduce the uncertainty of the results, future work will be focused on reducing the uncertainty of the line strength value. Since the dTDLAS results are in good agreement with the certified values (see Figure 4b), it can be concluded that the line strength used for the analysis is validated. The line strength value is taken from the HITRAN database [45]. This first validation of the line strength value was further assessed in a separate analysis to determine the line strength value from the data measured with the 300 µmol/mol gas mixture, resulting in a line strength value of (5.444 ± 0.076)·10−21 cm/molecule, k = 1, at 294.4 K for the NO2 line 2919.59 cm−1. It could be noted here that the accuracy of the measured line strength would be dependent on the accuracy of the certified gas mixture used. This line strength value (5.444 ± 0.076)·10−21 cm/molecule is in agreement with HITRAN’s [45] value of (5.425 ± 0.271)·10−21 cm/molecule, k = 1, at 296 K, further validating the HITRAN value [45]. In order to test the effect of a new line strength value with a reduced uncertainty, we used the line strength value 5.444·10−21 cm/molecule and the standard uncertainty (0.076·10−21 cm/molecule: 1.4% relative) to evaluate the uncertainty of the NO2 results (similar to Table 1), yielding relative uncertainty of 1.7% (k = 1). The uncertainty of the dTDLAS results currently stands at 5.1%, and a 1.7% relative uncertainty for the NO2 amount fraction results will fulfill the requirements of portable emission measurement systems (PEMS) of less than 2% for NO2. Regarding the line strength, we estimate that we can achieve an uncertainty better than 1.7% or 0.2% in future line strength measurements, similarly to in our previous works [51,52], which would lead to a corresponding improvement in the NO2 amount fraction uncertainty. One of the critical points in achieving an OGS is ensuring direct traceability to the SI. The traceability of the NO2 amount fraction results in this work was addressed via the traceability of the input quantities on the right-hand side of Eqn.1, such as quantities ptotal, T, L (SI-traceable to respective PTB standards). As presented, the validated line strength ST value was taken from the HITRAN database [45] and SI-traceability was not reported. Therefore, in addition to a reduced uncertainty in the line strength value, new SI-traceable NO2 line strength measurements are planned in separate experiments similar to [51,52].
This work shows that an OGS for NO2 amount fraction measurements is feasible. An OGS (providing direct NO2 amount fraction results traceable to the SI) for NO2 can complement the current standard reference method (based on CLD), where NO2 is “only” indirectly measured (note: the concept of indirect measurement is not the issue here, but rather the conversion efficiency of NO2 to NO and maintaining a minimum residence time). By providing amount fraction results that are directly traceable to the SI, an OGS can also complement (use in the place of) calibrate reference gases both in the laboratory and in the field. An OGS will add the advantage of simultaneous in situ measurement and side-by-side field NO2 calibrations of other analyzers.

4. Conclusions

We presented a new dTDLAS NO2 instruments are designed to be operated as an optical gas standard (OGS) and can thus perform absolute (“calibration free”) NO2 amount fraction measurements directly, i.e., without any chemical conversion which is needed in the current standard reference method CLD. Nitrogen dioxide amount fractions in the range of 100–1000 µmol/mol were delivered by the instrument and are in good agreement with certified reference values attributed to the reference gas mixtures. The relative uncertainty of the results is currently at 5.1% (k = 1) and is almost completely (96%) dominated by the inaccurate molecular reference data. The relative repeatability of the results at 982.6 µmol/mol is 0.1% and the instrument response time ranges from 0.5 s with a 10-scan average to 0.06 s in the single scan mode. Future work will focus on measuring a new SI-traceable NO2 line strength value to further reduce the line strength uncertainty influence and thus the instrument accuracy. The instruments’ detection limit of 0.825 µmol/mol makes it well suited for direct measurements of the NO2 concentration in car exhaust gases. Therefore, further plans are to extend the capability of the instrument for simultaneous measurements of NO2 and NO concentrations (in both atmospheric, i.e., less than 1 µmol/mol level, and car exhaust gas species).

Author Contributions

Experiment concept: J.A.N., Z.Q., V.E. Experiment realization and data collection: J.A.N., Z.Q. Data evaluation and discussion: J.A.N., Z.Q., O.W., V.E. Paper conceptualization writing—review and editing: J.A.N., Z.Q., O.W. and V.E. All authors have read and agreed to the published version of the manuscript.

Funding

This work was internally funded.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Acknowledgments

The authors would like to acknowledge the support from the PTB working group 3.42 (Spectrometric Gas Analysis) in the development of the instrument.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Salonen, H.; Salthammer, T.; Morawska, L. Human exposure to NO2 in school and office indoor environments. Environ. Int. 2019, 130, 104887. [Google Scholar] [CrossRef] [PubMed]
  2. EMPIR Project 16ENV02 Metrology for Nitrogen Dioxide (MetNO2). Available online: http://empir.npl.co.uk/metno2/ (accessed on 29 April 2021).
  3. European Standard: Ambient Air—Standard Method for the Measurement of the Concentration of Nitrogen Dioxide and Nitrogen Monoxide by Chemiluminescence, EN 14211:2012. Available online: https://www.beuth.de/en/standard/din-en-14211/146720023 (accessed on 27 May 2021).
  4. Miller, C.C. Chemiluminescence Analysis and Nitrogen-Dioxide Measurement. Lancet 1994, 343, 300–301. [Google Scholar] [CrossRef]
  5. Steinbacher, M.; Zellweger, C.; Schwarzenbach, B.; Bugmann, S.; Buchmann, B.; Ordóñez, C.; Prevot, A.S.H.; Hueglin, C. Nitrogen oxide measurements at rural sites in Switzerland: Bias of conventional measurement techniques. J. Geophys. Res. Atmos. 2007, 112, D11307. [Google Scholar] [CrossRef] [Green Version]
  6. EMPIR Project 19ENV09 Improved Vehicle Exhaust Quantification by Portable Emission Measurement Systems Metrology (MetroPEMS). Available online: https://www.metropems.ptb.de/home/ (accessed on 29 April 2021).
  7. Nwaboh, J.A.; Qu, Z.; Werhahn, O.; Ebert, V. Interband cascade laser-based optical transfer standard for atmospheric carbon monoxide measurements. Appl. Opt. 2017, 56, E84–E93. [Google Scholar] [CrossRef]
  8. Buchholz, B.; Böse, N.; Ebert, V. Absolute validation of a diode laser hygrometer via intercomparison with the German national primary water vapor standard. Appl. Phys. B 2014, 116, 883–899. [Google Scholar] [CrossRef]
  9. Pogany, A.; Wagner, S.; Werhahn, O.; Ebert, V. Development and Metrological Characterization of a Tunable Diode Laser Absorption Spectroscopy (TDLAS) Spectrometer for Simultaneous Absolute Measurement of Carbon Dioxide and Water Vapor. Appl. Spectrosc. 2015, 69, 257–268. [Google Scholar] [CrossRef]
  10. Qu, Z.; Nwaboh, J.; Werhahn, O.; Ebert, V. Towards a dTDLAS-Based Spectrometer for Absolute HCl Measurements in Combustion Flue Gases and a Better Evaluation of Thermal Boundary Layer Effects. Flow Turbul. Combust. 2021, 106, 533–546. [Google Scholar] [CrossRef]
  11. Wagner, S.; Klein, M.; Kathrotia, T.; Riedel, U.; Kissel, T.; Dreizler, A.; Ebert, V. Absolute, spatially resolved, in situ CO profiles in atmospheric laminar counter-flow diffusion flames using 2.3 μm TDLAS. Appl. Phys. B 2012, 109, 533–540. [Google Scholar] [CrossRef] [Green Version]
  12. Wu, K.; Li, F.; Cheng, X.; Yang, Y.; Lin, X.; Xia, Y. Sensitive detection of CO2 concentration and temperature for hot gases using quantum-cascade laser absorption spectroscopy near 4.2 μm. Appl. Phys. B 2014, 117, 659–666. [Google Scholar] [CrossRef]
  13. Foltynowicz, A.; Schmidt, F.M.; Ma, W.; Axner, O. Noise-immune cavity-enhanced optical heterodyne molecular spectroscopy: Current status and future potential. Appl. Phys. B 2008, 92, 313–326. [Google Scholar] [CrossRef]
  14. Henderson, B.; Khodabakhsh, A.; Metsälä, M.; Ventrillard, I.; Schmidt, F.M.; Romanini, D.; Ritchie, G.A.D.; Hekkert, S.T.L.; Briot, R.; Risby, T.; et al. Laser spectroscopy for breath analysis: Towards clinical implementation. Appl. Phys. B 2018, 124, 161. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  15. McCurdy, M.R.; Bakhirkin, Y.; Wysocki, G.; Lewicki, R.; Tittel, F.K. Recent advances of laser-spectroscopy-based techniques for applications in breath analysis. J. Breath Res. 2007, 1, 014001. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  16. Chao, X.; Jeffries, J.B.; Hanson, R.K. Absorption sensor for CO in combustion gases using 2.3 µm tunable diode lasers. Meas. Sci. Technol. 2009, 20, 115201. [Google Scholar] [CrossRef]
  17. Goldenstein, C.S.; Spearrin, R.M.; Jeffries, J.B.; Hanson, R.K. Wavelength-modulation spectroscopy near 2.5 μm for H2O and temperature in high-pressure and -temperature gases. Appl. Phys. B 2014, 116, 705–716. [Google Scholar] [CrossRef]
  18. Graf, M.; Scheidegger, P.; Kupferschmid, A.; Looser, H.; Peter, T.; Dirksen, R.; Emmenegger, L.; Tuzson, B. Compact and lightweight mid-infrared laser spectrometer for balloon-borne water vapor measurements in the UTLS. Atmos. Meas. Tech. 2021, 14, 1365–1378. [Google Scholar] [CrossRef]
  19. Viciani, S.; Montori, A.; Chiarugi, A.; D’Amato, F. A Portable Quantum Cascade Laser Spectrometer for Atmospheric Measurements of Carbon Monoxide. Sensors 2018, 18, 2380. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  20. Sur, R.; Ding, Y.; Jackson, R.B.; Hanson, R.K. Tunable laser-based detection of benzene using spectrally narrow absorption features. Appl. Phys. B 2019, 125, 195. [Google Scholar] [CrossRef]
  21. Liu, N.; Linguang, X.; Sheng, Z.; Tianbo, H.; Lei, Z.; Dianming, W.; Jingsong, L. Sensitive detection of atmospheric N2O isotopomers using a quantum cascade laser based spectrometer. J. Quant. Spectrosc. Radiat. Transf. 2019, 236, 106587. [Google Scholar] [CrossRef]
  22. Sobanski, N.; Tuzson, B.; Scheidegger, P.; Looser, H.; Kupferschmid, A.; Iturrate, M.; Pascale, C.; Hüglin, C.; Emmenegger, L. Advances in High-Precision NO2 Measurement by Quantum Cascade Laser Absorption Spectroscopy. Appl. Sci. 2021, 11, 1222. [Google Scholar] [CrossRef]
  23. Diemel, O.; Honza, R.; Ding, C.-P.; Böhm, B.; Wagner, S. In situ sensor for cycle-resolved measurement of temperature and mole fractions in IC engine exhaust gases. Proc. Combust. Inst. 2019, 37, 1453–1460. [Google Scholar] [CrossRef]
  24. Kebabian, P.L.; Herndon, S.C.; Freedman, A. Detection of Nitrogen Dioxide by Cavity Attenuated Phase Shift Spectroscopy. Anal. Chem. 2005, 77, 724–728. [Google Scholar] [CrossRef] [PubMed]
  25. Sur, R.; Peng, W.Y.; Strand, C.; Spearrin, R.M.; Jeffries, J.B.; Hanson, R.K.; Bekal, A.; Halder, P.; Poonacha, S.P.; Vartak, S.; et al. Mid-infrared laser absorption spectroscopy of NO2 at elevated temperatures. J. Quant. Spectrosc. Radiat. Transf. 2017, 187, 364–374. [Google Scholar] [CrossRef]
  26. Hundt, P.M.; Müller, M.; Mangold, M.; Tuzson, B.; Scheidegger, P.; Looser, H.; Hüglin, C.; Emmenegger, L. Mid-IR spectrometer for mobile, real-time urban NO2 measurements. Atmos. Meas. Tech. 2018, 11, 2669–2681. [Google Scholar] [CrossRef] [Green Version]
  27. Horii, C.V.; Zahniser, M.S.; Nelson, D.D., Jr.; McManus, J.B.; Wofsy, S.C. Nitric acid and nitrogen dioxide flux measurements: A new application of tunable diode laser absorption spectroscopy. In Application of Tunable Diode and Other Infrared Sources for Atmospheric Studies and Industrial Processing Monitoring II; SPIE: Denver, CO, USA, 1999; Volume 3758, pp. 152–161. [Google Scholar] [CrossRef]
  28. Mazurenka, M.I.; Fawcett, B.L.; Elks, J.M.F.; Shallcross, D.E.; Orr-Ewing, A.J. 410-nm diode laser cavity ring-down spectroscopy for trace detection of NO2. Chem. Phys. Lett. 2003, 367, 1–9. [Google Scholar] [CrossRef]
  29. Pan, Y.; Dong, L.; Yin, X.; Wu, H. Compact and Highly Sensitive NO2 Photoacoustic Sensor for Environmental Monitoring. Molecules 2020, 25, 1201. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  30. Dooly, G.; Fitzpatrick, C.; Lewis, E. Hazardous exhaust gas monitoring using a deep UV based differential optical absorption spectroscopy (DOAS) system. J. Phys. Conf. Ser. 2007, 76, 012021. [Google Scholar] [CrossRef]
  31. Jiang, Y.; Yang, T.; Liu, X.; Liu, Y. Automobile exhaust simulation detection system based on photoacoustic spectroscopy. In Proceedings of the ICOSM 2020: Optoelectronic Science and Materials, Hefei, China, 25–27 September 2020; Volume 11606, p. 1160608. [Google Scholar] [CrossRef]
  32. Yamamoto, Y.; Sumizawa, H.; Yamada, H.; Tonokura, K. Real-time measurement of nitrogen dioxide in vehicle exhaust gas by mid-infrared cavity ring-down spectroscopy. Appl. Phys. B 2011, 105, 923–931. [Google Scholar] [CrossRef]
  33. Li, Z.; Hu, R.; Xie, P.; Chen, H.; Liu, X.; Liang, S.; Wang, D.; Wang, F.; Wang, Y.; Lin, C.; et al. Simultaneous measurement of NO and NO2 by a dual-channel cavity ring-down spectroscopy technique. Atmos. Meas. Tech. 2019, 12, 3223–3236. [Google Scholar] [CrossRef] [Green Version]
  34. ISO. ISO Guide 98-3, Guide to the Expression of Uncertainty in Measurement; International Organization for Standardization: Geneva, Switzerland, 2008; ISBN 9267101889. [Google Scholar]
  35. Zacharof, N.; Tietge, U.; Franco, V.; Mock, P. Type approval and real-world CO2 and NOx emissions from EU light commercial vehicles. Energy Policy 2016, 97, 540–548. [Google Scholar] [CrossRef]
  36. Werhahn, O.; Petersen, J.C. (Eds.) TILSAM Technical Protocol V1_2010-09-29. 2010. Available online: http://www.euramet.org/fileadmin/docs/projects/934_METCHEM_Interim_Report.pdf (accessed on 29 April 2021).
  37. Nwaboh, J.A.; Persijn, S.; Arrhenius, K.; Bohlén, H.; Werhahn, O.; Ebert, V. Metrological quantification of CO in biogas using laser absorption spectroscopy and gas chromatography. Meas. Sci. Technol. 2018, 29, 095010. [Google Scholar] [CrossRef]
  38. The NIST Standard Reference Photometer for Ozone Measurement Traceability. Available online: https://www.nist.gov/programs-projects/nist-standard-reference-photometer-ozone-measurement-traceability (accessed on 29 April 2021).
  39. Witzel, O.; Klein, A.; Wagner, S.; Meffert, C.; Schulz, C.; Ebert, V. High-speed tunable diode laser absorption spectroscopy for sampling-free in-cylinder water vapor concentration measurements in an optical IC engine. Appl. Phys. B 2012, 109, 521–532. [Google Scholar] [CrossRef]
  40. Klein, A.; Witzel, O.; Ebert, V. Rapid, Time-Division Multiplexed, Direct Absorption- and Wavelength Modulation-Spectroscopy. Sensors 2014, 14, 21497–21513. [Google Scholar] [CrossRef] [Green Version]
  41. Witzel, O.; Klein, A.; Meffert, C.; Schulz, C.; Kaiser, S.A.; Ebert, V. Calibration-free, high-speed, in-cylinder laser absorption sensor for cycle-resolved, absolute H2O measurements in a production IC engine. Proc. Combust. Inst. 2015, 35, 3653–3661. [Google Scholar] [CrossRef]
  42. Witzel, O.; Klein, A.; Meffert, C.; Wagner, S.; Kaiser, S.; Schulz, C.; Ebert, V. VCSEL-based, high-speed, in situ TDLAS for in-cylinder water vapor measurements in IC engines. Opt. Express 2013, 21, 19951. [Google Scholar] [CrossRef] [PubMed]
  43. Nwaboh, J.A.; Qu, Z.; Li, G.; Kim, M.E.; Petersen, J.C.; Balslev-Harder, D.; Werhahn, O.; Ebert, V. Diode laser spectrometer for NO2 quantification: Absolute laser spectroscopic and direct NO2 concentration measurements for atmospheric monitoring within the EMPIR project MetNO2. In Proceedings of the EGU General Assembly 2020, Online. 4–8 May 2020. EGU2020-19259. [Google Scholar] [CrossRef]
  44. Whiting, E.E. An empirical approximation to the Voigt profile. J. Quant. Spectrosc. Radiat. Transf. 1968, 8, 1379–1384. [Google Scholar] [CrossRef]
  45. HITRAN Database. Available online: https://hitran.org/ (accessed on 29 April 2021).
  46. Ngo, N.H.; Lisak, D.; Tran, H.; Hartmann, J.-M. An isolated line-shape model to go beyond the Voigt profile in spectroscopic databases and radiative transfer codes. J. Quant. Spectrosc. Radiat. Transf. 2013, 129, 89–100. [Google Scholar] [CrossRef]
  47. Ngo, N.H.; Lisak, D.; Tran, H.; Hartmann, J.-M. An isolated line-shape model to go beyond the Voigt profile in spectroscopic databases and radiative transfer codes. J. Quant. Spectrosc. Radiat. Transf. 2013, 129, 89–100, Erratum in J. Quant. Spectrosc. Radiat. Transf. 2014, 134, 105. [Google Scholar] [CrossRef]
  48. Tennyson, J.; Bernath, P.F.; Campargue, A.; Csaszar, A.G.; Daumont, L.; Gamache, R.R.; Hodges, J.T.; Lisak, D.; Naumenko, O.V.; Rothman, L.S.; et al. Recommended isolated-line profile for representing high-resolution spectroscopic transitions (IUPAC Technical Report). Pure Appl. Chem. 2014, 86, 1931–1943. [Google Scholar] [CrossRef] [Green Version]
  49. Werle, P.; Mücke, R.; Slemr, F. The limits of signal averaging in atmospheric trace-gas monitoring by tunable diode-laser absorption spectroscopy (TDLAS). Appl. Phys. B 1993, 57, 131–139. [Google Scholar] [CrossRef]
  50. Deutscher Kalibrierdienst (DKD). Available online: https://www.dkd.eu/ (accessed on 29 April 2021).
  51. Werwein, V.; Brunzendorf, J.; Li, G.; Serdyukov, A.; Werhahn, O.; Ebert, V. High-resolution Fourier transform measurements of line strengths in the 0002-0000 main isotopologue band of nitrous oxide. Appl. Opt. 2017, 56, E99–E105. [Google Scholar] [CrossRef]
  52. Pogany, A.; Ott, O.; Werhahn, O.; Ebert, V. Towards traceability in CO2 line strength measurements by TDLAS at 2.7 µm. J. Quant. Spectrosc. Radiat. Transf. 2013, 130, 147–157. [Google Scholar] [CrossRef]
Figure 1. Schematic of the dTDLAS instrument for direct NO2 measurements (Det: detector).
Figure 1. Schematic of the dTDLAS instrument for direct NO2 measurements (Det: detector).
Applsci 11 05361 g001
Figure 2. (a) Typical waves at 108 hPa signals measured using the spectrometer in Figure 1; (b) absorbance data derived from the data in Figure 2a. With the eight NO2 lines shown (in red) in the line strength plot (right y axis), eight Voigt profiles are fitted to the data in order to derive the line areas.
Figure 2. (a) Typical waves at 108 hPa signals measured using the spectrometer in Figure 1; (b) absorbance data derived from the data in Figure 2a. With the eight NO2 lines shown (in red) in the line strength plot (right y axis), eight Voigt profiles are fitted to the data in order to derive the line areas.
Applsci 11 05361 g002
Figure 3. (a) Repeated NO2 measurements: plot of NO2 OGS amount fraction results vs. measurement number; (b) Allan deviation: plot of the xNO2 Allan deviation as a function of time.
Figure 3. (a) Repeated NO2 measurements: plot of NO2 OGS amount fraction results vs. measurement number; (b) Allan deviation: plot of the xNO2 Allan deviation as a function of time.
Applsci 11 05361 g003
Figure 4. (a) Plot of the NO2 OGS amount fraction results at different levels as a function of time (at Δt = 86 s, 1σ for 108.4, 305.0 and 983.5 are 6.2, 9.9 and 9.6, respectively); (b) dTDLAS NO2 amount fraction xNO2 results as a function-certified values of gas mixtures.
Figure 4. (a) Plot of the NO2 OGS amount fraction results at different levels as a function of time (at Δt = 86 s, 1σ for 108.4, 305.0 and 983.5 are 6.2, 9.9 and 9.6, respectively); (b) dTDLAS NO2 amount fraction xNO2 results as a function-certified values of gas mixtures.
Applsci 11 05361 g004
Figure 5. Simulation: NO2 in exhaust gas matrix.
Figure 5. Simulation: NO2 in exhaust gas matrix.
Applsci 11 05361 g005
Table 1. Uncertainty budget of the NO2 amount fraction results.
Table 1. Uncertainty budget of the NO2 amount fraction results.
ParameterValueRelative Standard
Uncertainty (k = 1)/%
Index (% Individual Contribution)
Pressure108.5 hPa0.200.10
Temperature294.3 K0.10≤0.05
Path length77.4 cm0.13≤0.05
Line strength
(HITRAN value [40])
5.425·10−21 cm/molecule5.0096.00
Line area0.0010999414 cm−11.003.80
NO2 concentration (xNO2) result982 µmol/mol5.10 (combined
uncertainty)
Table 2. Summary of NO2 OGS instrument capabilities.
Table 2. Summary of NO2 OGS instrument capabilities.
QuantityValue
Current concentration range (NO2)100–1000 µmol/mol
Optimal precision0.825 µmol/mol at
Δt = 86 s
GUM compliant total relative uncertainty5.1% (k = 1)
Relative repeatability at 1000 µmol/mol and at 100 µmol levels0.1% and 0.5%
Time resolution
(max = single scan to 10 scan average)
0.06 * sec to 0.5 s
* >acquisition time of 3.5 ms as extra time is needed to save the data.
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Nwaboh, J.A.; Qu, Z.; Werhahn, O.; Ebert, V. Towards an Optical Gas Standard for Traceable Calibration-Free and Direct NO2 Concentration Measurements. Appl. Sci. 2021, 11, 5361. https://0-doi-org.brum.beds.ac.uk/10.3390/app11125361

AMA Style

Nwaboh JA, Qu Z, Werhahn O, Ebert V. Towards an Optical Gas Standard for Traceable Calibration-Free and Direct NO2 Concentration Measurements. Applied Sciences. 2021; 11(12):5361. https://0-doi-org.brum.beds.ac.uk/10.3390/app11125361

Chicago/Turabian Style

Nwaboh, Javis A., Zhechao Qu, Olav Werhahn, and Volker Ebert. 2021. "Towards an Optical Gas Standard for Traceable Calibration-Free and Direct NO2 Concentration Measurements" Applied Sciences 11, no. 12: 5361. https://0-doi-org.brum.beds.ac.uk/10.3390/app11125361

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop